← Product Code [FDS](/productcode/FDS) · K093779

# CLEARPATH UGI (K093779)

_Easyglide , Ltd. · FDS · Mar 9, 2010 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K093779

## Device Facts

- **Applicant:** Easyglide , Ltd.
- **Product Code:** [FDS](/productcode/FDS.md)
- **Decision Date:** Mar 9, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The ClearPath-UGI System is intended for irrigating or cleaning the upper digestive tract and evacuating the irrigation fluid, blood and bile in the upper GI tract during endoscopic procedures. It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

## Device Story

ClearPath UGI is a gastroscope accessory designed to improve procedural visualization. The system consists of a disposable irrigator head and a reusable control cabinet. The irrigator attaches to a standard endoscope via silicone bands. The control cabinet utilizes a peristaltic pump to manage water flow and vacuum pressure, drawing from the facility's standard medical vacuum system. Operated by trained medical personnel in clinical settings, the device enables irrigation and evacuation of fluids, blood, and bile from the upper GI tract. It does not interfere with endoscope functionality. The device benefits patients by maintaining a clear visual field during endoscopic procedures, potentially increasing reliability and efficiency.

## Clinical Evidence

Bench testing only. Testing demonstrated that the operating performance of the ClearPath UGI is equivalent to the predicate for flow rate (up to 300 ml/min), irrigation pressure (up to 30 psi), and suction pressure (approximately 0.5 Bar).

## Technological Characteristics

System comprises a disposable irrigator and reusable control cabinet. Materials in patient contact are identical to predicate K091305. Peristaltic pump provides irrigation; facility wall suction provides vacuum. Connectivity is standalone. No software modifications were made.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Easy Glide – ClearPath ([K091305](/device/K091305.md))
- Karl Storz video gastroscope ([K060907](/device/K060907.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary Page 1 of 2

7-Dec-09

Kog3779

EasyGlide Ltd. 30 Ha'Ella St. Kfar Truman, 73150 Israel

Tel - 011-972-528565644 Fax - 011-972-776201003

| Official Contact:          | Izhak Fabian - CEO    |
|----------------------------|-----------------------|
| Proprietary or Trade Name: | ClearPath UGI         |
| Common/Usual Name:         | Gastroscope accessory |

MAR - 9 ŽOID

Device:

ClearPath UGI

CFR 876.1500

FDS - endoscope and accessories

| Predicate Devices: | K091305 – Easy Glide – ClearPath       |
|--------------------|----------------------------------------|
|                    | K060907 - Karl Storz video gastroscope |

### Device Description:

Classification Name/Code:

The ClearPath UGI's main purpose is to improve procedure reliability by improving procedural visualization during endoscopic procedures.

The ClearPath UGI utilizes a suction / irrigation head for the purpose of irrigating or cleaning the upper digestive tract.

The ClearPath UGI is composed of two major units:

- . The disposable Irrigator and
- . the reusable Control cabinet.

The disposable Irrigator is attached to a standard endoscope by means of silicone bands and does not hinder the endoscope's functionality nor affects the procedure sequence in any way.

The Control cabinet supplies the water flow and vacuum control using a peristaltic pump. The standard medical facility vacuum system is the vacuum source.

### Indications for Use:

The ClearPath-UGI System is intended for irrigating or cleaning the upper digestive tract and evacuating the irrigation fluid, blood and bile in the upper GI tract during endoscopic procedures.

It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

# 510(k) Summary

{1}------------------------------------------------

Page 2 of 2 7-Dec-09

**Patient Population:**

Individuals undergoing procedures where a gastroscope is used.

### Environment of Use:

Hospitals, clinics, and doctors' offices.

### Summary of substantial equivalence:

We have performed testing to demonstrate that the operating performance of the ClearPath UGI is equivalent to the predicate for flow rate, irrigation and vacuum pressures.

|                                        | Proposed<br>ClearPath UGI                      | ClearPath<br>EasyGlide                         |
|----------------------------------------|------------------------------------------------|------------------------------------------------|
| 510(k) number                          |                                                | K091305                                        |
| Air / Water pressure<br>specifications | Up to 30 Psi                                   | Up to 30 Psi                                   |
| Flow rate                              | Up to 300 ml/min                               | Up to 300 ml/min                               |
| Suction<br>specifications              | Standard wall suction<br>Approximately 0.5 Bar | Standard wall suction<br>Approximately 0.5 Bar |

The proposed ClearPath UGI is only a modification of the predicate ClearPath in that the disposable tubing set is shorter in length and the irrigation tip has been redesigned. No modification have been made to the controller or software.

The ClearPath UGI is viewed as substantially equivalent to the predicate devices because:

### Indications -

- . Identical to predicate - K060907 - Karl Storz
#### Technology -

- Identical technology used K091305 EasyGlide ClearPath .
#### Materials -

- The materials in patient contact are identical to the predicate device, ClearPath K091305. .
#### Environment of Use -

- Identical to predicate K091305 EasyGlide ClearPath .
#### Performance specifications -

- Irrigation pressures, flow rate and suction pressures are identical to the predicate -. K091305 EasyGlide ClearPath

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with three lines forming its body and wings, symbolizing service to the nation. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" encircles the eagle, indicating the department's name and country.

9 2010 MAR

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

EasyGlide Ltd. % Mr. Paul Dryden President ProMedic, Incorporated 24301 Woodsage Drive BONITA SPRINGS FL 34134

K093779 Re:

> Trade/Device Name: ClearPath UGI Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDS Dated: December 8, 2009 Received: December 9, 2009

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related

{3}------------------------------------------------

# Page 2 -

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/SafetyReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Statement

Page 1 of 1

510(k) Number:

( 09 3779 (To be assigned)

Device Name:

ClearPath UGI

Indications for Use:

The ClearPath-UGI System is intended for irrigating or cleaning the upper digestive tract and evacuating the irrigation fluid, blood and bile in the upper GI tract during endoscopic procedures.

It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

Prescription Use XX (Part 21 CFR 801 Subpart D)

.

or

Over-the-counter use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hul Remu

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

---

**Source:** [https://fda-staging.innolitics.com/device/K093779](https://fda-staging.innolitics.com/device/K093779)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
