K093758 · Bg Medicine, Inc. · OSX · Nov 17, 2010 · Clinical Chemistry
Device Facts
Record ID
K093758
Device Name
BGM GALECTIN -3
Applicant
Bg Medicine, Inc.
Product Code
OSX · Clinical Chemistry
Decision Date
Nov 17, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1117
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K093758 · Nov 17, 2010
BGM GALECTIN -3
Bg Medicine, Inc.
Banked EDTA-plasma samples from the HF-ACTION clinical trial; Banked EDTA-plasma samples from an independent European multi-center clinical study
Retrospective analysis of banked clinical trial samples was used to validate the clinical effectiveness of prognostic cut-off values and to demonstrate the association of galectin-3 levels with mortality and hospitalization endpoints in chronic heart failure patients.
Heart Failure: A Controlled Trial Investigating Outcomes of Exercise Training (HF-ACTION); Retrospective analysis of banked samples from a prospective clinical trial; Follow-up/Duration: Median follow-up of approximately 30 months
895 chronic heart failure patients with left ventricular dysfunction (NYHA class II, III, or IV); Sample Size: 895; Number of Sites: Multi-center
Independent European multi-center clinical study; Retrospective analysis of banked samples; Follow-up/Duration: Median follow-up > 18 months
582 chronic heart failure patients (NYHA class II, III, and IV); Sample Size: 582; Number of Sites: Multi-center
Not applicable for this study
Composite of all-cause mortality and hospitalization for HF
Indications for Use
The BGM Galectin-3 Assay is an in vitro diagnostic device that quantitatively measures galectin-3 in serum or plasma by enzyme-linked immunosorbant assay (ELISA) on a microtiter plate platform. BGM Galectin-3 Assay is indicated to be used in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure (HF).
Device Story
In vitro diagnostic ELISA; measures galectin-3 protein levels in human serum or EDTA plasma. Input: patient serum/plasma sample; process: microtiter plate coated with rat anti-mouse galectin-3 capture antibody; incubation with sample; addition of HRP-labeled mouse anti-human galectin-3 tracer antibody; addition of TMB substrate; colorimetric reaction stopped with sulfuric acid; output: absorbance read at 450 nm via microtiter plate reader. Used in clinical laboratories; results interpreted by physicians alongside clinical evaluation to assess prognosis of chronic heart failure patients. Higher galectin-3 levels correlate with increased risk of all-cause mortality and heart failure-related hospitalization. Complements natriuretic peptide testing.
Clinical Evidence
Clinical validation study used 895 banked EDTA-plasma samples from the HF-ACTION trial (chronic HF patients, NYHA class II-IV). Primary endpoints: all-cause mortality, cardiovascular mortality, and hospitalizations. Results: Galectin-3 levels >17.8 ng/mL significantly associated with increased hazard of death/hospitalization (HR 1.35-1.46 for composite endpoints, p<0.01) after adjusting for age, gender, NYHA class, LVEF, diabetes, and smoking. Median follow-up 30 months. Bench testing established precision (total CV 4.2-12.0%), linearity (1.4-94.8 ng/mL), and LoQ (1.32 ng/mL).
Technological Characteristics
Sandwich ELISA on microtiter plate. Solid phase: rat monoclonal anti-mouse galectin-3 antibody. Tracer: HRP-labeled mouse monoclonal anti-human galectin-3 antibody. Detection: TMB substrate, colorimetric reader at 450 nm. Sample types: serum, EDTA-plasma. Measuring range: 1.4-94.8 ng/mL. No high-dose hook effect up to 500 ng/mL. Validated per CLSI guidelines (EP5-A2, EP6-A, EP17-A, EP7-A).
Indications for Use
Indicated for use as an aid in assessing the prognosis of patients diagnosed with chronic heart failure (HF). For prescription use only.
Regulatory Classification
Identification
The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.
Special Controls
*Classification.* Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”
Predicate Devices
Triage B-Type Natriuretic Peptide (BNP) Test (K080269)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k093758
B. Purpose for Submission:
New device
C. Measurand:
Galectin-3
D. Type of Test:
Quantitative enzyme-linked immunosorbant assay (ELISA)
E. Applicant:
BG Medicine, Inc.
F. Proprietary and Established Names:
BG Medicine, Inc. Galectin-3 Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1117, Test, Natriuretic Peptide
2. Classification:
Class II
3. Product code:
OSX, Galectin-3 in vitro diagnostic assay
4. Panel:
75, Chemistry
{1}
# H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The BGM Galectin-3 Assay is an in vitro diagnostic device that quantitatively measures galectin-3 in serum or plasma by enzyme-linked immunosorbant assay (ELISA) on a microtiter plate platform. BGM Galectin-3 Assay is indicated to be used in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure (HF).
3. Special conditions for use statement(s):
Prescription use only
4. Special instrument requirements:
96-well microtiter plate reader capable of reading at $450\mathrm{nm}$
# I. Device Description:
The Galectin-3 Assay is an in vitro diagnostic device that contains the microtiter plate, reagents, controls and standards required to perform analyses on serum or EDTA plasma samples. The Galectin-3 Assay contents are listed in the table below:
| Quantity | Name | Description |
| --- | --- | --- |
| 1 plate | Plate | Ready-to-use microtiter plate coated with rat monoclonal anti- mouse galectin-3 monoclonal antibody |
| 1 bottle | Assay diluent | Phosphate buffered saline with 1% bovine serum albumin |
| 1 bottle | TMB substrate | Tetramethyl benzidine (15 mL) |
| 1 bottle | Stop solution | 0.5 M sulfuric acid (10 mL) |
| 2 bottles | Wash buffer concentrate | 0.5M tris buffered saline (2 x 50 mL); (10X concentrate) |
| 1 bottle | Tracer concentrate | Horseradish peroxidase (HRP) labeled mouse monoclonal anti- human galectin-3 monoclonal antibody (0.45 mL) |
| 2 vials | Standard | Recombinant human galectin-3 12 ng per vial (lyophilized) |
| 2 vials | Low Quality Control (QC)* | Low QC material. Recombinant human galectin-3 in protein matrix (lyophilized) |
| 2 vials | High Quality | High QC material, assayed. Recombinant human |
| | | galectin-3 in protein matrix (lyophilized) |
| 2 vials | Low Quality | Low QC material. Recombinant human galectin-3 in protein matrix (lyophilized) |
{2}
Contains processed human plasma tested negative or nonreactive for anti-HIV-1/2, anti-HCV and HBsAg when tested by an FDA approved method.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Triage B-Type Natriuretic Peptide (BNP) Test
2. Predicate 510(k) number(s): k080269
3. Comparison with predicate:
| | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications for Use | indicated for use in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure | aid in the diagnosis and assessment of severity of heart failure, an aid in the risk stratification of patients with heart failure, and an aid in the risk stratification of patients with acute coronary syndromes. |
| Sample type | Serum and plasma | Whole blood and plasma |
| Instrumentation | Microtiter plate reader | Triage® Meter |
| Technology | ELISA | Lateral flow immunoassay |
| Cut-off | galectin-3 risk categories: • galectin-3 greater than 25.9 ng/mL • galectin-3 between 17.8 and 25.9 ng/mL • galectin-3 less than or equal to 17.8 ng/mL | 100 pg/mL |
| Assay range | 1.4 to 94.8 ng/mL | 5-5000 pg/mL |
{3}
4
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline
CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach: Approved Guideline
CLSI EP7-A: Interference Testing in Clinical Chemistry; Approved Guideline
CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline
CLSI C28-A2: How to Define and Determine Reference Intervals in the Clinical Laboratory
L. Test Principle:
BGM Galectin-3 is a microtiter plate-based ELISA for the quantitative determination of galectin-3 levels in human serum and plasma. The Galectin-3 Assay utilizes two monoclonal antibodies against galectin-3. One rat monoclonal anti-mouse galectin-3 antibody is coated onto the surface of the wells in a microtiter plate and serves as the capture antibody to bind galectin-3 molecules in samples, while the other mouse monoclonal anti-human galectin-3 antibody is provided in solution and functions as the tracer antibody for detecting galectin-3 molecules bound to the capture antibody. The microtiter plate is ready to use.
Standards, quality control materials, and patient samples are introduced into the wells and incubated for 60 minutes. During this incubation, the galectin-3 present in the standards and samples is bound to the capture antibody coated onto the well surface. A subsequent wash step removes any unbound galectin-3.
The tracer antibody, a horseradish peroxidase (HRP) labeled anti-galectin-3 antibody, is then introduced into the well and incubated for 60 minutes. During this time, an antibody-antigen-antibody complex is formed.
After a wash step to remove any unbound tracer antibody, the tetramethyl benzidine substrate is added, yielding a blue color in the presence of HRP. The color development is stopped after 20 minutes by the addition of sulfuric acid, changing the color to yellow that can be read at an absorbance of 450 nm. The test results of the samples are read from the calibration curve. The absorbance is proportional to the galectin-3 levels in the samples.
{4}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision of the Galectin-3 Asay was assessed in an evaluation according to CLSI EP5-A2 Guidelines. EDTA plasma pools spanning a range of galectin-3 concentrations were analyzed in duplicate with 2 runs per day over 20 days using one reagent lot, two operators and one microtiter plate reader. Estimates of within-run, run-to-run, day-to-day and total precision were calculated. The results are summarized in the table below:
| Test specimen (n) | Mean (ng/m L) | Within-run | | Run to run | | Day to day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV% | SD | CV% | SD | CV% | SD | CV% |
| 1 (160) | 6.1 | 0.3 | 5.7 | 0.6 | 10.5 | 0.0 | 0.0 | 0.7 | 12.0 |
| 2 (168) | 17.6 | 0.4 | 2.1 | 0.7 | 3.8 | 0.5 | 2.8 | 0.9 | 5.1 |
| 3 (160) | 20.7 | 0.7 | 3.4 | 1.4 | 6.7 | 0.3 | 1.7 | 1.6 | 7.7 |
| 4 (168) | 26.3 | 0.6 | 2.2 | 0.8 | 3.0 | 0.5 | 2.1 | 1.1 | 4.2 |
| 5 (168) | 46.2 | 1.1 | 2.4 | 1.6 | 3.6 | 0.5 | 1.1 | 2.0 | 4.4 |
| 6 (160) | 72.2 | 2.4 | 3.3 | 4.3 | 6.0 | 2.9 | 4.0 | 5.7 | 8.0 |
Additional precision studies were performed at three CLIA-certified clinical laboratories. Three EDTA plasma pools spanning a range of galectin-3 concentrations were analyzed in quadruplicate with 2 runs over 20 days for Site 1, 17 days for Site 2 and 18 days for Site 3 using two reagent lots and 3 different microtiter plate readers (one at each site) and a total of four operators. Pooled human plasma was divided and spiked to yield 3 concentrations spanning the measurement range. The results are summarized in the table below:
| Site | Test specimen | n | Mean (ng/mL) | SD (Total) | CV% (Total) |
| --- | --- | --- | --- | --- | --- |
| Site 1 | Low | 160 | 5.96 | 0.46 | 7.66 |
| | Medium | 160 | 20.12 | 1.20 | 5.96 |
| | High | 156 | 68.32 | 9.97 | 14.59 |
| Site 2 | Low | 132 | 6.72 | 0.63 | 9.36 |
| | Medium | 136 | 21.60 | 1.55 | 7.17 |
| | High | 136 | 75.50 | 12.75 | 16.89 |
| Site 3 | Low | 140 | 6.28 | 0.59 | 9.46 |
| | Medium | 144 | 21.22 | 1.19 | 5.60 |
| | High | 140 | 71.47 | 6.25 | 8.75 |
{5}
b. Linearity/assay reportable range:
Linearity was calculated based on CLSI guideline EP6-A, Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach. The claimed measurement range of the assay is 1.4 to 94.8 ng/mL. Low and high serum samples were mixed in different fractional parts to create an 11-level dilution series of samples with equally spaced concentrations of galectin-3 ranging from 1.4 to 94.8 ng/mL. Samples were tested in duplicate. Linearity of BGM Galectin-3 was demonstrated and supported the claimed measurement range. The regression equation for the linear regression studies is y = 0.9905x - 0.41. The percent recoveries for each dilution level ranged from 95.5 to 101.5 %.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The Galectin-3 kit includes two lyophilized quality control materials (C1 and C2). The controls are assayed materials that are supplied with target values and QC ranges. Lot-specific target values and ranges have been assigned with multiple operators over multiple days. The lower control (C1) is formulated with a target galectin-3 concentration of 18.4 ng/mL and is designed to monitor the performance of patient specimens near the clinical cutoff level of 17.8 ng/mL and specimens in the lower or normal range corresponding to the lower end of the standard curve. The higher control (C2) is formulated with approximately 69 ng/mL galectin-3 and is designed to monitor the performance of the assay for patient specimens with higher galectin-3 values and the upper end of the standard curve. The controls are comprised of recombinant galectin-3 in a protein matrix. The controls are stable for 12 months. After use, remaining reconstituted controls may be stored for a maximum of 10 days at 2-8°C.
Galectin-3 is calibrated with a set of seven standards that are prepared by the user by serially diluting a reconstituted vial of the standard (S1). The seven standards once prepared are stable for a maximum of 10 days at 2-8°C. Two vials of lyophilized standard (S1) are supplied with each kit. The calibrators (standard S1) are stable for 12 months un-opened. After reconstitution, remaining standard may be stored for a maximum of 10 days at 2-8°C.
The calibrator (SI) is prepared from recombinant human galectin-3 protein expressed in E. coli. A stock solution of galectin-3 is prepared and the concentration is determined spectrophotometrically in-house. From the stock solution, 10 working calibrators are prepared and used to derive a calibration curve for value assignment of the lyophilized calibrator materials (standard S1) which are supplied with assay kit.
6
{6}
d. Detection limit:
LoB, LoD and LoQ were determined according to the CLSI guideline EP 17-A. The limit of blank (LoB) was determined as the 95th percentile value of forty-eight (48) replicate measurements of the BGM Galectin-3 Assay Buffer. The limit of detection (LoD) was determined as LoD = LoB + cβ SDs, where SDs is the pooled standard deviation from four (4) serum samples with different levels of galectin-3, each of which was measured in thirty-two (32) replicates and β is the 95th percentile of the standard Gaussian distribution corrected for the degree of freedom. The Limit of quantitation (LoQ) was specified as the lowest galectin-3 concentration that is above the LoD and corresponds to a coefficient of variation (CV) of no more than 10.4% in the four serum samples.
Limit of Blank (LoB): LoB = 0.86 ng/mL
Limit of Detection (LoD): LoD = 1.13 ng/mL
Limit of Quantitation (LoQ): LoQ = 1.32 ng/mL
The lower limit of the measuring range was set at 1.4 ng/mL based on the linearity studies (see section M[b] above).
e. Analytical specificity:
Galectin-3 was evaluated for the effects of potential interfering substances, both endogenous and exogenous, according to the recommendations of the CLSI EP7-A guideline. In the studies, potential interferents were added to EDTA plasma samples with levels of galectin-3 at approximately 12, 22 and 40 ng/mL. Conjugated bilirubin (up to 16.8 mg/dL), unconjugated bilirubin (up to 40.3 mg/dL), albumin (BSA, up to 12 g/dL), triglycerides (up to 3000 mg/dL), cholesterol (up to 747 mg/dL), and creatinine (up to 5 mg/dL) do not show any significant interference with the assay (defined by the sponsor as ≥10 % interference). Purified hemoglobin (up to 500 mg/dL) does not show significant interference in BGM Galectin-3; however, packed blood cell lysate does show interference ≥ 10%. Human anti-mouse antibodies (HAMA) and rheumatoid factor (RF) above 50 IU/mL cause significant positive interference with BGM Galectin-3. High levels of gamma globulins (> 2.5 g/dL) may cause false elevation in galectin-3 levels. The following table is included in the labeling:
{7}
| Potential interfering substance | Result of interference study based on an interference acceptance limit of +/- 10% |
| --- | --- |
| Conjugated bilirubin | No significant interference up to 16.8 mg/dL |
| Unconjugated bilirubin | No significant interference up to 40.3 mg/dL |
| Albumin | No significant interference up to 12 g/dL |
| Triglycerides | No significant interference up to 3000 mg/dL |
| Cholesterol | No significant interference up to 747 mg/dL |
| Creatinine | No significant interference up to 5 mg/dL |
| Purified hemoglobin | No significant interference up to 500 mg/dL |
| Whole blood lysate | Hemolyzed specimens should not be used with BGM Galectin-3™ |
| Human anti-mouse antibodies (HAMA) | Specimens from patients with HAMA should not be used with BGM Galectin-3™ |
| Rheumatoid Factor (RF) | Interference seen at levels > 50 IU/mL |
| Gamma globulins | Interference seen at levels ≥ 2.5 g/dL |
BGM Galectin-3 measurements were not significantly affected (defined by the sponsor as $\geq 10\%$ interference) when tested in the presence of 34 common pharmaceutical substances; including heart failure drugs.
Common Drugs That Did Not Show Significant Interference with BGM Galectin-3:
| Acetaminophen | Carvedilol | Dopamine | Lisinopril | Quinidine |
| --- | --- | --- | --- | --- |
| Acetylsalicylic acid | Captopril | Enalaprilat | Losartan | Ramipril |
| Amlodipine | Chloramphenicol | Furosemide | Lovastatin | Spironolactone |
| Ampicillin | Diclofenac | Hydrochlorothiazide | Methyldopa | Theophylline |
| Ascorbic Acid | Digoxin | Ibuprofen | Metoprolol | Verapamil |
| Atenolol | Diltiazem | Indomethacin | Naproxen | Warfarin |
| Caffeine | Disopyramide | Lidocaine | Nifedipine | |
Studies were performed which demonstrated that there is no high dose hook effect at galectin-3 levels up to $500\mathrm{ng / mL}$. High concentrations of galectin-3 were spiked into different matrices (serum and EDTA plasma) and analyzed in duplicate. At spiked concentrations of 250 and $500\mathrm{ng / mL}$, readings were outside the optical range of the plate reader indicating values greater than the measuring range of the assay.
{8}
f. Assay cut-off:
See Clinical Cutoff in M (4) below
2. Comparison studies:
a. Method comparison with predicate device:
See clinical studies in M (3) and M (4) below
b. Matrix comparison:
The BGM Galectin-3 assay has been validated for use with EDTA plasma or serum samples. The equivalence of these sample matrices were demonstrated in a study of forty-nine (49) matched natural serum and EDTA-plasma samples with values spanning the measurement range. The samples had values ranging from 5.45 to 90.61 ng/mL. The regression equation for the study was y = 0.96x + 0.72.
3. Clinical studies:
a. Clinical Sensitivity:
See 3(c) below
b. Clinical specificity:
See 3(c) below
c. Other clinical supportive data (when a. and b. are not applicable):
Cutoff values were determined in an independent, controlled multi-center clinical study in Europe enrolling NYHA I, II, III and IV subjects (see section M [4] Clinical Cutoff below for further details).
To validate the clinical effectiveness of the cut-off values for the BGM Galectin-3 assay, galectin-3 levels were measured in a set of 895 banked EDTA-plasma samples from chronic heart failure participants in the United States and Canada in a controlled multi-center clinical study, the Heart Failure: A Controlled Trial Investigating Outcomes of Exercise Training (HF-ACTION) study. The HF-ACTION study involved 2,331 chronic HF patients with left ventricular dysfunction and with NYHA class II, III or IV symptoms. The average age of the 895 participants whose galectin-3 levels were assessed in the clinical validation study was 58 years, 29% were female, and 36% were non-white. A statistical analysis (sensitivity analysis) was performed comparing the set of 895 HF-ACTION subjects having evaluable galectin-3 values with all other HF-ACTION participants (1436), and it was found that
{9}
the clinical validation results based on the evaluable set of 895 subjects were robust and representative of the larger study population. In this sensitivity analysis, galectin-3 values were imputed conservatively for the 1436 remaining patients in the dataset based on the probability of the assay categorizing a patient into a high or low risk group. The difference in survival curves for the risk groups remained statistically significant, indicating that the results on the evaluable subset (895) were robust and representative of the entire study population. The median follow-up time was approximately 30 months. Chronic heart failure participants were categorized based on galectin-3 risk categories defined below:
The derived galectin-3-dependent risk categories are as follows:
- galectin-3 greater than 25.9 ng/mL
- galectin-3 between 17.8 and 25.9 ng/mL
- galectin-3 less than or equal to 17.8 ng/mL
For the clinical validation study, Cox regression models were used to evaluate the association of baseline galectin-3 levels in 895 chronic HF patients with the endpoints of: (i) composite of all-cause mortality and all-cause hospitalization, (ii) cardiovascular mortality, (iii) composite of cardiovascular mortality and heart failure-related hospitalization, and (iv) all-cause mortality. Galectin-3 levels were found to be significantly associated with increased risk of each of these endpoints in Cox regression models. Galectin-3 remained significantly associated with increased risk upon adjustment for baseline risk factors of age, gender, NYHA functional classification, left ventricular ejection fraction, diabetes status, and smoking status. The following figures and tables display Kaplan Meier curves for the composite endpoint of all-cause mortality or all-cause hospitalization, by baseline galectin-3 category and cumulative probabilities for events for the endpoints of the composite of all-cause mortality and all-cause hospitalization, cardiovascular mortality, and the composite of cardiovascular mortality and heart failure-related hospitalization in the clinical validation study, by baseline galectin-3 category at time points of 6, 12, 24 and 36 months after baseline.
10
{10}
# All Cause Mortality and All-Cause Hospitalization
Hazard Ratios for All-Cause Mortality and All-Cause Hospitalization Events for Chronic HF Subjects in the Clinical Validation Study
| | Hazard Ratio (95% CI, p value) | | |
| --- | --- | --- | --- |
| Galectin-3 Category | ≤ 17.8 ng/mL | 17.8-25.9 ng/mL | > 25.9 ng/mL |
| Number of Subjects | 647 | 170 | 78 |
| Galectin-3* | 1.0 | 1.35
(1.10-1.65, p= 0.004) | 1.46
(1.11-1.92, p= 0.006) |
*adjusted for baseline risk factors: age, gender, NYHA functional classification, left ventricular ejection fraction, diabetes status, and smoking status
Kaplan-Meier Curves for the Composite Endpoint of All-Cause Mortality or All-Cause Hospitalization, for chronic HF Subjects in the Clinical Validation Study, by Baseline Galectin-3 Level

Months From Baseline to Death/Hospitalization
{11}
Cumulative Probability of Event for the Composite Endpoint of All-Cause Mortality and All-Cause Hospitalization, at Various Time Points and By Baseline Galectin-3 Level, for chronic HF Subjects in the Clinical Validation Study

Cumulative Probability (with 95% Confidence Intervals) of Event for the Composite Endpoint of All-Cause Mortality and All-Cause Hospitalization, at Various Time Points and By Baseline Galectin-3 Level, for HF Subjects in the Clinical Validation Study
| | Cumulative Probability of All-Cause Mortality and All-Cause Hospitalization (95% CI) by Galectin-3 Category and Time Point (in percent) | | | |
| --- | --- | --- | --- | --- |
| Galectin-3 Category | 6 months | 12 months | 24 months | 36 months |
| ≤ 17.8 ng/mL | 24.2% (21.1%-27.7%) | 39.4 (35.7-43.3) | 58.9 (55.0-62.9) | 69.8 (65.8-73.7) |
| 17.8-25.9 ng/mL | 39.5 (32.5-47.3) | 58.1 (50.8-65.7) | 74.7 (67.7-81.1) | 81.7 (74.9-87.6) |
| > 25.9 ng/mL | 44.2 (33.9-55.9) | 55.8 (45.2-67.1) | 79.8 (69.9-88.2) | 82.1 (72.0-90.0) |
{12}
Hazard Ratios for Cardiovascular Mortality Events for chronic HF Subjects in the Clinical Validation Study
| | Hazard Ratio (95% CI, p value) | | |
| --- | --- | --- | --- |
| Galectin-3 Category | ≤ 17.8 ng/mL | 17.8-25.9 ng/mL | > 25.9 ng/mL |
| Number of Subjects | 647 | 170 | 78 |
| Galectin-3* | 1.0 | 1.91
(1.28-2.86, p= 0.002) | 2.33
(1.43-3.80, p< 0.001) |
*adjusted for baseline risk factors: age, gender, NYHA functional classification, left ventricular ejection fraction, diabetes status, and smoking status

Cumulative Probability of Event for the Endpoint of Cardiovascular Mortality, at Various Time Points and By Baseline Galectin-3 Level, for chronic HF Subjects in the Clinical Validation Study
{13}
Cumulative Probability (with 95% Confidence Intervals) of Event for the Cardiovascular Mortality, at Various Time Points and By Baseline Galectin-3 Level, for chronic HF Subjects in the Clinical Validation Study
| | Cumulative Probability of Cardiovascular Mortality (95% CI) by Galectin-3 Category and Time Point (in percent) | | | |
| --- | --- | --- | --- | --- |
| Galectin-3 Category | 6 months | 12 months | 24 months | 36 months |
| ≤ 17.8 ng/mL | 0.9% (0.4%-2.1%) | 2.4(1.4-3.9) | 7.0(5.2-9.3) | 12.0(9.4-15.2) |
| 17.8-25.9 ng/mL | 4.8(2.4-9.3) | 7.2(4.1-12.3) | 17.9(12.7-24.9) | 26.8(20.0-35.5) |
| > 25.9 ng/mL | 2.6(0.6-9.9) | 9.1(4.4-18.1) | 23.6(15.0-36.0) | 30.9(20.4-45.0) |
# Cardiovascular Mortality and Heart Failure Related Hospitalization
Hazard Ratios for Cardiovascular Mortality and Heart Failure-Related Hospitalization Events for chronic HF Subjects in the Clinical Validation Study
| | Hazard Ratio (95% CI, p value) | | |
| --- | --- | --- | --- |
| Galectin-3 Category | ≤ 17.8 ng/mL | 17.8-25.9 ng/mL | > 25.9 ng/mL |
| Number of Subjects | 647 | 170 | 78 |
| Galectin-3* | 1.0 | 1.51
(1.14-2.00, p= 0.004) | 1.70
(1.19-2.42, p= 0.004) |
*adjusted for baseline risk factors: age, gender, NYHA functional classification, left ventricular ejection fraction, diabetes status, and smoking status
{14}
Cumulative Probability of Event for the Composite Endpoint of Cardiovascular Mortality and Heart Failure-Related Hospitalization, at Various Time Points and By Baseline Galectin-3 Level, for chronic HF Subjects in the Clinical Validation Study

Cumulative Probability (with 95% Confidence Intervals) of Event for Cardiovascular Mortality and Heart Failure-Related Hospitalization, at Various Time Points and By Baseline Galectin-3 Level, for HF Subjects in the Clinical Validation Study
| | Cumulative Probability of Cardiovascular Mortality and Heart Failure-Related Hospitalization (95% CI) by Galectin-3 Category and Time Point (in percent) | | | |
| --- | --- | --- | --- | --- |
| Galectin-3 Category | 6 months | 12 months | 24 months | 36 months |
| ≤ 17.8 ng/mL | 7.2% (5.4%-9.5%) | 14.0 (11.5-17.0) | 23.8 (20.6-27.4) | 31.4 (27.6-35.6) |
| 17.8-25.9 ng/mL | 13.8 (9.4-20.1) | 25.3 (19.4-32.7) | 40.6 (33.4-48.7) | 48.5 (40.5-57.2) |
| > 25.9 ng/mL | 15.6 (9.2-25.8) | 22.2 (14.4-33.2) | 50.2 (38.9-62.8) | 54.8 (42.9-67.6) |
{15}
# All Cause Mortality
Kaplan-Meier Curves for the Endpoint of All-Cause Mortality, for chronic HF Subjects in the Clinical Validation Study, by Baseline Galectin-3 Level

Months From Baseline to Death
Cumulative Probability (with $95\%$ Confidence Intervals) of Event for the Endpoint of All-Cause Mortality, at Various Time Points and By Baseline Galectin-3 Level, for chronic HF Subjects in the Clinical Validation Study
| | Cumulative Probability of All-Cause Mortality Event (95% CI) by Galectin-3 Category and Time Point (in percent) | | | |
| --- | --- | --- | --- | --- |
| Galectin-3 Category | 6 months | 12 months | 24 months | 36 months |
| ≤ 17.8 ng/mL | 1.2% (0.6%-2.5%) | 3.3 (2.1-5.0) | 8.7 (6.7-11.3) | 15.3 (12.4-18.8) |
| 17.8-25.9 ng/mL | 5.3 (2.8-10.0) | 8.9 (5.5-14.4) | 22.0 (16.3-29.4) | 30.5 (23.4-39.1) |
| > 25.9 ng/mL | 2.6 (0.6-9.9) | 9.1 (4.4-18.1) | 26.6 (17.5-39.1) | 35.5 (24.5-49.5) |
{16}
Hazard Ratios for All-Cause Mortality Events for chronic HF Subjects in the Clinical Validation Study
| | Hazard Ratio (95% CI, p value) | | |
| --- | --- | --- | --- |
| Galectin-3 Category | ≤ 17.8 ng/mL | 17.8-25.9 ng/mL | > 25.9 ng/mL |
| Number of Subjects | 647 | 170 | 78 |
| Galectin-3* | 1.0 | 1.84
(1.28-2.64, p= 0.001) | 2.06
(1.31-3.23, p= 0.002) |
*adjusted for baseline risk factors: age, gender, NYHA functional classification, left ventricular ejection fraction, diabetes status, and smoking status.
4. Clinical cut-off:
To determine cutoff values for the BGM Galectin-3 assay in HF, galectin-3 levels were measured in 582 banked EDTA-plasma samples from an independent, controlled multi-center clinical study conducted in Europe enrolling NYHA class II, III and IV subjects. The average age of patients was 71 years, and 38% were female. The median follow-up time was greater than 18 months, and the primary end point was the composite of all-cause mortality and hospitalization for HF. Two cutoff values were derived that in turn defined three ordered galectin-3 categories. The derived galectin-3-dependent risk categories are as follows:
- galectin-3 greater than 25.9 ng/mL
- galectin-3 between 17.8 and 25.9 ng/mL
- galectin-3 less than or equal to 17.8 ng/mL
Galectin-3 and natriuretic peptides are measures of separate and distinct biological processes. Each marker provides independent and complementary information on the prognosis of patients with chronic heart failure.
The following additional interpretive information is provided in the labeling which provides information on the use of galectin-3 and NT-proBNP together.
Event Rates at 6, 12, 24 and 36 Months for the Composite Endpoint of All-Cause Mortality and All-Cause Hospitalization, by Galectin-3 Category and NT-proBNP level, for HF Subjects in the Clinical Validation Study:
{17}
18
| | Galectin-3 ≤ 17.8 ng/mL and NT-proBNP ≤ median | Galectin-3 ≤ 17.8 ng/mL and NT-proBNP > median
or
Galectin-3 > 17.8 ng/mL and NT-proBNP ≤ median | Galectin-3 > 17.8 ng/mL and NT-proBNP > median |
| --- | --- | --- | --- |
| Event rate at 6 months | 19.4% | 31.8% | 42.7% |
| Event rate at 12 months | 32.0% | 50.0% | 58.0% |
| Event rate at 24 months | 55.3% | 71.1% | 85.7% |
| Event rate at 36 months | 76.1% | 85.8% | 93.0% |
5. Expected values/Reference range:
Subjects included in the reference range study were drawn from a large ongoing prospective observational study recruiting subjects without previous cardiac disease. Galectin-3 levels were measured in 1,099 banked plasma samples (EDTA) from the population of apparently healthy subjects without known heart disease but that otherwise resemble, by age and gender distribution, the HF patient population. Specimens were from women between the ages of 60 and 80 years (n=575) and men between the ages of 55 and 80 (n=524). This reference population comprised individuals of different ethnic background, as follows: Black or African-American (n=307, 27.9%), Caucasian (n=691, 62.9%), Hispanic (n=42, 3.8%), Asian or Pacific Islander (n=30, 2.7%), and not specified (n=29, 2.6%). All subjects had detectable galectin-3 levels (min-max, 3.2 - 94.6 ng/mL) within the measuring range of BGM Galectin-3 (1.4 to 94.8 ng/mL). The table below summarizes the galectin-3 distribution results. The 97.5 percentile of the galectin-3 distribution from this reference population is 26.2 ng/mL. Samples with values above 26.2 ng/mL were in the upper 2.5th percentile.
| Percentile | Galectin-3 (ng/mL) |
| --- | --- |
| 2.5th | 5.4 |
| 5th | 6.3 |
| 25th | 9.7 |
| 50th | 12.4 |
| 75th | 15.6 |
| 90th | 19.0 |
| 95th | 22.1 |
| 97.5th | 26.2 |
{18}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.