K093736 · DePuy Orthopaedics, Inc. · LZO · Mar 10, 2011 · Orthopedic
Device Facts
Record ID
K093736
Device Name
DEPUY CORAIL HIP SYSTEM, REVISION STEM
Applicant
DePuy Orthopaedics, Inc.
Product Code
LZO · Orthopedic
Decision Date
Mar 10, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The subject Corail Revision Stem is intended for cementless use as part of the femoral component in a total hip arthroplasty.
Device Story
The Corail Revision Stem is a modular, monolithic titanium alloy femoral component for total hip arthroplasty. It features a grit-blasted surface plasma-sprayed with hydroxyapatite. The stem includes a 12/14 AMT (Articul/eze Mini Taper) trunnion, allowing compatibility with previously cleared DePuy femoral heads (metal or ceramic) and acetabular systems. Available in 9 lengths and multiple sizes, the device is implanted by orthopedic surgeons in a clinical/hospital setting. It replaces the femoral head/neck to restore joint function and alleviate pain in patients with degenerative or traumatic hip conditions. The device is intended for cementless fixation.
Clinical Evidence
No clinical testing was required. Substantial equivalence was demonstrated through non-clinical bench testing, including range of motion, proximal fatigue, and ceramic head compatibility testing, alongside characterization of the hydroxyapatite coating.
Technological Characteristics
Monolithic titanium alloy stem; grit-blasted surface; plasma-sprayed hydroxyapatite coating (155 ± 35 microns); 12/14 AMT (Articul/eze Mini Taper) trunnion; available in 9 lengths and sizes 10-20; cementless fixation.
Indications for Use
Indicated for cementless use in patients with severely painful/disabled joints due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia; avascular necrosis of the femoral head; acute traumatic fracture of the femoral head or neck; failed previous hip surgery (reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement, or total hip replacement); or ankylosis.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
# 长 093736
510 (k) Summary
:
:
MAR 1 0 2011
(As required by 21 CFR 807.92 and 21 CFR 807.93)
| NAME OF SPONSOR: | DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>Warsaw, Indiana 46582<br>Establishment Registration Number: 1818910 |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) CONTACT: | Rhonda Myer<br>Senior Regulatory Affairs Associate<br>Telephone: (574) 371-4927<br>Facsimile: (574) 371-4987<br>Electronic Mail: Rmyer7@its.jnj.com |
| DATE PREPARED: | March 1, 2011 |
| PROPRIETARY NAME: | DePuy Corail® Hip System, Revision Stem |
| COMMON NAME: | Hydroxyapatite-coated hip prosthesis |
| CLASSIFICATION: | Class III per 21 CFR 888.3330: Hip joint<br>metal/metal semi-constrained, with an<br>uncemented acetabular component, prosthesis<br>(KWA)<br>Class II per 21 CFR 888.3353: Hip joint<br>metal/ceramic/polymer semi-constrained<br>cemented or nonporous uncemented prosthesis<br>(LZO)<br>Class II per 21 CFR 888.3353: Hip joint<br>metal/ceramic/polymer semi-constrained<br>cemented or nonporous uncemented prosthesis<br>(MEH)<br>Class II per 21 CFR 888.3360: Hip joint<br>femoral (hemi-hip) metallic cemented or<br>uncemented prosthesis (LWJ) |
| DEVICE PRODUCT CODE AND<br>DESCRIPTION: | KWA: prosthesis, hip, semi-constrained (metal<br>uncemented acetabular component)<br>LZO: prosthesis, hip, semi-constrained,<br>metal/ceramic/polymer, cemented or non-porous,<br>uncemented<br>MEH: prosthesis, hip, semi-constrained,<br>uncemented, metal/polymer, non-porous,<br>calcium-phosphate<br>LWJ: prosthesis, hip, semi-constrained,<br>metal/polymer, uncemented |
リ
{1}------------------------------------------------
## SUBSTANTIALLY EQUIVALENT DEVICES:
DePuy Corail Hip Stem, K042992 DePuy KAR™ Hip Stem, K961732
## DEVICE DESCRIPTION:
The Corail Revision Stem is part of a modular prosthesis system for use in total hip replacement. It mates with DePuy femoral heads which articulate with DePuy acetabular systems with either a metal or UHMPE liner. The Corail Hip System, Revision Stem is a line addition to the Corail hip stem family. The stems are available in 9 lengths in sizes 10-20 in standard and high-off versions. The subject titanium alloy hip stems are monolithic, grit blasted stems plasma sprayed with hydroxyapatite powders on the entire stem length. The stems feature a 12/14 AMT (Articul/eze Mini Taper) Taper trunnion which are compatible with previously cleared metal and ceramic DePuy femoral heads with corresponding 12/14 tapers and offsets up to +13 mm.
## INDICATIONS AND INTENDED USE:
#### Indications:
The DePuy Corail Revision Stem is indicated for cementless use in the treatment of:
- I. A severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, rheumatoid arthritis or congenital hip dysplasia;
- 2. Avascular necrosis of the femoral head;
- 3. Acute traumatic fracture of the femoral head or neck;
- 4. Failed previous surgery, including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthoplasty, or other total hip replacement; and
- 5. Certain cases of ankylosis.
#### Intended Use:
The subject Corail Revision Stem is intended for cementless use as part of the femoral component in a total hip arthroplasty.
#### Summary of Technologies/Substantial Equivalence:
The substantial equivalence of the subject Corail Revision Stem is demonstrated by similarities in intended use, indications for use, materials, geometry, design and performance as compared to the predicate devices. The changes presented in this 510(k) do not present new issues of safety or effectiveness as the technological characteristics are basically the same as the predicates. Based on the materials, geometry, mechanical testing and indications for use, the DePuy Corail Revision Hip System is considered to be substantially equivalent to the predicate devices.
#### Non-clinical Testing:
{2}------------------------------------------------
Non-clinical testing and analysis were provided, including bench testing and coating characterization. Bench testing included Range of Motion, proximal fatigue and ceramic head compatibility testing. The nonporous, calcium phosphate coating was applied by plasma spraying to a nominal thickness of 155 ± 35 microns from hydroxyapatite powders. The calcium phosphate coating is applied by a different vendor and was characterized per FDA's "510(k) Information needed for Hydroxyapatite Coated Orthopedic Implants." All of the observed results indicate that the Corail Revision Hip System is substantially equivalent to devices currently marketed. Therefore, the subject device is as safe, as effective, and performs at least as safely and effectively as legally marketed predicates.
#### Clinical Testing:
No clinical testing was required to demonstrate substantial equivalence.
#### Conclusion:
The subject Corail Revision Stem is substantially equivalent to the predicate devices identified in this premarket notification.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three curved lines representing its wings or body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DePuy Orthopaedics. Inc. % Ms. Rhonda Myer Senior Regulatory Affairs Associate 700 Orthopaedic Drive Warsaw, Indiana 46581-0988
## MAR 1 0 2011
Re: K093736
Trade/Device Name: DePuy Corail Hip System, Revision Stem Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: Class III Product Code: KWA, LZO, MEH, LWJ Dated: February 17, 2011 Received: February 18, 2011
ቦ
Dear Ms. Myer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 -- Ms. Rhonda Myer
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
A.S. B. r2 h
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
093736 510 (k) Number (if known):
Device Name: DePuy Corail® Hip System, Revision Stem
### Indications for Use:
The DePuy Corail Revision Stem is indicated for cementless use in the treatment of:
- 1. A severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, or rheumatoid arthritis, or congenital hip dysplasia;
- 2. Avascular necrosis of the femoral head;
- 3. Acute traumatic fracture of the femoral head or neck;
- 4. Failed previous hip surgery, including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthoplasty, or other total hip replacement; and
- 5. Certain cases of ankylosis.
| Prescription Use | X |
|-----------------------------|--------|
| (Part 21 CFR 801 Subpart D) | AND/OR |
| Over-The-Counter Use | |
| (21 CFR 807 Subpart C) | |
(Please do not write below this line. Continue on another page if needed.)
(Division Sign Off) for M. Melkerson
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k093736
DePuy Orthopaedics, Inc.
Corail Hip System, Revision Stem Traditional 510(k)
Page 31 of 364
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.