MEDIGUARD NITRILE EXAMINATION GLOVE- BLUE (TESTED FOR USE WITH CHEMOTHERAPY DRUGS)

K093726 · Medline Industries, Inc. · LZA · Feb 22, 2010 · General Hospital

Device Facts

Record IDK093726
Device NameMEDIGUARD NITRILE EXAMINATION GLOVE- BLUE (TESTED FOR USE WITH CHEMOTHERAPY DRUGS)
ApplicantMedline Industries, Inc.
Product CodeLZA · General Hospital
Decision DateFeb 22, 2010
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The Mediguard Powder-Free Nitrile Examination Glove (tested for use with chemotherapy drugs) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Story

Disposable nitrile examination glove; worn by healthcare personnel to provide barrier protection against contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy agents. Output is physical barrier protection. Used in clinical settings. Benefit: reduces risk of cross-contamination and exposure to hazardous chemotherapy drugs during handling.

Clinical Evidence

Bench testing only. Permeation testing performed to determine breakthrough detection times (BDT) for 12 chemotherapy drugs. Results show BDT >240 minutes for most agents, with specific warnings provided for Carmustine (1.10 min) and Thiotepa (8.13 min).

Technological Characteristics

Material: Nitrile rubber. Form factor: Disposable examination glove. Powder-free. Non-sterile. Barrier protection principle.

Indications for Use

Indicated for use as a disposable patient examination glove to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs. Contraindicated for use with Carmustine and Thiotepa.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Matt Clausen Regulatory Affairs Specialist Medline Industries, Incorporated One Medline Place Mundelein, Illinois 60060 FEB 2 8 2010 Re: K093726 Trade/Device Name: Mediguard Powder Free Nitrile Examination Glove (Blue) -Tested for use with Chemotherapy Regulation Number: 21CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: 1 Product Code: LZA Dated: January 26, 2010 Received: January 29, 2010 Dear Mr. Clausen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2- Mr. Clausen Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" . (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Susan Tunney Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ ## Indications for Use ## 510(k) Number (if known): KO93726 Device Name: Mediguard Powder-Free Nitrile Examination Glove (Blue) - Tested for use with Chemotherapy Drugs Indications For Use: The Mediguard Powder-Free Nitrile Examination Glove (tested for use with chemotherapy drugs) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The tested chemotherapy drugs and their breakthrough detection times are as follows: | Chemotherapy Drug | Average BDT | |----------------------------|-------------| | 5-Fluorouracil | >240 min. | | Etoposide (Toposar) | >240 min. | | Cyclophosphamide (Cytoxan) | >240 min. | | Carmustine | 1.10 min. | | Thiotepa | 8.13 min. | | Paclitaxel (Taxol) | >240 min. | | Doxorubicin Hydrochloride | >240 min. | | Dacarbazine (DTIC) | >240 min. | | Cisplatin | >240 min. | | Ifosfamide (Ifex) | >240 min. | | Mitoxantrone | >240 min. | | Vincristine Sulfate | >240 min. | WARNING: Do not use with Carmystine and Thio-Te (Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K093726 Prescription Use (Part 21 CFR 801 Subpart D) .. OR Over-the-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 9
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...