← Product Code [LLZ](/productcode/LLZ) · K093703

# 3DI (K093703)

_Shina Systems , Ltd. · LLZ · Jan 19, 2010 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K093703

## Device Facts

- **Applicant:** Shina Systems , Ltd.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jan 19, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

3Di is a software package of PACS workstation for handling multimodality (CT, XA, MR, PET, SPECT & Ultrasound) images, which are using DICOM protocol. It includes volume rendering, Multi-planar reconstruction (MPR) and viewing of the inner and outer surfaces of organs as well as within their walls. 3Di is intended for use as an interactive tool for assisting professional Radiologists, Cardiologists and specialists to reach their own diagnosis, by providing tools of communication, clinics networking, WEB Serving, image viewing, image manipulation, 2D/3D image visualization, image processing, reporting and archiving. This product is not intended for use with or for diagnostic interpretation of Mammography images. The 3Di indications for use are processing of Cardiac CT angiography studies, including coronaries analysis, cardiac functional assessment and CT colonoscopy.

## Device Story

3Di is a PACS workstation software for multimodality medical image management. It ingests DICOM-compliant images from CT, XA, MR, PET, SPECT, and Ultrasound. The system provides tools for volume rendering, multi-planar reconstruction (MPR), and visualization of internal/external organ surfaces. Operated by radiologists, cardiologists, and specialists in clinical settings, it facilitates image viewing, manipulation, processing, reporting, and archiving. The software assists clinicians in reaching diagnoses by providing specialized clinical applications for cardiac CT angiography (including coronary analysis and functional assessment) and CT colonoscopy. It functions as a network-based tool for communication and clinical networking. The output is displayed to the clinician for diagnostic review and decision-making, potentially improving diagnostic efficiency and patient care through advanced visualization.

## Clinical Evidence

Bench testing only. Performance was validated by comparing 3Di to the Philips Brilliance predicate device. Evaluated parameters included general image reformatting functionality across modalities, reliability of orientation annotations, measurement accuracy, image quality, and results of cardiac analysis graphs and colon analysis.

## Technological Characteristics

Software-based PACS workstation. Supports DICOM protocol for multimodality imaging (CT, XA, MR, PET, SPECT, Ultrasound). Features include volume rendering, MPR, and 2D/3D visualization. Designed to meet ISO 14971 safety standards and 21 CFR 820.30 quality system regulations. Network-based architecture for image viewing, processing, and archiving.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- CardioCT ([K070226](/device/K070226.md))
- Philips Brilliance ([K090462](/device/K090462.md))
- Visage PACS/CS ([K082269](/device/K082269.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness (in accordance to 21 CFR 807,87(b))

Device Name Proprietary Device Name: 3Di

#### Establishment Name and Registration Number of Submitter

JAN 1 9 2010

Name: Shina Systems Ltd. Corresponding Official: Dan Laor Sireni 6, Haifa 32972, Israel TEL: 972-4-8246632

#### Device Classification

Product Code: CFR section: Panel Identification: Device Description: Classification:

LLZ 892.2050 Radiology Picture archiving and communications system II

## Reason for 510(k) Submission

Traditional 510(k) Submission

# Identification of Legally Marketed Predicate Devices

K070226 CardioCT K090462 Philips Brilliance K082269 Visage PACS/CS

#### Device Description

3Di is a PACS device which enables users to access medical images over a network and to utilize 3Di's image visualization tools to review the images. It provides the following functions: Web server, patient browser, PACS capabilities, multi-modality viewing, CT Cardiac and Colonoscopy clinical applications.

## Intended use and indications for Use

3Di is a software package of PACS workstation for handling multimodality (CT, XA, MR, PET, SPECT & Ultrasound) images, which are using DICOM protocol. It includes volume rendering, Multi-planar reconstruction (MPR) and viewing of the inner and outer surfaces of organs as well as within their walls. 3Di is intended for use as an interactive tool for assisting professional Radiologists, Cardiologists and specialists to reach their own diagnosis, by providing tools of communication, clinics networking, WEB Serving, image viewing, image manipulation, cora of image visualization, image processing, reporting and archiving. This product is not intended for use with or for diagnostic interpretation of Mammography images. The 3Di indications for use, are processing of Cardiac CT angiography studies, including coronaries analysis, cardiac functional assessment and CT colonoscopy

#### Safety & Effectiveness

The device has been designed, verified and validated complying to 21CFR 820.30 regulations. The device has been designed to meet the requirements of ISO 14971 Safety standard. Its performance has been validated by comparison to the performance of the Philips Brilliance predicate device. The following functions have been compared: General functionality of the images reformatting for the different imaging modalities, Reliability of orientation annotations displayed over the images, correctness of measurements, Image quality, Cardiac analysis Graphs and Results and Colon analysis results.

- Charter 6

age i of 2

{1}------------------------------------------------

The comparison demonstrates that the results of the two devices are very similar and substantial equivalent in terms of performance.

# Substantial Equivalency

It is Shina System opinion that the 3Di is substantially equivalent in terms of safety and effectiveness to the predicate devices.

#### 3Di submission · Chapter 6

page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is written in a serif font to the right of the eagle. The text is in all caps and is horizontally aligned.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

## JAN 1 9 2010

Shina Systems, Ltd. % Mr. Dan Laor Quality & Regulatory Advisor 6 Sireni, Haifa, 32972 ISRAEL

Re: K093703

Trade/Device Name: 3Di Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 24, 2009 Received: December 1, 2009

Dear Mr. Laor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indives for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, increase of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 2017 Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish fruther announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other madou cocs afth Act Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Donald J. Trump

Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K093703

Device Name: 3Di

Indications For Use:

3Di is a software package of PACS workstation for handling multimodality (CT, XA, MR, PET, SPECT & Ultrasound) images, which are using DICOM protocol. It includes volume rendering, Multi-planar reconstruction (MPR) and viewing of the inner and outer surfaces of organs as well as within their walls.

3Di is intended for use as an interactive tool for assisting professional Radiologists, Cardiologists and specialists to reach their own diagnosis, by providing tools of communication, clinics networking, WEB Serving, image viewing, image manipulation, 2D/3D image visualization, image processing, reporting and archiving. This product is not intended for use with or for diagnostic interpretation of Mammography images.

The 3Di indications for use are processing of Cardiac CT angiography studies, including coronaries analysis, cardiac functional assessment and CT colonoscopy.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR . . .

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dmall SHR

(Division Sien-Off) Division of Radiological Devices

Page 1 of 1

510(k) Number K093703

---

**Source:** [https://fda-staging.innolitics.com/device/K093703](https://fda-staging.innolitics.com/device/K093703)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).