K093694 · Geratherm Medical AG · DWJ · Sep 20, 2010 · Cardiovascular
Device Facts
Record ID
K093694
Device Name
GERATHERM UNIQUERESC
Applicant
Geratherm Medical AG
Product Code
DWJ · Cardiovascular
Decision Date
Sep 20, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Geratherm UniqueResc+ is a thermal regulating system for adult patients in order to prevent hypothermia in preclinical settings or treatments on-site. It is intended for use while at the rescue scene, during patient transport by air or ground to a medical facility or e.g. during disaster operation in first aid tents. The device should be used by appropriately trained healthcare professionals.
Device Story
Geratherm UniqueResc is a microprocessor-controlled, active warming blanket system for preventing/treating hypothermia in adult patients. Used in preclinical settings, rescue scenes, and during transport by trained healthcare professionals. System consists of a flexible, lightweight, conductive heating blanket with an integrated control unit. Power options include AC/DC adapter, 15VDC Li-Ion battery, or 12VDC-28VDC onboard power. Conductive material provides uniform heating across approximately 95% of the blanket surface. Benefits include noiseless operation, portability, and improved heating efficiency compared to previous models. Healthcare providers apply the blanket to the patient to maintain body temperature, aiding in clinical stabilization during transport or emergency care.
Clinical Evidence
Bench testing only. The device underwent successful testing for biocompatibility, electromagnetic compatibility, functionality, and safety according to international standards. No clinical data provided.
Technological Characteristics
Microprocessor-controlled active warming system. Conductive heating material. Integrated control unit. Power: 100-240VAC, 15VDC Li-Ion battery, or 12-28VDC onboard power. Flexible, lightweight form factor. Class II device (21 CFR 870.5900).
Indications for Use
Indicated for adult patients to prevent hypothermia in preclinical settings, on-site treatments, rescue scenes, patient transport (air/ground), and disaster operations. Contraindicated for children.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}------------------------------------------------
SEP 2 0 2010
K093694
":"}}
## 510(k) Summary as required by section 807.92(c)
9/17/2010
Submission Applicant: Geratherm Medical AG Thomas Robst Fahrenheitstraße 1 98716 Geschwenda / Germany Establishment Registration Number: 3003591690
Phone: +49 36205 98-0 Fax: +49 36205 98-116 E-mail: t.robst@geratherm.com
Application Correspondent/Contact: RG Medical Diagnostics Ron LeTourneau 28351 Beck Rd., Suite G5 Wixom, MI 48393 / USA
Phone: 888-596-9498 Fax: 248-504-6733 E-mail: rglt@rgmd.com
Trade Name: Geratherm UniqueResc*
Common Name: Warming systems / blankets
Classification Name: Thermal regulating systems (21 CFR 870.5900, Product code DWJ)
Substantial Equivalence Warming System: 510(k)-Number: K051097, GERATHERM SOS-RESCUE BLANKET, RG ENT., INC., 28351 Beck Rd., Wixom, MI 48393
#### Description of the Device:
Geratherm UniqueResc* has been developed for the prevention and treatment of hypothermia. The warming blankets are light-weight, comfortable and easily transportable.
1
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "Geratherm Medical Diagnostic Systems". The word "Geratherm" is in a bold, sans-serif font, with a registered trademark symbol to the right of the "m". Below the main title, the words "Medical Diagnostic Systems" are written in a smaller, sans-serif font.
The use of several blankets per patient is possible. The flexibility of the blankets makes it possible to warm the patient even during complicated rescue operating conditions.
Geratherm UniqueResc is suitable for use with adult patients of all sizes. The blankets should not be used for children.
Geratherm UniqueResc should be operated only by qualified medical professionals.
The benefits of Geratherm UniqueResc+ are:
- · Active warming applied to the patient from superior side.
- · State-of-the-art conductive material provides exceptionally uniform heating.
- · Microprocessor-controlled
- · Functions noiselessly
- · Simple and easy to use
- Highly durable and easy to clean
- · The optional Geratherm AC~DC Power Supply Adapter (see section 12.3 of the instruction manual for detailed description) functions with all alternating current mains power sources (100~240VAC, 50~60 Hz). Other optional power sources include 15VDC 2590 Li-Ion Battery and 12VDC-28VDC Onboard Power.
#### Indications for Use:
ﺮ
Geratherm UniqueResc+ is a thermal regulating system for adult patients in order to prevent hypothermia in preclinical settings or treatments on-site. It is intended for use while at the rescue scene, during patient transport by air or ground to a medical facility or e.g. during disaster operation in first aid tents. The device should be used by appropriately trained healthcare professionals.
### Comparison with Predicate Device:
The Geratherm UniqueReso is identical in intended use, indications for use, target population, where it is used, method of operation, safety characteristics (both devices met mainly the similar standards), biocompatibility, used material (whereby UniqueResc has improved material characteristics), sterilization instructions and several technical characteristics. However, the main difference between the new and old device is that UniqueResc is not a full body blanket. Geratherm has developed, regarding to effectiveness and handling, smaller blankets than the SOS-Rescue Blanket. The blanket's weight has also been reduced. Thereby, the blankets can be easily folded and require
{2}------------------------------------------------
# Geratherm®
considerably less storage space, e.g. in rescue vehicles. The control unit of UniqueResc' is integrated in the blanket. Furthermore, UniqueResc has more control safety features than the P.D. With the development of UniqueResc , appears a very efficient rescue blanket which surface can be heated to approximately 95%. Compared to the predicate device (only 1/3 of the surface is heated) it is an enormous improvement of the performance.
#### Summary:
The substantial equivalence of the UniqueResc* R130 and the referenced and approved earlier warming system, Gerathern SOS-Rescue Blanket (K051097), is based upon substantially similar methods for warming patients.
The predicate device provides warmth to patients through the transmission of heat from the surface of the device to the patient. The R130 and R150 similarly provide the transmission of heat from the surface of the unit to the patient.
The presented data that was conducted on the Geratherm UniqueResot shows in its results, and in comparison to the predicate device, that the product is safe and effective for its intended use. The product components which are covered by this 510(k) premarket notification have been successfully tested for biocompatibility, electromagnetic compatibility, functionality and safety according to international standards.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Geratherm Medical AG c/o Mr. Ronald G. LeTourneau · 28351 Beck Road, Suite G-5 Wixom, Michigan 48393
SEP 2 0 2010
Re: K093694
Trade/Device Name: Geratherm UniqueResc* Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal regulating system Regulatory Class: II Product Code: DWJ Dated: September 13, 2010 . Received: September 14, 2010
Dear Mr. LeTourneau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Ronald G. LeTourneau
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director . Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
7
Enclosure
{5}------------------------------------------------
### Indication for Use
510(k) Number (if known):
Device Name: Geratherm UniqueResc
Indications for Use:
Geratherm UniqueResc is a thermal regulating system for adult patients in order to prevent hypothermia in preclinical settings or treatments on-site. It is intended for use while at the rescue scene, during patient transport by air or ground to a medical facility or e.g. during disaster operation in first aid tents. The device should be used by appropriately trained healthcare professionals.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page I of 1
n Slan-Off) on of Cardiovaso r Devices 510(k) Number 109
Ko93694
SEP 2 0 2010
SEP 1 7 5.J'3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.