K093683 · Instrumentarium Dental, Palodex Group OY · MUH · Oct 6, 2010 · Dental
Device Facts
Record ID
K093683
Device Name
ORTHOPANTOMOGRAPH OP300
Applicant
Instrumentarium Dental, Palodex Group OY
Product Code
MUH · Dental
Decision Date
Oct 6, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
The Orthopantomograph® OP300 dental panoramic, cephalometric and cone beam computed tomography x-ray device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D x-ray images as well as xray projection images of an examined volume for the reconstruction of a 3D view. The device must only be operated and used by dentists and other legally qualified professionals.
Device Story
Orthopantomograph® OP300 is an extraoral dental X-ray system; captures 2D panoramic, cephalometric, and TMJ images; performs cone beam rotations to acquire projection images for 3D reconstruction. Operated by dentists/qualified professionals in clinical settings. Device transforms X-ray projections into 3D volumes via viewing software. Output viewed by clinicians to support diagnostic decision-making regarding teeth, jaw, and TMJ anatomy. Optional Volumetric Tomography device integration available. Benefits include enhanced visualization of dental structures for improved clinical assessment.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and bench testing comparisons to legally marketed predicate devices.
Technological Characteristics
Extraoral digital X-ray system; produces 2D panoramic, cephalometric, and 3D cone beam images. Employs pulsing and continuous power modes. Supports optional Volumetric Tomography. Connectivity via 3D viewing software. Intended for professional dental use.
Indications for Use
Indicated for dental radiographic examinations of teeth, jaw, and TMJ areas in patients requiring 2D panoramic, cephalometric, or 3D cone beam computed tomography imaging. Must be operated by dentists or legally qualified professionals.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
K093683.
Image /page/0/Picture/1 description: The image shows the word "INSTRUMENTARIUM" in a simple, sans-serif font. To the left of the word is a logo that appears to be a stylized "i" inside of a circle. The text and logo are black against a white background. The image is clear and easy to read.
# 510(k) Summary
OCT - 6 2010
Date:
April 29., 2010
## The name of the 510(k) owner:
Instrumentarium Dental, PaloDEx Group Oy Nahkelantie 160 04300 Tuusula, Finland
Tel: +358 10 270 2000 Fax: +358 9 851 4048
Contact person: Mr. Matti Tulikoura, Tel +358 400 609 507
## United States Sales Representative (U.S. Designated agent):
INSTRUMENTARIUM DENTAL INC. 1245 W. Canal Street, Milwaukee, Wisconsin 53233, USA
Tel: +1 414 747 1030, 800 558 6120 Fax: +1 414 481 8665
Contact Person: Mr. Frank Kashinski, Tel +1 414 747 6315
#### Trade name:
Orthopantomograph® OP300
#### Common name:
Dental panoramic, cephalometric and cone beam computed tomography x-ray device
### Classification name:
System, x-ray, extraoral source, digital (21 CFR 872.1800, product code MUH)
#### Predicate device:
Planmeca Promax 3D (K060328), SCANORA® 3D (K073350) and Veraviewepocs 3D (K073696).
### Description:
Orthopantomograph® OP300 is an extraoral source dental x-ray device, which produces conventional digital 2D panoramic, cephalometric and TMJ x-ray images as well as digital x-ray projection images taken during cone beam rotations around a patient's head. The projection images are reconstructed to be viewed in 3D by a 3D viewing software. The name Orthopantomograph® OP300 is used when the device can produce panoramic, TMJ and optionally 3D images. The name Orthoceph® OC300 is used when the cephalometric option for cephalometric x-ray images is added to the device. Furthermore, the device can be equipped with the Volumetric Tomography device (K063773).
#### Instrumentarium Dental
Nahkelantie 160, 04300 Tuusula Finland Street address: P.O. Box 20, F1-04301 Tuusula, Finland Postal adoress: +358 10 270 2000 Tel. +358 10 270 2230 Fax
Bank
SWIFT Nordea Bank NDEAFIHH Account: F19015963000046864
PaloDEx Group Oy. Tuusula, FINLAND Business ID 1981046-8 FI19810468 VAT
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "INSTRUMENTARIUM" in a simple, sans-serif font. To the left of the word is a logo that appears to be a stylized letter "i" inside of a circle. The logo is black and white, with the "i" being white and the rest of the circle being black. The text is all in uppercase letters.
Image /page/1/Picture/1 description: The image shows a black and white drawing of a diagonal line. The line is thick and appears to be drawn with a marker or pen. There is also a small dot and a short vertical line near the top right of the image.
## Intended use:
The Orthopantomograph® OP300 dental panoramic, cephalometric and cone beam computed tomography x-ray device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D x-ray images as well as xray projection images of an examined volume for the reconstruction of a 3D view. The device must only be operated and used by dentists and other legally qualified professionals.
# Technological characteristics:
The Orthopantomograph® OP300 is similar to the predicate devices Planmeca Promax 3D & Soredex Scanora 3D and otherwise similar to the Morita Veraviewepocs 3D predicate device by offering options for conventional cephalometric x-ray images and cone beam 3D images using pulsing and continuous power mode.
## Substantial Equivalence:
We consider Orthopantomograph® OP300 is as safe and effective as the prodicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by a series of curved lines.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
OCT 0 6 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Matti Tulikoura Regulatory Manager Instrumentarium Dental, PaloDEx Group OY Nahkelantie 160, Tuusula 04300 FINLAND
Re: K093683
Trade/Device Name: Orthopantomograph® OP300 .. Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system -------Regulatory Class: II Product Code: MUH Dated: September 3, 2010 Received: September 8, 2010
Dear Mr. Tulikoura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
signature
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "INSTRUMENTARIUM" in all capital letters. To the left of the word is a logo. The logo is a circle with a white swan-like figure inside of it, surrounded by a black background.
# Indications for Use
510(k) Number (if known): JK 093683
Device Name: Orthopantomograph® OP300
Indications for Use:
The Orthopantomograph® OP300 dental panoramic, cephalometric and cone beam computed tomography x-ray device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D x-ray images as well as x-ray projection images of an examined volume for the reconstruction of a 3D view. The device must only be operated and used by dentists and other legally qualified professionals.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Dovice Evaluation (ODE) () ( ) ( ) ( ) ( )
Mr. L. O'Hern for David C. Brown
(Division Sign-Off)
610K K093683
Instrumentarium Dental Street address: Nahkelantie 160, 04300 Tuusula Finland Postal address: P.O. Box 20, Fl-04301 Tuusula, Finfand +358 45 7882 2000 l el +358 45 7802 2506
Bank
SWIFT Nordea Bank NDEAFIHH
Account: F19015963000046864 PaloDEx Group Oy, Tuusula, FINLAND
Business ID 1981046-8 VAT F:19810468
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.