GAMBRO WATER PURIFICATION UNIT, MODELS WRO 300 AND WRO 300 H
Applicant
Gambro Renal Products, Inc.
Product Code
FIP · Gastroenterology, Urology
Decision Date
Jul 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
The Gambro WRO 300 Water Purification Unit is intended to be used as a dialysis accessory to produce water through reverse osmosis for one hemodialysis equipment. The WRO 300 can be connected to hemodialysis equipment used both in hospitals and in home environments, in conjunction with appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S.) standards. The Gambro WRO 300 H Water Purification Unit is intended to be used as a dialysis accessory to produce water through reverse osmosis for one hemodialysis equipment. The WRO 300 H can be connected to hemodialysis equipment used both in hospitals and in home environments, in conjunction with appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S.) standards.
Device Story
Water purification units for hemodialysis; utilize reverse osmosis (RO) to produce purified water for single dialysis machine. Input: feed water (pretreated via carbon/softener/particle filters). Operation: high-pressure pump forces water through polyamide thin-film composite RO membrane; dissolved salts, bacteria, pyrogens flushed as reject water. WRO 300 H includes heat disinfection; both support chemical disinfection. Used in hospitals or home environments; operated by clinicians or patients. Output: purified water for dialysis fluid preparation. Benefits: maintains low microbiological levels in flow path; ensures water quality for hemodialysis.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: RO membrane performance (rejection of salts, bacteria, endotoxins), output flow, water pressure, drain flow, electrical safety, EMC, and biocompatibility (ISO 10993-5, ANSI/AAMI RD62, ISO/FDIS 13959). Heat disinfection validation confirmed 5-log microbial reduction and maintenance of membrane integrity.
Indicated for use as a dialysis accessory to produce reverse osmosis water for one hemodialysis machine in hospital or home settings. Used in conjunction with pre/post-treatment units to meet AAMI/Federal water standards.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Gambro WRO 300 Water Purification System For Hemodialysis (K042797)
Submission Summary (Full Text)
{0}------------------------------------------------
K093608
PAGE 1 OF 6
Traditional 510(k) WRO 300 and WRO 300 H
JUL 2 9 2010
### 510(K) SUMMARY
| Submitter's Name | Gambro Renal Products, Inc. |
|-----------------------------------|------------------------------------------------------------------|
| Address | 14143 Denver West Parkway, Suite 400<br>Lakewood, Colorado 80401 |
| Establishment Registration Number | 2087532 |
| Contact Person | Kae Miller, Regulatory Affairs Manager |
| Telephone Number | 303.222.6724 |
| Fax Number | 303.222.6916 |
| Date of Summary | July 23, 2010 |
| Device | |
|----------------------|------------------------------------------------------------------------------|
| Name of the Device | WRO 300<br>Catalogue Number: 107364<br>WRO 300 H<br>Catalogue Number: 107365 |
| Common or Usual Name | Water Purification Unit |
| Classification Name | Subsystem, Water treatment |
| Device Class | II |
| Product Code | 78FIP |
| Regulation Number | 876.5665 |
:
:
.
Page 19 of 54 Revised July 23, 2010
Gambro CONFIDENTIAL
..
{1}------------------------------------------------
K093608
PAGE 2 OF 6
Traditional 510(k) WRO 300 and WRO 300 H
#### 510(k) SUMMARY, continued
| | Legally Marketed Device (Predicate Device) |
|---------------------|-----------------------------------------------------------|
| Name of the Device | Gambro WRO 300 Water Purification System For Hemodialysis |
| Catalogue Number | 107364 |
| Classification Name | Subsystem, Water treatment |
| Device Class | II |
| Product Code | 78FIP |
| Regulation Number | 876.5665 |
| 510(k) number | K042797 |
Image /page/1/Picture/5 description: The image shows two water softener systems. The systems are gray and have a digital display on the front. The systems are standing next to each other on a dark surface. The background is a light gray color.
| | WRO 300 | WRO 300 H |
|-----------------------|-----------------------------------------------------------|--------------------------------|
| Weight | 29 kg | 33 kg |
| Height | 563 mm | 563 mm |
| Depth | Max 520 mm<br>Footprint 380 mm | Max 520 mm<br>Footprint 380 mm |
| Width | Max 205 mm<br>Footprint 185 mm | Max 205 mm<br>Footprint 185 mm |
| Internal fluid volume | Approximately 3.5 liters excluding the product water loop | |
The Gambro WRO 300 and WRO 300 H Water Purification Units are dialysis accessories that produce water through reverse osmosis for one hemodialysis equipment.
They are both designed to maintain the low microbiological level in their flow path by the use of regular disinfection as a regular maintenance. Both WRO 300 and WRO 300 H Water Purification Units are designed with chemical disinfection capability. The WRO 300 H Water Purification Unit is also designed with heat disinfection capability.
WRO 300 or WRO 300 H Water Purification Unit is intended for use in conjunction with one dialysis machine, provided that the input flow and pressure demands correspond to the output of the WRO 300 or WRO 300 H Water Purification Unit.
WRO 300 and WRO 300 H Water Purification Units utilize the Reverse Osmosis Principle. A high pressure pump forces the water through the RO membrane. The product water is further distributed to the dialysis machine.
Page 20 of 54 Revised July 23, 2010
Gambro CONFIDENTIAL
{2}------------------------------------------------
K093608
PAGE 3 OF 6
Traditional 510(k) WRO 300 and WRO 300 H
Reverse osmosis (RO) is a membrane process that is the most widely used technique for purification of water for dialysis. When the feed water is in contact with the semipermeable membrane (the most vital part of the system) and a high pressure is applied, water will flow through the membrane to the product water side. Most of the other constituents (dissolved salts, particles, bacteria and pyrogens) will remain on the feed water side of the membrane and be flushed to drain as reject water. The membrane material used for the WRO 300 and WRO 300 H Water Purification Units are Polyamide Thin-Film Composite.
.
An acceptable quality of the feed water is required. The feed water is usually pretreated with such as active carbon filters, softener, and particle filters before it is supplied to the WRO 300 or WRO 300 H Water Purification Unit. Depending on the local water quality and regulations different pretreament equipment may be required.
Gambro WRO 300 and WRO 300 H Water Purification units contain the same software package (version P4.2) which is intended for controlling water production and disinfection / clanning of the flow path of the WRO.
The software also supports Heat Disinfection in the WRO 300 H Water Purification Unit.
#### INDICATIONS FOR USE WRO 300:
The Gambro WRO 300 Water Purification Unit is intended to be used as a dialysis accessory to produce water through reverse osmosis for one hemodialysis equipment.
The WRO 300 can be connected to hemodialysis equipment used both in hospitals and in home environments, in conjunction with appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S.) standards.
#### INDICATIONS FOR USE WRO 300 H:
The Gambro WRO 300 H Water Purification Unit is intended to be used as a dialysis accessory to produce water through reverse osmosis for one hemodialysis equipment.
The WRO 300 H can be connected to hemodialysis equipment used both in hospitals and in home environments, in conjunction with appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S.) standards.
Page 21 of 54 Revised July 29, 2010
Gambro CONFIDENTIAL
{3}------------------------------------------------
Traditional 510(k)
# 510(k) SUMMARY, contin
DEVICE COMPARISON TABLE
in the following table the WRO 300 H are compared with the predicate device WRO 300 (K0427)
| | PREDICATE<br>WRO 300 (K042797)<br>SW1.0 | WRO 300<br>SW P4.2 | WRO 300 H<br>SW P4.2 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Gambro WRO 300 Water Purification<br>System is intended to be used as a dialysis<br>accessory device in conjunction with one<br>dialysis machine to produce water used to<br>prepare and dilute dialysis concentrate to<br>form dialysis fluid by using the reverse<br>osmosis concept. | The Gambro WRO 300 Water<br>Purification Unit is intended to be used<br>as a dialysis accessory to produce water<br>through reverse osmosis for one<br>hemodialysis equipment.<br><br>The WRO 300 can be connected to<br>hemodialysis equipment used both in<br>hospitals and in home environments, in<br>conjunction with appropriate pre and<br>post treatment units, as a part of a water<br>treatment system designed to meet<br>current AAMI and Federal (U.S.)<br>standards. | The Gambro WRO 300 H Water Purification<br>Unit is intended to be used as a dialysis<br>accessory to produce water through reverse<br>osmosis for one hemodialysis equipment.<br><br>The WRO 300 H can be connected to<br>hemodialysis equipment used both in hospitals<br>and in home environments, in conjunction with<br>appropriate pre and post treatment units, as a<br>part of a water treatment system designed to<br>meet current AAMI and Federal (U.S.)<br>standards. |
| Feed Water<br>Supply | <i>Input:</i> 3.0 l/min required<br><i>Pressure:</i> 0.2 to 0.8 MPa<br><i>Temperature:</i> +5 to +30°C<br><i>Hardness:</i> recommendation < 1 °dH<br>(20 ppm as CaCO3). | <i>Input:</i> Min. 3.0 l/min required<br><i>Pressure:</i> 0.15 to 0.8 MPa during<br>operation<br><i>Temperature:</i> +5 to +30°C<br><i>Hardness:</i> recommendation < 1 °dH | <i>Input:</i> Min. 3.0 l/min required<br><i>Pressure:</i> 0.15 to 0.8 MPa during<br>operation<br><i>Temperature:</i> +5 to +30°C<br><i>Hardness:</i> recommendation < 0.3 °dH |
Page 22 of 54
Revised July 23, 201
mbro CONFIDENTL
## K093608
PAGE 4 OF 6
ਟ
{4}------------------------------------------------
| Gambro Renal Products, Inc. |
|--------------------------------------|
| 14143 Denver West Parkway, Suite 400 |
| Lakewood, Colorado 80401 |
Traditional 510(k)
WRO 300 and WRO 300 H
| | PREDICATE<br>WRO 300 (K042797) | WRO 300 | WRO 300 H |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| | SW1.0 | SW P4.2 | SW P4.2 |
| | Output flow: | | |
| | Minimum 1.1 l/min at 10°C.<br>Product water loop: Maximum 20 meters (2<br>x 10m). Designed for flexible, rein-forced<br>tubing 8 mm x 2.5 mm. | Output flow:<br>Minimum 1.1 l/min at 10°C.<br>Product water loop: Maximum 20<br>meters (2 x 10m). Designed for<br>flexible, rein-forced tubing. | Output flow:<br>Minimum 1.2 l/min at 10°C.<br>Product water loop: Maximum 2 meters<br>(2 x 1m). Designed for flexible, rein-<br>forced tubing. |
| Product water | Product water pressure: 0.12 to 0.6 MPa<br>during RUN mode. | Product water pressure: 0.12 to 0.6<br>MPa during RUN mode. | Product water pressure: 0.12 to 0.6 MPa during<br>RUN mode. |
| | Rejection rates: | Rejection rates: | Rejection rates: |
| | Total dissolved salts: >95%.<br>Bacteria and endotoxin: >99%. | Total dissolved salts: >96%.<br>Bacteria and endotoxin: >99%. | Total dissolved salts: >96%.<br>Bacteria and endotoxin: >99%. |
| Drain<br>requirements | Operation: 1.2 ±0.1 l/min.<br>Peak flow (rinse): 3.0 l/min | Operation: 1.2 ±0.1 l/min.<br>Peak flow (rinse): 3.0 l/min | Operation: 1.2 ±0.1 l/min.<br>Peak flow (rinse): 3.0 l/min |
| Reverse osmosis<br>membrane | Material:<br>Polyamide, thin film composite.<br>Configuration: Spiral wound.<br>pH-tolerance:<br>2-11 | Material:<br>Polyamide, thin film composite.<br>Configuration: Spiral wound.<br>pH-tolerance:<br>2-11 | Material:<br>Polyamide, thin film composite.<br>Configuration: Spiral wound.<br>pH-tolerance:<br>2-11 |
| Disinfection and<br>cleaning | Chemical disinfection.<br>Cleaning. | Chemical disinfection.<br>Cleaning. | Chemical disinfection.<br>Cleaning.<br>Heat disinfection. |
| Ambient | Temperature:<br>+ 10 to + 40 °C | Temperature:<br>+ 10 to + 40 °C | Temperature:<br>+ 10 to + 40 °C |
mbro CONFIDENTIA
Page 23 of 54
Revised July 23, 2016
K093608
PAGE 5 OF 6
こ
{5}------------------------------------------------
WRO 300 and WRO 300 H
#### 510(k) SUMMARY, continued
#### Assessment of non-clinical performance data
The non-clinical testing applied to the Gambro WRO 300 H Water Purification Units, equipped with the software version P4.2, in order to determine the substantial equivalence with the predicate device, included:
- verification and validation of the software changes, with static and dynamic software testing, and . regression testing,
- verification and validation of the functional, performance and safety requirements related to ●
- O the membrane life expectancy;
- user interface/control system o
- disinfection both chemical and heat 0
- physical ergonomics O
- maintenance and reliability ಂ
- compliance with international standard on electrical safety (IEC 60601-1),
- compliance with international standard on electromagnetic compatibility (IEC 60601-1-2).
The assessment was performed by internal and external independent personnel with the appropriate skills.
#### Assessment of clinical performance data
Not applicable for this submission.
#### Conclusion
Testing results of the R.O. Membrane (rejection of total dissolved salts, bacteria and endotoxins), output flow, water pressure, drain flow, electrical safety (in accordance to IEC 60601-1), electromagnetic compatibility (in accordance to IEC 60601-1-2), biocompatibility (in accordance with ISO 10993-5, ANSI/AAMI RD62, ISO/FDIS 13959, European Pharmacopoeia 5th ed.), and software validation demonstrate that the WRO 300 and WRO 300H with software version 4.2 meets all performance specifications equivalent to the predicate device.
In addition, testing conducted on the WRO 300H confirmed that heat disinfection did not affect safety or membrane performance, and performed equivalent to predicate device disinfection capability for the purpose of ensuring low microbiological level in the flow path. In addition to standards tests applicable for reverse osmosis, testing specific to the heat disinfection included:
- · Biocompability test (in accordance with ISO 10993-5, ANSI/AAMI RD62, ISO/FDIS 13959, European Pharmacopoeia 5th ed.) of extract from the hydraulic flow path after RO membrane
- . Validation of heat disinfection procedure in regards to temperature in flow path during Heat disinfection (coolest spot in flow path at least 80° C, with 5 log reduction for all positions)
- Evaluation of Microbial retention by the RO membrane
. Page 24 of 54 Revised July 23, 2010
#### Gambro CONFIDENTIAL
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN" written around the left side of the circle. Inside the circle is a stylized image of a person with their arms outstretched.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G60 Silver Spring, MD 20993-0002
Ms. Kae Miller Regulatory Affairs Manager Gambro Renal Products, Inc. 14143 Denver West Parkway, Suite 400 LAKEWOOD CO 80401
11111. 2 5 2010
Re: K093608 Trade/Device Name: WRO 300 and WRO 300 H Regulation Number: 21 CFR §876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: FIP Dated: July 26, 2010 Received: July 27, 2010
Dear Ms. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{7}------------------------------------------------
Page 2 -
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{8}------------------------------------------------
K093605
Traditional 510(k) WRO 300 and WRO 300 H
Indications for Use
510(k) Number (if known):
JUL 2 9 2010
Device Name: WRO 300
#### Indications for Use WRO 300:
The Gambro WRO 300 Water Purification Unit is intended to be used as a dialysis accessory to produce water through reverse osmosis for one hemodialysis equipment.
WRO 300 can be connected to hemodialysis equipment used both in home environments, in conjunction with appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S.) standards.
#### Device Name: WRO 300 H
#### Indications for Use WRO 300 H:
The Gambro WRO 300 H Water Purification Unit is intended to be used as a dialysis accessory to produce water through reverse osmosis for one hemodialysis equipment.
WRO 300 H can be connected to hemodialysis equipment used both in hospitals and in home environments, in conjunction with appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S.) standards.
Halul Leen
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devi 510(k) Number.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 18 of 54
#### Gambro CONFIDENTIAL
Attachment 1 - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.