Reduction or even elimination of pain in exposed cervical areas that do not require restoration. Alleviation or prevention of dentinal sensitivity after preparation of teeth to receive fixed prosthesis or restorations.
Device Story
Gluma Desensitizer PowerGel is a dental desensitizing agent applied by clinicians to hypersensitive dentine. The product is an aqueous, viscous gel containing glutaraldehyde, HEMA, colloidal silica, and pigments. It functions by precipitating plasma proteins, which reduces dentinal permeability and occludes peripheral dentinal tubules, thereby inhibiting fluid movement that causes sensitivity. The gel-like consistency prevents uncontrolled flow, reducing soft tissue irritation, while the green opaque color provides visual contrast for precise application. It is supplied in 1.8 mL syringes. The device is used in dental clinics to treat exposed cervical areas or teeth prepared for restorations. By occluding tubules, it provides pain relief and prevents sensitivity, improving patient comfort during and after dental procedures.
Clinical Evidence
No clinical trials were conducted; evidence is based on a clinical evaluation report per MEDDEV 2.7.1 and non-clinical bench testing. Biocompatibility was verified per EN ISO 10993-1. Physical properties (viscosity, composition) were verified against internal specifications. The benefit/risk ratio is judged positive based on the established safety profile of the predicate device and the performance of the active ingredients (glutaraldehyde/HEMA).
Technological Characteristics
Aqueous, viscous gel containing glutaraldehyde, HEMA, colloidal silica, and pigments. Viscosity: 1000-3000 mPas. Active substance content: 4.0-6.0%. Form factor: 1.8 mL syringe. Biocompatibility: ISO 10993-1 compliant. No electronic components or software.
Indications for Use
Indicated for patients with hypersensitive dentine, specifically for pain reduction in exposed cervical areas not requiring restoration and for alleviating/preventing dentinal sensitivity following tooth preparation for fixed prostheses or restorations. Contraindicated in cases of pulpitis, perforated pulp or dentine, or known/suspected allergy to the product.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
K093575
#### Technical File
MAR - 5 2010
Summary of safety and effectiveness of the desensitizing agent Gluma® Desensitizer PowerGei
#### 1. Submitter name
Heraeus Kulzer, LLC 300 Heraeus Way South Bend, Indiana 46614
Contact person: Cheryl V. Zimmerman Tel. +1 (574) 299 - 5444
Date summary prepared: March, 02, 2010
- 2. Name of the device:
Gluma Desensitizer PowerGel The product is Regulatory Class II and the product code KLE It is used as Desensitizer.
The GMDN number is 34782 and the description is Desensitizer, Dental bonding agent, polymer based, the UMDS number is 34782 and the description is Klebemittel.
### 3. Substantially equivalence
Gluma Desensitizer PowerGel (Project Name IGLU) is a revised version of the product Gluma Desensitizer (K962812). The main components and their ratio in IGLU and Gluma Desensitizer are similar. IGLU contains additionally filler particles and pigments to realize a green opaque coloured gel or paste. These changes will lead to an easier and safer handling of Gluma Desensitizer PowerGel when compared to Gluma Desensitizer.
#### 4. Description of the device
GLUMA Desensitizer PowerGel is used for treatment of hypersensitive dentine. The product should be applied only on dentin with a sufficient remaining pulpal wall. It achieves its effects by precipitation of plasma proteins, which reduces dentinal permeability and occludes the peripheral dentinal tubules. This inhibits
{1}------------------------------------------------
Seite 2
the flow of fluid through the tubules which is the cause of sensitivity. Gluma® Desensitizer PowerGel is a product for a simple and safe application. It's viscous gel like consistency leads to an easy and save application. Uncontrolled flowing from the tooth is avoided and subsequent irritation of soft tissue should be reduced. The opaque, green colour has a good visible contrast to the tooth and the mucosa so that the operator could easily see where he has applied the product.
The physical properties of Gluma® Desensitizer PowerGel especially the consistency were determined in accordance with internal specifications. The results have shown good and sufficient properties of Gluma® Desensitizer PowerGel according to the specification.
The product will be put in the market in 1,8 ml Syringe with a content of 1 g product.
The specification of the product is as followed:
Appearance Composition tested with GC Green, homogenous Gel, without impurities active substance 4,0 - 6,0 % Cross linking agent according to
reference Viscosity
1000 - 3000 mPas
### 5. Intended use
Gluma® Desensitizer PowerGel is an aqueous, viscous gel of Glutaraldehyd, HEMA, colloidal silica and pigments. It achieves its effect by precipitation of plasma proteins, which reduces the dentinal permeability and occludes the peripheral dentinal tubules. This occlution reduces the liquid movement within the tubules. This liquid movement is the cause of sensitivity.
The intended use of the product is as follows:
- > Reduction or even elimination of pain In exposed cervical areas that do not require restoration
- > Allevation or prevention of dentinal sensitivity after preparation of teeth to receive fixed prosthesis or restorations
Contra indication
Gluma Desensitizer PowerGel must not be used
{2}------------------------------------------------
Selte 3
- > If the necessary precautions cannot be taken or the stipulated application technique is not possible
- > In cases of known or suspected allergy to Gluma Desensitizer Powergel. If an allergy is suspected it is recommended to carry out an allergy test prior to treatment
- > If the pain is due to pulpitis
- > If the pulp is perforated or the dentine adjacent to the pulp is perforated.
### 6: Summaries
- a: See point 3
b: (1) Nonclinical- tests: In accordance with the Medical Device Directive 93/42/EWG and national European medical device legislation, any medical device is requested to be evaluated by the legal medical device manufacturer regarding its clinical performance and safety. This includes an evaluation of biocompatibility in accordance with EN ISO 10993-1.
The biological compatibility of Gluma® Desensitizer PowerGel was verified in accordance with the international standard.
The biocompatibility of Gluma® Desensitizer PowerGel in the aforementioned indication was documented in a biocompatibility evaluation report and the benefit/risk-relation has been judged as positive.
(2) Clinical Evaluation
In accordance with the medical Device directive 93/42/EWG and national European medical device legislation, any medical device is requested to be evaluated by the legal medical device manufacturer regarding its clinical performance and safety. This includes a clinical evaluation in accordance with MEDDEV 2.7.1., which is intended to critically evaluate the clinical benefits of the medical device in comparison to its potential risks. Therefore, any clinical evaluation is part of the compulsory risk management process according to EN ISO 14971, and critical findings must further be considered in the current risk management process of the medical device manufacturer responsible for the evaluated device.
On this background, the clinical evaluation was performed in order to comply with the current European medical device legislation, in particular with MEDDEV 2.7.1. This critical evaluation followed the procedures outlined in the corresponding clinical evaluation plan.
Gluma® Desensitizer PowerGel is a Desensitizer, which is generally classified as a class IIa medical device under the Medical Device Directive 93/42/EEC.
{3}------------------------------------------------
Seile 4
The material is used for the treatment of hypersensitive dentine, intended for long-term application.
Considering the evaluated scientific data and technical results for Gluma® Desensitizer PowerGel it is concluded that the product can be expected to exhibit the claimed technical performance and that potential undesirable clinical effects and risks seem well controlled and acceptable, when weighed against their benefits in dentistry. Therefore, a positive benefit versus risk ratio can be stated by the expert for Gluma® Desensitizer PowerGel, provided that the product applied in accordance with its intended use as outlined in the manufacturer's instruction for use.
Considering the evaluated scientific data for dental desensitizers in general as well as for comparator products, it is concluded, that Gluma® Desensitizer PowerGel will exhibit the claimed clinical performance when applied according to the intended use as follows:
Gluma® Desensitizer PowerGel is used for treatment of hypersensitive dentine. The product should be applied only on dentin with a sufficient remaining pulpal wall.
Due to the ingredients of Gluma® Desensitizer PowerGel, such as hydroxyethylmethacrylate and glutardialdehyde, warnings on a likely dermal, eye and respiratory irritating potential and on a likely dermal and respiratory sensitizing potential have to be included in the respective instructions for use.
Nevertheless, the potential undesirable clinical effects and risks seem well controlled and acceptable, when weighed against their benefits in dentistry. Therefore, a positive benefit versus risk ratio can be stated for Gluma® Desensitizer PowerGel, provided that the product is applied as outlined in the manufacturer's information for use.
The clinical evaluation report was prepared in accordance with MEDDEV 2.7.1 and followed the provisions of the corresponding clinical evaluation plan.
(3) Conclusion
A positive benefit versus risk ratio can be stated by the experts for Gluma® Desensitizer PowerGel, provided that the product applied in accordance with its intended use as outlined in the manufacturer's instruction for use for the clinical and the non-clinical test results.
{4}------------------------------------------------
### Heracus
Seite 5
### (c) 510 (k) summary
The risk potential of the Desensitizer Gluma® Desensitizer PowerGel was proved considering the current composition. All properties of the product were verified successfully.
The biological compatibility of the desensitizing material was investigated to evaluate the toxicological risk. A toxicological evaluation report has confirmed that the product Gluma® Desensitizer PowerGel meets the requirements of the DIN EN ISO 10993 standard. The results were discussed in a Biocompatibility Evaluation Report and the benefit/risk-relation has been judged as positive.
The physical properties of Gluma® Desensitizer PowerGel were determined in accordance with internal standards. The results have shown good properties of Gluma® Desensitizer to the specification.
Based on the results of the clinical evaluation report it is concluded that the product can be expected to exhibit the claimed technical performance and that potential undesirable clinical effects and risks seem well controlled and acceptable when weighted against their benefits in dentistry.
The risk analysis (according to DIN EN ISO 14971) was carried out for Gluma® Desensitizer PowerGel and showed that the application of Gluma® Desensitizer PowerGel could be considered to be safe.
Gluma® Desensitizer PowerGel meets all requirements relevant for dental desensitizing material in accordance with the Medical Device directive 93/42/EEC and national European medical device legislation. Based on the actual facts Gluma® Desensitizer PowerGel could be evaluated to be effective and safe with its intended use as outlined in the manufacturer's instruction for use.
Dormagen, 02.03.2010
i.A.
Dr. Marcus Hoffmann
i.A. Wagner
Annegrete Wegne
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of a stylized human figure, often referred to as the "Human Services Symbol," which is meant to represent care and assistance.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
MAR - 5 2010
Ms. Cheryl V. Zimmerman Manager, Quality Operations and Compliance Heraeus Kulzer, Incorporated 300 Heraeus Way South Bend, Indiana 46614
Re: K093575
Trade/Device Name: Gluma Desensitizer PowderGel Regulation Number: 21CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: II Product Code: LBH Dated: February 4, 2010 Received: February 12, 2010
Dear Ms. Zimmerman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2- Ms. Zimmerman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
Wh foc
Anthony D. Watson, B.S., M.S., M.B.A., Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use
510(k) Number (if known): ___K093575_
Gluma Desensitizer PowerGel Device Name: ___
Indications for use:
- Reduction or even elimination of pain in exposed cervical areas that do not . require restoration.
- Alleviation or reduction of dentine sensitivity after preparation of teeth to receive . fixed prosthesis or restorations.
Over-The-Counter Use Prescription Use × AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
RSBett MAS for Dr K.P. Mulhy
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093575
Page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.