K093563 · Philips Ultrasound, Inc. · IYN · Feb 1, 2010 · Radiology
Device Facts
Record ID
K093563
Device Name
IU22 ULTRASOUND SYSTEM AND TRANSDUCERS
Applicant
Philips Ultrasound, Inc.
Product Code
IYN · Radiology
Decision Date
Feb 1, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The iU22 Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (or 2-D), M-mode (including Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Tissue Doppler Imaging, Harmonics (Tissue and Contrast) and Elastography modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Ophthalmic Fetal Intra-operative Abdominal Laparoscopic Pediatric Small Organ Adult and Neonatal Cephalic Trans-rectal Trans-vaginal Musculoskeletal Urology Cardiac (Adult, Pediatric, Trans-esophageal) Fetal Echo Peripheral Vessel. The clinical environments where the iU22 3.7 system can be used include point-of-care areas in offices, clinical and hospital settings for diagnosis of patients.
Device Story
The iU22 is a diagnostic ultrasound system that uses a piezoelectric transducer array to convert electronic voltage pulses into ultrasonic pressure waves. These waves propagate through body tissues, and the system processes the resulting echoes to generate images and analyze fluid flow (Doppler). The system supports multiple imaging modes, including B-mode, M-mode, Doppler (Pulse Wave, Continuous Wave, Color, Tissue), Harmonics, and Elastography. It provides measurement and analysis packages to assist clinicians in diagnosing anatomical structures and blood flow. Used in offices, clinics, and hospitals, the device is operated by healthcare professionals. Output is displayed on the system monitor for clinical decision-making. The system is designed to provide diagnostic information across a wide range of clinical applications, including cardiac, fetal, and musculoskeletal imaging.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, including verification of acoustic output levels, electrical/physical safety, and material biosafety, confirming performance equivalent to previously cleared predicate devices.
Technological Characteristics
System utilizes piezoelectric transducer arrays for ultrasonic wave generation. Supports B-mode, M-mode, Doppler, Harmonics, and Elastography. Connectivity includes standard ultrasound imaging interfaces. Manufactured under established quality systems with materials evaluated for biosafety. Operates within Track 3 FDA acoustic output limits. Software includes analysis packages for anatomical measurement and fluid flow quantification.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in ophthalmic, fetal, abdominal, intra-operative, laparoscopic, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal, urological, cardiac, and peripheral vascular applications. Used in clinical, hospital, and point-of-care settings by healthcare professionals.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
510(k) Premarket Notification
### FEB - 1 2010 510(k) Summary of Safety and Effectiveness iU22 Diagnostic Ultrasound System
This summary of safety and effectiveness information is submitted in accordance with 21CFR §807.92
### 1. Submitter's name, address, telephone number, contact person. Philips Ultrasound, Inc. 22100 Bothell Everett Hwy Bothell, WA 98021-8431
Contact person: Nancy Burke, Regulatory Affairs Specialist Email: Nancy.Burke@philips.com Tel: (425) 487-7371 Fax: (425) 487-8666
Date prepared: November 3, 2009
## 2. Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Common/usual name: Diagnostic ultrasound system and transducers Proprietary name: iU22 Ultrasound System
These devices are classified as follows:
| Classification Name | 21 CFR Section | Product Code |
|------------------------------------------|----------------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90ITX |
As stated in 21 CFR, parts 892.1550, 892.1560 and 892.1570, each of these generic types of devices have been classified as Class II.
### 3. Substantially Equivalent Devices
Philips Ultrasound believes that the iU22 Ultrasound System is substantially equivalent to the following currently marketed devices:
| Product | 510(k) |
|----------------------|------------------|
| Philips iU22 | K042540 |
| Siemens Acuson S2000 | K072786 |
| Philips QLAB | K021966, K023877 |
{1}------------------------------------------------
### 4. Device Description and Technical Comparison to Predicate Devices
The subject of this 510(k) notification, the iU22 ultrasound system and transducer(s), function in a manner identical to all diagnostic ultrasound systems and transducers. The system circuitry generates an electronic voltage pulse. which is transmitted to the transducer. In the transducer, a piezo electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The Doppler functions of system process the Doppler shift frequencies from the echoes of moving targets such as blood to detect and graphically display the Doppler shifts of these tissues as flow.
The iU22 system gives the operator the ability to measure anatomical structures and offers analysis packages that provide information used by competent healthcare professionals to make a diagnosis.
#### 5. Intended Use
The iU22 Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (or 2-D), M-mode (including Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Tissue Doppler Imaging, Harmonics (Tissue and Contrast) and Elastography modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:
> Ophthalmic Fetal Intra-operative Abdominal Laparoscopic Pediatric Small Organ Adult and Neonatal Cephalic Trans-rectal Trans-vaginal Musculoskeletal Urology Cardiac (Adult, Pediatric, Trans-esophageal) Fetal Echo Peripheral Vessel
The clinical environments where the iU22 3.7 system can be used include point-ofcare areas in offices, clinical and hospital settings for diagnosis of patients.
#### 6. Conclusion
iU22 Ultrasound System and transducers is substantially equivalent in safety and effectiveness to the predicate devices identified above:
{2}------------------------------------------------
- The systems are indicated for the diagnostic ultrasonic imaging and fluid flow ♥ analysis.
- The systems have the same gray-scale and Doppler capabilities. .
- The systems have acoustic output levels below the Track 3 FDA limits. .
- The systems are manufactured under equivalent quality systems. .
- The systems are manufactured of materials with equivalent biosafety. The . materials have been evaluated and found to be safe for this application.
- The systems are designed and manufactured to the same electrical and physical . safety standards.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized depiction of three birds in flight, arranged in a vertical stack. The birds are silhouetted and appear to be soaring upwards.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
# FEB - 1 2010
Philips Ultrasound, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K093563
Trade/Device Name: iU22 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: January 12, 2010 Received: January 13, 2010
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the iU22 Diagnostic Ultrasound System, as described in your premarket notification:
#### Transducer Model Number
### L12-5
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/RevortaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely yours
signature
Donald St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
Philips Ultrasound, Inc. 510(k) Premarket Notification
510(k) Number (if known):
Device Name: iU22 Diagnostic Ultrasound System
Indications for Use:
The iU22 Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (or 2-D), M-mode (including Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Tissue Doppler Imaging, Harmonics (Tissue and Contrast) and Elastography modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:
> Ophthalmic Fetal Intra-operative Abdominal Laparoscopic Pediatric Small Organ Adult and Neonatal Cephalic Trans-rectal Trans-vaginal Musculoskeletal Urology Cardiac (Adult, Pediatric, Trans-esophageal) Fetal Echo Peripheral Vessel
The clinical environments where the iU22 3.7 system can be used include point-of-care areas in offices, clinical and hospital settings for diagnosis of patients.
Prescription Use XX (Part 21 CFR 801 Subpart D) OR
Over-the-Counter Use (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Amel J. R.
(Division Sign Off) Division of Radiological Devices
| 510(k) Number | K093563 |
|---------------|---------|
|---------------|---------|
Indications for Use Statement
{6}------------------------------------------------
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No: System: Intended Use:
iU22 Ultrasound System
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|----------------------------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|---------------------|---------------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | P | P | P | | P | Notes 1,3 | Notes, 5,6,8,10,12,13 |
| Fetal Imaging<br>& Other | Fetal (includes Echo) | P | P | P | | P | Notes 1,2,3 | Notes,<br>5,6,7,8,10,11,12,13 |
| | Abdominal (includes urology) | P | P | P | | P | Notes 1,2,3 | Notes,<br>5,6,7,8,9,10,11,12,13 |
| | Intra-operative<br>(Abdominal, cardiac, spine, Vascular) | P | P | P | E* | P | Notes<br>1,2,3,4 | Notes 5,6,8,10,12,13 |
| | Intra-operative (Neuro.) | P | P | P | | P | Notes 1, 3 | Notes 3,5,6,10,12,13 |
| | Laparoscopic | P | P | P | | P | Notes 1, 3 | Notes 8,10,12,13 |
| | Pediatric | P | P | P | E* | P | Notes 1,2, 3 | Notes 5,6,8,9,10,12,<br>13 |
| | Small Organ (breast, thyroid, testicle) | P | P | P | | P | Notes 1,2, 3 | Notes<br>5,6,8,9,10,11,12,13,15 |
| | Neonatal Cephalic | P | P | P | | P | Notes 1,3 | Notes 5,8,10,12,13 |
| | Adult Cephalic | P | P | P | | P | Notes 1,3,4 | Notes 10, 13 |
| | Trans-rectal | P | P | P | | P | Notes 1,3 | Notes 5,6,10,11,12, 13 |
| | Trans-vaginal | P | P | P | | P | Notes 1,2,3 | Notes<br>5,6,7,10,11,12,13 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Conventional) | P | P | P | | P | Notes 1,2,3 | Notes 5,6,8,10,12,13 |
| | Musculo-skel. (Superficial) | P | P | P | | P | Notes 1,2,3 | Notes 5,6,8,10,12,13 |
| | Intra-luminal | | | | | | | |
| | Other: Urology | P | P | P | | P | Notes 1,3 | Notes 5,6,10,12 |
| Cardiac | Cardiac Adult | P | P | P | P | P | Notes<br>1,2,3,4 | Notes 10,11,12,13,14 |
| | Cardiac Pediatric | P | P | P | P | P | Notes<br>1,2,3,4 | Notes 10,11,12,13,14 |
| | Trans-esophageal (Cardiac) | P | P | P | P | P | Notes<br>1,2,3,4 | Notes 10,11,12,13 |
| | Other (Fetal Echo) | P | P | P | P | P | Notes<br>1,2,3,4 | Notes 5,10,12,13 |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Notes<br>1,2,3,4 | Notes<br>2,3,5,6,8,9,10,12,13 |
| | Cerebral Vascular | P | P | P | | P | Notes 1, 2, 3 | Notes 5,6,8,9,10,12,13 |
N= new indication; P= previously cleared (K042540); E= added under Appendix E (*9/21/2005)
| Additional Comments: | |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| *Color Doppler includes Color Amplitude Doppler | Note 8: Extended Field of View (EFOV), AKA Panoramic<br>Imaging includes SonoCT imaging |
| Note 1: Combined modes include: B+PWD; B+Color;<br>B+Amplitude; B+M | Note 9: EFOV including Amplitude Doppler |
| Note 2: Combined modes include: B+M+Color | Note 10: Harmonic Imaging |
| Note 3: Combined modes include: B+Color+PWD;<br>B+Amplitude+PWD | Note 11: Contrast Imaging |
| Note 4: Combined modes include: B+CWD; B+Color+CWD;<br>B+Amplitude+CWD | Note 12: 3D Imaging |
| Note 5: SonoCT | Note 13: XRES |
| Note 6: Imaging for guidance of biopsy | Note 14: TDI |
| Note 7: Infertility monitoring of follicle development | Note 15: Elastography |
DHR
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off) Division of Radiological Devices
510(k) Number K093563
{7}------------------------------------------------
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
| 510(k) No: |
|---------------|
| System: |
| Transducer: |
| Intended Use: |
iU22 Ultrasound System
12-5
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|----------------------|-------------------------------|---|---------------------------------------------------|-----|-----|----------|---------------|------------------------|--|
| General | Specific | в | M | PWD | CWD | Colar | Combined* | Other | |
| (Track I only) | (Tracks & III) | | | | | Doppler* | (Spec.) | (Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | P | P | P | | P | Notes 1, 3 | Notes 5,6,8,10,12,13 | |
| | Abdominal | P | P | P | | P | Notes 1. 3 | Notes 5,6,8,10,12,13 | |
| | Intra-operative (cardiac) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging | Laparoscopic | | | | | | | | |
| & Other | Pediatric | P | P | p | | P | Notes 1, 3 | Notes 5,6,8,10,12,13 | |
| | Small Organ (Breast, thyroid, | P | P | ר | | p | Notes 1, 3 | Notes | |
| | testicle) | | | | | | | 5,6,8,10,11,12,13,15 | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | |
| | Musculo-skel. (Conventional) | P | P | P | | P | Notes 1, 2, 3 | Notes 5,6,8,10,12,13 | |
| | Musculo-skel. (Superficial) | D | P | P | | P | Notes 1, 2, 3 | Notes 5,6,8,10,12,13 | |
| | Intra-luminal | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | | |
| | Fetal Echo | | | | | | | | |
| Perpheral | Peripheral vessel | p | P | 0 | | P | Notes 1, 2, 3 | Notes 5,6,8,9,10,12,13 | |
| Vessel | Cerebral Vascular | P | P | ് | | p | Notes 1, 2, 3 | Notes 5,6,8,9,10,12,13 | |
N= new indication; P= previously cleared (K030455), E= Added under appendix E.
#### Additional Comments:
| *Color Doppler includes Color Amplitude Doppler | Note 8: Extended Field of View (EFOV), AKA Panoramic<br>Imaging includes SonoCT imaging | | | | |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------|--|--|--|--|
| Note 1: Combined modes include: B+PWD; B+Color;<br>B+Amplitude; B+M | Note 9: EFOV including Amplitude Doppler | | | | |
| Note 2: Combined modes include: B+M+Color | Note 10: Harmonic Imaging | | | | |
| Note 3: Combined modes include: B+Color+PWD;<br>B+Amplitude+PWD | Note 11: Contrast Imaging | | | | |
| Note 4: Combined modes include: B+CWD; B+Color+CWD;<br>B+Amplitude+CWD | Note 12: 3D Imaging | | | | |
| Note 5: SonoCT | Note 13: XRES | | | | |
| Note 6: Imaging for guidance of biopsy | Note 14: TDI | | | | |
| Note 7: Infertility monitoring of follicle development | Note 15: Elastography | | | | |
signature
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off) Division of Radiological Devices
510(k) Number K093563
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.