K093546 · Togo Medikit Co., Ltd. · FOZ · Dec 17, 2009 · General Hospital
Device Facts
Record ID
K093546
Device Name
SUPERCATH 5 MODEL SP120-XX-XX, SP125-XX-XX
Applicant
Togo Medikit Co., Ltd.
Product Code
FOZ · General Hospital
Decision Date
Dec 17, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
The SUPERCATH 5 is intended to access a vein or artery and to administer fluids. The SUPERCATH 5 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks.
Device Story
Intravascular catheter for venous/arterial access and fluid administration; used in clinical settings by healthcare professionals. Features include radiopaque polyurethane catheter, check valve to reduce blood flashback, and side-notch needle for visual confirmation of vessel entry. Safety mechanism: button-activated coiled spring retracts metallic introducer needle into hub to prevent needlestick injuries. Available in 14G-24G sizes. Clinician inserts catheter; observes flashback; presses button to retract needle; manages fluid delivery. Benefits include reduced risk of accidental needlesticks and controlled blood flashback during insertion.
Clinical Evidence
Bench testing and simulated clinical use study. Simulated use study involved 8 healthcare professionals performing >500 tests across 5 catheter sizes to verify the needle retraction safety mechanism. No failures observed. Modifications were determined via risk analysis to have no impact on the safety feature performance.
Technological Characteristics
Polyurethane catheter; radiopaque; ethylene oxide sterilized; single-use. Features: check valve, side-notch introducer needle, spring-loaded needle retraction mechanism. Available in 14G, 16G, 18G, 20G, 22G, 24G. Complies with ISO and USP standards.
Indications for Use
Indicated for patients requiring venous or arterial access for fluid administration. Suitable for short-term use (<30 days). No specific age or gender contraindications provided.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
CONFIDENTIAL
SUPERCATH 5 PREMARKET NOTIFICATION 510(k)
DEC 1 7 2009
## Section 6- 510(k) Summary
# a. Owner/Company name, address TOGO MEDIKIT CO., LTD. 17148-6 Aza Kamekawa, Oaza Hichiya Hyuga City, Miyazaki, 883-0062, Japan
Toshimitsu Hayashi Manager, Quality Assurance Department Phone: 011-81-982-53-8000 Fax: 011-81-982-53-8008 qc@togomedikit.co.jp Email:
## b. Contact/Application Correspondent
Fumiaki Kanai, Ph.D. President and CEO MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan
011-81-3-3818-8577 Phone: 800-714-4641 (USA Toll-free) Fax: fkanai@micinternational.com Email:
# c. Date prepared
,
October 5, 2009
## d. Name of device
SUPERCATH 5 Trade Name: Intravascular Catheter Common Name: Catheter, Intravascular, Therapeutic, Short-term less than 30 days Classification Name: Classification Regulation: 21 CFR 880.5200
TOGO MEDIKIT CO., LTD. OCTOBER 5, 2009
SECTION 6 6-2 OF 6-6
{1}------------------------------------------------
K093546
SUPERCATH 5 PREMARKET NOTIFICATION 510(k) CONFIDENTIAL
#### e. Predicate devices
The SUPERCATH 5 is substantially equivalent to the following legally marketed devices:
| 510(k): | k013800 |
|---------------|-------------------------------|
| Trade name: | Insyte Autoguard™ IV Catheter |
| Product code: | FOZ |
The same trade name with the device for the application; 510(k): k081953 SUPERCATH 5 Trade name: Product code: FOZ
#### f. Description of the device
The device is an intravascular catheter known by the trade name SUPERCATH 5 (PROPOSED). It is available in four models, with a check valve, with a side-notch needle, with a check valve and a side-notch needle, and without both a check valve and a notched needle. (See below Figure 1).
The SUPERCATH 5 (PROPOSED) is intended to access a vein or artery and to administer fluids. The SUPERCATH 5 (PROPOSED) is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks.
The SUPERCATH 5 (PROPOSED) catheter hub has a built-in check valve which together with the bealthcare professional's finger pressure on the blood vessel, assists to reduce blood flashback when the metallic introducer needle is withdrawn following blood vessel puncture. The built-in check valve is not intended to stop bleeding completely. Pressing the button on the needle hub activates the coiled spring in the hub, retracting the metallic introducer needle into the needle hub.
The SUPERCATH 5 (PROPOSED) introducer needle has a side-notch to provide rapid visual continention of vessel entry. When the introducer needle is inserted into the vein, blood Commination of vesser caside of the and returns down along the inside of the catheter tube.
The SUPERCATH 5 (PROPOSED) is available in 14G, 18G, 20G, 20G, 22G and 24G.
Image /page/1/Figure/13 description: The image shows a diagram of a catheter tube and needle hub. The catheter tube is a thin, hollow tube that is inserted into a vein or artery to administer fluids or medications. The needle hub is the part of the catheter that connects to the syringe or IV line. The catheter tube is labeled on the left side of the image, and the needle hub is labeled on the right side of the image.
Introducer needle
Catheter hub
TOGO MEDIKIT CO., LTD. OCTOBER 5, 2009
SECTION 6 6-3 OF 6-6
Figure 1. SUPERCATH 5
{2}------------------------------------------------
CONFIDENTIAL
## g. Modification description
The SUPERCATH 5 (PROPOSED) contains the following modifications as compared to the SUPERCATH 5 (k081953):
- 1. Addition of 14G and 16G sizes.
- 2. Modify 22G and 24G introducer needle to include a side-notch to provide rapid visual confirmation of vessel entry.
To evaluate the any effects of the above changes on safety and effectiveness, risk analysis (see Section 18 Risk Analysis) and performance testing (see Section 21 Performance Testing - Bench) were performed. In conclusion the risk analysis and performance testing demonstrated that the above listed modifications did not raise any new safety or effectiveness concerns.
## h. Indications for Use
### Indication for Use
The SUPERCATH 5 is intended to access a vein or artery and to administer fluids. The SUPERCATH 5 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks.
#### Statement of substantial equivalence i.
The characteristics of the SUPERCATH 5 (PROPOSED) are similar to those of the predicate devices described in Item e above. The similarities are:
- . Same Intended Use
- Same Catheter Material (Polyurethane) .
- Radiopaque ●
- Flashback Visualization ●
- Needle Stick Injury Prevention Feature ●
- Check Valve .
- Ethylene Oxide Sterilized .
- Single Sterile Wrapped .
- Multiple Gauge Sizes and Needle Lengths .
TOGO MEDIKIT CO., LTD. OCTOBER 5, 2009
{3}------------------------------------------------
K093546
## SUPERCATH 5 PREMARKET NOTIFICATION 510(k)
#### CONFIDENTIAL
| Table 1. Comparison of Factor/Component<br>Factor/Component | SUPERCATH 5<br>(PROPOSED) | Insyte Autoguard™<br>IV Catheter<br>(k013800) | SUPERCATH 5<br>(k081953) |
|-------------------------------------------------------------|---------------------------|-----------------------------------------------|--------------------------|
| Same Intended Use | Yes | Yes | Yes |
| Polyurethane Catheter | Yes | Yes | Yes |
| Radiopaque Catheter | Yes | Yes | Yes |
| Flashback Visualization | Yes | Yes | Yes |
| Needle Stick Injury Prevention Feature | Yes | Yes | Yes |
| Check Valve | Yes | No | Yes |
| Ethylene Oxide Sterilized | Yes | Yes | Yes |
| Single Sterile Wrapped | Yes | Yes | Yes |
| Multiple gauge sizes and needle lengths | Yes | Yes | Yes |
| Side-Notch needle | Yes | Yes | No |
Table 1, Comparison of Factor/Comnonent
The SUPERCATH 5 (PROPOSED) has the same intended use and similar technological che Coriences as the "Insyte Autoguard™ IV Catheter" (k013800) and the "SUPERCATH 5" (k081953). Similar components and materials are used in the SUPERCATH 5 (PROPOSED) as in the "SUPERCATH 5" (k081953).
#### Risk Analysis i.
The SUPERCATH 5 (PROPOSED) was evaluated in accordance with ISO14971:2007. The risk management of the device was deemed satisfactory. See Section 18 Risk Analysis.
## k. Bench Testing
Bench tests were performed to ensure the safety and effectiveness of the SUPERCATH 5 (PROPOSED), verify conformity to the recognized standards and demonstrate substantial equivalence to the predicate device(s). All SUPERCATH 5 (PROPOSED) samples were equivale with the ISO and USP standards and demonstrated substantial equivalence to the predicate device(s).
#### Simulated Clinical Use l.
Shirdlated Chilital Coo In accordance while is a great go 2005, the sharp needle injury prevention feature of the SUPERCATH 5 (k081953) was tested.
According to the risk analysis (See Section 18 Risk Analysis), the modifications to the SUPERCATH 5 (k081953), of 1) the additional gauge sizes of 14G and 16G and 2) the Sorthernitt > (18001953), or endle of the 22G and 24G, have no effect on the sharp needle Side-noten on the Introducer novel of the SUPERCATH 5 (PROPOSED). Therefore, Togo Mediliti has included the exact same report submitted with the SUPERCATH 5 (k081953). (See Section 23 Performance Testing - Simulated Clinical Use.)
{4}------------------------------------------------
K093546
## SUPERCATH 5 PREMARKET NOTIFICATION 510(k)
CONFIDENTIAL
The purpose of the investigation was to test if the sharp needle injury prevention feature of the SUPERCATH 5 (K081953) worked properly under a simulated use clinical environment. As per recommendation under the "Guidance", eight (8) health care professional volunteers who routinely use this type of catheter in their clinical practice were recruited.
The primary evaluation parameter was to test if the metal introducer needle completely retracted into the plastic needle hub after firmly pressing the safety button. This mechanism is the sharp needle injury prevention feature of the SUPERCATH 5 (k081953) (PREDICATE) and the SUPERCATH 5 (PROPOSED).
No failures were observed in over 500 tests, under different test conditions and five (5) different catheter sizes of the SUPERCATH 5 (PREDICATE). Thus, the requirement of statistical significance of the safety feature was confirmed for the sharp needle injury prevention feature of the SUPERCATH 5 (PREDICATE).
Therefore, Togo Medikit believes that the same results apply to the identical safety feature of the SUPERCATH 5 (PROPOSED), since the modifications to the SUPERCATH 5 (k081953), of 1) the additional gauge sizes of 14G and 16G and 2) the side-notch on the introducer needle of the 22G and 24G, have no effect on the sharp needle injury prevention feature of the SUPERCATH 5 (PROPOSED) (See Section 18 Risk Analysis.)
### m. Conclusion
Based on the above discussion and enclosed sections regarding substantial equivalence to predicate devices, Togo Medikit concludes that the SUPERCATH 5 (PROPOSED) is substantially equivalent to the "Insyte Autoguard" IV Catheter" (k013800) and the "SUPERCATH 5" (k081953) and does not raise any new questions regarding safety or effectiveness.
SECTION 6 6-6 OF 6-6
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract symbol to the right, which resembles three human figures connected by flowing lines. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Togo Medikit Company, Limited C/o Fumiaki Kanai, Ph.D. President and CEO MIC International 4-1-17 Hongo Bunkyo-Ku, Tokyo 113-0033 JAPAN
DEC 1 7 2009
Re: K093546
Trade/Device Name: SUPERCATH 5 Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: October 5, 2009 Received: November 17, 2009
### Dear Dr. Kanai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2- Dr. Kanai
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH /CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Susan Runner
Anthony D. Watson, B.S., M.S., M.B.A.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
# Indications for Use
510(k) Number (If known): K093546
Device Name: SUPERCATH 5
#### Indication for Use
The SUPERCATH 5 is intended to access a vein or artery and to administer fluids. The The Borekering o is mondou to de use and for short-term use (less than 30 days), is Sor EKCATI 9-18-dongied 10. "Single in the end to reduce accidental needlesticks.
Prescription Use
(Per 21 CFR 801 Subpart D)
AND/OR
Over-the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093546
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.