K093477 · Soteira, Inc. · NDN · Dec 8, 2011 · Orthopedic
Device Facts
Record ID
K093477
Device Name
SHIELD KYPHOPLASTY SYSTEM
Applicant
Soteira, Inc.
Product Code
NDN · Orthopedic
Decision Date
Dec 8, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Soteira Shield Kyphoplasty System (SKS) is intended to provide control of cement flow during injection of PMMA bone cement that has been cleared for use in vertebral augmentation for the treatment of acute, persistently painful (after a minimum of 6 weeks of conservative care), stable, anterior column osteoporotic compression fractures (wedge or concave) of the vertebrae at levels T4 - L5 in the adult spine.
Device Story
The Shield Kyphoplasty System is a surgical kit used for vertebral augmentation. It includes a permanent cement director implant, a delivery system, and access instruments to create a cavity within the vertebral body. The system is used in conjunction with commercially available PMMA bone cement. Operated by physicians in a clinical setting, the device provides controlled cement flow during injection to stabilize osteoporotic compression fractures. By directing the cement, the system aims to reduce the risk of cement leakage compared to standard vertebroplasty. The output is the stabilized vertebral body, which helps alleviate pain and restore function in patients with spinal fractures.
Clinical Evidence
Clinical evidence included a pilot study and a prospective randomized study comparing the Shield Kyphoplasty System to vertebroplasty. Total of 69 subjects (102 levels) treated with the subject device and 28 subjects (38 levels) with the control. Primary endpoints included pain and functional improvement, which showed equivalent initial and sustained benefits at one year. The subject device demonstrated fewer asymptomatic leaks than the control. Adverse events were similar between groups, including death, adjacent/distant level fractures, and refracture.
Technological Characteristics
System includes a permanent cement director implant, delivery system, and access instruments. Materials are biocompatible per ISO 10993. The device is MR-conditional per ASTM F2503-05. The system is a mechanical device used for manual delivery of PMMA bone cement; it does not utilize electronic energy sources or software algorithms.
Indications for Use
Indicated for adult patients with acute, persistently painful, stable, anterior column osteoporotic compression fractures (wedge or concave) of the vertebrae at levels T4-L5, following at least 6 weeks of conservative care.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
## 510(k) Summary
DEC - 8 2011
| Owner's Name: | Soteira, Inc. |
|------------------------|-----------------------------------------------------------------------------------------|
| Address: | 14 Tech Circle<br>Natick, MA 01760 |
| Telephone Number: | (508) 651-2611 |
| Fax Number: | (508) 651-3611 |
| Contact Person: | John V. Hamilton, Director of Spinal Programs |
| Subject Device Name: | Shield Kyphoplasty System |
| Trade Name: | Shield Kyphoplasty System |
| Classification Name: | Polymethylmethacrylate (PMMA) Bone Cement |
| Product Code: | NDN |
| Regulation: | 21 CFR 888.3027 |
| Device Classification: | Class II device |
| Predicate Devices: | Stryker SpinePlex™ Bone Cement (K032945)<br>Kyphon Kyphx® Inflatable Bone Tamp (K041454 |
### Device Description
The Soteira Shield Kyphoplasty System consists of a cement director (permanent implant), a delivery system (used to place the cement director within the vertebral body) and instruments that are used to achieve percutaneous access to the vertebral body and to create a cavity into which the cement director implant will be placed. The system is intended for use with commercially available PMMA bone cement.
#### Indications for Use
The Soteira Shield Kyphoplasty System (SKS) is intended to provide control of cement flow during injection of PMMA bone cement that has been cleared for use in vertebral augmentation for the treatment of acute, persistently painful (after a minimum of 6 weeks of conservative care), stable, anterior column osteoporotic compression fractures (wedge or concave) of the vertebrae at levels T4 - L5 in the adult spine.
#### Performance Testing
Pre-clinical testing included cadaveric bench testing, biocompatibility and MRI compatibility. The cadaveric bench tests consisted of static and cyclic mechanical testing of osteoporotic vertebral bodies treated with the Soteira Shield Kyphoplasty System and vertebroplasty. The Soteira Shield Kyphoplasty System demonstrated equivalent behavior under both test conditions. Biocompatibility testing was performed in accordance with ISO 10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing demonstrated the combination of the Soteira Shield implant and PMMA bone cement is biocompatible. The implant and bone cement have been tested for MRI compatibility and have been determined to be "MR-conditional" in accordance with the ASTM standard F2503-05 Standard Practice for Marking Medical Devices and Other Item for Safety in the Magnetic Resonance Environment.
Clinical testing consisted of a pilot study and a prospective randomized study using vertebroplasty as the control device. All subjects were diagnosed with osteoporotic compression fractures that had undergone conservative care for a minimum of 6 weeks or had
{1}------------------------------------------------
been hospitalized for pain. A total of 69 subjects and 102 levels were treated with the Soteira Shield Kyphoplasty System and 28 subjects and 38 levels were treated with the control device. Pain and functional testing demonstrated equivalent initial improvements and sustained benefits out to one year. The Soteira Shield Kyphoplasty System showed less assymptomatic leaks than the control (vertebroplasty). For the Soteira device adverse events included ; death, adjacent and distant level fractures, incomplete filling, refracture and retreatment of the treated levels while the control exhibited; death, adjacent and distant level fractures,and , refracture of a treated level
## Conclusion
The Soteira Shield Kyphoplasty System has been demonstrated, through labeling, descriptive characteristics and performance testing data, to be substantially equivalent to the predicate devices.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three parallel lines forming its body and wings. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Soteria, Inc. % Delphi Medical Device Consulting, Inc. Ms. Pamela Papineau 5 Whitcomb Avenue Ayer, Massachusetts 01432
DEC 0 8 2011
Re: K093477
Trade/Device Name: Shield Kyphoplasty System Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN Dated: September 15, 2011 Received: October 24, 2011
Dear Ms. Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability. warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Pamela Papineau
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm.
Sincerely yours,
F-R
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K093477
Device Name:
Shield Kyphoplasty System
Indications for Use:
The Soteira Shield Kyphoplasty System (SKS) is intended to provide control of cement flow during injection of PMMA bone cement that has been cleared for use in vertebral augmentation for the treatment of acute, persistently painful (after a minimum of 6 weeks of conservative care), stable, anterior column osteoporotic compression fractures (wedge or concave) of the vertebrae at levels T4 - L5 in the adult spine.
Prescription Use X (Per 21 CFR 801 Subpart D)
OR
Over-the -Counter Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K093477
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.