OMNIGUIDE BEAMPATH CO2 MARK III WAVEGUIDE FIBER WITH LOW PROFILE/LOW LOSS TIP
K093451 · Omniguide, Inc. · GEX · Nov 20, 2009 · General, Plastic Surgery
Device Facts
Record ID
K093451
Device Name
OMNIGUIDE BEAMPATH CO2 MARK III WAVEGUIDE FIBER WITH LOW PROFILE/LOW LOSS TIP
Applicant
Omniguide, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The OmniGuide Beam Path CO2 Mark III WaveGuide Fiber with Low Profile/Low Loss Tip is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy. The indications for use for which the delivery system is used for are dependent upon the cleared indications for use of the laser system and those laser system accessories to which it is attached.
Device Story
Flexible fiber assembly accessory for CO2 laser systems; delivers CO2 laser energy (10.6 um wavelength) for minimally invasive surgery. Features photonic bandgap reflector lining hollow core; includes 14 mm internally coated stainless steel distal tip. Tip configuration reduces heating, enabling continuous wave mode operation. Used in OR/clinic settings by physicians; attaches to OmniGuide Laser Adapter. Enables incision, excision, ablation, vaporization, and coagulation of soft tissues. Benefits patient via minimally invasive access to surgical sites.
Clinical Evidence
Bench testing only. Evaluated laser power output and beam quality per FDA 1995 Laser Guidance. No clinical trials conducted; device utilizes same technology and intended use as predicate devices.
Technological Characteristics
Optical fiber assembly; 1-2 m length. Photonic bandgap reflector lining hollow core; 14 mm internally coated stainless steel distal tip. Transmits 10.6 um CO2 laser wavelength. Single-use, sterile. Standalone accessory for CO2 laser systems.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues (including intra-oral) across general, plastic, oral/maxillofacial, dental, dermatological, gynecological, ENT, gastroenterological, neurosurgical, urological, and pulmonological specialties. Used in open or endoscopic minimally invasive procedures (e.g., laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy, bronchoscopy).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
OmniGuide Beam Path CO2 Mark III WaveGuide Fiber (K070157)
OmniGuide BeamPath CO2 Mark III WaveGuide Fiber With Low Profile Tip (K073313)
Submission Summary (Full Text)
{0}------------------------------------------------
# 1093451
Image /page/0/Picture/1 description: The image shows the logo for OmniGuide. The logo consists of two concentric circles, with the word "OmniGuide" written in a sans-serif font across the center of the circles. The text is black, and the circles are also black.
NOV 2 0 2009
## 510(k) Summary
| Submitter: | OmniGuide, Inc.<br>One Kendall Square, Building 100 3rd Floor<br>Cambridge, MA 02139 |
|-------------------------------|-----------------------------------------------------------------------------------------|
| Contact Person:<br>Telephone: | Douglas W. Woodruff<br>617-551-8404 |
| Fax: | 617-551-8445 |
| Proprietary Name: | OmniGuide Beam Path CO2 Mark III WaveGuide Fiber<br>with Low Profile/Low Loss Tip |
| Common Name: | CO2 Laser Powered Surgical Instrument |
| Classification: | 878.4810 |
| Product Code: | GEX |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology. |
#### Substantial Equivalence Claimed To:
K070157, OmniGuide Beam Path CO2 Mark III WaveGuide Fiber K073313, OmniGuide BeamPath CO2 Mark III WaveGuide Fiber With Low Profile Tip
{1}------------------------------------------------
#### Description :
The OmniGuide OmniGuide BeamPath CO2 Mark III WaveGuide Fiber with Low Profile/Low Loss Tip is an accessory for CO2 laser systems. It consists of a flexible fiber assembly that delivers CO2 laser energy that enables minimally invasive surgery. OmniGuide Beam Path CO2 Mark III WaveGuide Fiber with Low Profile/Low Loss metal tip is supplied sterile and is intended for single procedure use in conjunction with the OmniGuide Laser Adapter.
#### Intended Use:
The OmniGuide Beam Path CO2 Mark III WaveGuide Fiber with Low Profile/Low Loss Tip is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy.
The indications for use for which the delivery system is used for are dependent upon the cleared indications for use of the laser system and those laser system accessories to which it is attached.
#### Summary of Technological Characteristics:
The device consists of an optical fiber assembly. The main functional component of the fiber assembly is a photonic bandgap reflector lining its hollow core that reflects and thereby guides CO2 laser energy and an internally coated, stainless steel waveguide distal tip. The fiber assembly is 1 to 2 m long and transmits at the CO2 laser emission wavelength of 10.6 um. The fiber has 14 mm long, internally coated, stainless steel tip that is coated internally and acts as an efficient continuation of the waveguide. The new tip configuration reduces tip heating allowing the fiber to be used in continuous wave mode without heating the tip.
#### Performance Data:
Non-clinical Performance Data: The OmniGuide Beam Path CO2 Mark III WaveGuide Fiber with Low Profile /Low Loss Tip performance characteristics have been evaluated through testing and analysis of laser power output and beam quality. This type of testing complies with the respective section of the FDA Guidance on the Content and Organization of a Premarket Notification for a Medical Laser (1995) and is similar to the predicate device tests. The performance of OmniGuide Beam Path CO2 Mark III WaveGuide Fiber with Low Profile/Low Loss Tip and the related parameters of the predicate devices are comparable.
Clinical Performance Data: Formal clinical trials were not deemed necessary as the device is using the same technology and intended use as the predicate devices.
Image /page/1/Picture/10 description: The image shows the text "Page 15 of 72". This indicates that the image is a page from a document. The document has a total of 72 pages. The current page being displayed is page 15.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a black and white graphic of what appears to be a water droplet. The droplet is connected to a larger, irregular black shape on the left side of the image. The droplet is round and smooth, while the shape on the left is jagged and uneven.
Conclusions Drawn from Tests and Analysis: The intended use and major performance parameters (energy transmission levels and beam quality) of the OmniGuide Beam Path CO2 Mark III WaveGuide Fiber with Low Profile/Low Loss Tip are similar or equivalent to the characteristics of above mentioned legally marketed devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized representation of an eagle or bird-like figure, with its wings spread, symbolizing the department's mission related to health and human welfare.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
OmniGuide, Inc. % Regulatory Technology Services, LLC Mr. Mark Job 1394 25th Street NW Buffalo, Minnesota 55313
Re: K093451
Trade/Device Name: OmniGuide BeamPath CO2 Mark III WaveGuide Fiber with Low Profile/Low Loss Tip Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 4, 2009 Received: November 5, 2009
NOV 2 0 2009
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Mark Job
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Sincerely yours,
Mark N. Mcllhinney
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known): Not yet assigned
#### Device Name: OmniGuide BeamPath CO2 Mark III WaveGuide Fiber with Low Profile/Low Loss Tip
#### Indications For Use:
The OmniGuide Beam Path CO2 Mark III WaveGuide Fiber with Low Profile/Low Loss Tip is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical snecialti os of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatlogy, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endosconic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy.
The indications for use for which the delivery system is used for are dependent upon the cleared indications for use of the laser system and those laser system accessories to which it is attached.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neilheblen for mta
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
510(k) Number K093451
Page 13 of 72
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.