The intended use of the DGH 6000 is the measurement of AL, ACD, LT of the human eye. The DGH 6000 is also intended to calculate the optical power of an IOL that is to be implanted during cataract surgery. The DGH 6000 is intended to be used solely by qualified medical professionals. Clinical consideration and judgment should be applied when using the DGH 6000.
Device Story
DGH 6000 Scanmate A is a USB-connected A-Mode pulsed echo ultrasound biometer. Device captures ultrasonic echoes from human eye structures; processes signals to measure axial length, anterior chamber depth, and lens thickness. Software integrates established IOL calculation formulas to determine required IOL power for cataract surgery. Operated by qualified medical professionals in clinical settings; output displayed on PC monitor. Healthcare providers use measurements and calculated IOL values to guide surgical planning. Benefits include accurate ocular biometry for improved IOL selection during cataract procedures.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including phantom comparative tests and adherence to electrical safety and acoustic output standards.
Technological Characteristics
A-Mode pulsed echo ultrasound; USB-connected transducer (DGH 6006); PC-based display. Measures AL, ACD, LT. Software includes IOL calculation formulas. Complies with EN 60601-1, EN 60601-1-2, NEMA UD-2, and NEMA UD-3.
Indications for Use
Indicated for ophthalmic patients undergoing cataract surgery requiring measurement of axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) for IOL power calculation.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Quantel Medical AXIS II Ultrasonic Biometer (K000554)
{0}------------------------------------------------
# TECHNOLOGY, INC.
## 510(k) Summary (21 CFR 807.92(c)) 510(k) K093371
Submitter:
Contact Person:
DGH Technology, Inc. 110 Summit Drive, Suite B Exton, PA 19341
M. Luther Detweiler Vice President Regulatory Affairs and Quality Assurance
Telephone: 610-594-9100 Fax: 610-594-0390 Email: Lou@dghkoi.com
Date Summary Prepared:
March 12th, 2010
Device Trade Name:
DGH 6000 Scanmate A.
Device Common Name:
Ultrasound A-Scan biometer.
Device Classification:
Device: System, Imaging, Pulsed Echo, Ultrasonic Panel: Radiology Product Code: IYO Device Class: II Regulation Number: 21 CFR 892.1560
Device: Diagnostic Ultrasonic Transducer Panel: Radiology Product Code: ITX Device Class: II Regulation Number: 21 CFR 892.1570
110 Summit Drive, Suite B Exton, PA 19341
PA 19341
Phone: (610) 594-9100 (800) 722-3883
Fax: (610) 594-0390
APR - 5 2010
.
.
.
.
.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for DGH Technology, Inc. The letters "DGH" are in large, bold, black font on the left side of the image. To the right of the letters is a stylized graphic that appears to be a horizontal line with curved extensions above and below it. Above the graphic, the words "TECHNOLOGY, INC." are printed in a smaller font.
Legally Marketed Predicate Device(s):
Description of Device:
The DGH 6000 A-Scan is a USB plug-in device that uses A-Mode, pulsed echo ultrasound technology to measure the axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) of the human eye. The device includes formulas to calculate the implanted IOL power, using the axial length measurement.
Quantel Medical AXIS II Ultrasonic Biometer (K000554)
DGH 3000A Ultrasonic A-Scan (K872726)
Intended Use of the Device:
The intended use of the DGH 6000 is the measurement of AL, ACD, LT of the human eye. The DGH 6000 is also intended to calculate the optical power of an IOL that is to be implanted during cataract surgery. The DGH 6000 is intended to be used solely by qualified medical professionals. Clinical consideration and judgment should be applied when using the DGH 6000.
#### Technological Characteristics:
Axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) measurements are obtained using the same technology as the predicate device, DGH 3000A. The technology is based on ultrasonic pulse echo technology, whereby short bursts of ultrasonic energy are transmitted and the resulting echoes are captured, amplified, filtered and processed. Specific time distances in the captured echo peaks are then measured and converted into distance information. Digital signal processing algorithms validate individual peaks and proper probe alignment. The resulting measurements and waveforms are then displayed on a PC monitor for the medical (ophthalmic) professional user. In addition, the software assists the user in calculating replacement intraocular lens values based on established IOL calculation formulas.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for DGH Technology, Inc. The letters "DGH" are in a bold, sans-serif font. To the right of the letters is a stylized graphic that resembles a wing or a stylized line. Above the graphic, the words "TECHNOLOGY, INC." are printed in a smaller font.
#### Performance Tests:
The following tests were performed to demonstrate substantial equivalence:
(A) Non-Clinical Tests
- . Comparative test block (phantom) tests
- . EN 60601-1:1990 Medical Electrical Equipment Part 1 General Requirements for Electrical Safety
- . EN 60601-1-2:2002 Medical Electrical Equipment Collateral Standard: Electromagnetic Compatibility
- . NEMA Standard Publication UD-2 2004: Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- . NEMA Standard Publication UD-3 2004: Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment.
(B) Clinical Tests
Since the DGH 6000 Scanmate-A uses the same technology as existing devices, clinical tests are not required.
#### Conclusions:
We have reviewed the results of the performance tests and have determined that the DGH 6000 is substantially equivalent, in safety and efficacy, to the legally marketed predicate device.
Phone: (610) 594-9100 (800) 722-3883
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
### APR - 5 2010
DGH Technology, Inc. % Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories, Inc. 333 Pfingsten Road NORTHBROOK IL 60062
Re: K093371
Trade/Device Name: DGH 6000 Scanmate A Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed Doppler imaging system Regulatory Class: II Product Code: IYO and ITX Dated: March 23, 2010 Received: March 24, 2010
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the DGH 6000 Scanmate A, as described in your premarket notification:
#### Transducer Model Number
#### DGH 6000
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely yours,
Donald J. St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K093371
Device Name: DGH 6000 Scanmate A
Indications for Use: The intended use of the DGH 6000 is the measurement of AL. ACD, LT of the human eye. The DGH 6000 is also intended to calculate the optical power of an IOL that is to be implanted during cataract surgery. The DGH 6000 is intended to be used solely by qualified medical professionals. Clinical consideration and judgment should be applied when using the DGH 6000.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
KMR
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Page 1 of 3
{6}------------------------------------------------
## . Attachment 3
## Indications for Use Form
.. ...........................................................................................................................................................................
13
510(k) K093371 System: DGH 6000 Scanmate A Transducer: DGH 6006
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|------------------------------|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | A-Mode |
| Ophthalmic | Ophthalmic | | | | | | | P |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | |
| Vessel | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K K093371
{7}------------------------------------------------
## Indications for Use Form
#### 510(k) K093371 Transducer: DGH 6006
Intended Use: The intended use of the DGH 6006 transducer is the measurement of AL, ACD, LT of the human eye. The DGH 6006 transducer is intended to be used solely by qualified medical professionals. Clinical consideration and judgment should be applied when using the DGH 6006 transducer.
| Clinical Application | | Mode of Operation | | | | | | | |
|------------------------------|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------|--|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | A-Mode | |
| Ophthalmic | Ophthalmic | | | | | | | P | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (Specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | | |
| Vessel | Other (Specify) | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
(Division Sign-Off)
Division of Radiological Office of In Vitro Diagnostic Device Ev
510K K093371
3/3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.