DELPHI INFUSION PUMP/ DELPHI ADMINISTRATION SET MODEL IV-01000/ MODEL IV-02XXXX (WHERE X IS ANY NUMBER BETWEEN 0-9)
K093332 · Delphi Medical Systems · FRN · Dec 9, 2009 · General Hospital
Device Facts
Record ID
K093332
Device Name
DELPHI INFUSION PUMP/ DELPHI ADMINISTRATION SET MODEL IV-01000/ MODEL IV-02XXXX (WHERE X IS ANY NUMBER BETWEEN 0-9)
Applicant
Delphi Medical Systems
Product Code
FRN · General Hospital
Decision Date
Dec 9, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Delphi Infusion Pump is intended for use in the controlled delivery of epidural and IV solutions.
Device Story
Large volume infusion pump; delivers IV/epidural solutions via rotary peristaltic mechanism. Inputs: user-programmed flow rates, volumes, and drug parameters via touch screen/keypad. Operates in continuous, intermittent (dose/multi-step), TPN, and PCA modes. Features Master Drug Library for dose limits; safety suite for clinician-defined alarms/alerts. Used in clinical settings by healthcare professionals; portable for patient transport. Output: controlled fluid delivery; visual/auditory status and alarm feedback. Benefits: reduces medication errors via drug library and improved UI; ensures safe, accurate delivery across diverse therapies.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing demonstrated equivalence to predicate in flow accuracy, safety, and EMI/EMC compliance.
Technological Characteristics
3-roller rotary peristaltic pump; 3-phase brushless motor with hall/flag sensors. Up/downstream occlusion sensors; ultrasonic air-in-line sensor. Large color touch screen; alphanumeric keypad. Battery backup. Dedicated disposable administration sets with safety slider/roller clamp for free-flow protection. Software includes Master Drug Library and safety suite for alarm/alert configuration.
Indications for Use
Indicated for controlled delivery of epidural and IV solutions, including lipids, blood/blood products, saline, and TPN, in clinical environments under supervision of licensed healthcare professionals.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Spectrum and Spectrum with Master Drug Library (K042121)
Submission Summary (Full Text)
{0}------------------------------------------------
**DELPHI**
Medical Systems
Kq9332 (R.1 of 4)
5725 Delphi Drive Troy, Mi 48098-2815 USA
DEC - 9 2009
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Section V | 510(k) Summary (As required by section 807.92(c)) | |
|---------------------------------|--------------------------------------------------------------------------------------------------|-------------|
| Submitter: | Delphi Medical Systems<br>5725 Delphi Drive<br>Troy MI 48098 | |
| Contact: | Paul T. Kotnik., Infusion Product Architect & Sr.<br>Regulatory Affairs Engineer<br>FDA CDRH DMC | |
| Telephone: | (248) 813-2242 | NOV 27 2009 |
| Fax: | (248) 813-4799 | |
| Date Prepared: | July 2009 | Received |
| DEVICE NAME AND CLASSIFICATIONS | | |
| Type of Submission: | Traditional 510(k) | |
| Proprietary Name: | Delphi Infusion Pump / Delphi Infusion Sets<br>Model IV-01000 / Model IV-02XXXX (X = 0-9) | |
| Common Name: | Infusion Pump / Infusion Set | |
| Classification Name: | Pump, Infusion / Set, Administration,<br>Intravascular | |
| Product Code: | FRN / FPA | |
General Hospital Medical Specialty / Panel:
Class II Device Classification:
21 CFR 868.5440 / 21 CFR 880.5440 Regulation Number:
> Delphi Medical Systems, IVantage Infusion Pump, 510(k) # K953896* Sigma International, Spectrum and Spectrum with Master Drug Library, 510(k) # K042121
# Indications For Use:
Predicate Device:
The Delphi Infusion Pump is intended for use in the controlled delivery of epidural and IV solutions.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "DELPHI" in a bold, sans-serif font. The letters are black and appear to be slightly distressed, with some minor imperfections or variations in the ink density. The word is presented in all capital letters and is horizontally oriented.
K493332 (0.25A4
5725 Delphi Drive Troy, MI 48098-2815 USA
### Intended Use:
The pump shall deliver medications in a safe and accurate manner over a broad range of flow rates and parameters. The pump shall be compatible with lipids, blood and blood products, saline, Total Parenteral Nutrition (TPN), or any IV medication. This pump is intended to operate in a clinically appropriate environment under the supervision of a licensed health care professional
# Device Description:
The Delphi Infusion Pump is a large volume infusion pump customizable to meet the specific needs of multiple care areas and deliver a wide range of therapies in a safe and effective manner. For each area of care, clinically appropriate alarms, alerts, and performance specifications can be established. With the available Safety Suite, the pharmacy can define default and safe operating ranges for all drugs being infused. There are four basic infusion modes available plus epidural delivery.
## Continuous Mode
These are infusions which deliver a defined volume to be infused (VTBI) at a constant flow rate. Optionally, a weight based approach can be used to enter the dose rate. The ability to infuse secondary infusions as well as keep vein open (KVO) settings are available. After infusion has begun. infusion rate can be changed temporarily using Titrate or for the duration of the infusion using Rate Change.
### Intermittent Mode (Dose Mode and Multi-Step Mode)
This infusion mode allows the user to either repeat the same dose a programmable number of times (Dose Mode) or to program a series of doses at varying rates and durations (Multi-Step Mode).
#### Total Parenteral Nutrition (TPN) Mode
This infusion mode allows the user to ramp the delivery rate up at the beginning of an intravenous nutrition delivery and ramp down the delivery rate at the end.
# Patient Controlled Analgesia (PCA) Mode
This infusion mode allows the patient to initiate doses of medication on an as-needed basis within preset limits. The clinician is given the ability to program a loading dose, basal rate, and controlled bolus doses.
#### Epidural Delivery
Allows access to Continuous, PCA, and Intermittent Epidural Modes and is indicated by a yellow background screen.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "DELPHI" in a bold, sans-serif font. The letters are large and closely spaced, filling most of the frame. The text appears to be a logo or brand name, presented in a simple, impactful way. The overall impression is one of clarity and strength.
фa3332 (p.3 of 4)
5725 Delphi Drive Trov. MI 48098-2815 USA
The Delphi Infusion Pump uses a rotary peristaltic pumping mechanism with dedicated disposable administration sets. Each administration set includes a precision-molded cassette designed for fast, safe, and secure single-handed pump loading. A safety slider switch on the cassette, along with a roller clamp. provides redundant free-flow protection when the tubing is unloaded. The user interface was designed using extensive input from clinical and human factors experts to offer a unique combination of a large touch screen display and a keypad for alpha-numeric data entries. The pump is designed for use on AC power with battery backup during power failure and is available during patient ambulation and transport. The portable light weight single-channel design allows for a dedicated display ("one screen, one drug") and improved utilization and increased flexibility in device placement.
Delphi Administration Sets are available in a variety of configurations containing components such as needleless Y-Sites, filters, drip chambers, PE-lined tubing, etc and a common cassette to interface with the Delphi Infusion Pump.
The Delphi Infusion pump will include the option of a Master Drug Library which provides customers the ability to set defined dose limits specific to their individual formulary. Dose limits may be attached to specific care areas and instances where dose limits are exceeded can be recorded. Biomedical setting options include the ability to set clinically appropriate alarms, alerts and performance specifications.
# Technological Characteristics:
The Delphi Infusion Pump uses a 3-roller rotary peristaltic pump to deliver fluids from a container to a patient. As the motor shaft rotates, the rollers on the pump rotate and the compressed portion of the tubing translates across the inner circumference of the cassette housing. This results in a rotary peristaltic pumping action. At a minimum, one of the rollers is always in contact with the infusion set to prevent free-flow during infusion and when not infusing.
The pump mechanism contains both an up/down stream occlusion sensors and an ultrasonic sensor for air detection. The pump can be programmed to detect occlusion and air sensitivity at high, low, and auto-sensing. The roller mechanism uses a 3-phase brushless motor with hall sensors and a flag sensor for redundant sensing of motor rotation. The Delphi Infusion Pump detects a variety of system alarms to monitor for proper operation of the device, Cassette detection, Occlusion, and Air-in-Line alarms.
The pump uses a large color touch screen and alphanumeric keypad to view operational status and program infusion. Each kevpress is identified by an auditory feedback to the user.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "DELPHI" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The overall impression is clean and professional.
K49332(l.4ofu)
5725 Delphi Drive Troy, MI 48098-2815 USA
The infusion sets contain a cassette that is custom designed to interface with the Delphi Infusion Pump and has an integrated free-flow protection that occludes the tubing.
The technological characteristics of the Delphi Infusion Pump are substantially equivalent to the predicate device noted in the submission and these technologies are well established for Delphi Infusion Pumps.
## Substantial Equivalence
Both predicate device and Delphi Infusion Pump are substantially equivalent in performance and are intended to be used for the same purpose. The two products are different from each other only by User Interface, battery duration, size, and weight. Larger user interface and custom software reduces entry error by a clinician. Also, the Delphi Infusion Pump contains more alarms and indicators for the user than the predicate device. The Delphi Infusion Pump is larger in size than the predicate device but is still able to be used in a typical clinical environment.
# Performance Requirement
Bench performance testing has also demonstrated that the Delphi Infusion Pump is substantially equivalent in performance as the predicate device in flow accuracy and in construction when compared using performance standards for infusion pumps.
Furthermore, the Delphi Infusion Pump will also be tested for safety, EMI/EMC, and will bear a safety marked from a Nationally Recognized Testing Laboratory, which will be substantially equivalent to the safety, EMI/EMC testing, and safety mark on the predicate device.
## Conclusion:
The Delphi Infusion Pump is substantially equivalent in function, technological characteristics, and intended use to other devices currently in commercial distribution, namely the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract image of a bird-like figure with three heads, which is the department's official symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
DEC - 9 2009
Delphi Medical Systems C/O Mr. Ned Devine Responsible Third Party Official Underwriter Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
Re: K093332
Trade/Device Name: Delphi Infusion Pump Regulation Number: 21 CFR 880,5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: November 27, 2009 Received: November 27, 2009
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2- Mr. Rongero
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
K for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Medical Systems 5725 Delphi Drive Troy, M1 48098-2815 USA
# Section IV: Indications for Use Statement
Applicant: Delphi Medical Systems
X493332 510(k) Number:
Device Name: Delphi Infusion Pump
Indications for Use:
The Delphi Infusion Pump is intended for use in the controlled delivery of epidural and IV solutions.
Prescription Use Only: _ _ X
Over-the-Counter Use: ________________________________________________________________________________________________________________________________________________________
ノ
.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
(in them ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: __ k p 93333 2 ________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.