K093306 · Nonin Medical, Inc. · DQA · Mar 4, 2010 · Cardiovascular
Device Facts
Record ID
K093306
Device Name
SINGLE-PATIENT USE DISPOSABLE SENSOR
Applicant
Nonin Medical, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Mar 4, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Nonin's Models 6500MA and 6500SA Single-Patient Use Disposable Pulse Oximeter Sensors are indicated for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients who are well or poorly perfused, weighing greater than 60 pounds (30 kilograms). It is intended for use in environments including operating room, surgical recovery, critical care, emergency room, long-term care, home use and mobile environments.
Device Story
The 6500MA (wrap-around) and 6500SA (interlocking) are single-patient use disposable transmittance pulse oximeter sensors. They utilize optical components (red 660 nm, infrared 910 nm) to measure SpO2 and pulse rate when connected to Nonin-branded pulse oximeters. The sensors consist of a lamination of foams (polyurethane, polyester, polyethylene) and a malleable wire for positioning on a patient's finger. The device is operated by clinicians or patients in various settings, including ORs, critical care, and home environments. The oximeter monitor processes the optical signals to provide real-time SpO2 and pulse rate data. This output assists healthcare providers in assessing patient oxygenation status, enabling timely clinical interventions. The disposable nature of the sensors helps reduce the risk of cross-contamination.
Clinical Evidence
The device underwent bench and clinical testing to demonstrate compliance with ISO 9919:2005 (accuracy) and IEC 60601-1:1998/ISO 10993-1:2003 (biocompatibility). Clinical testing confirmed SpO2 accuracy of ±2 digits for both normal and low perfusion conditions, improving upon the predicate's ±3 digits.
Technological Characteristics
Fingertip transmittance sensor; optical components (660 nm red, 910 nm infrared); materials include polyurethane, cross-linked polyester foam, polyethylene, and polyester with adhesive. Compatible with Nonin-branded pulse oximeters. Non-sterile. Operating temperature 0 to +40°C.
Indications for Use
Indicated for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients weighing >30 kg (60 lbs), regardless of perfusion status. Intended for use in clinical and home environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
K093306
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# 510(k) Summary
| Submitter: | Nonin Medical, Inc. | MAR - 4 2010 |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Contact Person: | Lori M. Roth<br>Clinical/Regulatory Specialist<br>Nonin Medical, Inc.<br>13700 1st Ave. North<br>Plymouth, MN 55441-5443 | |
| Date Prepared: | October 22, 2009 | |
| Trade Name: | Oximeter Sensor | |
| Classification Name:<br>and Number: | Class II, 21 CFR 870.2700 | |
| Product Code: | 74 DQA | |
| Predicate Device(s): | Nonin's 7000 sensor as cleared in the following 510(K)<br>submissions: Model 7500 (K07128 cleared on July 12, 2007),<br>Model LS1-9R LifeSense (K063752 cleared on May 4, 2007), Model<br>9600 (K023044 cleared on July 23, 2003), Model 2500A (K050056<br>cleared on June 21, 2005), and Model 2500 (K002690 cleared on<br>October 11, 2000). | |
| Device Description: | The 6500MA (wrap-around model) and 6500SA (interlocking<br>model) are fingertip single-patient use disposable, transmittance<br>sensors. They are comprised of a lamination of two foams<br>(patient contact side and external side) with the optical<br>components and a malleable wire within the lamination. The<br>optical components are identical to the currently marketed Model<br>7000 single-patient use disposable sensor. The sensors are<br>compatible with all Nonin-branded pulse oximeters. | |
| Intended Use: | Nonin's Models 6500MA and 6500SA Single-Patient Use<br>Disposable Pulse Oximeter Sensors are indicated for non-invasive<br>spot-checking and/or continuous monitoring of adult and<br>pediatric patients who are well or poorly perfused, weighing<br>greater than 60 pounds (30 kilograms). It is intended for use in<br>environments including operating room, surgical recovery, critical<br>care, emergency room, long-term care, home use and mobile<br>environments. | |
{1}------------------------------------------------
### Functional and Safety Testing:
Nonin's Model 6500 sensor series have successfully undergone both bench and clinical testing in order to demonstrate that it meets the requirements of ISO 9919:2005 Clause 50 Accuracy of Operating Data, Clause 102 section 102.2 Labeling, and IEC 60601-1:1998 (ISO 10993-1:2003) Clause 48 Biocompatibility.
### Substantial Equivalence:
| | Predicate Device<br>Model 7000 Sensor Series | Subject Device<br>Model 6500 Sensor Series |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Patient Population: | Adult/Pediatric (weighing > 30<br>kilograms) | Same |
| Sensor Application Site: | Fingers | Same |
| Patient Use/Reuse: | Disposable | Same |
| Sterility: | Non-sterile | Same |
| Measurement Technique: | Fingertip transmittance sensor | Same |
| SpO2 Accuracy (Arms) (70-<br>100%): | $\pm$ 3 digits | $\pm$ 2 digits |
| SpO2 Low Perfusion Accuracy<br>(Arms) (70-100%): | $\pm$ 3 digits | $\pm$ 2 digits |
| Pulse Rate Accuracy (Arms)<br>(18-300 BPM): | $\pm$ 3 digits | Same |
| Low Perfusion Pulse Rate<br>Accuracy (Arms) (40-240<br>BPM): | $\pm$ 3 digits | Same |
| Red:<br>Infrared: | 660 nm @ 0.8 mW maximum<br>average power<br>910 nm @ 1.2 mW maximum<br>average power | Same |
| Operating: | 0 o to +40o C (32o F to 104o F) | Same |
| Storage/Transportation: | -30o to +50o C (-22o F to 122o F) | -30o to +70o C (-22o F to 158o F) |
| Operating:<br>Storage/Transportation: | 10 to 90% non-condensing<br>10 to 95% non-condensing | Same |
| Sensor Housing: | Microfoam | Polyurethane, cross-linked<br>polyester foam,<br>polyethylene/polyurethane,<br>polyester with adhesive |
#### Conclusion:
Nonin's Model 6500 sensor series is substantially equivalent to Nonin's Model 7000 sensors when used with Nonin-branded Pulse Oximeters monitors.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The words are arranged on a single line and are centered horizontally.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Lori Roth Clinical/Regulatory Specialist Nonin Medical, Incorporated 13700 15t Avenue North Plymouth, Minnesota 55441
MAR - 4 2010
Re: K093306
Trade/Device Name: Model 6500 Sensor Series Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: February 1, 2010 Received: February 4, 2010
Dear Ms. Roth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 – Ms. Roth
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
far
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{4}------------------------------------------------
## Indications for Use Statement
| 510(k) Number (if known) | K093306 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Nonin Medical, Inc. Model 6500 Sensor Series |
| Indications for Use | Nonin's Models 6500MA and 6500SA Single-Patient Use Disposable Pulse Oximeter Sensors are indicated for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients who are well or poorly perfused, weighing greater than 60 pounds (30 kilograms). It is intended for use in environments including operating room, surgical recovery, critical care, emergency room, long-term care, home use and mobile environments. |
Prescription Use _ X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultetus
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.