EXTENSION SET (STANDARDBORE TUBING), EXTENSION SET (MINIBORE TUBING), IV ADMINISTRATION SET ( STANDARD
K093271 · Victus, Inc. · FPA · Jan 27, 2010 · General Hospital
Device Facts
Record ID
K093271
Device Name
EXTENSION SET (STANDARDBORE TUBING), EXTENSION SET (MINIBORE TUBING), IV ADMINISTRATION SET ( STANDARD
Applicant
Victus, Inc.
Product Code
FPA · General Hospital
Decision Date
Jan 27, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To administer IV fluids/medication to a patient's vascular system.
Device Story
Single-use, sterile, non-pyrogenic intravascular administration and extension sets; used to deliver IV fluids or nutritional solutions to patient vascular systems via catheter venous sites. Gravity-controlled flow; components include non-DEHP PVC tubing, flow regulators (GVS SPA), filters (0.22µ or 1.2µ), check valves, and needle-free injection sites/connectors. Needle-free components reduce needlestick injury risk. Operated by healthcare providers in clinical settings. Device acts as a conduit between fluid source and patient; flow regulator manages infusion rate; filter removes particulates from solution. Extension sets function as accessories to other administration sets.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing was performed to verify the performance of the administration and extension sets.
Technological Characteristics
Single-use, sterile, non-pyrogenic IV administration/extension sets. Materials: non-DEHP PVC tubing. Components: 0.22µ or 1.2µ filters, flow regulators, check valves, needle-free connectors. Gravity-fed flow. Sterilization method not specified. No software or electronic components.
Indications for Use
Indicated for the administration of intravenous fluids or medications to a patient's vascular system via a catheter venous site under gravity-controlled flow. Suitable for patients requiring IV therapy or nutritional solutions.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
K093271
| 5. 510(K) Summary or 510(K) Statement | | | | | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---------------------|--------------------------------------|----------------------------------|
| Submitted by: | Ileana Yanes | | | | |
| | Victus, Inc. | | | | |
| | 4918 S.W. 74 Court | | | | |
| | Miami, FL 33155 | | | | |
| | Tel: (305) 663 - 2129 ext. 102 | | | | |
| | Fax: (305) 663 - 1843 | | | | |
| Date of Summary: | October 13, 2009 | | | | |
| Device Name | (1) Victus Intravascular (I. V.) Administration Sets (with 0.22 µ Filter and/or Flow Regulator) and (2) Victus I.V. Extension Sets (with 0.22 µ Filter and/or Flow Regulator) | | | | |
| Common Name | Intravascular Administration Set | | | | |
| Classification Name | <table><tr><th>Regulation Number</th><th>Classification Name</th></tr><tr><td>21 C.F.R §880.5440<br/>ProCode 80 FPA</td><td>Intravascular Administration Set</td></tr></table> | Regulation Number | Classification Name | 21 C.F.R §880.5440<br>ProCode 80 FPA | Intravascular Administration Set |
| Regulation Number | Classification Name | | | | |
| 21 C.F.R §880.5440<br>ProCode 80 FPA | Intravascular Administration Set | | | | |
| Predicate Devices | Victus IV Administration Sets (K023469; K030246)<br>Millipore/NP Medical IVEX Filter (K832952)*<br>GVS SPA Flow Regulator (K061115)*<br>Halkey-Roberts (Robertsite Needleless Injection Site: (K002689)*<br>Leventon S.A. Administration Set and Extension Set with Flow Regulator with and without "Y" Connector or Needleless Access Connector (K952810)<br>Churchill Medical Systems Extension Set with 0.22μ & 1.2μ filter (K884370), (K894954), and (K021395), and Administration Set (K896333) and (K011336)<br>ICU Medical, Inc. Administration Set and Extension Set with Needle-Free straight connector and Needle-Free Injection Site (K964435) and (K970855)<br>Hospira, Inc. Microbore Extension Set (K052722)<br>McGAW, Inc. (B.Braun Medical) Administration Set (K932165)<br>Kendall McGAW Laboratories, Inc. Administration Set with 1.2µ filter (K896292)<br>AMSINO INTL., INC. Administration Set with Flow regulator (K973107)<br>*Note: Components manufactured by these companies will be used in the manufacture of the VICTUS I.V. Administration Sets and VICTUS I.V. Extension Sets. | | | | |
JAN 2 7 2010
•
and the state of the state Carlos Concession
{1}------------------------------------------------
tended
This submission provides information on the following chances and additions to its IV Administration Set product line:
(1) substitutes a Flow Regulator (GVS SPA; K061115) for the Roller Clamp component in this system for specific applications (Note: The Roller Clamp will continue to be used in some versions of this system);
(2) replaces the PVC tubing (non DEHP) with PVC tubing (non-DEHP) in a systems; and
(3) substitutes Halkey-Roberts Needle-Free connector for the CLAVE® connector: and
(4) adds a Filter (typically 0.22u) to some configurations of the IV Administration Set and Extension Sets; (other filter sizes such as 1.2u may be used); and
(5) adds check valve in some configurations.
In addition, the I. V. Extension Set (with Minibore tubing with/without Filter Flow Regulator, Needle-Free straight connector and Standard (latex free) or Needle-Free Injection Site) will be marketed as an accessory for use with other I. V. Administration Sets.
None of the additions/substitutions to the device design affect safety and effectiveness of the Victus IV Administration Sets and Extension Sets.
The Victus I.V. Administration and Extension Sets are single use, sterile, non-pyrogenic devices used to administer filtered intravenous solutions and/or nutritional solution to a patient's vascular system via a catheter venous site under gravity-controlled flow. When used, the purpose of the filter is for filtering solution during intravascular administration. The device may include a needle-free valve injection site/connector which eliminates the use of needle to access the set during IV administration and aids in the prevention of needlestick injuries.
To administer IV fluids/medication to a patient's vascular system.
The device may include a needle-free valve injection site/connector which eliminates the use of needles to access the set during IV administration and aids in the prevention of needlestick injuries.
When included, the purpose of the filter is for filtering solution (I.V. Solution and/or Nutritional fluid) during intravascular administration.
{2}------------------------------------------------
The IV Extension Sets & IV Sets have the same technological characteristics As the legally marketed predicate Churchill Medical Systems Extension Sets with 0.22u and 1.2u filter (K884370), (K894954) and Administration Set (K896333) and (K011336), Hospira, Inc. Microbore Extension Set (K052722), Leventon S.A. Administration Set and Extension Set with Flow Regulator (K952810), McGAW, Inc. Administration Set with 1.2u filter (K932165), Kendall McGAW Laboratories, Inc. Administration Set with 1.2u filter (K896292), ICU Medical Extension Sets with Needle-Free straight connector (K964435) and (K970855) and AMSINO INTL, Inc. Administration Set with Flow Regulator (K973107). This system incorporates PVC tubing that does not contain DEHP. The Flow Regulator that will be incorporated will be supplied by GVS SPA and is cleared under K061115.
The Technological characteristics of the Victus Sets are substantiallyequivalent to the referenced predicates.
The VICTUS IV Extension Set and IV Administration Set have undergone Mechanical testing to verify performance.
## Contract Manufacturer and Contract Sterilizer:
Plásticos y Materias Primas S.A. de C.V. (PyMPSA),
Juan de la Barrera No. 3609-2
Alamo Industrial 44700 Guadalajara, Jal. Mexico.
Establishment Registration Number: 9710643
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure embracing a globe, composed of three curved lines.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Ileana Yanes Manager of Regulatory Affairs and Quality Assurance Victus, Incorporated 4918 Southwest 74th Court Miami, Florida 33155
JAN 2 7 2010
Re: K093271
Trade/Device Name: VICTUS Intravascular (IV) Administration Sets VICTUS Intravascular (IV) Extension Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA
Dated: January 19, 2010
Received: January 21, 2010
Dear Ms. Yanes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Yanes
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); ' and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K093271
Device Name: YICTUS Intravascular (IV) Administration Sets and
VICTUS Intravascular (IV) Extension Sets
Indications for Use:
To administer IV fluids/medication to a patient's vascular system.
Prescription Use _ X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) .
Concurrence of CDRH, Office of Device Evaluation (ODE)
Charles B
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K09327/
Page __ of __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.