K093251 · Omniguide, Inc. · GEX · Oct 29, 2009 · General, Plastic Surgery
Device Facts
Record ID
K093251
Device Name
OMNIGUIDE BEAMPATH FELS 25A CO2 LASER SYSTEM
Applicant
Omniguide, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 29, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
The OmniGuide BeamPath® FELS 25A, CO2 Laser System intended to be used in conjunction with the OmniGuide BeamPath Adapter and the OmniGuide BeamPath fiber and is intended to be used for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues.
Device Story
The OmniGuide BeamPath FELS 25A is a table-top CO2 laser system emitting 10.6 um wavelength light at up to 30 watts continuous wave. It utilizes an RF-excited laser tube. Laser energy is delivered to the surgical site via an adapter and hollow waveguide fiber. The device is operated by a physician in a surgical setting. The CO2 laser energy is strongly absorbed by water in biological tissue, providing high cutting precision (0.1-0.2 mm depth) and coagulation capabilities. The console features a screen for parameter display and control. The system is used for soft tissue surgery across multiple specialties. It benefits patients by providing precise tissue ablation and coagulation with minimal thermal damage to adjacent structures.
Clinical Evidence
No clinical data. Bench testing only, including evaluation of laser power output and beam quality, confirming compliance with FDA Guidance on the Content and Organization of a Premarket Notification for a Medical Laser (1995).
Technological Characteristics
30W max power CO2 laser; 10.6 um wavelength; RF-excited laser tube. Materials: machined aluminum, stainless steel, standard optics. Delivery: hollow waveguide fiber via adapter. Form factor: table-top console. Connectivity: none. Sterilization: not specified (passive components).
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues in dermatology, general surgery, gynecology, head & neck surgery, neurosurgery, oral surgery, orthopedic surgery, otorhinolaryngology, pediatric surgery, plastic & reconstructive surgery, podiatry, and urology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
OmniGuide
# ん093251
## OCT 2 9 2009
### 510(k) Summary
| Submitter: | OmniGuide, Inc.<br>One Kendall Square, Building 100 3rd Floor<br>Cambridge, MA 02139 |
|-------------------------------|-----------------------------------------------------------------------------------------|
| Contact Person:<br>Telephone: | Douglas W. Woodruff<br>617-551-8404 |
| Fax: | 617-551-8445 |
| Proprietary Name: | OmniGuide BeamPath® FELS 25A, CO2 Laser System |
| Common Name: | CO2 Laser Powered Surgical Instrument |
| Classification: | 878.4810 |
| Product Code: | GEX |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology. |
Substantial Equivalence Claimed To:
K063698 C-Las CO2 Laser System
{1}------------------------------------------------
#### Description:
The OmniGuide BeamPath FELS 25A, CO2 Laser System, is a Carbon dioxide (CO2) Laser emitting light at a wavelength of 10.6 um. It has an RF excited laser tube which produces a maximum continuous wave (CW) output power of up to 30 watts. Laser energy (CO2), with a wavelength of 10.6 um, has a variety of attractive features for surgical use. Because it is strongly absorbed by water (the main constituent of biological tissue) it does not penetrate into tissue beyond the point of application of the laser energy. This small depth of penetration (0.1-0.2 mm) leads to high cutting precision and minimal thermal damage to adjacent tissue. In addition, the CO2 wavelength has excellent coagulation capabilities
The laser console is a table top console which can be easily transported from one surgery room to the other. The controls for the physician are part of the console and the parameters are clearly displayed on a large screen display. The laser light is guided through different mirrors and focused via an attached adapter is the interface allowing an OmniGuide BeamPath Fiber or equivalent to attach to the laser.
Materials used are mainly machined aluminum or stainless Steel and standard optics for the transmission or the reflection of the CO2 laser wave length. Only the laser energy from the OmniGuide BeamPath FELS 25A, CO2 Laser System comes in contact with the patient. The energy is guided through a hollow waveguide such as the OmniGuide BeamPath hollow waveguide liber. It is worth noting that all components (mirrors, lenses and the fiber) that the CO2 light travels though are passive and do not alter the wayelength or any other of the fundamental properties or beneficial attributes of the CO2 Laser.
#### Indications for Use:
The OmniGuide BeamPath FELS 25A, CO2 Laser System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues in the following specialties:
- . Dermatology
- General Surgery .
- Gynecology .
- Head & Neck Surgery .
- Neurosurgery .
- Oral Surgery ●
- Orthopedic Surgery .
- Otorhinolaryngology .
- Pediatric Surgery ●
- Plastic & Reconstructive Surgery
- Podiatry .
{2}------------------------------------------------
- · Urology
#### Intended Use:
The OmniGuide BeamPath® FELS 25A, CO2 Laser System intended to be used in conjunction with the OmniGuide BeamPath Adapter and the OmniGuide BeamPath fiber and is intended to be used for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues.
#### Summary of Technological Characteristics:
The device is a 30 watt maximum power CO2 laser, emitting wavelength of 10.6 um. The System incorporates an adapter that allows the connection to CO2 waveguide fibers such as the OmniGuide BeamPath WaveGuide Fiber, to transmit the laser encrgy to the surgical site.
#### Performance Data:
Non-clinical Performance Data: The OmniGuide BeamPath® FELS 25A, CO2 Laser System performance characteristics have been cyaluated through testing and analysis of laser power output and beam quality. This type of testing complies with the respective section of the FDA Guidance on the Content and Organization of a Premarket Notification for a Medical Laser (1995) and is similar to the predicate device tests. The performance of OmniGuide BeamPath® PELS 25A, CO2 Laser System and related parameters of the predicate device are the same.
Clinical Performance Data: Formal clinical trials were not deemed necessary as the device is using the same technology and intended use as predicate device.
Conclusions Drawn from Tests and Analysis: The intended use and major performance parameters of the OmniGuide BeamPath® FELS 25A, CO2 Lascr System are equivalent to the characteristics of above mentioned legally marketed device.
{3}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
OmniGuide, Inc. % Regulatory Technology Services, LLC Mr. Mark Job 1394 25th Street, Northwest Buffalo, Minnesota 55313
OCT 2 9 2009
Re: K093251
Trade/Device Name: OmniGuide BeamPath® FELS 25A, CO2 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: October 15, 2009 Received: October 16, 2009
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Mark Job
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
11 093251 Not vet assigned 510(k) Number (if known):
OmniGuide BeamPath® FELS 25A, CO2 Laser System Device Name:
Indications for Use:
The OmniGuide BeamPath FELS 25A, CO2 Laser System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues in the following specialties:
- . Dermatology
- General Surgery .
- Gynecology .
- Head & Neck Surgery .
- . Neurosurgery
- Oral Surgery .
- Orthopedic Surgery �
- Otorhinolaryngology .
- Pediatric Surgery ●
- Plastic & Reconstructive Surgery .
- Podiatry .
- Urology .
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nil R.P. Gyle- fir mxm
Page 1 of __ 1_
(Division Sign-Of Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K093251
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.