K093208 · Discus Dental, LLC · KIF · Feb 12, 2010 · Dental
Device Facts
Record ID
K093208
Device Name
SIMPLISEAL
Applicant
Discus Dental, LLC
Product Code
KIF · Dental
Decision Date
Feb 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Indications for Use
SimpliSeal is a root canal sealer for the permanent sealing of root canals following established endodontic procedures and may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points). SimpliSeal is intended for use by qualified healthcare personnel trained in its use.
Device Story
SimpliSeal is an epoxy-based resin root canal sealer; delivered via dual-barrel syringe; used by dentists/endodontists in clinical settings. Device functions as a sealant for root canals; applied in conjunction with gutta-percha points during endodontic procedures. Provides permanent sealing of root canal space; aids in successful obturation. Benefits patient by ensuring effective root canal seal; preventing microleakage. No complex electronics or software involved.
Clinical Evidence
Bench testing only. Testing included physical property evaluation per ISO 6876 (setting time, flow, film thickness, solubility, dimensional change) and biocompatibility testing per ISO 10993 (cytotoxicity, acute systemic toxicity, irritation, sensitivity, genotoxicity). All results were within specification.
Technological Characteristics
Epoxy-based resin root canal sealer. Premixed, two-part paste in a dual-barrel plastic syringe. Physical properties comply with ISO 6876. Biocompatibility complies with ISO 10993 standards. No electronic components, software, or energy sources.
Indications for Use
Indicated for permanent sealing of root canals following established endodontic procedures in patients requiring root canal therapy; used in conjunction with auxiliary materials like gutta-percha points. Intended for use by trained healthcare personnel.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{0}------------------------------------------------
K 69.3208
# FEB 1 2 2010
, 1
# 510(k) Summary - SimpliSeal
- Contact Person: 1.
B.
Clark D. vonAhsen Associate, Regulatory Affairs Discus Dental, LLC 8550 Higuera Street Culver City, CA 90232 310.845.8216 - phone 310.845.8647 - fax clarkv(@discusdental.com
May 28, 2009 Date of Summary Preparation:
- 2. Name of Medical Device
| Proprietary Name: | SimpliSeal |
|----------------------|---------------------------------|
| Common/Usual Name: | Root Canal Sealer |
| Classification Name: | Resin, Root Canal Filling (KIF) |
- Substantial Equivalence Determination: 3.
Discus Dental, LLC believes that SimpliSeal is substantially equivalent to the following commercially marketed products:
| Predicate Device | Company | 510(k) No. |
|------------------|----------------------|------------|
| Adseal | Meta BioMed Co., LTD | K042769 |
| AH-Plus | Dentsply Intl. | K960548 |
#### Description of Medical Device: 4.
SimpliSeal is a root canal sealer delivered in a dual-barrel syringe. This epoxy based resin sealer is easy to use with gutta-percha points and can also be used with established and more recent root canal sealing techniques. SimpliSeal with colaonent obturation of root canals when combined with obturation points.
{1}------------------------------------------------
- 5. Intended Use:
SimpliSeal is a root canal sealer for the permanent sealing of root canals following established endodontic procedures and may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points).
SimpliSeal is intended for use by qualified healthcare personnel trained in its use.
- Discus Dental, LLC believes that SimpliSeal is substantially equivalent to the 6. following commercially marketed products:
| | Discus Dental<br>--- | Meta BioMed Co.,<br>LTD<br>--- | Dentsply Intl.<br>--- |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SimpliSeal | Adseal | AH-Plus |
| Labeling | Permanent root canal<br>sealer | Permanent root<br>canal sealer | Permanent root<br>canal sealer |
| Intended Use | SimpliSeal is a root<br>canal sealer for the<br>permanent sealing of<br>root canals following<br>established<br>endodontic<br>procedures and may<br>be used in<br>conjunction with the<br>auxiliary materials in<br>the root canal (i.e.<br>gutta percha points).<br>Intended for use by<br>qualified healthcare<br>personnel trained in<br>its use. | For permanent<br>sealing of root<br>canals following<br>established<br>endodontic<br>procedures and<br>may be used in<br>conjunction with<br>the auxiliary<br>materials in the<br>root canal (i.e.<br>gutta-percha<br>points).<br>Intended for use by<br>qualified<br>healthcare<br>personnel trained<br>in its use. | AH PLUS Root<br>Canal Sealer is<br>used for permanent<br>sealing of root<br>canals following<br>established<br>endodontic<br>procedures.<br>Intended for use by<br>qualified<br>healthcare<br>personnel trained<br>in its use. |
| Similar Physical<br>Properties | ISO 6876 fluidity,<br>working time, film<br>thickness,<br>radiopacity, solubility<br>& disintegration | ISO 6876 fluidity,<br>working time, film<br>thickness,<br>radiopacity,<br>solubility &<br>disintegration | ISO 6876 fluidity,<br>working time, film<br>thickness.<br>radiopacity,<br>solubility &<br>disintegration |
| Design,<br>Construction,<br>Components | Premixed, two part<br>paste, packaged in<br>tow component<br>plastic syringe ready<br>to be dispensed and<br>mixed. | Premixed, two part<br>paste, packaged in<br>tow component<br>plastic syringe<br>ready to be<br>dispensed and<br>mixed. | Premixed, two part<br>paste, packaged in<br>tow component<br>plastic syringe<br>ready to be<br>dispensed and<br>mixed. |
{2}------------------------------------------------
The following physical tests have been completed on SimpliSeal:
Setting time - per ISO 6876 Flow - per ISO 6876 Film thickness – per ISO 6876 Solubility -- per ISO 6876 Dimensional change following setting - per ISO 6876
. The following biocompatibility tests have been completed on SimpliSeal:
Cytotoxicity - ISO 10993-5 Acute Systemic Toxicity (oral) - ISO 10993-11 Irritation - ISO 10993-10 . Sensitivity - ISO 10993-10 Genotoxicity - ISO 10993-3
All testing is within specification and the device performs as designed. The results demonstrate that the device is safe, effective and performs as well as or better than products that are already legally marketed.
### END OF 510(k) SUMMARY
:
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract shapes that resemble birds in flight or flowing ribbons.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB 1 2 2010
Mr. Clark D. Von Ashsen Associate, Regulatory Affairs Discus Dental, L.L.C. 8550 Higuera Street Culver City, California 90232
Re: K093208
Trade/Device Name: SimpliSeal Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: II Product Code: KIF Dated: January 21, 2010 Received: January 22, 2010
Dear Mr. Von Ashen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 -- Mr. Von Ashen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
hm far
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and Radiological Health
### Enclosure
{5}------------------------------------------------
#### E. Indications for Use Statement
510(k) Number (if known): _K0643208
Device Name: SimpliSeal
Indications for Use:
SimpliSeal is a root canal sealer for the permanent sealing of root canals following established endodontic procedures and may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points).
SimpliSeal is intended for use by qualified healthcare personnel trained in its use.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR :
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
\underline{RSBetz Das for Dr. K. P. Mully}
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093208
Page | of |
D - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.