← Product Code [ORW](/productcode/ORW) · K093189

# ACTIPROTECT UC N95 RESPIRATOR (K093189)

_Glaxosmithkline Consumer Healthcare · ORW · Jul 2, 2010 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K093189

## Device Facts

- **Applicant:** Glaxosmithkline Consumer Healthcare
- **Product Code:** [ORW](/productcode/ORW.md)
- **Decision Date:** Jul 2, 2010
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 880.6260
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

Actiprotect™ UC N95 Respirator is a single use, disposable respirator coated with Virucoat™ on the outer layer (active ingredient: citric acid, 1.8%, a pH lowering agent) and is not an antiviral drug. Actiprotect kills (inactivates) 99.99% of Influenza A viruses (tested against Influenza A subtypes H1N1 (including 2009 pandemic strain)) within one minute of contact with the surface of the respirator. In vitro (laboratory) tests have demonstrated 99.99% kill (inactivation) on the surface of the outer layer of the respirator when tested in vitro against the following influenza A viruses (tested against Influenza A subtypes (and strains): H1N1(JPN/35/2007, JPN/36/2007, including 2009 pandemic strain: NYMC X-179A), H2N2 (A2/JP/305/57), H3N2 (Hong Kong 8/68, JPN/12/2007) and including bird flu subtypes: H5N1 (VNH5N1-PR8/CDCRG), H5N9 (Turkey A/Wisc/68, Myna A/Mass/71), and Influenza B strains (JPN/85/2007, JPN/128/2007, JPN/143/2007)) under tested contact conditions. No clinical studies have been conducted comparing the ability of an uncoated N95 respirator and this coated N95 respirator to protect the wearer from influenza infection. Actiprotect™ is intended for use by the general public in public health medical emergencies. Effectiveness tested against specific Influenza A and Influenza A pandemic strains. It is intended to cover the nose and mouth of the wearer to help reduce wearer exposure to pathogenic biological airborne particulates. Actiprotect™ UC N95 Respirator also helps to protect the wearer from splash and spray of body fluids.

## Device Story

Single-use, flat-fold N95 filtering facepiece respirator; consists of multiple non-woven fabric layers. Outer layer coated with Virucoat™ (1.8% citric acid) to inactivate 99.99% of specific Influenza A and B viruses upon contact. Filtration achieved via mechanical entrapment and electrostatic surface attraction. Secured to user's face via two synthetic elastic straps. Intended for general public use during public health medical emergencies. Provides physical barrier against airborne particulates and body fluid splashes. No clinical studies confirm superior protection against infection compared to uncoated respirators.

## Clinical Evidence

No clinical efficacy studies conducted. Safety evidence limited to Human Repeated Insult Patch Tests (HRIPT) on respirator materials, which showed no evidence of sensitization or irritation. Efficacy claims based on in vitro laboratory testing demonstrating 99.99% inactivation of specific Influenza A and B strains within one minute of contact.

## Technological Characteristics

Flat-fold, multi-layer non-woven fabric respirator. Outer layer treated with 1.8% citric acid (Virucoat™). Filtration via mechanical entrapment and electrostatic forces. Secured with two synthetic elastic straps. Tested to ASTM F1862 (fluid resistance), NIOSH 42 CFR 84.181 (particulate filtration), MIL M36954C/ASTM F2101 (bacterial filtration), 16 CFR 1610 (flammability), and NIOSH 42 CFR 84.180 (breathing resistance).

## Regulatory Identification

A filtering facepiece respirator for use by the general public in public health medical emergencies is a device that is a disposable half-facepiece non-powered air-purifying particulate respirator intended for use to cover the nose and mouth of the wearer to help reduce wearer exposure to pathogenic biological airborne particulates during a public health medical emergency. The device is made of polymeric materials and is intended to fit closely to the face and to function by filtering particulate material.

## Special Controls

*Classification.* Class II (special controls). The special controls are:(1) Certification by the National Institute for Occupational Safety and Health (NIOSH) as a non-powered air-purifying particulate respirator with a minimum filtration efficiency classification of N95, in accordance with 42 CFR part 84.
(2) The FDA guidance document entitled: “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Filtering Facepiece Respirator for use by the General Public in Public Health Medical Emergencies.” See § 880.1(e) for information on obtaining a copy of this guidance document.

## Predicate Devices

- Actiprotect™ UF N95 Respirator ([K081923](/device/K081923.md))
- 3M 8670F Respirator for Use by the General Public in Public Health Medical Emergencies ([K062070](/device/K062070.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUL - 2 2010

#### 510(k) Summary 6.

#### 6.1 Applicant and Correspondent

| Name:           | GlaxoSmithKline Consumer Healthcare                               |
|-----------------|-------------------------------------------------------------------|
| Address:        | 1500 Littleton Road<br>Parsippany, NJ 07054-3884                  |
| Contact Person: | Zinatara A. Manji, M.S., Pharm.D.<br>Director, Regulatory Affairs |
| Phone:          | (973)-889-2322                                                    |
| Fax:            | (973)-889-2501                                                    |
| E-mail:         | zinatara.a.manji@gsk.com                                          |

Date of Preparation: July 1, 2010

#### 6.2 Manufacturer:

Sperian Protection Armor SAS ZI de la Gare 22940 Plaintel France

#### 6.3 Name of Device:

| Trade/Proprietary/Model Name: | Actiprotect™ UC N95 Respirator<br>Double Strap, Flat fold                                                                                 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:                  | Filtering Facepiece Respirator                                                                                                            |
| Classification Name:          | Filtering Facepiece Respirator for Use by<br>General Public in Public Health Medical<br>Emergencies (21 CFR 880.6260) Product Code<br>ORW |

{1}------------------------------------------------

#### Devices to Which New Device is Substantially Equivalent: 6.4

Actiprotect™ UF N95 Respirator (K081923), GlaxoSmithKline Consumer Healthcare

3M 8670F Respirator for Use by the General Public in Public Health Medical Emergencies (K062070), 3M Health Care

#### 6.5 Device Description:

Actiprotect™ UC N95 Respirator is a double strap, flat-fold style filtering facepiece respirator. It consists of multiple layers of non-woven fabric. A filtration layer provides a barrier to particles of various sizes through mechanical entrapment in a tortuous pathway and by surface attraction by electrostatic forces. The outer layer contains a coating to capture and inactivate viruses. The layers of non-woven fabric are ultrasonically sealed together at the outer perimeter. Two synthetic elastic straps are stapled or welded to the edge of the respirator and are used to secure the respirator to the user's face.

#### Statement of Intended Use 6.6

Actiprotect™ UC N95 Respirator is a single use, disposable respirator coated with Virucoat™ on the outer layer (active ingredient: citric acid, 1.8%, a pH lowering agent) and is not an antiviral drug. Actiprotect kills (inactivates) 99.99% of Influenza A viruses (tested against Influenza A subtypes H1N1 (including 2009 pandemic strain)) within one minute of contact with the surface of the respirator. In vitro (laboratory) tests have demonstrated 99.99% kill (inactivation) on the surface of the outer layer of the respirator when tested in vitro against the following influenza A viruses (tested against Influenza A subtypes (and strains): H1N1(JPN/35/2007, JPN/36/2007, including 2009 pandemic strain: NYMC X-179A), H2N2 (A2/JP/305/57), H3N2 (Hong Kong 8/68, JPN/12/2007) and including bird flu subtypes: H5N1 (VNH5N1-PR8/CDCRG), H5N9 (Turkey A/Wisc/68, Myna A/Mass/71), and Influenza B strains (JPN/85/2007, JPN/128/2007, JPN/143/2007)) under tested contact conditions. No clinical studies have been conducted comparing the ability of an uncoated N95 respirator and this coated N95 respirator to protect the wearer from influenza infection. Actiprotect™ is intended for use by the general public in public health medical emergencies. Effectiveness tested against specific Influenza A and Influenza A pandemic strains. It is intended to cover the nose and mouth of the wearer to help reduce wearer exposure to pathogenic biological airborne 

{2}------------------------------------------------

particulates. Actiprotect™ UC N95 Respirator also helps to protect the wearer from splash and spray of body fluids.

#### Summary of the Technological Characteristics 6.7

Actiprotect™ UC N95 Respirator has the same intended use/indications as the predicate devices, Actiprotect™ UF N95 Respirator and 3M 6870F Respirator for Use by the General Public in Public Health Medical Emergencies. Actiprotect™ UF N95 Respirator offers the protection of a coating on its outermost surface to capture and inactivate influenza viruses on contact with the surface. Actiprotect™ UC N95 Respirator is intended for use by the general public in public medical health emergencies.

Actiprotect UF N95 Respirator has been tested for and passed standardized tests for fluid penetration resistance, particulate filtration efficiency, bacterial filtration efficiency (BFE), virus filtration efficiency (VFE), flammability, and breathing resistance.

The materials of construction used in Actiprotect UC N95 Respirator are identical to those of the predicate device, Actiprotect UF N95 Respirator. The device as a whole has been demonstrated to be biocompatible by cytotoxicity and human repeated insult patch testing.

### 6.8 Brief discussion of the nonclinical tests submitted. referenced, or relied on in the premarket notification submission for a determination of substantial equivalence:

- . Fluid Penetration Resistance - ASTM F1862
- . Particulate Filtration Efficiency - NIOSH 42 CFR 84.181
- . Bacterial Filtration Efficiency - MIL M36954C, AST M F2101
- . Flammability - 16 CFR 1610
- Breathing Resistance NIOSH 42 CFR 84.180 ●

{3}------------------------------------------------

### Brief discussion of the clinical tests submitted, referenced, or 6.9 relied on in the premarket notification submission for a determination of substantial equivalence

Human Repeated Insult Patch Tests were conducted using samples of the respirator materials held under occlusive patches. There were no adverse reactions reported during the studies. The outcome of these tests revealed no evidence of sensitization or irritation.

# 6.10 Conclusions drawn from the nonclinical and clinical tests

Cytotoxicity and human repeated insult patch tests plus a toxicological review of data on the product's ingredient indicates the product is safe for use in the intended application. Bench testing demonstrated the product's efficacy. Actiprotect™ UC N95 Respirator has been shown to be substantially equivalent to the predicate devices.

{4}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Zinatara Manji Director, Regulatory Affairs Glaxosmithkline Consumer Healthcare 1500 Littleton Road Parsippany, New Jersey 07054-3884

JUL-- 2 2010

Re: K093189

Trade/Device Name: Actiprotect™ UC N95 Respirator, Double Strap Flat Fold Regulation Number: 21 CFR 880.6260 Regulation Name: Filtering Facepiece Respirator for use by the General Public in Public Health Medical Emergencies Regulatory Class: II Product Code: ORW Dated: June 18, 2010 Received: June 21, 2010

Dear Ms. Manii:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Ms. Manji

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

### 510(k) Number (if known): K093189

Device Name: Actiprotect™ UC N95 Respirator, Double Strap Flat Fold For Use by the General Public in Public Health Medical Emergencies

Indications For Use:

Actiprotect™ UC N95 Respirator is a single use, disposable respirator coated with Virucoat™ on the outer layer (active ingredient: citric acid, 1.8%, a pH lowering agent) and is not an antiviral drug. Actiprotect kills (inactivates) 99.99% of Influenza A viruses (tested against Influenza A subtypes H1N1 (including 2009 pandemic strain)) within one minute of contact with the surface of the respirator. In vitro (laboratory) tests have demonstrated 99.99% kill (inactivation) on the surface of the outer laver of the respirator when tested in vitro against the following influenza A viruses (tested against Influenza A subtypes (and strains): H1N1(JPN/35/2007, JPN/36/2007, including 2009 pandemic strain: NYMC X-179A), H2N2 (A2/JP/305/57), H3N2 (Hong Kong 8/68, JPN/12/2007) and including bird flu subtypes: H5N1 (VNH5N1-PR8/CDCRG), H5N9 (Turkey A/Wisc/68, Myna A/Mass/71), and Influenza B strains (JPN/85/2007, JPN/128/2007, JPN/143/2007)) under tested contact conditions. No clinical studies have been conducted comparing the ability of an uncoated N95 respirator and this coated N95 respirator to protect the wearer from influenza infection. Actiprotect™ is intended for use by the general public in public health medical emergencies. Effectiveness tested against specific Influenza A and Influenza A pandemic strains. It is intended to cover the nose and mouth of the wearer to help reduce wearer exposure to pathogenic biological airborne particulates. Actiprotect™ UC N95 Respirator also helps to protect the wearer from splash and spray of body fluids.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital 
Division of Anesthesion Devices Division of Anisanomical Devices

K09318

Doc ID 0900233c80861fca

510(k) Number:

8

---

**Source:** [https://fda-staging.innolitics.com/device/K093189](https://fda-staging.innolitics.com/device/K093189)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).