← Product Code [FRO](/productcode/FRO) · K093188

# AFM ULTRA AG DRESSINGS (K093188)

_Milliken & Co. · FRO · Feb 17, 2010 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K093188

## Device Facts

- **Applicant:** Milliken & Co.
- **Product Code:** [FRO](/productcode/FRO.md)
- **Decision Date:** Feb 17, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

The AFM Ultra Ag Dressings are indicated for management of partial thickness burns, incisions, skin grafts, donor sites, lacerations, and Stage I-IV dermal ulcers (vascular, venous, pressure and diabetic).

## Device Story

AFM Ultra Ag Dressings are sterile, single-use wound dressings for moist wound management. Device consists of 4 layers: occlusive synthetic top layer, polyurethane foam, hotmelt adhesive, and silver-containing knitted composite fabric (silver-coated nylon, lycra, and polyester). Silver provides antimicrobial barrier within dressing. Used in clinical settings for various wound types. Benefits include effective microbial barrier and moist wound environment maintenance. Safety and efficacy supported by biocompatibility and antimicrobial testing.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including biocompatibility testing per ISO recommendations, silver extraction analysis, zone-of-inhibition testing, and antimicrobial barrier efficacy testing.

## Technological Characteristics

4-layer dressing: occlusive synthetic top layer, polyurethane foam, hotmelt adhesive, and silver-coated nylon/lycra/polyester knitted composite fabric. Silver provides antimicrobial barrier. Sterile, single-use. No software or electronic components.

## Predicate Devices

- Milliken Silver Wound Dressing ([K051445](/device/K051445.md))
- Contreet Foam Adhesive/Nonadhesive ([K022416](/device/K022416.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K093188

# Traditional 510(k) Summary for · AFM Ultra Ag Dressings

- 1. SPONSOR
FEB 1 7 2010

Milliken Chemical 920 Milliken Rd. Spartanburg, SC 29303 Contact Person: John D. Bruhnke, Ph.D.

Telephone: 864-503-2844 Date Prepared: February 16, 2010

## 2. DEVICE NAME

Proprietary Name: AFM Ultra Ag Dressings Common/Usual Name: Wound Dressing Classification Name: Dressing, Wound, Drug

## 3. PREDICATE DEVICES

- . Milliken Silver Wound Dressing (K051445)
- Contreet Foam Adhesive/Nonadhesive (K022416) .

#### 4. DEVICE DESCRIPTION

The AFM Ultra Ag Dressings are sterile, single-use wound care dressings for use in moist wound management. The dressings are comprised of 4 layers, each performing a specific function; an occlusive synthetic top layer, a polyurethane foam layer, a hotmelt adhesive and a layer of a silver-containing knitted composite fabric.

#### 5. INTENDED USE

The AFM Ultra Ag Dressings are indicated for management of partial thickness burns, incisions, skin grafts, donor sites, lacerations, and Stage I-IV dermal ulcers (vascular, venous, pressure and diabetic).

## 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The technological characteristics of AFM Ultra Ag Dressings and the predicate products are substantially equivalent in that they are all dressings suitable for use on pressure sores, leg ulcers, post-operative wounds, superficial wounds and abrasions.

{1}------------------------------------------------

K093188
2/2

AFM Ultra Ag Dressings and the predicate devices are also similar in that they all contain silver which provides an effective barrier to microbial activity in the dressing itself. AFM Ultra Ag Dressings are manufactured from a silver coated nylon, lycra and polyester fabric laminated onto a polyurethane foam whereas the predicate Contreet Foam Adhesive/Nonadhesive (K022416) dressing is made from silverimpregnated polyurethane foam, and the Milliken Silver Wound Dressings (K051445) is made from the same silver coated nylon, lycra and polyester fabric. The differences between AFM Ultra Ag Dressings and the predicate devices include construction details and slightly different silver concentrations, which are minor and do not affect safety and effectiveness of the device, as demonstrated by the biocompatibility and efficacy testing.

#### 7. PERFORMANCE TESTING

Biocompatibility testing was performed in accordance with the International Organization for Standardization recommendations. Results of the biocompatibility tests demonstrate that the device is suitable for its intended use. Antimicrobial testing was performed which showed that AFM Ultra Ag Dressings provide an effective microbial barrier to the dressing itself. Milliken believes that the data included in this submission including the technical characteristics, physical properties, silver extraction, zone-of-inhibition and antimicrobial testing demonstrates that AFM Ultra Ag Dressings are substantially equivalent in design, function and intended use to Milliken Silver Wound Dressings (K051445) and Contreet Foam Adhesive/ Nonadhesive (K022416).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three lines representing its wings and a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the circle, with "DEPARTMENT OF HEALTH & HUMAN" at the top and "SERVICES - USA" at the bottom.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Milliken & Company % Medical Device Consultants, Inc. Ms. Mary McNamara-Cullinane, RAC 49 Plain Street North Attleboro, Massachusetts 02760

FEB 1 7 2010

Re: K093188 Trade/Device Name: AFM Ultra Ag Dressings Regulatory Class: Unclassified Product Code: FRO Dated: January 21, 2010 Received: January 25, 2010

Dear Ms. McNamara-Cullinane:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Mary McNamara-Cullinane, RAC

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/McdicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Aabaro Bnulhup

Mark N. Malkerson

Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K093188

# Indications for Use

510(k) Number (if known): K093188

Device Name: AFM Ultra Ag Dressings

Indications for Use:

AFM Ultra Ag Dressings are indicated for management of partial thickness burns, incisions, skin grafts, donor sites, lacerations, and Stage I-1V dermal ulcers (vascular, venous, pressure and diabetic).

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR .

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David Keene for MXUY

Division Sion-C Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K093188.

---

**Source:** [https://fda-staging.innolitics.com/device/K093188](https://fda-staging.innolitics.com/device/K093188)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com