WONDFO ONE STEP BUPRENORPHINE URINE TEST, OXYCODONE URINE TEST, PROPOXPHENE URINE TEST, MULTIPLE DRUG OF ABUSE URINE TES
Applicant
Guangzhou Wondfo Biotech Co., Ltd.
Product Code
DJG · Clinical Toxicology
Decision Date
Aug 25, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3650
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K093175 · Aug 25, 2010
WONDFO ONE STEP BUPRENORPHINE URINE TEST, OXYCODONE URINE TEST, PROPOXPHENE URINE TEST, MULTIPLE DRUG OF ABUSE URINE TES
Guangzhou Wondfo Biotech Co., Ltd.
Clinical urine specimens from Shenzhen Drug Addiction Recovery Center; Clinical urine samples of varying concentrations
Clinical urine samples were used to evaluate the accuracy and performance of the Wondfo drug test strips by comparing results against GC-MS confirmation.
Clinical urine specimens; Method comparison; Accuracy study
Clinical urine samples from a drug addiction recovery center and other clinical sources; Sample Size: 240 total (80 per drug); Number of Sites: 1
GC-MS (Gas Chromatography-Mass Spectrometry)
Percent agreement with GC-MS results
Indications for Use
Wondfo One Step Buprenorphine Urine Test Strip, Wondfo One Step Oxycodone Urine Test Strip, and Wondfo One Step Propoxyphene Urine Test Strip are intended for the qualitative determination of Buprenorphine, Oxycodone, d-Propoxyphene at the specific cut-off concentration in human urine. They are intended for healthcare professionals in a central laboratory setting only and that it is not for use in point-of-care settings.
Device Story
Single-use lateral flow immunochromatographic test strips for qualitative detection of buprenorphine, oxycodone, or propoxyphene in human urine. Operation involves dipping strip into urine sample; competitive binding between drug in sample and drug-labeled conjugate (with chromagen) occurs on strip. Presence of drug in sample inhibits binding to test line, resulting in absence of line (positive result); absence of drug allows binding, forming line (negative result). Includes procedural control line (mouse monoclonal anti-IgG) to verify sample migration and strip integrity. Intended for central laboratory use by healthcare professionals; results read visually by operator. Provides preliminary screening; requires confirmatory testing via GC/MS or LC/MS for clinical decision-making.
Clinical Evidence
Bench testing only. Precision/reproducibility assessed using spiked urine samples across three lots and three operators. Method comparison performed against GC/MS reference method using clinical urine samples. Analytical specificity evaluated against structurally related and unrelated compounds. No clinical data provided.
Technological Characteristics
Lateral flow immunochromatographic assay. Uses mouse monoclonal antibodies for competitive binding. Qualitative visual readout. No instrumentation required. Internal procedural control (mouse monoclonal anti-IgG) included. Storage 4-30°C. Shelf life 24 months.
Indications for Use
Indicated for qualitative detection of buprenorphine (10 ng/mL), oxycodone (100 ng/mL), or d-propoxyphene (300 ng/mL) in human urine. Intended for use by healthcare professionals in central laboratory settings only; not for point-of-care use. Provides preliminary results requiring confirmation by GC/MS or LC/MS.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
Acon BUP One Step Buprenorphine Test Strip and Acon BUP One Step Buprenorphine Test Device (k060466)
Acon BUP One Step Oxycodone Test Strip and Acon BUP One Step Oxycodone Test Device (k033047)
Acon BUP One Step Propoxyphene Test Strip and Acon BUP One Step Propoxyphene Test Device (k040445)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k093175
B. Purpose for Submission:
New device
C. Measurand:
Buprenorphine, Propoxyphene, and Oxycodone in human urine
D. Type of Test:
Qualitative immunochromatographic test for drugs of abuse in human urine
E. Applicant:
Guangzhou Wondfo Biotech Co., Ltd.
F. Proprietary and Established Names:
Wondfo One Step Buprenorphine Urine Test
Wondfo One Step Oxycodone Urine Test
Wondfo One Step Propoxyphene Urine Test
G. Regulatory Information:
1. Regulation section:
| Regulation | Classification | Product Code | Device Name |
| --- | --- | --- | --- |
| 21 CFR §862.3650 Opiate test system | II | DJG | Buprenorphine |
| 21 CFR §862.3650 Opiate test system | II | DJG | Oxycodone |
| 21 CFR §862.3700 Propoxyphene test system | II | JXN | Propoxyphene |
4. Panel:
Toxicology (91)
{1}
H. Intended Use:
1. Intended use(s):
See below.
2. Indication(s) for use:
a. Wondfo One Step Buprenorphine Urine Test
The Wondfo One Step Buprenorphine Urine Test Strip is intended for the qualitative determination of buprenorphine in human urine at the cut-off concentration of 10 ng/ml.
The assay is in strip format. The device is intended for healthcare professionals in a central laboratory setting only and not for use in point-of-care settings. For in vitro diagnostic use.
This assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy (GC/MS) is the recommended confirmatory method.
b. Wondfo One Step Oxycodone Urine Test
The Wondfo One Step Oxycodone Urine Test Strip is intended for the qualitative determination of oxycodone in human urine at the cut-off concentration of 100 ng/ml.
The assay is in strip format. The device is intended for healthcare professionals in a central laboratory setting only and is not for use in point-of-care settings. For in vitro diagnostic use.
This assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy (GC/MS) is the recommended confirmatory method.
c. Wondfo One Step Propoxyphene Urine Test
The Wondfo One Step Propoxyphene Urine Test Strip is intended for the qualitative determination of d-propoxyphene in human urine at the cut-off concentration of 300 ng/ml.
The assay is in strip format. The device is intended for healthcare professionals in a central laboratory setting only and is not for use in point-
{2}
of-care settings. For in vitro diagnostic use.
This assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy (GC/MS) is the recommended confirmatory method.
3. Special conditions for use statement(s):
The devices are for prescription use.
The Wondfo One Step Buprenorphine, Oxycodone, and Propoxyphene Urine Tests provide only preliminary analytical test results. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Liquid chromatography/mass spectrometry (LC/MS) or gas chromatography/mass spectrometry (GC/MS) are the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
These devices are designated for central laboratory use only.
Tests for buprenorphine cannot distinguish between abused drugs and certain prescribed medications.
4. Special instrument requirements:
There are no instrument requirements. The devices are visually read single-use devices.
I. Device Description:
The devices are for use in human urine. The Wondfo Buprenorphine, Oxycodone, and Propoxyphene assays are single-test test strips.
For the test strips, users dip the strip into the urine and the reaction is initiated by movement of the sample through the test strip.
The test strips include a procedural control for each test to indicate that the assay is working correctly. Negative and positive results are determined by the presence or absence of a line in the test portion of the device. Users read the results at a specified time after the addition of the urine sample to the device.
J. Substantial Equivalence Information:
{3}
4
1. Predicate device name(s):
| Predicate Device Name | Predicate 510(k) |
| --- | --- |
| | |
| Acon BUP One Step Buprenorphine Test Strip and Acon BUP One Step Buprenorphine Test Device | k060466 |
| Acon BUP One Step Oxycodone Test Strip and Acon BUP One Step Oxycodone Test Device | k033047 |
| Acon BUP One Step Propoxyphene Test Strip and Acon BUP One Step Propoxyphene Test Device | k040445 |
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicates |
| Intended Use | Same | For the qualitative determination of buprenorphine, oxycodone, or propoxyphene in human urine. |
| Calibrator | Same | Buprenorphine (buprenorphine) Oxycodone (oxycodone) Propoxyphene (propoxyphene) |
| Methodology | Same | Competitive binding, lateral flow immunochromatographic assays based on the principle of antigen antibody immunochemistry. |
| Type Of Test | Same | Immunoassay principles that rely on antigen-antibody interactions to indicate positive or negative result |
| Specimen Type | Same | Human urine |
| Cut Off Values | Same | Buprenorphine (buprenorphine):10 ng/ml Oxycodone (oxycodone): 100 ng/ml Propoxyphene (propoxyphene): 300 ng/ml |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended testing sites | For healthcare professionals in central laboratories | For healthcare professionals in point-of-care sites |
| Configurations | Strip | Strip, Cartridge |
K. Standard/Guidance Document Referenced (if applicable):
None referenced.
{4}
5
L. Test Principle:
Each assay (buprenorphine, oxycodone, propoxyphene) employs lateral flow immunochromatographic technology.
Buprenorphine, oxycodone, or propoxyphene are detected in human urine by competitive binding between mouse monoclonal antibodies to one of its respective drug and drug-labeled conjugate (containing a chromagen). Binding of the drug in the sample causes the absence of a line at the test area, i.e., a positive result. When drug is not present in the sample, the drug-labeled conjugate binds at the test line, resulting in formation of a line, i.e., a negative result. The absence or presence of the line is determined visually by the operator.
The device also has an internal process control (mouse monoclonal anti-IgG) which indicates that an adequate volume of sample has been added and that the immunochromatographic strip is intact.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision was assessed with three lots of Wondfo test strips tested by three users over five consecutive days. Seven separate normal urine samples were spiked with the target drug at concentrations of 0, -50% cutoff, -25% cutoff, cutoff, +25% cutoff, +50% cutoff and +200% cutoff. Target levels were confirmed by GC/MS. All samples were blinded to the users and tested randomly. Each operator performed a total of 75 samples for each concentration for each of the three drugs. Results were as follows:
| Buprenorphine | Buprenorphine concentration (ng/ml) | % of Cutoff | Results #Neg./#Pos. | Agreement |
| --- | --- | --- | --- | --- |
| Lot I: | 0 | Negative | 75/0 | 100% |
| | 5 | -50% | 75/0 | 100% |
| | 7.5 | -25% | 62/13 | 82.7% |
| | 10 | Cutoff | 9/66 | 88% |
| | 12.5 | +25% | 4/71 | 94.7% |
| | 15 | +50% | 0/75 | 100% |
| | 20 | +100% | 0/75 | 100% |
| Lot II: | 0 | Negative | 75/0 | 100% |
| | 5 | -50% | 75/0 | 100% |
{5}
| | 7.5 | -25% | 63/12 | 84% |
| --- | --- | --- | --- | --- |
| | 10 | Cutoff | 8/67 | 89.3% |
| | 12.5 | +25% | 3/72 | 96% |
| | 15 | +50% | 0/75 | 100% |
| | 20 | +100% | 0/75 | 100% |
| Lot III: | 0 | Negative | 75/0 | 100% |
| | 5 | -50% | 75/0 | 100% |
| | 7.5 | -25% | 61/14 | 81.3% |
| | 10 | Cutoff | 9/66 | 88% |
| | 12.5 | +25% | 2/73 | 97.3% |
| | 15 | +50% | 0/75 | 100% |
| | 20 | +100% | 0/75 | 100% |
| Oxycodone | Oxycodone concentration (ng/ml) | % of Cutoff | Results #Neg./#Pos. | Agreement |
| --- | --- | --- | --- | --- |
| Lot I: | 0 | Negative | 75/0 | 100% |
| | 5 | -50% | 75/0 | 100% |
| | 75 | -25% | 63/12 | 84% |
| | 100 | Cutoff | 10/65 | 86.7% |
| | 125 | +25% | 3/72 | 96% |
| | 150 | +50% | 0/75 | 100% |
| | 200 | +100% | 0/75 | 100% |
| Lot II: | 0 | Negative | 75/0 | 100% |
| | 5 | -50% | 75/0 | 100% |
| | 75 | -25% | 64/11 | 85.3% |
| | 100 | Cutoff | 11/64 | 85.3% |
| | 125 | +25% | 4/71 | 94.7% |
| | 150 | +50% | 0/75 | 100% |
| | 200 | +100% | 0/75 | 100% |
| Lot III: | 0 | Negative | 75/0 | 100% |
| | 5 | -50% | 75/0 | 100% |
| | 75 | -25% | 63/12 | 84% |
| | 100 | Cutoff | 9/66 | 88% |
| | 125 | +25% | 2/73 | 97.3% |
| | 150 | +50% | 0/75 | 100% |
| | 200 | +100% | 0/75 | 100% |
{6}
| Propoxyphene | Propoxyphene concentration (ng/ml) | % of Cutoff | Results #Neg./#Pos. | Agreement |
| --- | --- | --- | --- | --- |
| Lot I: | 0 | Negative | 75/0 | 100% |
| | 150 | -50% | 75/0 | 100% |
| | 225 | -25% | 65/10 | 86.7% |
| | 300 | Cutoff | 9/66 | 88% |
| | 375 | +25% | 6/69 | 92% |
| | 450 | +50% | 0/75 | 100% |
| | 600 | +100% | 0/75 | 100% |
| Lot II: | 0 | Negative | 75/0 | 100% |
| | 150 | -50% | 75/0 | 100% |
| | 225 | -25% | 64/11 | 85.3% |
| | 300 | Cutoff | 11/64 | 85.3% |
| | 375 | +25% | 4/71 | 94.7% |
| | 450 | +50% | 0/75 | 100% |
| | 600 | +100% | 0/75 | 100% |
| Lot III: | 0 | Negative | 75/0 | 100% |
| | 150 | -50% | 75/0 | 100% |
| | 225 | -25% | 54/21 | 72% |
| | 300 | Cutoff | 64/11 | 85.3% |
| | 375 | +25% | 9/66 | 88% |
| | 450 | +50% | 0/75 | 100% |
| | 600 | +100% | 0/75 | 100% |
b. Linearity/assay reportable range:
Not applicable - these devices are qualitative only.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Real-time stability, accelerated stability and shipping studies were performed on three lots each of Wondfo One-Step Buprenorphine, Oxycodone and Propoxyphene test strips. Stability data supports the sponsor's recommended storage conditions of 4-30° C for 24 months.
These devices have internal process controls. A colored line appearing in the control region confirms that sufficient sample volume has been applied and that the sample has migrated correctly on the test strip. Users are informed that the test is invalid if a line fails to appear in the control region. External
{7}
controls are not supplied with this device.
d. Detection limit:
Analytical performance of the device around the cutoff is described in the precision section See 1.a. above.
e. Analytical specificity:
**Structurally-related compounds:**
Buprenorphine, Oxycodone and Propoxyphene were each tested in duplicate for structurally related compounds using 3 lots of test strips. Structurally related compounds at 100,000 ng/mL were added to drug-free urine. Any compounds demonstrating interference were serially diluted and tested to determine the lowest concentration that produced a positive result. Cross-reactivity was calculated by dividing the concentration at which the compound yielded a positive result by the designated cut-off concentration and multiplying by 100%. Other drugs or drug metabolites not listed may interfere with Buprenorphine, Oxycodone or Propoxyphene assays and cause false results. The results are summarized below:
Buprenorphine
| Compounds | Response equivalent to cutoff in ng/mL | Cross Reactivity (%) |
| --- | --- | --- |
| Buprenorphine-3-D-Glucuronide | 15 | 66.67 |
| Norbuprenorphine | 20 | 50 |
| Norbuprenorphine-3-D-Glucuronide | 200 | 5 |
Oxycodone
| Compounds | Response equivalent to cutoff in ng/mL | Cross Reactivity (%) |
| --- | --- | --- |
| Dihydrocodeine | 20,000 | 0.5 |
| Codeine | 100,000 | 0.1 |
| Hydromorphone | 100,000 | 0.1 |
| Morphine | >100,000 | Not detected |
| Acetylmorphine | >100,000 | Not detected |
| Buprenorphine | >100,000 | Not detected |
| Ethylmorphine | >100,000 | Not detected |
Propoxyphene
| Compounds | Response equivalent to cutoff in ng/mL | Cross Reactivity (%) |
| --- | --- | --- |
| d-Norpropoxyphene | 300 | 100 |
{8}
9
# Structurally unrelated compounds:
To evaluate the interference with buprenorphine, oxycodone and propoxyphene from non-related compounds, the interference studies were conducted. The compounds were added to either drug-free urine or urine containing buprenorphine, oxycodone or d-propoxyphene with concentrations at 50% below, 25% below, 25% above and 50% above cutoff level at a high target concentration of 100,000 ng/mL. Three lots of each test strip were used in the study. In addition, buprenorphine was evaluated for interference from tramadol. No negative or positive interference was observed for the following compounds:
| Acetophenetidin | Deoxycorticosterone | Meprobamate | Quinine |
| --- | --- | --- | --- |
| Nalidixic acid | Dextromethorphan | Methoxyphenamine | Ranitidine |
| Acetylsalicylic acid | Diclofenac | Nalidixic acid | Salicylic acid |
| Aminopyrine | Diflunisal | Naloxone | Serotonin |
| Amoxicillin | Digoxin | Naltrexone | Sulfamethazine |
| Ampicillin | Diphenhydramine | Naproxen | Sulindac |
| L-Phenylephrine | L-Ephedrine | Niacinamide | Tetracycline |
| Apomorphine | Ecgonine methylester | Nifedipine | Tetrahydrocortisone, |
| Aspartame | Ethyl-p-aminobenzoate | Norethindrone | 3-Acetate |
| Atropine | β-Estradiol | D-Norpropoxyphene | Morphine-3-β-Dglucuronide |
| Benzilic acid | Estrone-3-sulfate | Noscapine | Tetrahydrozoline |
| Benzoic acid | Erythromycin | D,L-Octopamine | Thiamine |
| Benzphetamine | Fenoprofen | Oxalic acid | Thioridazine |
| Bilirubin | Furosemide | Oxolinic acid | D,L-Tyrosine |
| Deoxycorticosterone | Gentisic acid | Oxymetazoline | Tolbutamide |
| Caffeine | Hemoglobin | Papaverine | Triamterene |
| Chloralhydrate | Hydralazine | Penicillin-G | Trifluoperazine |
| Chloramphenicol | Hydrochlorothiazide | Perphenazine | Trimethoprim |
| Chlorothiazide | Hydrocortisone | Phenelzine | Tyramine |
| D,L-Chlorpheniramine | O-Hydroxyhippuric acid | L-Phenylephrine | D,L-Tryptophan |
| Chlorpromazine | 3-Hydroxytyramine | β-Phenylethylamine | Urine acid |
| Chlorquine | D, L-Isoproterenol | Phenylpropanolamine | Verapamil |
| Cholesterol | Isoxsuprine | Prednisone | Labetalol |
| Clonidine | D,L-Propanolol | Zomepirac | Quinidine |
| Cortisone | Ketoprofen | L-Cotinine | Tramadol (Buprenorphine only) |
| D-Pseudoephedrine | Creatinine | Loperamide | |
# Specific Gravity:
The effect of specific gravity on the Wondfo One-Step Buprenorphine, Oxycodone and Propoxyphene devices was determined by evaluating 11 urine samples with specific gravities ranging from 1.000-1.040 and spiking them with the target drugs at 50% below the cutoffs and 50% above the cutoffs. Testing was performed in duplicate on 3 lots of test strips. No interference from specific gravity was observed at concentrations of 1.000 to 1.035.
# pH:
{9}
The effect of pH on the Wondfo One-Step Buprenorphine, Oxycodone and Propoxyphene devices was determined by spiking a negative urine pool with the target drugs at 50% below the cutoffs and 50% above the cutoffs. pH was adjusted in increments of 1 pH unit across the range from 3.0-10.0 and each urine was tested in duplicate with three lots of test strips. No interference from pH was observed at pH levels of 4.0 to 9.0.
f. Assay cut-off:
Cutoff concentrations are:
| Device | Cut-off Concentration |
| --- | --- |
| Wondfo One-Step Buprenorphine | 10 ng/mL |
| Wondfo One-Step Oxycodone | 100 ng/mL |
| Wondfo One-Step Propoxyphene | 300 ng/mL |
Cutoff studies were performed for buprenorphine, oxycodone and propoxyphene using a combination of clinical and spiked samples for each drug (n=150 per drug). The testing protocol was identical for each drug.
25 clinical samples were collected for each of the three drugs. Concentrations of buprenorphine, oxycodone or propoxyphene in the samples were determined by GC/MS. An additional 125 drug free negative samples were obtained for each drug and spiked with either buprenorphine, oxycodone or propoxyphene at -50% cutoff, -25% cutoff, cutoff, +25% cutoff, and +50% cutoff. Absence of any drugs in the negative urines and concentrations after spiking were confirmed by GC/MS.
5 clinical samples and 25 spiked samples were tested at each concentration for each drug in replicates of 30 using three strip lots (n=90) and 3 operators. Results are summarized below:
Buprenorphine
| Lot | -50% cutoff (pos/neg) | -25% cutoff (pos/neg) | Cutoff (pos/neg) | +25% cutoff (pos/neg) | +50% cutoff (pos/neg) |
| --- | --- | --- | --- | --- | --- |
| 1 | 0/90 | 11/79 | 83/7 | 88/2 | 90/0 |
| 2 | 0/90 | 9/81 | 82/8 | 88/2 | 90/0 |
| 3 | 0/90 | 11/79 | 87/3 | 88/2 | 90/0 |
Oxycodone
| Lot | -50% cutoff (pos/neg) | -25% cutoff (pos/neg) | Cutoff (pos/neg) | +25% cutoff (pos/neg) | +50% cutoff (pos/neg) |
| --- | --- | --- | --- | --- | --- |
| 1 | 0/90 | 7/83 | 80/10 | 88/2 | 90/0 |
| 2 | 0/90 | 9/81 | 83/7 | 86/4 | 90/0 |
| 3 | 0/90 | 13/77 | 80/10 | 88/2 | 90/0 |
{10}
11
# Propoxyphene
| Lot | -50% cutoff (pos/neg) | -25% cutoff (pos/neg) | Cutoff (pos/neg) | +25% cutoff (pos/neg) | +50% cutoff (pos/neg) |
| --- | --- | --- | --- | --- | --- |
| 1 | 0/90 | 10/80 | 80/10 | 86/4 | 90/0 |
| 2 | 0/90 | 6/84 | 83/7 | 86/4 | 90/0 |
| 3 | 0/90 | 11/79 | 86/4 | 89/1 | 90/0 |
## 2. Comparison studies:
### a. Method comparison with predicate device:
Method comparison studies for buprenorphine, oxycodone and propoxyphene were evaluated in the same manner using all natural urine samples using one lot of test strips and three operators. 10 samples were negative for all three drugs. 40 samples ranging from less than 50% cutoff to the cutoff and 40 samples ranging from cutoff to greater than 50% cutoff were evaluated. The test strip results were compared to GC/MS.
Discrepant samples were analyzed by GC/MS. Metabolites detected for buprenorphine were buprenorphine and nor-buprenorphine. Oxycodone detected only oxycodone, and propoxyphene detected both d-propoxyphene and d nor-propoxyphene. The results obtained by each user with the Wondfo One-Step devices, including discrepant results compared to GC/MS are below.
## Buprenorphine
### Viewer A:
| Wondfo Result | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 1 | 16 | 20 |
| Negative | 20 | 19 | 4 | 0 |
% agreement among positives is 90% (95% Confidence Interval 74.5% - 100%)
% agreement among negatives is 97.5% (95% Confidence Interval 82.0% - 100%)
| % of the cutoff | Concentration GC/MS (ng/mL) Total result | Viewer A result |
| --- | --- | --- |
| -50% cutoff to cutoff | 8.9 | + |
| Cutoff to +50% cutoff | 10.5 | - |
| | 11.5 | - |
| | 12 | - |
| | 12.9 | - |
### Viewer B:
{11}
12
| Wondfo Result | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 2 | 16 | 20 |
| Negative | 20 | 18 | 4 | 0 |
% agreement among positives is 90% (95% Confidence Interval 74.5% - 100%)
% agreement among negatives is 95% (95% Confidence Interval 79.5% - 100%)
| % of the cutoff | Concentration GC/MS (ng/mL) Total result | Viewer B result |
| --- | --- | --- |
| -50% cutoff to cutoff | 7.0 | + |
| | 8.9 | + |
| Cutoff to +50% cutoff | 10.5 | - |
| | 11.5 | - |
| | 12.0 | - |
| | 12.9 | - |
# Viewer C:
| Wondfo Result | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 0 | 16 | 20 |
| Negative | 20 | 20 | 4 | 0 |
% agreement among positives is 90% (95% Confidence Interval 74.5% - 100%)
% agreement among negatives is 100% (95% Confidence Interval 84.5% - 100%)
| % of the cutoff | Concentration GC/MS (ng/mL) Total result | Viewer C result |
| --- | --- | --- |
| Cutoff to +50% cutoff | 10.5 | - |
| | 11.5 | - |
| | 12.0 | - |
| | 12.9 | - |
# Oxycodone
# Viewer A:
| Wondfo Result | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 1 | 15 | 23 |
| Negative | 20 | 19 | 2 | 0 |
% agreement among positives is 95% (95% Confidence Interval 79.5% - 100%)
% agreement among negatives is 97.5% (95% Confidence Interval 82.0% - 100%)
{12}
13
| % of the cutoff | Concentration GC/MS (ng/mL) | Viewer A result |
| --- | --- | --- |
| -50% cutoff to cutoff | 94.2 | + |
| Cutoff to +50% cutoff | 110.1 | - |
| | 112.3 | - |
## Viewer B:
| Wondfo Result | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 2 | 14 | 23 |
| Negative | 20 | 18 | 3 | 0 |
% agreement among positives is 92.5% (95% Confidence Interval 77.0% - 100%)
% agreement among negatives is 95% (95% Confidence Interval 79.5% - 100%)
| % of the cutoff | Concentration GC/MS (ng/mL) | Viewer B result |
| --- | --- | --- |
| -50% cutoff to cutoff | 94.2 | + |
| | 94.5 | + |
| Cutoff to +50% cutoff | 109.3 | - |
| | 110.1 | - |
| | 112.3 | - |
## Viewer C:
| Wondfo Result | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 0 | 13 | 23 |
| Negative | 20 | 20 | 4 | 0 |
% agreement among positives is 90% (95% Confidence Interval 74.5% - 100%)
% agreement among negatives is 100% (95% Confidence Interval 84.5% - 100%
| % of the cutoff | Concentration GC/MS (ng/mL) | Viewer C result |
| --- | --- | --- |
| Cutoff to +50% cutoff | 109.3 | - |
| | 110.1 | - |
| | 112.3 | - |
| | 118.7 | - |
## Propoxyphene:
Viewer A:
{13}
14
| Wondfo Result | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 2 | 17 | 19 |
| Negative | 26 | 12 | 4 | 0 |
% agreement among positives is 90% (95% Confidence Interval 74.5% - 100%)
% agreement among negatives is 95% (95% Confidence Interval 79.5% - 100%)
| % of the cutoff | Concentration GC/MS (ng/mL) | Viewer A result |
| --- | --- | --- |
| -50% cutoff to cutoff | 289.2 | + |
| | 290.3 | + |
| Cutoff to +50% cutoff | 308.2 | - |
| | 318.2 | - |
| | 320.8 | - |
| | 342.7 | - |
# Viewer B:
| Wondfo Result | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 1 | 18 | 19 |
| Negative | 26 | 13 | 3 | 0 |
% agreement among positives is 92.5% (95% Confidence Interval 77.0% - 100%)
% agreement among negatives is 97.5% (95% Confidence Interval 82.0% - 100%)
| % of the cutoff | Concentration GC/MS (ng/mL) | Viewer B result |
| --- | --- | --- |
| -50% cutoff to cutoff | 289.2 | + |
| Cutoff to +50% cutoff | 308.2 | - |
| | 320.8 | - |
| | 342.7 | - |
# Viewer C:
| Wondfo Result | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 0 | 16 | 19 |
| Negative | 26 | 14 | 5 | 0 |
% agreement among positives is 87.5% (95% Confidence Interval 72.0% - 100%)
% agreement among negatives is 100% (95% Confidence Interval 84.5% - 100%)
{14}
| % of the cutoff | Concentration GC/MS (ng/mL) | Viewer C result |
| --- | --- | --- |
| Cutoff to +50% cutoff | 308.2 | - |
| | 310.2 | - |
| | 318.2 | - |
| | 320.8 | - |
| | 342.7 | - |
b. Matrix comparison:
These devices are intended for use with human urine only.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.