ENDOSCOPE REPROCESSOR OER-PRO

K093106 · Olympus Medical Systems Corporation · FEB · Feb 12, 2010 · Gastroenterology, Urology

Device Facts

Record IDK093106
Device NameENDOSCOPE REPROCESSOR OER-PRO
ApplicantOlympus Medical Systems Corporation
Product CodeFEB · Gastroenterology, Urology
Decision DateFeb 12, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K093106 · Feb 12, 2010ENDOSCOPE REPROCESSOR OER-PROOlympus Medical Systems CorporationIn-use clinical testingIn-use testing was conducted to evaluate the cleaning and high-level disinfection efficacy of the OER-Pro reprocessor in a clinical environment.In-use testing; Cleaning efficacy; High-level disinfection

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
In-use cleaning validation; In-use clinical testingEndoscopes and accessoriesNot applicable for this studyReduction of protein and hemoglobin levels
In-use disinfection validation; In-use clinical testingEndoscopes and valvesNot applicable for this studyRecovery of viable microorganisms

Indications for Use

The OER-Pro is intended for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories. Safe use requires detergent and an FDA-cleared high-level disinfectant/sterilant that Olympus has validated to be efficacious and compatible with the materials of the OER-Pro and Olympus flexible endoscopes and their accessories. Use of a detergent or high-level disinfectant/sterilant that has not been validated by Olympus may be ineffective and can damage the OER-Pro components and the endoscopes being reprocessed. Endoscopes must be subject to cleaning by the user prior to reprocessing; however, use of the OER-Pro enables the user to perform modified manual cleaning of the endoscope prior to automated cleaning and high-level disinfection in the OER-Pro.

Device Story

Automated endoscope reprocessor; one-basin design; processes two endoscopes simultaneously. Inputs: endoscopes/accessories; detergent; FDA-cleared high-level disinfectant. Process: leak test; ultrasonic cleaning; high-level disinfection; rinse; alcohol/air drying. Features: RFID reader for scope/user ID; built-in printer for cycle results; 0.2-micron air/water filters. Sensors monitor fluid levels, temperature, and pressure. Operated by healthcare personnel in clinical settings. Output: high-level disinfected endoscopes. Benefits: standardized, automated reprocessing; reduced manual cleaning steps; validated disinfection efficacy.

Clinical Evidence

Bench testing only. Validated using EndoQuick detergent and Aldahol III High-Level Disinfectant. Testing included process parameter verification, cleaning efficacy (protein/hemoglobin reduction), high-level disinfection (6-log reduction of M. terrae), full-cycle effectiveness, self-disinfection, water line disinfection, and toxicological evaluation of residues. No clinical data.

Technological Characteristics

Automated washer/disinfector; one-basin immersion system; ultrasonic cleaning; 0.2-micron bacteria-retentive air/water filters. Connectivity: RFID antenna for ID tracking; built-in printer. Sensors: fluid level, temperature, pressure. Materials: compatible with Olympus flexible endoscopes and validated disinfectants. Sterilization: high-level chemical disinfection.

Indications for Use

Indicated for cleaning and high-level disinfection of heat-sensitive Olympus flexible endoscopes and accessories in healthcare facilities.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K093106 Page 1 of 3 510(k) Premarket Notification ENDOSCOPE REPROCESSOR OER-Pro ## 510(k) Summary FEB 1 2 2010 Date Prepared: February 3, 2010 o Applicant OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, Japan Establishment Registration No: 8010047 o Official Correspondent Laura Storms-Tyler OLYMPUS AMERICA INC. 3500 Corporate Parkway, Center Valley, PA 18034-0610 Phone: (484) 896-5688 Fax: (484) 896-7128 Email: Laura.storms-tyler@olympus.com Establishment Registration No: 2429304 . Manufacturer AIZU OLYMPUS CO., LTD. 500 Aza-Muranishi, Ooaza-lidera, Monden-cho, Aizuwakamatsu-shi, Fukushima, Japan 965-8520 Establishment Registration No: 9610595 #### ්‍යා Device Identification Applicant Information ្ញ | • | Device Trade Name | ENDOSCOPE REPROCESSOR OER-Pro | |---|------------------------|------------------------------------------------| | • | Common Name | Endoscope washer/disinfector | | • | Class | II | | • | Regulation Number/Name | 876.1500 Endoscope and accessories | | • | Product Code | FEB - Accessories, Cleaning, For Endoscope | | • | Classification Panel | Gastroenterology/Urology | | • | Performance Standard | None established under Section 514 of FD&C Act | #### ប្រ Predicate Device (PD) | • Device Trade Name | Endoscope Disinfector DSD-91E | |---------------------|-------------------------------| | 510(k) Number | K914145 | | • Manufacturer | MediVators, Inc. | #### OLYMPUS AMERICA INC. 3500 CORPORATE PARKWAY, P.O. BOX 610, CENTER VALLEY, PA 13034-0610 IELEBHONE (484) 800-2000 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "OLYMPUS" in a bold, sans-serif font. The letters are black and appear to have a slightly rough or textured edge, giving them a somewhat distressed look. The word is presented in all capital letters and is horizontally oriented. RJSJC Page 2 of 3 #### Device Description The OER-Pro Endoscope Reprocessor is an automated endoscope reprocessor intended for highlevel disinfection of Olympus flexible cudoscopes and its accessories, utilizing both a detergent and FDA cleared high-level disinfectant validated by Olympus to be efficacious and compatible with the materials of the OER-Pro and Olympus flexible endoscopes and accessories. The OER-Pro is a one-basin automatic endoscope reprocessor that performs leak test, cleaning, disinfection, rinse, and alcohol flush to render a high-level disinfected endoscope and accessories. The OER-Pro utilizes an immersion method for cleaning, disinfecting, and rinsing of endoscope and accessory external surfaces, and connectors for endoscope channel cleaning, disinfecting, and rinsing. Two endoscopes, with several exceptions, can be reprocessed simultaneously in the basin during one reprocessing cycle. The OER-Pro's cleaning cycle includes ultrasonic cleaning, which helps remove debris and dirt from endoscope surfaces, The OER-Pro enables the user to perform modified manual cleaning of the endoscope prior to automated cleaning and high-level disinfection in the OER-Pro. The modified manual cleaning process climinates several steps including the syringe flushing into the endoscope channels. The OER-Pro is capable of fully automated detergent/disinfectant solution dispensing and alcohol/air drying of endoscope channels. The 0.2-micron air/water filters are bacteria retentive and produce suitable rinse water and air for reprocessing. Built-in sensors detect fluid levels, fluid temperature, air/fluid pressure, and the operating states of the components within the OER-Pro. The OER-Pro is also equipped with a RFID (Radio-Frequency Identification) function. With a built-in antenna, the OER-Pro is capable of reading user and scope ID data from the proprietary ID tag/chip. The scope/user ID information and each reprocessing result can be printed out with a built-in printer. #### D Indications for Use The OER-Pro is intended for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories. Safe use requires detergent and an FDAcleared high-level disinfectant/sterilant that Olympus has validated to be efficacious and compatible with the materials of the OER-Pro and Olympus flexible endoscoves and their accessories. Use of a detergent or high-level disinfectant/sterilant that has not been validated by Olympus may be ineffective and can damage the OER-Pro components and the endoscopes being reprocessed. Endoscopes must be subject to cleaning by the user prior to reprocessing; however, use of the OER-Pro cnables the user to perform modified manual cleaning of the endoscope prior to automated cleaning and high-level disinfection in the OER-Pro. #### Comparison to Predicate Device បា The OER-Pro is equivalent in indications and operational principles to the DSD-91E. Both devices are intended for high-level disinfection of flexible endoscopes, use an immersion system, flush water/disinfectant to endoscope channels via connectors, and utilize a detergent and FDAcleared liquid chemical germicide. Both devices can reprocess two endoscopes simultancously. The major difference is that the OER-Pro has one basin that can accommodate two endoscopes, whereas the DSD-91E has two basins that reprocess one endoscope in each basin, {2}------------------------------------------------ # 11 093106 Image /page/2/Picture/1 description: The image contains the word "OLYMPUS" in a bold, sans-serif font. The letters are black and have a slightly distressed or textured appearance, giving them a rugged or vintage feel. The word is presented in a straightforward, horizontal arrangement. Page 3 of 3 #### ם Performance Data The OER-Pro has been tested following the requirements in the FDA guidance document titled "Guidance on Premarket Notification [510(k)] Submissions for Automated Endoscope Washers, Washer/Disinfectors, and Disinfectors Intended for Use in Health Care Facilities", issued in August 1993. The OER-Pro was validated with EndoQuick detergent and Aldahol III High-Level Disinfectant and Sterilant. Test reports provided in this premarket notification include: ### Process Parameter Test The OER-Pro was tested to demonstrate that the device performs as intended. The test results showed that the OER-Pro achieves and maintains the specified physical process parameters, including detection of the defined fault conditions and execution of automatic response/ processing following fault detection, #### Validation Testing - Cleaning The OER-Pro was tested to evaluate its ability to clean endoscopes in both simulated and in-use conditions. The test results demonstrate that the OER-Pro effectively reduced protein and hemoglobin levels in all sample sites. ### Validation Testing - High-Level Disinfection The OER-Pro was tested to evaluate its ability to high-level disinfect endoscopes and valves in both simulated and in-use conditions. The simulated use testing demonstrated a 6 Logio reduction of M.terrae at all inoculated sites was achieved after reprocessing in the OER-Pro's disinfection cycle. In-use testing demonstrated no viable microorganisms were recovered from endoscopes and valves following reprocessing in the OER-Pro. ### Validation Testing - Full Cycle The OER-Pro was tested to evaluate its effectiveness for full cycle reprocessing including both cleaning and disinfection under simulated use conditions. The simulated use testing demonstrated that OER-Pro effectively cleaned and achieved hisinfection for Olympus endoscopes and valves. ### Simulated-Use Testing - Self-Disinfection Simulated-use testing was performed to validate self-disinfection of the OER-Pro. Testing demonstrated that a greater than 6 log reduction in M. terrae was achieved for all sample locations after completion of routine reprocessing of endoscopes within the OER-Pro. #### Simulated-Use Testing - Water Line Disinfection The simulated-use testing was performed to validate disinfection of the OER-Pro water line piping which does not contact high-level disinfectant during routine reprocessing. The test result showed that a greater than 6 log reduction in M. terrae was achieved for all sample locations affer completion of the water line disinfection procedure. ### Toxicological Evaluation of Residues The safety of residual chemicals remaining on endoscopes after reprocessing in the QER-Pro was evaluated. The test results showed that the OER-Pro reprocessing cycle removes detergent and disinfectant residues to non-toxic levels. ### Conclusion The information and performance data presented in this premarket notification support the claim that the OER-Pro is substantially equivalent to the predicate device. > QLYMPUS AMERICA INC. 3500 CORPORATE PARKWAY. P.O. BOX 610, CENTER VALLEY, PA 18034-0610 TELEPHONE (484) 836-5000 {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's logo, which features a stylized human figure embracing another. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Olympus Medical Systems Corporation C/O Ms. Laura Storms-Tyler Olympus America Incorporated 3500 Corporate Parkway Center Valley, Pennsylvania 18034-0610 Re: K093106 Trade/Device Name: Endoscope Reprocessor OER-Pro Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FEB Dated: January 13, 2010 -Received: January 15, 2010 FEB 1 2 2010 Dear Ms. Storms-Tyler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ### Page 2 - Ms. Storms-Tyler Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. . You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerelv vours. Authen W.m Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: ENDOSCOPE REPROCESSOR OER-Pro Indications For Use: The OER-Pro is intended for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories. Safe use requires detergent sonsitive of cleared high-level disinfectant/sterilant that Olympus has validated to be efficacions and compatible with the materials of the OER-Pro and Olympus flexible endoscopes and their accessories. Use of a detergent or high-level disinfectant/sterilant that has not been validated by Olympus may be ineffective and can damage the OER-Pro components and the endoscopes being reprocessed. Endoscopes must be subject to cleaning by the user prior to reprocessing; however, use of the OER-Pro enables the user to perform modified manual cleaning of the endossope prior to automated cleaning and high-level disinfection in the OER-Pro. Prescription Use______________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Elizabeth F. Clamin-Wells Page 1 of 1 (Division Sian-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices K093106 510(k) Number:
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...