K093075 · Aragon Surgical, Inc. · GEI · Jan 21, 2010 · General, Plastic Surgery
Device Facts
Record ID
K093075
Device Name
ARAGON SURGICAL RF SYSTEM TELEO INSTRUMENT
Applicant
Aragon Surgical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 21, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aragon Surgical Teleo Instrument is a dedicated bipolar electrosurgical instrument intended for use in general surgical and gynecologic procedures where ligation and division of vessels is desired. The instrument creates a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue. The indications for use include general surgical procedures (including urologic, vascular, thoracic, and thoracoscopic), and gynecologic surgery procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectorny, adhesiolysis, bowel resections and oophorectomy etc., or any procedure where vessel ligation (cutting and sealing), tissue grasping, and dissection is performed. The device can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument. The Aragon Surgical Teleo Instrument has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
Device Story
Aragon Surgical Teleo Instrument is a bipolar electrosurgical device used in general and gynecologic surgery. It functions as part of the Aragon Surgical RF System, consisting of a generator and the Teleo instrument. The device grasps tissue between its jaws, applies bipolar RF energy to seal vessels up to 7mm, and utilizes an integrated manual cutting blade to divide tissue. Operated by surgeons in clinical settings, the device provides hemostasis and tissue division, reducing the need for sutures or clips. It is provided sterile for single use.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, materials, function, and pre-clinical evaluation.
Technological Characteristics
Bipolar electrosurgical instrument; single-use; sterile. Features jaws for tissue grasping, RF energy delivery for coagulation, and an integrated mechanical cutting blade. Compatible with Aragon Surgical Generator. No software or complex algorithms described.
Indications for Use
Indicated for general surgical (urologic, vascular, thoracic, thoracoscopic) and gynecologic procedures requiring vessel ligation and division, tissue grasping, and dissection. Suitable for vessels up to 7mm and tissue bundles fitting within jaws. Contraindicated for tubal sterilization or tubal coagulation procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
K093075
Aragon Surgical Inc.
Special Premarket Notification: Teleo Instrument
### SECTION 5
### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
# REGULATORY AUTHORITY
JAN 2 1 2010
Safe Medical Devices Act of 1990, 21 CFR 807.92
## COMPANY NAME/CONTACT
Alan Curtis Aragon Surgical, Inc. 1810 Embarcadero Road, Suite B Palo Alto, CA 94303
### NAME OF DEVICE
| Trade Name: | Aragon Surgical RF System<br>Teleo Instrument |
|------------------------|-------------------------------------------------------------------------------------|
| Common Name: | Electrosurgical System |
| Device Product Code: | GEI |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories (21 CFR 878.4400) |
| Device Panel: | General Surgery/Restorative Devices |
| Device Classification: | Class II |
## PREDICATE DEVICES
Aragon Surgical RF System/L2 Instrument (K090306) .
## DEVICE DESCRIPTION
The Aragon Surgical RF Systems consists of two components: the Aragon Surgical Generator and the Aragon Surgical Teleo Instrument. The Aragon Teleo Instrument is provided as sterile and is intended for single use. The device is capable of vessel sealing, blunt dissection, grasping and dividing tissue enclosed within its jaws.
{1}------------------------------------------------
.
## INDICATION FOR USE STATEMENT
The Aragon Surgical Teleo Instrument is a dedicated bipolar electrosurgical instrument intended for use in general surgery and gynecologic procedures where ligation and division of vessels is desired. The instrument creates a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
The indications for use include general surgical procedures (including urologic, vascular, thoracic, and thoracoscopic), and gynecologic surgery procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectorny, adhesiolysis, bowel resections and oophorectomy etc., or any procedure where vessel ligation (cutting and sealing), tissue grasping, and dissection is performed.
The device can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
The Aragon Surgical Teleo Instrument has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
#### SUBSTANTIAL EQUIVALENCE COMPARISON
## Comparison to Predicate Devices
### Technological Characteristics
The technological characteristics of the Aragon Surgical Teleo Instrument are identical to those of the cited predicate electrosurgical device. K090306. This device is equivalent in terms of design, materials, and principal of operation. The modification between the Aragon Surgical Teleo Instrument and the predicate device does not raise new issues regarding safety or effectiveness.
{2}------------------------------------------------
# Indications for Use
Substantial equivalence is also supported for the Aragon Surgical Teleo Instrument by the predicate devices cleared for general surgery and gynecologic procedures where ligation and division of vessels is desired.
## CONCLUSION
Based on the design, materials, function, intended use, and pre-clinical evaluation, the Aragon Surgical Teleo Instrument is substantially equivalent to devices currently marketed under the Federal Food, Drug and Cosmetic Act. In addition, the Aragon Surgical Teleo Instrument raises no new safety or effectiveness issues. Therefore, safety and effectiveness are reasonably assured, justifying 510(k) clearance.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
JAN 9 1 2010
Aragon Surgical Incorporation % Mr. Alan Curtis, RAC Vice President, Regulatory/Clinical & Quality Affairs 1810 Embarcadero Road, Suite B Palo Alto. California 94303
Re: K093075
Trade/Device Name: Aragon Surgical RF System Teleo Instrument Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: January 8, 2010 Received: January 11. 2010
Dear Mr. Curtis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 - Mr. Alan Curtis
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21.CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You-may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Milburn
..
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### SECTION 4
# INDICATIONS FOR USE STATEMENT
510(k) Number: K093075
Device Name:
Aragon Surgical RF System Teleo Instrument
Indications for Use:
The Aragon Surgical Teleo Instrument is a dedicated bipolar electrosurgical instrument intended for use in general surgical and gynecologic procedures where ligation and division of vessels is desired. The instrument creates a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
The indications for use include general surgical procedures (including urologic, vascular, thoracic, and thoracoscopic), and gynecologic surgery procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectorny, adhesiolysis, bowel resections and oophorectomy etc., or any procedure where vessel ligation (cutting and sealing), tissue grasping, and dissection is performed.
The device can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
The Aragon Surgical Teleo Instrument has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
or
Prescription Use X (Per 21 CFR 801.109)
Over-The-Counter Use
Concurrence of CDRH, Office of Device Evaluation (ODE)
FOR M. MELKERSON
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K093075
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.