K093072 · Dentalez, Inc. · EFB · Dec 28, 2009 · Dental
Device Facts
Record ID
K093072
Device Name
STARCARE HANDPIECE MAINTENANCE UNIT
Applicant
Dentalez, Inc.
Product Code
EFB · Dental
Decision Date
Dec 28, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Indications for Use
The StarCare Handpiece Maintenance Unit is intended for the internal cleaning. i.e. purging of old lubricant, and the lubrication of air-driven high speed handpieces, electric micromotor attachments and turbines used in dentistry. StarCare Handpiece Maintenance Unit should be used only after the handpieces have been externally cleaned and prior to sterilization.
Device Story
Self-contained dental maintenance unit; cleans and lubricates high-speed handpieces, electric micromotor attachments, and turbines. Input: manual connection of handpieces to unit ports. Operation: internal reservoirs supply cleaner and lubricant via magnetic valves; pressurized air dispenses fluids; unit detects connection presence to activate specific stations; final compressed air burst purges excess debris and lubricant. Used in dental clinics; operated by dental staff. Output: serviced, lubricated handpieces ready for sterilization. Benefits: automates maintenance, ensures consistent lubrication, and removes internal debris, extending instrument life.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Self-contained system with dual reservoirs for cleaner and lubricant. Features magnetic valve-controlled delivery, pressurized air dispensing, and automated connection detection. Includes indicator lights for station status. Operates on compressed air. No software or electronic processing described.
Indications for Use
Indicated for internal cleaning and lubrication of air-driven high-speed dental handpieces, electric micromotor attachments, and turbines. Used by dental professionals post-external cleaning and pre-sterilization. Prescription only.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
K093072
DEC 2 8 2009
510(k) Premarket Notification StarCare Handpiece Maintenance Unit Section 5
510(k) Summary
## Section 5: 510(k) Summary
Company:
DentalEZ Inc., StarDental Division Owner/operator number 2520265
#### Contact Person:
Dale Braas, Engineering/Quality Manager Kay Engle, Regulatory Affairs Supervisor DentalEZ Inc., StarDental Division 1816 Colonial Village Lane Lancaster, PA 17601 Phone: (717) 291-1161 Fax: (717) 391-2757
#### Proprietary/Trade Name:
StarCare Handpiece Maintenance Unit
#### Classification:
Handpiece, Air-powered, Dental (21 C.F.R. § 872.4200, Product code EFB)
#### Predicate Device:
Statmatic, manufactured by SciCan LTD. Ontario, Canada (K073319) Lubrina, manufactured by J. Morita USA, Inc., Irvine, CA. (K070074) Assistina, manufactured by A-Dec Incorporated, Newburg, OR (K010127)
StarCare Handpiece Maintenance Unit has similar intended uses, similar methods of operation and similar technology to the previously cleared predicate devices. Although minor differences exist between the StarCare Handpiece Maintenance Unit and predicate devices, they do not present any new questions of safety or effectiveness.
#### Device Description:
The StarCare Handpiece Maintenance Unit is used for the cleaning, i.e. purging of old lubricant, and lubrication of lubricated high speed handpieces, electric micromotor attachments and handpiece turbines. The unit consists of two connections for high speed handpieces, one connection for electric micromotor
{1}------------------------------------------------
attachments and one connection for handpiece turbine cleaning. The unit supplies cleaner and lubricant to the connection only when a high speed handpiece or electric micromotor attachment is present. The handpiece turbine cleaning function is independent of the high speed handpiece and electric micromotor attachment cleaning and lubrication function.
Two reservoirs within the unit supply the cleaner and lubricant to the individual stations. These reservoirs are clearly marked regarding the contents of the reservoir. The cleaner and lubricant are delivered to the high speed handpiece and electric micromotor attachment connection via tubes controlled by the activation of magnetic valves. The liquid is dispensed by pressurized air to the high speed handpiece and electric micromotor attachment.
#### Intended Use:
The StarCare Handpiece Maintenance Unit is intended for the internal cleaning. i.e. purging of old lubricant, and the lubrication of air-driven high speed handpieces, electric micromotor attachments and turbines used in dentistry.
StarCare Handpiece Maintenance Unit should be used only after the handpieces have been externally cleaned and prior to sterilization.
### Technological Characteristics:
The StarCare Handpiece Maintenance Unit is a self-contained cleaning and lubrication system that can clean and lubricate up to two high speed handpieces and one electric micromotor attachment at the same time. Indicator lights on the front panel of the unit, show which handpiece station is currently being serviced. The cleaner and lubricant are pumped from separate reservoirs located within the unit. The unit detects if a high speed handpiece or electric micormotor attachment is connected and delivers cleaner and lubricant only to the occupied stations. After cleaning and lubrication cycle a burst of compressed air is forced into the high speed handpiece or electric micromotor attachment to purge the instrument of excess cleaner, lubricant and debris.
The handpiece turbine cleaning function of the StarCare Handpiece Maintenance Unit is independent of the high speed handpiece and electric micromotor attachment cleaning and lubrication. With the front door of the unit oven, the operator holds the handpiece in place during the handpiece turbine cleaning process. After the cleaning and lubrication cycle a burst of compressed air is forced into the chuck to purge the excess cleaner, lubricant and debris. An indicator light on the front panel illuminates during the cleaning process.
{2}------------------------------------------------
The following table summarizes the comparison of the StarCare Handpiece Maintenance Unit to the predicate devices for various technological characteristics.
| Technological Characteristics | Predicate Device Comparison<br>conclusion |
|-----------------------------------------------------|-------------------------------------------|
| Indication for use | Similar |
| Target population | Identical |
| Design | Similar |
| Materials | Similar |
| Performance | Similar |
| Sterility | Identical |
| Biocompatibility | Identical |
| Mechanical safety | Identical |
| Chemical safety | Identical |
| Energy used and/or delivered | Identical |
| Compatibility with environment<br>and other devices | Identical |
| Where used | Identical |
| Standards met | Similar |
| Electrical Safety | Similar |
## Substantial Equivalence:
The determination of substantial equivalence is based on the premise that the proposed device and the predicate devices have the same intended use, similar technology and design.
{3}------------------------------------------------
Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The eagle is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Ms. Kay Engle Regulatory Affairs Supervisor DentalEZ Incorporated StarDental Division 1816 Colonial Village Lane Lancaster, Pennsylvania 17601
# JAN 2 1 2010
Re: K093072
Trade/Device Name: StarCare Handpiece Maintenance Unit Regulation Number: 21CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB Dated: September 29, 2009 Received: September 30, 2009
Dear Ms. Engle:
This letter corrects our substantially equivalent letter of December 28, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Engle
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known):
Device Name: StarCare Handpiece Maintenance Unit
Indications for Use:
The StarCare Handpiece Maintenance Unit is intended for the internal cleaning, i.e purging of old lubricant, and the lubrication of air-driven high speed handpieces, electric micromotor attachments and turbines used in dentistry.
StarCare Handpiece Maintenance Unit should be used only after the handpieces have been externally cleaned and prior to sterilization.
CAUTION: Federal law restricts this device to sale by or on the order of a dentist.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
RSBER DDS for Dr. K.P. Mulry (Acting)
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
**510(k) Number:** K093072
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.