← Product Code [GEI](/productcode/GEI) · K093008

# HALO ABLATION CATHETER, MODEL 90-9100 (K093008)

_Barrx Medical, Incorporated · GEI · Jan 8, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K093008

## Device Facts

- **Applicant:** Barrx Medical, Incorporated
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jan 8, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K093008 · Jan 8, 2010 | HALO ABLATION CATHETER, MODEL 90-9100 | Barrx Medical, Incorporated | Published clinical literature; Clinical case studies | Clinical data from published literature and case studies were used to support the safety and effectiveness of the device for the treatment of Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP), and to provide guidance for optimal device settings. | GAVE; Radiation Proctitis; Literature review; Case studies |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical data from publications and case studies; Literature review and case study analysis | Patients with Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP) | Not applicable for this study | Safety and effectiveness for coagulation of bleeding sites in the gastrointestinal tract |

## Indications for Use

The HALO® System (including HALO® Ablation Catheter model 90-9100 and HALO® Energy Generator model 90-9000) is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).

## Device Story

HALO System comprises HALO Ablation Catheter (model 90-9100) and HALO Energy Generator (model 90-9000). Device functions as electrosurgical coagulation system for GI tract. Physician operates catheter endoscopically to deliver thermal energy to mucosal layer of GI tract; energy coagulates bleeding/non-bleeding sites. System provides controlled ablation to treat conditions like GAVE and Radiation Proctitis. Output is thermal coagulation of tissue; allows physician to treat vascular ectasia and lesions. Benefits include effective hemostasis and treatment of mucosal pathologies. Used in clinical/endoscopic settings.

## Clinical Evidence

Clinical data provided via publications and case studies to support expanded indications for GAVE and Radiation Proctitis. Data demonstrates that the HALO Ablation Catheter is at least as safe and effective as the predicate device for coagulation of GI bleeding sites. No new bench testing was conducted as the device is technologically identical to the previously cleared version.

## Technological Characteristics

Electrosurgical coagulation system. Comprises ablation catheter and energy generator. Principle of operation: thermal cautery/coagulation of mucosal tissue. Identical in construction, energy density, and operation to predicate HALO90 and HALO360 systems. No changes to hardware or software since previous clearance (K083737).

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- HALO90 Ablation Catheter model 90-9100 ([K060169](/device/K060169.md), [K062723](/device/K062723.md), [K083737](/device/K083737.md))
- HALO360 Ablation Catheter ([K050168](/device/K050168.md), [K062225](/device/K062225.md), [K083711](/device/K083711.md))
- Stellartech Coagulation Catheter ([K013139](/device/K013139.md), [K042909](/device/K042909.md), [K050831](/device/K050831.md))
- Microvasive Gold Probe ([K970278](/device/K970278.md))
- Olympus Heat Probe ([K982289](/device/K982289.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

093008
Page 1 of 4

# 510(k) SUMMARY BARRX's HALO" System

JAN - 8 2010

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

BARRX Medical, Inc. 540 Oakmead Parkway Sunnyvale, CA 94085 Phone: (408) 328-7300 Facsimile: (408) 328-7395 Contact: Viorica Filimon Date Prepared: June 2009

# Name of Device and Name/Address of Sponsor

HALO® System (including HALO® Ablation Catheter model 90-9100 and HALO® Energy Generator model 90-9000) BARRX Medical, Inc. 540 Oakmead Parkway Sunnyvale, CA 94085

# Common or Usual Name

Electrosurgical Coagulation System

# Classification Name

Electrosurgical Cutting or Coagulation Device

### Predicate Devices

HALO90 Ablation Catheter model 90-9100 (K060169, K062723 and K083737) HALO360 Ablation Catheter (K050168, K062225, K083711) Stellartech Coagulation Catheter (K013139, K042909, K050831) Microvasive Gold Probe (K970278) Olympus Heat Probe (K982289)

# Intended Use / Indications for Use

The HALO® System (including HALO® Ablation Catheter model 90-9100 and HALO® Energy Generator model 90-9000) is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus.

{1}------------------------------------------------

K093008
Page 2 of 4

Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).

### Technological Characteristics

The HALO® Ablation Catheter model 90-9100 operates in conjunction with HALO® Energy Generator model 90-9000. There are no changes to the HALO® Ablation Catheter or HALO® Energy Generator implemented since the devices were cleared by K083737, K062723 and K062441. The product is also identical in principle of operation and energy density to the treatment site with the predicate devices HALO360 Ablation Catheter, Stellartech Coagulation Catheter. HALO® is substantially equivalent in construction with the predicate devices HALO360 Ablation Catheter, Stellartech Coagulation Catheter and Microvasive Gold Probe. The product is similar in performance with Olympus Heat Probe. The similarities in performance and intended use with the other predicate devices are presented in Table 2.

#### Performance Data

The HALO® Ablation Catheter model 90-9100 is technologically identical to the cleared predicate device HALO® Ablation Catheter model 90-9100 cleared by K083737 and K062723. Accordingly, no performance testing was conducted.

Clinical data was provided to FDA to support the changes in the Instructions for Use associated with the use in the coagulation of bleeding sites for the HALO® Ablation Catheter for the treatment of GAVE and Radiation Proctitis. Clinical data demonstrated that, when used in accordance with those instructions, the HALO® Ablation Catheter used for the treatment of GAVE and Radiation Proctitis respectively is at least as safe and effective as the cleared HALO® Ablation Catheter for the treatment of bleeding sites in the gastrointestinal tract .

### Substantial Equivalence

The HALO® Ablation Catheter with expanded indications for use is as safe and effective as the predicate devices HALO® Ablation Catheter, HALO360 Ablation Catheter, Microvasive Gold Probe, and Olympus Heat Probe. The HALO® has the same general intended use. general indications for use, technological characteristics, and principles of operation as the predicate devices.

The HALO® Ablation Catheter is currently cleared for the coagulation of bleeding sites in the gastrointestinal tract. In addition, physicians have elected to use the HALO® System within its indication for use for the coagulation of bleeding sites within the gastrointestinal tract, specifically gastric antral vascular ectasia (GAVE) and radiation proctitis/proctopathy (hereafter, collectively RP). GAVE and RP are similar in that they are both confined to the mucosal layer of the gastrointestinal tract, associated with hemorrhage and treated with coagulative

{2}------------------------------------------------

K0930008
Page 3 of 4

therapy. In this way, they are also very much akin to the other bleeding conditions (esophageal ulcers, Mallory-Weiss tears, arteriovenous malformations, angiomata, Dieulafoy lesions, and angiodysplasia) cited as examples in the HALO® System cleared indication for use.

HALO® Ablation Catheter is substantially equivalent with several devices: HALO® Ablation Catheter (K06169, K062723 and K083737), HALO360 Ablation Catheter (substantially equivalency established in K060169 and K06223), Stellartech Coagulation Catheter (substantial equivalency established in K013139, K042909, K050831), Microvasive Gold Probe (equivalency established by K013139 and K970278), and Olympus Heat Probe (K982289).

HALO® Ablation Catheter included in this submission has identical indications for use with the HALO90 Ablation Catheter (K060169, K062723 and K083737), HALO360 Ablation Catheter (K050168, K062225, K083711) and Stellartech Coagulation Catheter 2.0 (K013139, K042909, K050831) with the exception of Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis(RP).

Based on the substantial equivalence established historically between HALO® Ablation Catheter and the predicate devices Microvasive Gold Probe (K970278) and Olympus Heat Probe (982289) we seek the expansion of the indications for use for HALO90 Ablation Catheter to include the GAVE (or Watermelon Stomach) indications for use as the Microvasive Gold Probe:

Indications for use for HALO® Ablation Catheter current submission:

The HALO® Ablation Catheter model 90-9100 is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus.

Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).

Indications for use for Microvasive Gold Probe (K970278):

The Gold Probe is indicated for use in transendoscopic electrocautery of visible bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus, stomach, duodenum, and colon. The indications include peptic ulcers, Mallory-Weiss tears, Arteriovenous Malformations (AVM), Angiomata, Dieulafoy Lesions, bleeding polyp stocks, angiomata, Watermelon Stomach(1), Barrett's Esophagus, angiodysplasia, and esophageal tumors.

ω Definition: Watermelon Stomach is the popular name for Gastric Antral Vascular Ectasia (GAVE) — a condition in which the lining of the stomach bleeds, causing it to look like the characteristic stripes of a watermelon when viewed by Endoscopy.

Indications for use for Olympus Heat Probe (K982289):

The Olympus Heat Probe Unit HPU-20 has been designed for thermal cautery with the Olympus Heat Probes.

The Olympus Heat Probe has been specifically designed to thermo-cauterize bleeding sites within the gastrointestinal tract under endoscopic observation.

The addition of Physician's Instructions recommending specific treatment settings and selection criteria for the patients (Contraindications, Warnings and

{3}------------------------------------------------

093008 age 4 of

Cautions) for the treatment of bleeding sites in the gastrointestinal tract such as GAVE and RP is supported by clinical data derived from publications and case studies and does improve the device risk profile by providing guidance required for optimal use of the device. The risk profile of the cleared predicate device HALO®0 Ablation Catheter with the indication for use of coagulation of bleeding sites in the gastrointestinal sites is identical with the risk profile of the device in the current submission HALO® Ablation Catheter with the indication for use for coagulation of bleeding sites in the gastrointestinal tract such as Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). Thus, the HALO90 Ablation Catheter is substantially equivalent.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Ms. Viorica Filimon VP Quality/Regulatory Affairs BARRX Medical, Inc. 540 Oakmead Parkway SUNNYVALE CA 94085

JAN - 8 2010

Re: K093008

Trade/Device Name: HALO99 System (including HALO90 Ablation Catheter model 90-9100 and HALO® Energy Generator model 90-9000) Regulation Number: 21 CFR §878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: December 8, 2009 Received: December 9, 2009

Dear Ms. Filimon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), " it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical

{5}------------------------------------------------

Page 2 -

Enclosure

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdaggy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ycm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

#### Indications for Use Statement

510(k) Number (if known): K093008

Device Name: HALO® System (including HALO99 Ablation Catheter model 90-9100 and HALO90 Energy Generator model 90-9000)

Indications for Use:

The HALO® System (including HALO® Ablation Catheter model 90-9100 and HALO® Energy Generator model 90-9000) is indicated for use in the coagulation of bleeding and nonbleeding sites in the gastrointestinal tract including but not limited to the esophagus.

Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert Plumer

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devi % 93008
510(k) Number

Page of of

---

**Source:** [https://fda-staging.innolitics.com/device/K093008](https://fda-staging.innolitics.com/device/K093008)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
