K092990 · Advantageous Medical Devices, LLC · GZJ · Oct 23, 2009 · Neurology
Device Facts
Record ID
K092990
Device Name
AMD 6605 TENS/NMES
Applicant
Advantageous Medical Devices, LLC
Product Code
GZJ · Neurology
Decision Date
Oct 23, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The AMD 6605 TENS/NMES is an electrotherapy device that is used for the following: Symptomatic relief and management of chronic (long-term), intractable pain and an adjunctive treatment in the management of post-surgical pain and post-traumatic acute pain. The AMD 6605 TENS/NMES is indicated for along with the following: Relaxation of muscle spasms; Prevention or retardation of disuse atrophy; Increasing local blood circulation; Muscle re-education; Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis; Maintaining or increasing range of motion
Device Story
AMD 6605 is a 2-channel portable TENS/NMES stimulator; delivers electrical stimulation via self-adhesive electrodes to skin. Device features 10 preset and 10 user-adjustable programs; clinician can mask unwanted programs to lock device in specific settings, preventing patient alteration. Input: user-selected program/intensity; Output: bi-phase rectangular electrical pulses (1Hz–3.65KHz, 30us pulse width, max 70mA). Operated by patient at home or clinic under physician prescription. Healthcare provider uses device to manage pain and muscle function; output intensity is adjusted by patient for comfort. Benefits include pain relief, muscle re-education, and prevention of atrophy/thrombosis. Powered by 4 AAA batteries.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through risk analysis, electrical safety testing (IEC 60601-1, 60601-1-2, 60601-1-4, 60601-2-10), and biocompatibility testing (ISO 10993-1, 10993-10, 10993-12).
Indicated for patients 18 years and older for symptomatic relief of chronic, intractable, post-surgical, or post-traumatic pain (TENS); and for muscle spasm relaxation, prevention/retardation of disuse atrophy, increased local blood circulation, muscle re-education, post-surgical calf stimulation to prevent venous thrombosis, and maintenance/increase of range of motion (NMES). Contraindications: patients with pacemakers or heart conditions; do not use over carotid sinus nerves, larynx, or throat muscles; do not use trans-cerebrally; do not use on undiagnosed pain.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
# Submitter's Name: Advantageous Medical Devices, LLC
Address:
1728 Virginia Beach Blvd. Suite# 115 Virginia Beach, VA 23454
OCT 2 8 2009
| Contact: | John Herod |
|----------|----------------|
| Phone: | (757) 351-0832 |
| Fax: | (800) 715-4039 |
Summary Prepared: New - November 14, 2008
Trade Name:
# AMD 6605
Manufacture: Address:
Shenzhen Dongdixin Technology Co., LTD. No.3 Factory Building, Fanshenxusheng Industrial Estate Bao'an District, Shenzhen P.R. China. Zip: 518108
Common Name: For TENS function - TENS device For NMES function - Powered Muscle Stimulator
Classification Name: For TENS functions:
Transcutaneous Electrical Nerve Stimulator for Pain Relief. 21 CFR 882.5890, Class II, Product Code: GZJ (Prescription) or Stimulator, Nerve, Transcutaneous, For Pain Relief.
Classification Name: For NMES functions:
Powered Muscle Stimulator for re-education of Muscles 21 CFR 890.5850, Class II, Product Code: IPF; (Prescription) or Stimulator, Nerve, Transcutaneous, muscle re-education
Section 5- Page 1 of 7
{1}------------------------------------------------
# Predicate Device Identification:
21CFR:890.5850 Product Code: IPF / GZJ Device Class: II Legally Marketed Device: Elpha II 3000 Manufacturer: Danmeter A/S K#: K032954
#### Device Description:
AMD 6605 TENS portion: is designed to provide electrical stimulation, for the following:
· Symptomatic relief and management of chronic (long-term), intractable pain and as an adjunctive treatment in the management of post-surgical pain and post-traumatic acute pain.
The AMD 6605 NMES portion: is a portable Neuromuscular Muscular Electrical Stimulator device for the following:
- · Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- · Increasing local blood circulation
- · Muscle re-education
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis and
- · Maintaining or increasing range of motion
The AMD 6605 is a 2 channel TENS/NMES stimulator device that is used to help reduce pain, with each channel being isolated from the other. The AMD 6605 is a programmable device that come equipped with 10 preset programs along with 10 user programs. The user programs are adjustable and can be changed to best help the patient at the doctor's recommendation and prescription settings. The program modes are preset programs that a clinician can conveniently choose from should they desire. This device has a special masking program to make the unwanted programs unavailable to the patients while locking the device in the needed relief setting. This way the patient does not receive the knowledge or ability to change the doctor's prescription settings without a clinician's consent. The AMD 6605 is programmed to default to program P-10 which contains the most common setting.
The device gives the clinician ability to store frequencies or to choose from a set of many frequencies that allow quick and easy selection, for prescription of stimulation regimen that can be later stored in any of the many available memory slots. However once the program is set the patients does not have the ability to alter the program from what the doctor or licensed practitioner has deemed to be the most appropriate program for their patient's needs.
The AMD 6605 has the following specifications: There is pulse mode with a bi-phase retangular pulse, the pulse frequency is 1 Hz to 3.65KHz, the pulse width is 30us, and the pulse amplitude is 140 levels(less than 70mA peak into 1000 ohms load to each channel). The power supply of AMD 6605 is 6.0V DC; this is supplied by four AAA type batteries.
#### Section 5- Page 2 of 7
{2}------------------------------------------------
# AMD 6605 package is comprised of the following items:
1. One TENS/NMES unit powered by 4 AAA 1 .5V batteries.
2. Two UL industry standard wires for electrodes conforming to FDA standards,
3. Four standard commercially available round 50(mm) self adhesive electrodes.
4. One UL 110 battery rechargeable unit
5. Instruction manual.
6. Quick start instruction manual.
7. Full package carrying case.
8. Personal carrying device.
#### Waveform Description
See attached Waveform pages that show different waveforms that are available for different protocols as the physician/clinician sees to be most appropriate for the patient.
Note: In order to name these modes and variations, we simply label them as variations P1, P2, P3... P 10. All wave variations have same indications of use. See waveform charts enclosed.
Note: The AMD 6605 has been preset to default to the pre-programmed mode of P-1 0 and cannot be changed unless under the orders of the treating physician or medical practitioner. This was decided for simplicity in operation, training, and operation for the patient need only turn the unit on and turn the intensity up to the desired comfort level that is best relieving the pain symptoms. The characteristics of the device can be varied and changed however only if the other programs are unmasked or unhidden by the medical practitioner.
### Indications for use
#### Indications for TENS function:
· Symotomatic relief and management of chronic (long-term), intractable pain and an adjunctive treatment in the management of post-surgical pain or post-traumatic acute pain.
#### Indications for NMES function:
- · Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- · Increasing local blood circulation
- · Muscle re-education
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- · Maintaining or increasing range of motion
There is no clinical basis or benefit for selecting one variation over the other. It is up to the physician to choose the correct placement, regimen and program to be used for chronic or acute pain relief along with whatever muscle function therapy that is being delivered to the patient ..
It is AMD's belief that this should be for distribution by Rx only by order of a physician or a medical practitioner. The predicated devices that have been shown to be like products are marketed in the same manner.
This is a Class II medical device by FDA standards and believed to be the standard format that should be followed in classification and prescribing methods.
{3}------------------------------------------------
### Intended use For TENS function:
The AMD 6605 TENS/NMES is an electrotherapy device that is used for the following:
· Symotomatic relief and management of chronic (long-term), intractable pain and an adjunctive treatment in the management of post-surgical pain or post-traumatic acute pain.
#### Intended use for NMES function:
The AMD 6605 TENS/NMES is indicated for along with the following:
- · Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- · Increasing local blood circulation
- · Muscle re-education
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- · Maintaining or increasing range of motion
The AMD 6605 TENS/NMES brings together similar TENS and Powered Muscle Stimulator functions from a previously cleared for market device; namely the Elpha Model 113000.
The AMD 6605 TENS/NIMES device is based on and is virtually identical to the Predicate device.
### Declaration of Conformity
Attached documentation with all Declaration of conformity paperwork has been included in the 510(k) submission that proves that this device is in conformity and upheld to the highest standards. The standards codes can be found in the Non-Clinical Data of this report.
#### Summary of Safety and Effectiveness
The AMD 6605 are substantially equivalent to the following device Elpha Model II3000 Manufactured by Danmeter A/S. The treatment system is substantially equivalent because of similar indication for use and unit characteristics.
The AMD 6605 does not raise new issues of safety and effectiveness based on their similarities.
This product has continually proven to be safe and effective and demonstrates that the product perform as intended.
#### Basis for substantial equivalence
AMD 6605 has basic technological characteristics that are substantially equivalent to the TENS and the EMS (Powered Muscle Stimulator) predicate devices. They all have preset output energy levels selectable by depression of a button. All units store and recall the desired treatment as selected by the physician or medical practitioner. All units have shrouded cables that comply with FDA's Final Rule "Medical Devices: Establishment of Performance Standards for Electrode Lead Wires and Patient Cables".
The AMD 6605 is equivalent in function and intended use to the predicate device. Both offer similar user options and functions. The AMD 6605 offers a simplified user interface; however it does not have a frequency range as great as the predicate devices already in market.
Section 5- Page 4 of 7
{4}------------------------------------------------
# 510(k) Summary
,
.
·
·
.
,
.
| Comparison: | AMD 6605 | ELPHA II 3000 |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The AMD 6605 TENS/NMES is an electrotherapy<br>device that is used for the following:<br>☐ Symptomatic relief and management of chronic<br>(long-term), intractable pain and an adjunctive<br>treatment in the management of post-surgical<br>pain and post-traumatic acute pain. | ELPHA II 3000 TENS/NMES is an electrotherapy<br>device that is used for the following:<br>☐ Symptomatic relief and management of chronic<br>(long-term), intractable pain and an adjunctive<br>treatment in the management of post-surgical<br>pain and post-traumatic acute pain. |
| Indications for use: | The AMD 6605 TENS/NMES is indicated for<br>along with the following:<br>☐ Relaxation of muscle spasms<br>☐ Prevention or retardation of disuse atrophy<br>☐ Increasing local blood circulation<br>☐ Muscle re-education<br>☐ Immediate post-surgical stimulation of the calf<br>muscle to prevent venous thrombosis<br>☐ Maintaining or increasing range of motion | ELPHA II 3000 TENS/NMES is indicated for<br>along with the following:<br>☐ Relaxation of muscle spasms<br>☐ Prevention or retardation of disuse atrophy<br>☐ Increasing local blood circulation<br>☐ Muscle re-education<br>☐ Immediate post-surgical stimulation of the calf<br>muscle to prevent venous thrombosis<br>☐ Maintaining or increasing range of motion |
| | The AMD 6605 TENS/NMES is an electrotherapy<br>device that is used for the following:<br>☐ Symptomatic relief and management of chronic<br>(long-term), intractable pain and an adjunctive<br>treatment in the management of post-surgical<br>pain and post-traumatic acute pain. | ELPHA II 3000 TENS/NMES is an electrotherapy<br>device that is used for the following:<br>☐ Symptomatic relief and management of chronic<br>(long-term), intractable pain and an adjunctive<br>treatment in the management of post-surgical<br>pain and post-traumatic acute pain. |
| | The AMD 6605 TENS/NMES is indicated for<br>along with the following:<br>☐ Relaxation of muscle spasms<br>☐ Prevention or retardation of disuse atrophy<br>☐ Increasing local blood circulation<br>☐ Muscle re-education<br>☐ Immediate post-surgical stimulation of the calf<br>muscle to prevent venous thrombosis<br>☐ Maintaining or increasing range of motion | ELPHA II 3000 TENS/NMES is indicated for<br>along with the following:<br>☐ Relaxation of muscle spasms<br>☐ Prevention or retardation of disuse atrophy<br>☐ Increasing local blood circulation<br>☐ Muscle re-education<br>☐ Immediate post-surgical stimulation of the calf<br>muscle to prevent venous thrombosis<br>☐ Maintaining or increasing range of motion |
| Target population: | 18 yrs of age and older | 18 yrs of age and older |
| Human factors: | Identical | |
| Contraindications for<br>TENS/NMES: | Patients with pacemakers and heart conditions<br>should NOT use<br>Do NOT use stimulation over the carotids<br>sinus nerves, the larynx or throat muscles<br>Do NOT use stimulation trans-cerebrally.<br>Do NOT use on undiagnosed pain or until<br>etiology is established. | Patients with pacemakers and heart conditions<br>should NOT use<br>Do NOT use stimulation over the carotids<br>sinus nerves, the larynx or throat muscles<br>Do NOT use stimulation trans-cerebrally.<br>Do NOT use on undiagnosed pain or until<br>etiology is established. |
The above comparisons are identical in every way for intended use, inclicalions and contraintications, above applied and the predicate devices according to FDA 510(K) requirements. : .
| Design: | | |
|---------------------|-----------------------------------------------|-----------------------------------------------|
| Type | TENS/NMES product | TENS/NMES product |
| Number of channels | 2 channel (independently controlled) | 2 channel (independently controlled) |
| Electrodes | 2 electrodes per channel (4 total) | 2 electrodes per channel (4 total) |
| Electrode type | 2" round self-adhering reusable | 2" round self-adhering reusable |
| | Amgel based electrodes, FDA approved | Amgel based electrodes |
| Wave Form | Monophasic rectangular wave-form single phase | Monophasic rectangular wave-form single phase |
| Pulse width | 30 - 300uS | 300uS |
| Max. output current | 70mA | 100mA |
| Polarity | Red = plus, Black = minus | Red = plus, Black = minus |
| Power source | battery powered | battery powered |
| Batteries | AAA 4ea. Alkaline or Rechargeable | 9V 1ea. Alkaline or Rechargeable |
| Dimensions . | 141(L) x 60(W) X 18(H) mm | 114 x 60 x 31 mm |
| Weight | 115 grams (with batteries) | 158 (with batteries) |
.
Section 5 - Page 5 of 7
.
{5}------------------------------------------------
#### Performance:
| Locking feature | keyboard lock (safety feature) | keyboard lock (safety feature) |
|---------------------|-----------------------------------------|-----------------------------------------|
| Alarm | electrode alarm | electrode alarm |
| Level Adjustment | Intensity adjustment (by 0.5 mA) | Level Adjustment (adjust by 1 mA) |
| Treatment Timer | Treatment Timer with auto shut off | Treatment Timer with auto shut off |
| Auto treat & repeat | automatic setting & repeat of treatment | automatic setting & repeat of treatment |
The design of both the applied for device and the predicate device are substantially equivalent to each other in almost every way. The AMD 6605 has basic technological characteristics that are substantially equivalent to the TENS and NMES (Powered Muscle Stimulator) predicate devices. They all have presst output energy levels selectable by depression of a button. All units store and recall the desired treatment as selected by the physician or medical . practitioner. All units have biocompatible shrouded cables that comply with the FDA's Final Rule "Medical Devices: Establishment of Performance Standards for Electrode Lead Wires and Patient Cables". All of the units are power sourced by battery. All of the devices also deliver the stimulation inpulses through biocompatible electrodes that have a contact membrane that is made from an Amgel product (US mainufactured) or equivalent materials. The predicate device uses a 9 volt battery and the AMD 6605 uses 4ea. AAA alkaline batteries.
Both devices have two independently controlled channels, with a retangular wave form, a pulse width of 300 uS. The predicate device has a stronger output current (100mA) than the AMD 6605's (70mA). The AMD 6605 gives a muscle contraction that is not quite as intense as the predicate device and is considered an equivalent device in working on relieving pain and muscle re-education ability. We have evaluated both devices side by side matching the specifications and outputs as they relate to the AMD 6605
.
·
.
| | | Safety standards met: |
|-----------------------------------|----------------------------------|---------------------------|
| Biocompatibility: | ISO 10993-1:1997 | Information NOT Available |
| Biocompatibility: | ISO 10993-12:2007 | Information NOT Available |
| Biocompatibility: | ISO 10993-10:2002 | Information NOT Available |
| General Safety<br>Requirements: | IEC 60601-1:1988+A1:1991+A2:1995 | Information NOT Available |
| Electromagnetic<br>Compatibility: | IEC 60601-1-2:2001/A1:2004 | Information NOT Available |
| Safety Requirements: | IEC 60601-1-4:2000 | Information NOT Available |
| Declaration of<br>Conformity | IEC 60601-1-6:2004 | Information NOT Available |
Section 5 - Page 6 of 7
{6}------------------------------------------------
## Non-Clinical Data:
Various reports were used in determining the substantial equivalence of the AMD 6605 to the predicate device. Including an EC Test report, A Safety Test report and a Substantial Equivalents Report.
A review of these documents will show that the AMD 6605 is substantially equivalent to the Elpha II 3000 and has passed all required Safety testing. Both devices have similar programs and functionality.
1. Risk Analysis results demonstrate acceptable and mitigated potential hazards.
2. The device will meet the following standards:
IEC 6060 1-1 IEC 6060 I-1-2 IEC 6060 1-1-4:2000 IEC 60601-2-10 Amd.1 (2001-09) AAMI/ANSI/ISO 10993-12:2007 AAMI/ANSI.ISO 10993-10:2002 AAMI/ANSI/ISO 10993-1:2003(E)
#### Conclusion:
The device is designed and labeled and verified for performance and safety. The performance is equivalent to legally marketed predicate devices. Risk Analysis does not demonstrate any design or performance potential hazards that are not adequately mitigated.
Section 5- Page 7 of 7
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Advantageous Medical Device, LLC c/o Jeff D. Rongero Third Party Reviewer, Senior Project Engineer Underwriters, Laboratory Inc. 12 Laboratory Drive Research Triangle Park, NC 27709
OCT 2 3 2009
Re: K092990
Trade Name: AMD 6605 TENS/NMES Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ; IPF Dated: September 25, 2009 Received: September 28, 2009
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
# Page 2 - Jeff D. Rongero
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
# #
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications For Use
510(k) Number(K092990):
Device Name: AMD 6605
Indications for Use:
# TENS:
• Symptomatic relief and management of chronic (long-term), intractable pain and an adjunctive treatment in the management of post-surgical pain and post-traumatic acute pain
## NMES:
- Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- · Increasing local blood circulation
- Muscle re-education
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
· Maintaining or increasing range of motion
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
Page 1 of 1
510(k) Number K092991
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.