WEDGELOC 180X (STANDARD) AND (CORTICAL) SUTURE ANCHOR WITH OPTI-FIBER SUTURES
K092914 · Medshape Solutions · MBI · Dec 9, 2009 · Orthopedic
Device Facts
Record ID
K092914
Device Name
WEDGELOC 180X (STANDARD) AND (CORTICAL) SUTURE ANCHOR WITH OPTI-FIBER SUTURES
Applicant
Medshape Solutions
Product Code
MBI · Orthopedic
Decision Date
Dec 9, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MedShape Solutions, Inc., WedgeLoc™ 2.5mm Suture Anchor with Opti-Fiber™ Sutures are intended for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist and elbow in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial collateral Ligament Reconstruction
Device Story
WedgeLoc™ 2.5mm Suture Anchor with Opti-Fiber™ Suture is a bone fixation fastener; used by orthopedic surgeons in clinical settings (OR). Device functions as a mechanical anchor to secure sutures to bone during soft tissue repair procedures. Surgeon inserts anchor into pre-drilled bone site; anchor provides fixation for sutures used to reattach tendons or ligaments. Benefits include stable soft tissue-to-bone fixation, facilitating healing in shoulder, knee, foot/ankle, hand/wrist, and elbow surgeries. Output is the physical fixation of the suture to the bone, which the surgeon uses to complete the repair. No software or electronic components involved.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and design similarities to predicate devices.
Technological Characteristics
Suture anchor for bone fixation. Metallic fastener. Dimensions: 2.5mm. Intended for use with Opti-Fiber™ suture. Mechanical fixation principle. No software, electronics, or energy source.
Indications for Use
Indicated for fixation of suture to bone in shoulder, foot/ankle, knee, hand/wrist, and elbow for various orthopedic soft tissue repairs (e.g., rotator cuff, ligament reconstruction, tendon reattachment). No specific age or gender contraindications listed.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
WedgeLoc™ 180x Suture Anchor with Opti-Fiber™ Suture (K091202)
{0}------------------------------------------------
**MEDSHAPE**
SOLUTIONS
DEC - 9 2009
Special 510(k) Page 9 of 88
ગુજરાત
p ト
## Special 510(k) Summary
Special 510(k) Number: K092914
September 18th, 2009 Date Prepared:
This Special 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
- A. Submitter: MedShape Solutions, Inc. (MSS) 1575 Northside Drive, Suite 440 Atlanta, Georgia 30318 Registration #10026693
- B. Company Contact: Jack Griffis Vice President, Research & Development (404) 583-6889 (cell) (404) 249-9158 (fax) Jack. Griffis@MedShapeSolutions.com
- C. Device Information: Trade Name: WedgeLoc™ 2.5mm Suture Anchor with Opti-Fiber™ Suture Common Name: Suture Anchor
- Fastener, Fixation, Non-degradable, Soft Tissue D. Classification Name: HWC/MBI 21 CFR 888.3040
- E. Predicate Device(s): MSS, WedgeLoc™ 180x Suture Anchor with Opti-Fiber™ Suture, K091202 Arthrex PushLock™ 2.5mm Suture Anchor, K063479
- E. Labeling and Intended Use: No changes to the labeling or Instructions for Use have been made to the submitted information of the MedShape predicate per K091202.
The proposed WedgeLoc™ 2.5mm Suture Anchor and Opti-Fiber™ Suture has the same intended uses as our previously cleared predicate device in K091202. In particular, both devices are indicated for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist and elbow in the following procedures:
- Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Tenodesis. Acromio-Clavicular Separation Bicens Repair. Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
{1}------------------------------------------------
- Lateral Stabilization, Medial Stabilization, Achilles Tendon Foot/Ankle: Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair
- Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair. Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Ligament Hand/Wrist: Scapholunate Reconstruction, Ulnar Collateral Ligament Reconstruction. Radial Collateral Ligament Reconstruction
- Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial collateral Ligament Reconstruction
- G. Substantial Equivalence Summary:
The proposed WedgeLoc™ 2.5mm Suture Anchor is an extension of sizes offered and is substantially equivalent to the WedgeLoc™ 180x Suture Anchor with Opti-Fiber™ Suture, cleared under K091202, and the predicate Arthrex PushLock™ 2.5mm Suture Anchor, cleared under K063479, in which the basic features and intended uses are the same. In addition, the technological characteristics of the WedgeLoc™ 180X and the WedgeLoc™ 2.5mm Suture Anchor are equivalent.
Any differences between the WedgeLoc™ 2.5mm Suture Anchor and the MedShape predicate WedgeLoc™ 180x or Arthrex PushLock™ Suture Anchor are considered minor and do not raise questions concerning safety and effectiveness. Based on the information submitted. MedShape Sofutions. Inc. has determined that the proposed WedgeLoc™ 2.5mm Suture Anchor is substantially equivalent to the currently marketed device.
Jack Griffit Vice President. Research & Development
Image /page/1/Picture/11 description: The image shows the logo for Medshape Solutions. The word "MEDSHAPE" is in large, bold, sans-serif font. Below it, the word "SOLUTIONS" is in a smaller, sans-serif font. A line separates the two words.
1575 Northside Drive, Suite 440 Atlanta, GA 30318 404-249-9156 phone 404-249-9158 fax
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized image of an eagle with three lines representing its wings and tail feathers. The eagle is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
DEC - 9 2009
MedShape Solutions c/o Mr. Jack Griffis Vice President. Research & Development 1575 Northiside Drive, Suite 440 Atlanta, Georgia 30318
Re: K092914
Trade/Device Name: WedgeLoc™ 2.5mm Suture Anchor and Opti-Fiber™ Sutures Regulation Number: 21 CFR 888.3040 . Regulation Name: Smooth or threaded metallic bone fixation fastner Regulatory Class: Class II Product Code:MBI, HWC Dated: November 16, 2009 Received: November17, 2009
Dear Mr. Griffis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Jack Griffis
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdaggov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Jonata Jr
Image /page/3/Picture/7 description: The image contains two words, "for" and "R". The word "for" is written in cursive and is located on the left side of the image. The letter "R" is capitalized and is located on the right side of the image.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
## 510(k) Number:
Device Name: WedgeLoc™ 2.5mm Suture Anchor and Opti-Fiber™ Sutures
## Indications for Use:
The MedShape Solutions, Inc., WedgeLoc™ 2.5mm Suture Anchor with Opti-Fiber™ Sutures are intended for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist and elbow in the following procedures:
- Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Shoulder: Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair. Capsular Shift or Capsulolabral Reconstruction
- Lateral Stabilization, Medial Stabilization, Achilies Tendon Repair, Hallux Foot/Ankle: Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair
- Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction
- Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial collateral Ligament Reconstruction
Prescription Use × (Part 21 CFR 801 Subpart D)
510(k) Number K092914
.. '
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign Off) | Page 1 of 1 |
|-----------------------------------------------------------|-------------|
| Division of Surgical, Orthopedic, and Restorative Devices | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.