← Product Code [BSZ](/productcode/BSZ) · K092864

# AESPIRE 7900, AESPIRE VIEW (K092864)

_Datex-Ohmeda Inc., A Division of Instrumentarium C · BSZ · Jan 12, 2010 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K092864

## Device Facts

- **Applicant:** Datex-Ohmeda Inc., A Division of Instrumentarium C
- **Product Code:** [BSZ](/productcode/BSZ.md)
- **Decision Date:** Jan 12, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5160
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation. The devices are not suitable for use in a MRI environment.

## Device Story

Anesthesia system providing inhalation anesthesia and ventilatory support; utilizes 7900 microprocessor-based, electronically controlled, pneumatically driven ventilator. Inputs: medical gases (O2, air, N2O), vaporizer settings, user-defined ventilation parameters. Operation: system delivers set gas flows; ventilator manages breathing patterns via bellows; monitors inspired O2, airway pressure, exhaled volume. Aespire View variant adds color display, high-performance CPU, optional electronic Total Fresh Gas Flow (TFS) monitoring, and PCV-VG mode. Used in clinical settings by trained medical professionals. Output: controlled ventilation, gas delivery, real-time monitoring data on display. Healthcare providers use output to adjust anesthesia/ventilation, ensuring patient stability during surgery. Benefits: precise gas delivery, automated ventilation support, reduced risk of hypoxic mixtures.

## Clinical Evidence

No clinical data; modifications verified through non-clinical bench testing, including software validation, electrical safety, electromagnetic compatibility, and biocompatibility testing.

## Technological Characteristics

Microprocessor-based, pneumatically driven anesthesia ventilator. Features: Selectatec manifold for Tec 4/5/6/6+/7 vaporizers; RS-232 serial communication. Aespire View variant: color display, high-performance CPU, increased memory. Software v6.X supports PCV-VG and optional Total Flow Sensing. Biocompatibility testing performed on flow sensor module. Electrical safety/EMC testing completed.

## Regulatory Identification

A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).

## Predicate Devices

- GE Datex-Ohmeda Aespire 7900 Anesthesia Systems ([K050626](/device/K050626.md))
- GE Datex-Ohmeda Aisys ([K090233](/device/K090233.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, with some decorative flourishes extending from the circle at various points.

IC07
GE Healthcare

Datex-Ohmeda Inc. P.O. Box 7550 Madison, WI 53703 USA

JAN 1 2 2010

# Premarket Notification 510(k) Summary

As required by section 807.92 GE Datex-Ohmeda Aespire Anesthesia System

# GENERAL COMPANY INFORMATION as required by 807.92(a)(1) COMPANY NAME/ADDRESS/PHONE/FAX:

GE Healthcare Datex-Ohmeda, Inc. PO Box 7550 Madison, WI 53707 USA · Tel: 608-221-1551 Fax: 608-223-2476

### NAME OF CONTACT:

Ms. Adrienne Lenz, RAC Ms. Karla Krause (alternate)

#### DATE:

September 16, 2009

### DEVICE NAME as required by 807.92(a){2) TRADE NAME:

GE Datex-Ohmeda Aespire Anesthesia System

COMMON NAME:

Gas Machine, Anesthesia

#### CLASSIFICATION NAME:

Anesthesiology, 73 BSZ, 21 CFR 868.5160 Gas Machine, Anesthesia

General Electric Company
Datex-Ohmeda, Inc.

{1}------------------------------------------------

# NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)

The original GE Datex-Ohmeda Aespire 7900 Anesthesia Systems received 510(k) clearance under the submittal number K050626.

Both the GE Datex-Ohmeda Aespire 7900 Anesthesia System and the GE Datex-Ohmeda Aisys (most recent clearance K090233) are used as predicates.

#### DEVICE DESCRIPTION as required by 807.92(a)(4)

The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The systems are to be used only by trained and qualified medical professionals.

The Aespire 7900 and Aespire View supply set flows of medical gases to the breathing system. A large selection of frames, gases, and vaporizers are available to give the user control of the system configuration. They are available in trolley and pendant models, with two or three gases, two vaporizer positions and up to three cylinder connections. All models connect to oxygen and can additionally connect with up to two optional gases (air, N2O). The Aespire systems accept Tec 4, Tec 5, Tec 6, Tec 6+ and Tec 7 vaporizers on a Selectatec manifold. Safety features are designed to decrease the risk of hypoxic mixtures and complete power or sudden gas supply failures. The Aespire View family member provides optional electronic Total Fresh Gas Flow (TFS) monitoring. The Aespire View also features a color display, while the Aespire 7900 uses a monochromatic display.

The Datex-Ohmeda 7900 Anesthesia Ventilator is used in this family of Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control (VCV) Mode, Pressure Control (PCV) Mode (Optional), Synchronized Intermittent Mandatory Ventilation with Pressure Support Ventilation (SIMV/PSV) Mode, Pressure Support with Apnea Backup (PSVPro) Mode (Optional), Synchronized Intermittent Mandatory Ventilation with Pressure Control (SIMV-PC) Mode (Optional), and Pressure Control Ventilation - Volume Guaranteed (PCV-VG) mode (Optional on Aespire View variant only),

{2}------------------------------------------------

### INTENDED USE as required by 807.92(a)(5)

The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation. The devices are not suitable for use in a MRI environment.

## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)

This submission introduces the Aespire View variant to the family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator. The Aespire 7900 variant remains available. The GE Datex-Ohmeda Aespire View Anesthesia System includes the following significant changes compared to the Aespire 7900:

- Software version (6.X) which includes pressure control ventilation- volume guarantee . (PCV-VG) ventilation mode and Total Flow Sensing monitoring (optional)
- . Total Flow Sensor Module (optional)
- New display unit including a color display, high performance CPU and increased system ● memory

# SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)

The GE Datex-Ohmeda Aespire View Anesthesia System has been thoroughly tested through verification of specifications and validation, including software validation. Biocompatibility testing was conducted for the total flow sensor module. Electrical safety and electromagnetic compatibility testing were also completed.

### SUMMARY OF CLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(2)

The modifications to the family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator did not require clinical testing.

### CONCLUSION:

The summary above shows that, as compared to the predicate device, there are no new questions of safety and effectiveness for the introduction of the Aespire View to the family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator.

{3}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the edge. Inside the circle is a stylized symbol that resembles three bird heads or wings layered on top of each other. The image is in black and white and appears to be a scan or photocopy due to its slightly grainy texture.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

JAN 12 2010

JAN 1 2 2010

Ms. Adrienne Lenz Regulatory Affairs Director GE Healthcare Datex-Ohmeda, Incorporated P.O. Box 7550 3030 Ohmeda Drive Madison, Wisconsin 53707-7550

Re: K092864

Trade/Device Name: GE Datex-Ohmeda Aespire Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: December 18, 2009 Received: December 22, 2009

Dear Ms. Lenz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require, approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

#### Page 2 - Ms. Lenz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. · Director

Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): K.O G. Z86 Y 11 - 1 -

Device Name: GE Datex-Ohmeda Aespire Anesthesia System

Indications For Use:

The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or, pressure control ventilation. The devices are not suitable for use in a MRI environment.

Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schultheis

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K092864

---

**Source:** [https://fda-staging.innolitics.com/device/K092864](https://fda-staging.innolitics.com/device/K092864)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
