K092840 · Instrumed International, Inc. · HCZ · Dec 3, 2009 · Obstetrics/Gynecology
Device Facts
Record ID
K092840
Device Name
INSTRUMED FORCEPS OBSTETRICAL/GYNAECOLOGICAL
Applicant
Instrumed International, Inc.
Product Code
HCZ · Obstetrics/Gynecology
Decision Date
Dec 3, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
INSTRUMED gynaecological forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynaecological examination.
Device Story
Instrumed gynaecological forceps are reusable manual surgical instruments designed for use in vaginal and abdominal hysterectomy procedures. The device consists of two blades and handles, featuring a ratchet closure mechanism to adjust tension and various jaw designs (longitudinal grooves, teeth, or cross serrations) to securely hold tissue without slipping or laceration. Surgeons select specific models based on anatomical site and procedure requirements. The device is operated by physicians in a clinical or surgical setting to facilitate tissue manipulation during gynaecological examinations or surgeries. It is supplied non-sterile and requires sterilization before use.
Clinical Evidence
Bench testing only. Biocompatibility review and sterilization studies were successfully completed.
Technological Characteristics
Reusable manual surgical forceps constructed from ASTM F 899-07 standardized stainless steel. Features include ring handles with ratchet closure, various jaw configurations (grooves, teeth, cross serrations), and varying lengths/bends. Supplied non-sterile.
Indications for Use
Indicated for use in patients undergoing gynaecological examinations or procedures requiring the pulling, grasping, or compression of tissue.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
Mader Surgical Forceps
Meisterhand Surgical Forceps
Miltex Surgical Forceps
Multiple Surgical Forceps
Sparta Surgical Forceps
Vantage Surgical Forceps
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
## 510(k) Summary as required by section 807.92(c)
date prepared 10/07/2009
## Submission Applicant:
INSTRUMED INTERNATIONAL, INC. 626 Cooper Court Schaumburg, IL 60173
### Establishment Registration Number:
1421101
Official Correspondent:
Mr. Berndt Fetzer INSTRUMED INTERNATIONAL, INC. 626 Cooper Court Schaumburg, IL 60173
Phone: 847-908-0292
### Trade name:
Instrumed Forceps Gynaecological
### Common name:
Various Gynaecologic specialized manual forceps:
·
- Gynecological Forceps: Hysterectorny Clamp, Parametrium Clamp, Z-Clamp, Placenta Forceps
#### Classification name:
21 CFR PART 884 -- OBSTETRICAL AND GYNECOLOGICAL DEVICES Subpart E-- Obstetrical and Gynecological Surgical Devices
Sec. 884.4530 Obstetric-gynecologic specialized manual instrument. Product Code HCZ
### Regulation Description
Obstetric-gynecologic specialized manual instrument/forceps
## Substantial Equivalence Claims:
| HCZ | multiple FORCEPS, SURGICAL, GYNECOLOGICAL |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Preamendment | Applicant AESCULAP, INC. Specification Developer |
| Preamendment | Mader Surgical Forceps; Meisterhand Surgical Forceps; Miltex Surgical Forceps;<br>Multiple Surgical Forceps; Sparta Surgical Forceps; Vantage Surgical Forceps<br>Applicant MILTEX, INC. Manufacturer<br>Owner/Operator: INTEGRA LIFESCIENCES CORPORATION |
DEC - 3 2009
{1}------------------------------------------------
## 510(k)Summary
### Description of the Device:
Instrumed gynaecological forceps are heavy forceps which have special designed jaws to hold securely without slipping during vaginal and abdominal hysterectorny procedures, yet do not lacerate the tissue to which it has been applied to even when it is necessary to exert traction.
The Instrumed gynaecological forceps are reusable surgical instruments.
To ensure the multi-purpose use of this device, many different models are available. The differences can be as follows:
- length and bending of the forceps and jaws .
- ring handles with a ratchet closure to adjust the amount of tension applied. �
- jaw design, option of longitudinal grooves, teeth or cross serrations .
- the choice of jaw style depends on the surgeon's preference .
The surgeon chooses the gynaecological forceps based on the anatomy of the site and the designs desired are based on the type of the surgical procedure.
Instrumed gynaecological forceps are made of the ASTM F 899-07 standardized Stainless Steel.
The instruments are offered in non-sterile condition.
### Indications for Use:
INSTRUMED gynaecological forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynaecological examination.
#### Comparison with Predicate Device:
The results of non-clinical and bench testing indicates that the new device is completely comparable to the predicate devices. Biocompatibility review and sterilization studies were successfully cry onleted.
The instrumed product is similiar to the predicate device in terms of technical characteristics, design, Indications for Use, target population, where it is used, performance, biocompatibility, sterilization method, mechanical safety characteristics as well as sizes and configurations. Therefore it can be deemed substantially equivalent for its indicated use.
#### Summary
The presented data that was conducted on the INSTRUMED gynaecological forceps shows in its results and in comparison to the predicate devices that the products are absolutely safe and effective for their intended use and do not raise any new questions regarding safety and effectiveness. The used materials are well researched and do not raise new questions regarding safety and effectiveness of the finished product.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
## DEC - 3 2009
Mr. Michael Massong QA/RA Director Instrumed International, Inc. 626 Cooper Court SCHAUMBURG IL 60173
Re: K092840
Trade/Device Name: Instrumed Forceps Gynaecological Regulation Number: 21 CFR §884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: HCZ Dated: November 10, 2009 Received: November 10, 2009
Dear Mr. Massong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR. Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
| 510(k) Number (if known): | K09 2840 |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Instrumed Forceps Gynaecological |
| Indications For Use: | INSTRUMED gynaecological forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynaecological examination. |
| Prescription Use (Part 21 CFR 801 Subpart D) | X |
| | AND/OR |
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hulst Rums
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number.
> Page 1 of ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.