← Product Code [EKX](/productcode/EKX) · K092833

# ENDOMAX CORDLESS ENDODONTIC HANDPIECE (K092833)

_Dentamerica · EKX · Jan 7, 2010 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K092833

## Device Facts

- **Applicant:** Dentamerica
- **Product Code:** [EKX](/productcode/EKX.md)
- **Decision Date:** Jan 7, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4200
- **Device Class:** Class 1
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Endomax™ is a cordless handpiece used primarily for root canal enlargement. This application area extends to endodontic procedures using a root canal instrument which is intended by the manufacturer for use in the mechanical and rotary preparation of root canals.

## Device Story

Endomax is a cordless, battery-powered endodontic handpiece for root canal enlargement and mechanical/rotary preparation. System components include drive handpiece, rechargeable battery, contra-angle head, and charging station. Operated by clinicians in dental settings. Features include automatic start/stop and automatic torque reverse to enhance control, flexibility, and safety during procedures. Device provides portable operation, eliminating cord-related constraints. Clinician uses handpiece to drive rotary instruments; automatic functions adjust operation based on torque feedback to prevent instrument damage and improve treatment accuracy.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and design similarities to predicate devices.

## Technological Characteristics

Cordless, battery-powered rotary handpiece. Materials: brass, silicon, Type II plastic. Modular system: drive handpiece, rechargeable battery, contra-angle head, charging station. Features automatic start/stop and automatic torque reverse functionality.

## Regulatory Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

## Predicate Devices

- ENDO-MATE TC Cordless Handpiece ([K990682](/device/K990682.md))
- Tri Auto ZX Cordless Endodontic Handpiece ([K970339](/device/K970339.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(k) Summary

Submission Date: September 8, 2009

"** ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ;

Submitter's Information: Dentamerica Inc.® 18688 E. San Jose Ave. Industry, CA 91748 USA

Contact Person: Eric Huang, (Product Manager) Phone: 626-912-1388 Fax: 626-913-0510

JAN - 7 2010

### Establishment Registration Number: 2024949

Trade Name: Endomax™ Common Name: Cordless Endodontic Handpiece Regulation Number: 21 CFR 872.4200 Regulation Name: Handpiece, Direct Drive, AC-powered Regulatory Class: Class I Product Code: EKX

Performance Standard: None established under section 514 Reason for 510(k): New finished device Special controls: No applicable mandatory performance standards or special controls exist for this device.

### Substantially Equivalent Legally Marketed Devices (predicates):

ENDO-MATE TC Cordless Handpiece – K990682 Tri Auto ZX Cordless Endodontic Handpiece - K970339

5-1

{1}------------------------------------------------

## Indications for Use:

The Endomax" is a cordless handpiece used primarily for root canal enlargement. This application area extends to endodontic procedures using a root canal instrument which is intended by the manufacturer for use in the mechanical and rotary preparation of root canals.

# Technological Characteristics and Substantial Equivalence:

The Endomax" Cordless Endodontic Handpiece has a number of substantially equivalent (SE) predicate devices. These predicate devices use the same technological methods to perform mechanical and rotary preparation of root canals with a cordless handpiece. A modular electrical system consisting of a drive handpiece, a rechargeable battery, contraangle head and a charging station are the main components of the predicate devices. The materials used in this device are composed of brass, silicon and Type II plastic as in the substantially equivalent devices.

The cordless endodontic handpiece introduces the convenience of portable operation without a cumbersome cord. The benefits of this can be demonstrated clinically. First, a cordless handpiece offers greater control and flexibility. In addition, cordless endodontic handpieces offer automatic functions such auto start/stop and auto torque reverse, enabling the clinician to perform safer and more accurate endodontic treatment.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Mr. Eric Huang Manager Dentamerica 18688 East San Jose Avenue Industry, California 91748

JAN - 7 2010

Re: K092833

Trade/Device Name: Endomax Cordless Endodontic Handpiece Regulation Number: 21CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EKX Dated: December 14, 2009 Received: December 23, 2009

Dear Mr. Huang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2- Mr. Huang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRF1's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

th for

Anthony D. Watson, BS, MS, MBA Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K092833

Device Name: Endomax™ Cordless Endodontic Handpiece

Indications for Use:

The Endomax™ is a cordless handpiece used primarily for root canal enlargement. This application area extends to endodontic procedures using a root canal instrument which is intended by the manufacturer for use in the mechanical and rotary preparation of root canals. .

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Betz MS for Dr. K.P. Mulay (Acting)
Division Sign-Off

1.vision of Anesthesiology, General Hospital Tection Control, Dental Devices

092833 i10(k) Number:

---

**Source:** [https://fda-staging.innolitics.com/device/K092833](https://fda-staging.innolitics.com/device/K092833)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
