K092823 · STERIS Corporation · LGZ · Dec 18, 2009 · General Hospital
Device Facts
Record ID
K092823
Device Name
AMSCO WARMING CABINET
Applicant
STERIS Corporation
Product Code
LGZ · General Hospital
Decision Date
Dec 18, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Amsco Warming Cabinet is designed to raise the temperature of blankets, linens and sterile surgical irrigation solutions and IV solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications. The Amsco Warming Cabinet is designed to hold a combination of flasks and/or dry goods.
Device Story
Amsco Warming Cabinet stores and warms sterile IV solutions, surgical irrigation solutions, linens, and blankets. Device uses electric heater and fan blower for convection heating. User interface includes digital push-button keypad, power switch, and LCD temperature display. Features include temperature lock-out, door ajar indicator, over-temperature alarm (visual/audible), and data port for temperature retrieval. Used in hospitals and surgical outpatient centers. Healthcare staff operate device to ensure fluids/linens reach acceptable temperatures for patient care. Benefits include maintaining optimal temperature for clinical applications, reducing risk of hypothermia during procedures. Device configurations include single/double units, wall-mounted or free-standing/mobile, with stainless steel, ABS plastic, and laminated galvanized steel construction.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Heating: Electric heater and fan blower (convection). Materials: Stainless steel, ABS plastic, laminated galvanized steel. Dimensions: 18" or 24" depth. Connectivity: Data port for temperature retrieval. Controls: Digital push-button keypad, LCD display. Safety: Over-temperature alarm (visual/audible), automatic heater shut-off, temperature lock-out. Power: 110/120 Vac or 220/240 Vac, 50/60 Hz.
Indications for Use
Indicated for warming blankets, linens, sterile surgical irrigation solutions, and IV solutions for use in surgical, obstetrical, emergency, critical care, and other healthcare applications.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the text "Public Health Service" in a bold, sans-serif font. The text is arranged on a single line, with each word clearly legible. The overall appearance is simple and straightforward, focusing on conveying the name of the organization.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Robert F. Sullivan Senior Director FDA Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060
## JAN 1 0 2017
Re: K092823
Trade/Device Name: Amsco Warming Cabinet Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LGZ Dated: September 11, 2009 Received: September 14, 2009
Dear Mr. Sullivan:
This letter corrects our substantially equivalent letter of December 18, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
Page 2 - Mr. Robert F. Sullivan
CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name:
Amsco Warming Cabinet
Indications For Use:
The Ansco Warming Cabinet is designed to raise the temperature of blankets, lineus and sterile surgical irrigation solutions and IV solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Austin for Gz
(Division Sign-Off) Division of Anesthesiology, General Hospital Intection Control, Dental Devices
510(k) Number: 692823
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a handwritten text that appears to be a word or a combination of letters and symbols. The characters are written in a cursive style, with some letters connected to each other. The writing is dark and bold, contrasting with the white background, making it easily readable.
# STERIS.
DEC 1 8 2009
## 510(k) Summary For Amsco Warming Cabinet
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459
Contact:
Robert F. Sullivan Senior Director FDA Regulatory Affairs Telephone: 440-392-7695 Fax No: 440-357-9198
Summary Date:
September 11, 2009
STERIS Corporation • 5960 Heisley Road • Mentor. OH 44060-1834 USA • 440-354-2600
September 11, 2009
{4}------------------------------------------------
#### Device Name 1.
| Trade Name: | Amsco Warming Cabinet |
|----------------------|-----------------------------|
| Common/usual Name: | Warming Cabinet |
| Classification Name: | Warmer, Irrigation Solution |
#### 2. Predicate Device
STERIS Amsco Warming Cabinet (Pre-Amendment) Enthermics EC-7701 Fluid Warming Cabinet, K993797, January 20, 2000
Warmer, Cabinet
#### 3. Description of Device
The Amsco Warming Cabinet is designed to store and warm sterile IV solutions, surgical irrigation solutions, linens and/or blankets to an acceptable level for hospital and surgical outpatient center applications.
The upper compartment of the 18" (457mm) deep model holds up to 24(1-liter) liquid bags or bottles and 14 (1-liter) IV solution bags; the lower compartment holds up to 72 (1-liter) liquid bags or bottles.
The upper compartment of the 24" (610mm) deep model holds up to 30 (1-liter) surgical flasks; the lower compartment holds up to 90 (1-liter) surgical flasks.
#### 4. Intended Use
The Amsco Warming Cabinet is designed to raise the temperature of blankets, linens and sterile surgical irrigation solutions and IV solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications.
#### ડ. Description of Safety and Substantial Equivalence
The Amsco Warming Cabinet is nearly identical to the two predicate devices in all material respects. A table comparing the technological characteristics of the proposed Amsco Warming Cabinet to the predicates is provided in Table 5-1.
{5}------------------------------------------------
### STERIS TRADITIONAL 510(k) PREMARKET SUBMISSION STERIS Amsco Warming Cabinet
| Features | Technological Characteristics | | |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Amsco Warming Cabinet<br>(IV Solution Capability) | Amsco Warming Cabinet<br>(Pre-amendment) | EC-7701 Fluid Warming<br>Cabinet (K993797) |
| Intended Use | The Amsco Warming Cabinet<br>is designed to raise the<br>temperature of blankets, linens<br>and sterile surgical irrigation<br>solutions and IV solutions to<br>an acceptable level for various<br>surgical, obstetrical,<br>emergency, critical care and<br>other healthcare applications.<br>The Amsco Warming Cabinet<br>is designed to hold a<br>combination of flasks and/or<br>dry goods. | Amsco Warming Cabinet is for<br>heating flasked solutions,<br>blankets and similar clinical<br>articles. | The Enthermics Medical<br>Systems EC-7701 Fluid<br>Warming Cabinet is<br>designed to safely store<br>and warm irrigation fluids<br>or injection fluids in<br>accordance with the<br>recommended warming<br>temperatures and storage<br>times stated in the fluid<br>manufacturer's labeling. |
| Heating<br>System | Electric heater and fan blower<br>(Convection heating) | Steam heat or Electric heater<br>blanket and fan blower<br>(Convection heating) | Fully insulated<br>electrothermal cable array<br>(Convection heating) |
| Unit<br>Configuration | Single/Double | Single/Dual | Single |
| Unit Depth | 18" or 24" | 18" or 24" | |
| Model | Wall or Counter | Wall or Counter | Wall |
| Interior and<br>Exterior<br>Surfaces | Stainless Steel, ABS Plastic<br>and laminated galvanized steel | Stainless Steel | Stainless Steel |
| Installation | Free-Standing (mobile) or<br>Recessed | Open-Mounted or Recessed | Free-standing (mobile) or<br>Recessed |
| Door | Stainless Steel (Solid and<br>Glass) | Stainless Steel | Stainless Steel (Glass) |
| Cabinet<br>Storage<br>Capacity | 18" upper / single chamber -<br>3.2 cu ft - up to 24 (1-liter)<br>bottles<br><br>18" lower chamber - 8.5 cu ft<br>- up to 72 (1-liter) bottles<br><br>24" upper / single chamber -<br>4.3 cu ft - up to 30 (1-liter)<br>bottles<br><br>24" lower chamber - 11.6 cu ft<br>- up to 90 ( 1-liter) bottles | Dual Compartment Model -<br>Two shelves<br>15 flasks - 18" upper<br>45 flasks - 18" lower<br>20 flasks - 24" upper<br>60 flasks - 24" lower<br><br>Single Compartment Model -<br>One shelf<br>15 flasks - 18"<br>20 flasks - 24" | The cabinet is equipped<br>with three (3) white,<br>epoxy-coated wire baskets,<br>each with a 24 liter<br>capacity. |
| Features | | | |
| Cabinet<br>Volume | Amsco Warming Cabinet<br>(IV Solution Capability)<br>18" upper chamber - 3.1 cu ft<br>24" upper chamber - 4.2 cu ft<br>18" lower chamber - 8.9 cu ft<br>24" lower chamber - 12 cu ft | Amsco Warming Cabinet (Pre-amendment)<br>18" upper chamber - 3.1 cu ft<br>24" upper chamber - 4.2 cu ft<br>18" lower chamber - 8.9 cu ft<br>24" lower chamber - 12 cu ft | EC-7701 Fluid Warming<br>Cabinet (K993797) |
| Controls | Digital Push Button keypad /<br>power switch / Digital LCD<br>temperature display / mode<br>selection buttons / door ajar<br>indicator / Over-temperature<br>light for each compartment /<br>Data port for retrieval of<br>stored temperatures. | Thermostat / power switch /<br>fuse with indicating light / color<br>coded temperature selector | Electronic control consists<br>of 4 digit LED display,<br>on/off key, increase and<br>decrease keys, integrated<br>lock feature and a series of<br>prompt sequence<br>indicators. |
| Software | Unit contains software | Not Applicable | Not applicable |
| Temperature<br>Selection<br>Range | 90°F (32°C) to 160°F (71°C) | 95°F to 150°F (35°C to 65°C) | 90°F (32°C) to 150°F<br>(66°C) |
| Temperature<br>Lock | Temperature lock-out function<br>to prevent unauthorized<br>temperature changes. | Not Available | The device allows the user<br>to "lock" the mode (IRR or<br>INJ) and temperature<br>setting controls.<br>Available as an option |
| Door Lock | All configurations will be<br>equipped with either a manual<br>mechanical door lock or<br>optional electronic door lock<br>system for each compartment | Available by SSQ (Special<br>Sales Quote) only | Available as an option |
| Over<br>Temperature<br>Alarm Point | Visual and audible alarm if<br>unit has a chamber<br>temperature greater than 10°F<br>(5.5°C) above set temperature.<br>In the event of an over temp<br>condition, sensors<br>automatically turns off the<br>heater(s). | Visual alarm if unit has a<br>chamber temperature greater<br>than 12°F above set<br>temperature. In the event of an<br>over temp condition, sensors<br>automatically turns off the<br>heater(s). | Visual and audible alarm if<br>unit has a chamber<br>temperature greater than<br>10°F (5.5°C) above set<br>temperature. In the event<br>of an over temp condition,<br>the heating system shuts<br>down. |
| Voltage<br>Requirements | 110/120 Vac, 220/240 Vac<br>nominal, 50/60 HZ | Electric Model: 110/120 Vac,<br>220/240 Vac nominal, 50/60<br>HZ<br>Steam Model: 120 VAC single<br>phase | 125 Vac, 60 HZ, 1 ph |
## Table 5-1 Summary of the Proposed Device and Predicate Devices Technological Characteristics
{6}------------------------------------------------
## STERIS TRADITIONAL 610(k) PREMARKET SUBMISSION
STERIS Amsco Warming Cabinet
.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.