AUTO DIGITAL UPPER ARM BLOOD PRESSURE MONITOR, MODEL KP-7770

K092806 · K-Jump Health Co., Ltd. · DXN · Oct 9, 2009 · Cardiovascular

Device Facts

Record IDK092806
Device NameAUTO DIGITAL UPPER ARM BLOOD PRESSURE MONITOR, MODEL KP-7770
ApplicantK-Jump Health Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateOct 9, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

The device is intended to measure the systolic and diastolic blood pressure and pulse rate (heart rate) by using an inflating cuff wrapped around the arm. The device is indicated for use with adults. The device displays irregular pulse detection (IPD) icon on LCD screen when irregular pulse is detected during blood pressure measurement.

Device Story

Electronic arm blood pressure monitor; uses inflating cuff, bellows sensor, internal air pump, leakage/exhaust valves. Measures systolic/diastolic blood pressure and pulse rate; detects irregular pulses. Operated by patient or clinician; displays results on LCD. Uses MOMI or fuzzy logic algorithms to process pressure signals. Provides non-invasive monitoring for adult patients.

Clinical Evidence

No clinical data. Substantial equivalence based on technological characteristics and compliance with recognized consensus standards (AAMI/ANSI SP10, IEC 60601-1, IEC 60601-1-2, EN61000-4-2, EN61000-4-3).

Technological Characteristics

Electronic blood pressure monitor; inflating cuff, LCD, bellows sensor, air pump, valves, battery. Uses MOMI or fuzzy logic algorithms. Complies with AAMI/ANSI SP10 (2002/A1:2003), IEC 60601-1 (1988), IEC 60601-1-2 (2002), EN61000-4-2, and EN61000-4-3.

Indications for Use

Indicated for use in adults to measure systolic/diastolic blood pressure and pulse rate. Includes irregular pulse detection (IPD) feature.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K092806 # 510(k) Summary ### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Mr. Daniel C.M. Tseng K-jump Health Co., Ltd. No. 56 Wu Kung 5th Road Wu Ku Industrial Park Taipei Hsien Taiwan Phone: + 886 2 22991378 Facsimile: + 886 2 22991386 Date Prepared: August 31, 2009 #### Name of Device Arm Blood Pressure Monitor Model KP-7770 #### Name/Address of Sponsor K-jump Health Co., Ltd. No. 56 Wu Kung 5th Road Wu Ku Industrial Park Taipei Hsien Taiwan Phone: + 886 2 22991378 Facsimile: + 886 2 22991386 Contact Person: Jason Cheng Common or Usual Name Arm Blood Pressure Monitor #### Classification Name Class II § 870.1130; System, Measurement, Blood Pressure, Non-Invasive #### Predicate Device - K jump Health Co., Ltd.'s Arm Blood Pressure Monitor Smartlogic Model KP-7500 and Fuzzylogic Model KP-7500D, K083753. ### Intended Use/Indications for Use OCT = 9 2009 i of 3 {1}------------------------------------------------ The device is intended to measure the systolic and diastolic blood pressure and pulse rate (heart rate) by using an inflating cuff wrapped around the arm. The device is indicated for use with adults. The device displays irregular pulse detection (IPD) icon on LCD screen when irregular pulse is detected during blood pressure measurement. #### Technological Characteristics The device is an electronic blood pressure monitor. The device consists of an inflating cuff, a LCD display, a bellows sensor, an internal air pump, a leakage valve, an exhaust valve, a battery power resource, and keys for operation. #### Performance Data The Arm Blood Pressure Monitor Model KP-7770 complies with the following FDA recognized consensus standards, to the extent that these standards are applicable to this device: - . AAMI/ANSI SP10 (2002 / A1:2003); - IEC 60601-1 (1988); and . - IEC 60601-1-2 (2002). The KP-7770 also complies with the following additional standards- - . EN61000-4-2 (1995); and - EN61000-4-3 (2002). pg 2 of 3 {2}------------------------------------------------ Clinical performance of the modified device is remain unchanged; therefore another clinical trial for this device, KP-7770, is not required. #### Substantial Equivalence The KP-7770 has the same intended use and indications for use as K-Jump's Arm Blood Pressure Monitor Model KP-7500 and KP-7500D. The KP-7770 has the same technological characteristics as the KP-7500 except that the KP-7770 can use MOMI or fuzzy logic method to measure blood pressure rather than the KP-7500 which can only use smart logic method or the KP-7500D which can only use fuzzy logic method. Thus, the minor differences between the KP-7770 and the predicate devices are not new technological characteristics for arm blood pressure monitors. Therefore, these minor technological differences do not raise any new questions of safety or effectiveness. Accordingly, the KP-7770 is substantially equivalent to its predicate devices. pg 3 of 3 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 K-jump Health Co., Ltd. c/o Mr. Jason Cheng Assistant Manager, Product Planning Department No. 56 Wu Kung 5th Road, Wu Ku Industrial Park Taipei Hsien 248 TAIWAN OCT - 9 2009 Re: K092806 > Trade/Device Name: Auto Digital Upper Arm Blood Pressure Monitor, Model KP-7770 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: September 9, 2009 Received: September 11, 2009 Dear Mr. Cheng: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Mr. Jason Cheng Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. M. G. Williams Bram D. Zuckerman, M.D. Ser Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if Known):__K092806 Device Name: Arm Blood Pressure Monitor Model KP-7770 Indication for Use: The KP-7770 is intended to measure the systolic and diastolic blood pressure and pulse rate (heart rate) by using an inflating cuff wrapped around the arm. It is indicated for use in adults. The devices display irregular pulse detection (IPD) icon on LCD screen when irregular pulse is detected during blood pressure measurement. Prescription Use (Per 21 C.F.R. 801.109) OR Over-The-Counter Use X (Optional Format 1-2-96) ાં ર્ ### (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) . A. Allibenne (Division Sign-Off) Division of Cardiovascular Devices K092806 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...