← Product Code [KCT](/productcode/KCT) · K092745

# VIS-U-ALL LOW TEMPERATURE TYVEK STERILIZATION POUCH FOR ETHYLENE OXIDE, MODELS: SELF AND HEAT SEAL POUCHES, HEAT SEAL TU (K092745)

_STERIS Corporation · KCT · Nov 19, 2010 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K092745

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** Nov 19, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Vis-U-All Low Temperature Tyvek Sterilization Pouch for ethylene oxide has been qualified by STERIS Corporation as suitable for use by health care providers to enclose and seal other medical devices to be sterilized by ethylene oxide (ETO). The Vis-U-All Low Temperature Tyvek Sterilization Pouch for ethylene oxide is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.

## Device Story

Vis-U-All Low Temperature Tyvek Sterilization Pouch is a containment system for medical devices undergoing ethylene oxide (ETO) sterilization. Available as self-seal pouches, heat-seal pouches, or heat-seal tubing. Includes an integrated ISO 11140 Class 1 compliant chemical process indicator to distinguish between processed and unprocessed units. Used by healthcare providers in clinical settings; devices are placed inside, sealed, and subjected to ETO sterilization cycles (1 hour at 130°F or 4.5 hours at 100°F). The pouch maintains sterility during handling and storage. Benefits include effective sterilant penetration, microbial barrier protection, and visual confirmation of processing status.

## Clinical Evidence

Bench testing only. Studies included sterilant penetration, event-related sterility maintenance, microbial barrier properties, tensile/tear/puncture/seal strength, cytotoxicity (ISO 10993-5), sterilant residues (ISO 10993-7), and chemical indicator performance (ANSI/AAMI/ISO 11140-1). All tests passed, confirming the pouch maintains sterility and integrity post-ETO processing.

## Technological Characteristics

Tyvek/plastic film sterilization pouch. Features integrated ISO 11140 Class 1 chemical process indicator ink. Sterilization compatibility: 100% ETO. Materials tested for cytotoxicity per ANSI/AAMI/ISO 10993-5 and residues per ISO 10993-7. Available in various pouch and tubing dimensions.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- Vis-U-All Self Seal Pouch ([K070765](/device/K070765.md))
- Vis-U-All Heat Seal Pouch and Tubing ([K071087](/device/K071087.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "K692745 p 1 of 5" at the top. Below that is the word "STERIS" with a registered trademark symbol. At the bottom of the image is a logo consisting of several horizontal lines.

NOV 1 9 2010

# 510(k) Summary For Vis-U-All Low Temperature Tyvek Sterilization Pouch for Ethylene Oxide Sterilization

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459

Contact:

Robert F. Sullivan Senior Director FDA Regulatory Affairs Telephone:(440) 392-7695 Fax No: (440) 357-9198

Summary Date:

November 19, 2010

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{1}------------------------------------------------

### K092745/RC STERIS Response to 11/18/10 Request for Clarification Vis-U-All Low Temperature Tyvek Sterilization Pouch for Ethylene Oxide Sterilization

#### 1. Device Name

Vis-U-All Low Temperature Tyvek Sterilization Trade Name: Pouch for Ethylene Oxide Sterilization Sterilization Pouch Common/usual Name: Classification Name: Sterilization Wrap (21 CFR 880.6850) . КСТ Product Code:

#### 2. Predicate Device

- ' Vis-U-All Self Seal Pouch (K070765)
- I Vis-U-All Heat Seal Pouch and Tubing (K071087)

#### 3. Description of Device

The proposed Vis-U-All Low Temperature Tyvek Sterilization Pouch for Ethylene Oxide is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider using ethylene oxide (ETO). The proposed pouch is available as a self seal pouch, a heat seal pouch, or heat seal tubing.

#### 4. Intended Use

The Vis-U-All Low Temperature Tyvek Sterilization Pouch for ethylene oxide has been qualified by STERIS Corporation as suitable for use by health care providers to enclose and seal other medical devices to be sterilized by ethylene oxide (ETO). The Vis-U-All Low Temperature Tyvek Sterilization Pouch for ethylene oxide is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.

{2}------------------------------------------------

| STERIS Part # | Type               | Size (inches) |
|---------------|--------------------|---------------|
| 875037        | Pouch, Heat Seal,  | 3 x 7         |
| 875049        | Tyvek (low temp)   | 4 x 9         |
| 875412        |                    | 4 x 12        |
| 875422        |                    | 4 x 22        |
| 875610        |                    | 6 x 10        |
| 875812        |                    | 8 x 12        |
| 875115        |                    | 10 x 15       |
| 875118        |                    | 12 x 18       |
| 876037        | Pouch, Self Seal,  | 3 x 7         |
| 876049        | Tyvek (low temp)   | 4 x 9         |
| 876412        |                    | 4 x 12        |
| 876422        |                    | 4 x 22        |
| 876610        |                    | 6 x 10        |
| 876812        |                    | 8 x 12        |
| 876115        |                    | 10 x 15       |
| 876118        |                    | 12 x 18       |
| 872031        | Tubing, Heat Seal, | 3" x 100 ft   |
| 872041        | Tyvek (low temp)   | 4" x 100 ft   |
| 872061        |                    | 6" x 100 ft   |
| 872091        |                    | 9" x 100 ft   |
| 872141        |                    | 14" x 100 ft  |

Pouch and Tubing Sizes for STERIS Vis-U-All Low Temperature Products

The following are the validation test conditions:

- 1 hour exposure; at 130(±5) °F*, >30% RH using 100% ETO (750-790 mg/L);
- · 4.5 hour exposure; at 100(±5) °F*, >30% RH using 100% ETO (750-790 mg/L);

* ±5 °F is during sterilization phase following an equilibration period of 10% of exposure time.

The ethylene oxide process indicator is intended to be used by a health care provider with the Vis-U-All Low Temperature Tyvek Sterilization Pouches to distinguish between processed and unprocessed units.

{3}------------------------------------------------

K092745

p4 JR 5

#### Description of Safety and Substantial Equivalence 5.

The Vis-U-All Low Temperature Tyvek Sterilization Pouch models have been qualified by STERIS Corporation as suitable for use by health care providers to enclose and seal other medical devices to be sterilized using ethylene oxide. The predicates. Vis-U-All Self Seal Pouch (K070765) and Vis-U-All Heat Seal Pouch and Tubing (K071087) are also intended to enclose and seal medical devices to be sterilized. The proposed device's intended use is identical to the predicates, exceptingh thelow temperature sterilization modality to which the pouch is exposed: ethylene oxide (ETO) instead of hydrogen peroxide (VHP).

The proposed device is identical to the predicate devices in terms of physical and chemical properties, configurations and dimensions, air permeance and percent of surface perforations. The material composition of the proposed and predicate device is identical with the exception of the addition of an ISO 11140 Class 1 compliant, ethylene oxide chemical indicator on the proposed device.

As described in the next section, performance testing of the Vis-U-All Low Temperature Tyvek Sterilization Pouch for Ethylene Oxide demonstrated that the proposed product is qualified for use with ethylene oxide (ETO) sterilization and is as safe, as effective, and performs the same as the predicate devices.

#### 6. Performance Testing

The following table summarizes the non-clinical testing performed for both indicated sterilization cycles to demonstrate that the Vis-U-All Low Temperature Tyvek Sterilization Pouch for Ethylene Oxide is safe and effective. The performance testing has demonstrated that the proposed device is substantially equivalent to its predicates and raises no new questions of safety or effectiveness.

{4}------------------------------------------------

K092745

### K092745/RC STERIS Response to 11/18/10 Request for Clarification Vis-U-All Low Temperature Tyvek Sterilization Pouch for Ethylene Oxide Sterilization

| 12 | ে<br>><br>- | て<br>0 | 1<br>)<br>1 |
|----|-------------|--------|-------------|
|----|-------------|--------|-------------|

| Testing                                                                       | Results                                                                                                                                                                                                                                                                                                                                                                                 |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilant Penetration                                                         | PASS<br>Testing demonstrated that ethylene oxide can effectively penetrate the<br>pouches to sterilize the load.                                                                                                                                                                                                                                                                        |
| Event Related<br>Sterility<br>Maintenance                                     | PASS<br>All packs processed in an ethylene oxide sterilizer and subjected to<br>accelerated aging and handling maintained their performance (strength<br>and microbial resistance).                                                                                                                                                                                                     |
| Microbial Barrier<br>Properties                                               | PASS<br>No microbial growth occurred on test articles or negative control<br>pouches. Test articles from positive control pouches demonstrated<br>microbial growth                                                                                                                                                                                                                      |
| Tensile / Tear /<br>Puncture Resistance<br>/ Seal Peel Strength               | PASS<br>The differences in tensile properties (elongation, breaking force) of<br>processed and unprocessed Tyvek and plastic samples were not<br>statistically significant. All processed but un-aged pouches resulted in<br>clean peels and all processed pouches had acceptable burst strength,<br>indicating that ethylene oxide sterilization does not compromise pouch<br>sealing. |
| Cytotoxicity                                                                  | PASS<br>Following ethylene oxide sterilization, the Vis-U-All Low<br>Temperature Tyvek Pouch/Tubing plastic and Tyvek were not<br>cytotoxic.                                                                                                                                                                                                                                            |
| Sterilant Residues                                                            | PASS<br>Following sterilization and aeration, pouch materials were confirmed<br>to not retain harmful levels of ethylene oxide or its byproducts as<br>outlined in ISO 10993-7:1995.                                                                                                                                                                                                    |
| Chemical Indicator<br>Class 1 Compliance                                      | PASS<br>Chemical Indicators were validated against the ethylene oxide process<br>indicator requirements of ANSI/AAMI/ISO 11140-1:2005                                                                                                                                                                                                                                                   |
| Chemical Indicator<br>Toxicity                                                | PASS<br>The indicator ink on its substrate- either exposed to ethylene oxide or<br>unexposed - is not cytotoxic per the methodology and limits defined in<br>ANSI/AAMI/ISO10993-5:1999                                                                                                                                                                                                  |
| Chemical Indicator<br>Post-Processed<br>Color Stability                       | PASS<br>The post-processed chemical (process) indicator color is stable, after<br>exposure to ETO sterilization conditions, for at least one year of<br>storage at ambient conditions.                                                                                                                                                                                                  |
| Simulated Use                                                                 | PASS<br>All criteria of the study were met, demonstrating that, ethylene oxide<br>can penetrate through the Vis-U-All Low Temperature Tyvek Pouch<br>and Tubing to sterilize loads.                                                                                                                                                                                                     |
| Process Indicator<br>End Point Stability -<br>Aged Vis-U-All<br>Tyvek pouches | PASS<br>The chemical indicator on Vis-U-All Low Temperature Tyvek Pouches<br>stored at warehouse conditions still meet performance criteria.                                                                                                                                                                                                                                            |

{5}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the caduceus. The caduceus is depicted as a stylized bird with three tail feathers.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Marcia Benedict Director, Regulatory Affairs Steris Corporation 5960 Heisley Road Mentor, Ohio 44060

NOV 1 9 2010

Re: K092745

Trade/Device Name: Vis-U-All Low Temperature Tyvek Sterilization Pouch with Ethylene Oxide Process Indicator Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: November 11, 2010 Received: November 12, 2010

Dear Ms. Benedict:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

### Page 2 - Ms. Benedict

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,
fum I. Sated
for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology. General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{7}------------------------------------------------

K092745/S001 STERIS Response to 11/10/09 Request for Additional Information Vis-U-All Low Temperature Tyvek Sterilization Pouch for Ethylene Oxide Sterilization

## Indications for Use

NOV 1 9 2010

510(k) Number (if known): K092745

Device Name:

## Vis-U-All Low Temperature Tyvek Sterilization Pouch with Ethylene Oxide Process Indicator

Indications For Use:

The Vis-U-All Low Temperature Tyvek Sterilization Pouch for ethylene oxide has been qualified by STERIS Corporation as suitable for use by health care providers to enclose and seal other medical devices to be sterilized by ethylene oxide (ETO). The Vis-U-All Low Temperature Tyvek Sterilization Pouch for ethylene oxide is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.

Pouch and Tubing Sizes for STERIS Vis-U-All Low Temperature Products

| STERIS Part # | Type               | Size (inches) |
|---------------|--------------------|---------------|
| 875037        | Pouch, Heat Seal,  | 3 x 7         |
| 875049        | Tyvek (low temp)   | 4 x 9         |
| 875412        |                    | 4 x 12        |
| 875422        |                    | 4 x 22        |
| 875610        |                    | 6 x 10        |
| 875812        |                    | 8 x 12        |
| 875115        |                    | 10 x 15       |
| 875118        |                    | 12 x 18       |
| 876037        | Pouch, Self Seal,  | 3 x 7         |
| 876049        | Tyvek (low temp)   | 4 x 9         |
| 876412        |                    | 4 x 12        |
| 876422        |                    | 4 x 22        |
| 876610        |                    | 6 x 10        |
| 876812        |                    | 8 x 12        |
| 876115        |                    | 10 x 15       |
| 876118        |                    | 12 x 18       |
| 872031        | Tubing, Heat Seal, | 3" x 100 ft   |
| 872041        | Tyvek (low temp)   | 4" x 100 ft   |
| 872061        |                    | 6" x 100 ft   |
| 872091        |                    | 9" x 100 ft   |
| 872141        |                    | 14" x 100 ft  |

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of 2

510(k) Number:

May 05, 2010

{8}------------------------------------------------

The following are the validation test conditions:

- 1 hour exposure; at 130(±5) °F*, >30% RH using 100% ETO (750-790 mg/L); .
- 4.5 hour exposure; at 100(±5) ºF*, >30% RH using 100% ETO (750-790 mg/L); .

* ±5 °F is during sterilization phase following an equilibration period of 10% of exposure time.

The ethylene oxide process indicator is intended to be used by a health care provider with the Vis-U-All Low Temperature Tyvek Sterilization Pouches to distinguish between processed and unprocessed units.

Prescription Use AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use X (21 CFR 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sted N. Murphey D. Matz J. Ellevan-Williams
(Division Sign Off)

Anesthesiology, General Hospital Control, Dental Devices

510(k) Number: K092245

Page 2 of 2

---

**Source:** [https://fda-staging.innolitics.com/device/K092745](https://fda-staging.innolitics.com/device/K092745)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
