The Wrist Arthrodesis Nail System is intended for wrist arthrodesis. Specific indications include post-traumatic arthritis of the joints of the wrist, rheumatoid wrist deformities requiring restoration, complex carpal instability, post-septic arthritis of the wrist, severe unremitting wrist pain related to motion, brachial plexus nerve palsies, tumor resection, and spastic deformities.
Device Story
Intramedullary nailing platform for wrist arthrodesis; utilizes minimally invasive dorsal approach into third metacarpal and distal radius. System components: titanium (Ti 6AL-4V) distal radius nail, metacarpal nails, connectors, and unicortical screws. Device placed within medullary canal to reduce soft tissue irritation, subcutaneous prominence, and surgical incision size. Operated by trained physicians in clinical setting. Provides structural fixation for wrist fusion; benefits patient by minimizing soft tissue complications associated with traditional plate fixation.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via bench testing: static four-point bending (ASTM F82), dynamic four-point bending (ASTM F1264, 1,000,000 cycles), torque testing, and pull testing. Cadaver lab performed to confirm surgical procedure feasibility and repeatability.
Technological Characteristics
Materials: Medical grade titanium alloy (Ti 6AL-4V) and cobalt chrome. Intramedullary fixation rod design. Testing standards: ASTM F82 (static bending), ASTM F1264 (dynamic bending).
Indications for Use
Indicated for patients requiring wrist arthrodesis due to post-traumatic arthritis, rheumatoid deformities, complex carpal instability, post-septic arthritis, severe motion-related wrist pain, brachial plexus nerve palsies, tumor resection, or spastic deformities.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Skeletal Dynamics. The logo consists of a stylized four-petal flower-like symbol on the left, followed by the words "skeletal dynamics" on the right. The word "skeletal" is on the top line, and the word "dynamics" is on the bottom line.
# K092720
JUN 2 9 2010
### 510(k) Summary of Safety and Effectiveness Wrist Arthrodesis Nail System System
June 18, 2010
Submitter:
Skeletal Dynamics, LLC 8905 SW 87 Avenue Suite 102 Miami, FL 33176 Tel: (305) 596-7585 Contact: Ana M. Escagedo / Vice President Quality & Regulatory Affairs Email: aescagedo@skeletaldynamics.com
FDA Establishment Registration Number: Pending
Trade Name, Common Name, Classification: Device Trade Name: Wrist Arthrodesis Nail System IMPlate Wrist Arthrodesis Nails
Device Common or Usual Names: Intramedullary fixation rod
Product Code: HSB
Classification: Class II, 21 CFR 888.3020
Predicate Device: Synthes - LCP Wrist Fusion - K042355
#### Description of the Device:
The Wrist Arthrodesis Nail System is designed as an intramedullary nailing platform to address wrist arthrodesis procedures utilizing a minimally invasive dorsal approach into the third metacarpal and distal radius by trained physicians. The IMPLATE Wrist Arthrodesis Nails are secured within the intramedullary canals by means of bone screws, and then assembled into a completed construct using a Connector.
The IMPLATE Wrist Arthrodesis Nail System is comprised of a titanium (Ti 6AL-4V) Distal Radius Intramedullary Nail, various lengths of titanium (Ti 6AL-4V) Metacarpal Intramedullary Nails, multiple titanium (Ti 6AL-4V) nail Connectors in various lengths and angles, and Unicortical Screws of various lengths, Pins and System specific instrumentation
#### Intended Use:
The Wrist Arthrodesis Nail System is intended for wrist arthrodesis. Specific indications include post-traumatic arthritis of the joints of the wrist, rheumatoid wrist deformities requiring restoration, complex carpal instability, post-septic arthritis of the wrist, severe unremitting wrist pain related to motion, brachial plexus nerve palsies, tumor resection, and spastic deformities.
Page 1 of 2
{1}------------------------------------------------
#### Technological Characteristics:
The Wrist Arthrodesis Nail System has the foilowing similarities to the LCP Wrist Fusion manufactured and distributed by Synthes pursuant to K042355:
- The same indications for use and intended use .
- The same materials .
- The same load strength .
- Accommodate different size anatomies. �
#### Biocompatibility:
The materials selected for the Wrist Arthrodesis Nail System have a long history of safe use in the orthopedic industry. The materials are medical grade titanium alloy and cobalt chrome.
#### Comparison of Technological Modifications:
In contrast to the predicate device, the Wrist Arthrodesis Nail System was designed to be placed within the medullary canal. Placement within the medullary canal reduces the possibility of soft tissue irritation and subcutaneous prominence. Intramedullary placement also minimizes the size of the surgical incision.
#### Summary of Non-Clinical Testing:
The following tests were performed to demonstrate that the Wrist Arthrodesis Nail System is safe and effective:
- Static four point bending test in accordance with ASTM F82 was performed to determine . the bending strength of the subject device and compare the yield point and bending stiffness to that of the predicate. All subject devices tested passed the acceptance criteria without failure.
- A dynamic four point bending test in accordance with ASTM F1264 was performed to . determine the maximum run out value at 1,000,000 cycles. The testing confirmed the endurance value of 1,000,000 cycles without evidence of failure.
- A torque test was performed to challenge the strength of the connector. All subject . devices tested passed the acceptance criteria without evidence of failure.
- A pull test was performed to challenge the interconnection of the connector to the . metacarpal and radius nails. All subject devices passed the acceptance criteria without evidence of failure.
- A cadaver lab was performed to confirm the repeatability of the surgical procedure. The . lab documented that the procedure is feasible and repeatable.
#### Conclusion:
We believe the subject device is substantially equivalent to the predicate device and conclude that the subject device is as safe and effective as the predicate device.
Page 2 of 2
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Skeletal Dynamics, LLC % Ms. Ana M. Escagedo Vice President Quality & Regulatory Affairs 8905 Southwest 87 Avenue, Suite 102 Miami, Florida 33176
JUN 2 9 2010
Re: K092720
Trade/Device Name: Wrist Arthrodesis Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Code: HSB Dated: June 22, 2010 Received: June 23, 2010
Dear Ms. Escagedo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Ms. Ana M. Escagedo
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Sincerely, yours,
Barbary Brielund
Mark N. Melke Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): Y. O9 27 ZO
Device Name: Wrist Arthrodesis Nail System
Indications For Use:
The Wrist Arthrodesis Nail System is intended for wrist arthrodesis. Specific indications include posttraumatic arthritis of the joints of the wrist, rheumatoid wrist deformities requiring restoration, complex carpal instability, post-septic arthritis of the wrist, severe unremitting wrist pain related to motion, brachal plexus nerve palsies, tumor resection, and spastic deformities.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ulla Dneun
Orthopedic. ivision of Surg and Restorative Devices
510(k) Number K092730
Page li of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.