CT2011 SonicStimu is a single-channel combination device for therapeutic ultrasound and TENS. System includes console, control/display panel, AC-DC adaptor, and electrodes. Microprocessor generates pro-modulated low-frequency monophasic electrical pulses and 1 MHz ultrasound waves. User interface features LCD, control buttons, and audio feedback for adjusting signal intensities and operation modes. Device used for electrical stimulation, ultrasound, or combination therapy. Operator adjusts power settings via interface. Output provides therapeutic stimulation and ultrasound energy to target tissues. Benefits include pain management and muscle spasm reduction.
Clinical Evidence
Bench testing only. No clinical data was conducted or relied upon for this determination.
Technological Characteristics
System includes microprocessor-controlled console, LCD, and audio feedback. Ultrasound transducer uses aluminum; shell uses ABS plastic. Operates at 1 MHz ultrasound frequency. Electrical output: constant voltage, pro-modulated low-frequency monophasic pulses. Powered by 15V/1A DC. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and IEC 60601-2-5. Dimensions: 14.1 x 2.5 x 3.8 inches. Weight: 0.84 lbs.
Indications for Use
Indicated for patients requiring pain relief, reduction of muscle spasm, or treatment of joint contractures via therapeutic ultrasound, and for patients requiring symptomatic relief of chronic intractable pain, post-traumatic pain, or post-surgical pain via transcutaneous electrical nerve stimulation (TENS).
Regulatory Classification
Identification
An ultrasound and muscle stimulator for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies. The device also passes electrical currents through the body area to stimulate or relax muscles.
{0}------------------------------------------------
K092627
ﺎ
JAN 1 4 2010
Caretalk®
File No: WMI-04-CT2011-FDA-05 Version: 1.1
)
## 510(k) SUMMARY
CT2011 SonicStimu, K (
| Submitter's Name: | Shenzhen Dongdixin Technology Co., Ltd |
|-------------------|--------------------------------------------------------------------------------------------------------|
| Address: | No.3bldg. xiliyangguang industrial estae<br>xilixiaobaimang Nanshan district<br>Shenzhen, CHINA 518108 |
| Telephone: | +86(755) 27652471 |
| Fax: | +86(755) 27652674 |
| Contact: | Zhigang Zhao |
| Date Prepared: | Aug, 8 2009 |
Section 05-Page 1 of 5
{1}------------------------------------------------
File No: WMI-04-CT2011-FDA-05 Version: 1.1
#### 1. Proposed Device:
Trade Name: CT2011 SonicStimu Pain Relief Device Classification Name: Ultrasound and Muscle Stimulator 21 CFR 890.5860 Class II Product Code: IMG, GZJ
#### Predicate Device: 2.
21 CFR 890.5860 Product Code: IMG, GZJ, LIH Device Class: II Legally Marketed Device: Sonicator Plus 940, ME940 Manufacturer: Mettler Electronics Corp 510(k) Number: K071137
#### 3. Description of Proposed Device:
The CT2011 SonicStimu Pain Relief Device is comprised of following main components:
- A system console including software and control electronics;
- A control and display panel;
- Device accessories including AC-DC Adaptor, Electrode 40*90mm, Lead for adapter, Lead for electrical.
The CT2011 is a single channel combination unit for therapeutic ultrasonic and transcutaneous electrical nerve stimulation. The microprocessor in CT2011 provides pro-modulated low frequency and monophasic electrical pulse waveforms with enhanced reliability and ease of using of the device. In the same time it offers I MHz ultrasound treatment also.
The user interface friendly comprises keys, LCD and audio feedback. The LCD provides operator information about operation mode and signal intensities. The user through control buttons to make adjusting power for ultrasound and stimulation. The CT2011 can be used as electrical stimulation or ultrasound therapeutic equipment separately or combination for both.
Section 05-Page 2 of 5
{2}------------------------------------------------
### File No: WMI-04-CT2011-FDA-05 Version: 1.1
### Proposed Device Intended Use Statement: 4.
Device Name: CT2011 SonicStimu Pain Relief Device
## Proposed Device Indications for Use:
Therapeutic Ultrasound
- 1. Pain relief
- 2. Reduction of muscle spasm
- 3. Joint contractures
## Transcutaneous Electrical Nerve Stimulation
- 1. Symptomatic relief of chronic intractable pain
- 2. Post-traumatic pain
- 3. Post-surgical pain
#### Biocompatibility Certification: 5.
Electrodes to be provided with this device are from the manufacturer Top-Rank Health Care Equipment Co., Ltd (K070612) who submitted in 2007.
The shell of device is used ABS material; this material has passed Biocompatibility testing in Jiangsu TUV Product Service Ltd. Shanghai Branch. Identification No: 080960.
Ultrasound transducer attached to a metal surface and patient contact through the metal, this material is aluminum that passed biocompatibility testing in School of Radiation Medicine and Public Health Soochow University. The testing report No: SRPSU-2008-0544, SRPSU-2008-0545, SRPSU-2008-0546,
### Technological Characteristics and Substantial Equivalence 6.
Both the CT2011 SonicStimu Pain Relief Device and the Predicate device Stimulator. have the same intended use and fundamental technology under the same product code. A side-by-side comparison of the CT2011 SonicStimu Pain Relief Device and the cited predicate devices is included in the 510(k) submission. The CT2011 SonicStimu Pain Relief Device is substantially equivalent to the technological features as the predicate devices.
{3}------------------------------------------------
### File No: WMI-04-CT2011-FDA-05 Version: 1.1
| 1 | 510K# | K | K071137 |
|----|---------------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| 2 | Device Name | CT2011 | Sonicator Plus940 |
| 3 | Manufacturer | Shenzhen Dongdixin Technology Co.,<br>Ltd. | Mettler Electronics |
| 4 | Power Source | DC 15V/1A | AC Line |
| | -Method of Line current isolation | Reinforced insulation | Reinforced insulation |
| | - Patient Leakage Current (μA) | | |
| | -Normal condition | <1 | >50 |
| | -Single fault condition | >50 | >50 |
| 5 | Number of Output Modes | 15 | 8 |
| 6 | Number of Output Channels | 1 | 4 |
| | Synchronous Or Alternating | N/A | 1&2 or 3&4 |
| 7 | Constant Current or Constant Voltage | Constant Voltage | Constant Current |
| 8 | Software/Firmware/Microprocessor<br>Control | Yes | Yes |
| 9 | Automatic Overload Trip | Yes | Yes |
| | Automatic Over Current Trip | Yes | Yes |
| 10 | Automatic No Load contact Trip | Yes | Yes |
| 11 | Automatic Shut off | No | Yes |
| 12 | Patient Override Control? | No | No |
| 13 | Indicator Display | | |
| | -On/Off Status | Yes | Yes |
| | -Voltage/Current Level? | Yes | Yes |
| | -Low Battery indicator | N/A | N/A |
| 14 | Timer Display | 0-30 minutes | 0-60 minutes |
| 15 | Standards | ISO14971, UL60601-1, IEC 60601-1, IEC<br>60601-1-2, IEC 60601-2-10, MDD<br>93/42/EEC, Annex II | ISO14971, UL2601-1, IEC 60601-1,<br>IEC 60601-1-2, IEC 60601-2-10, MDD<br>93/42/EEC, Annex II |
| 16 | Compliance with 21 CFR 898 | Yes | Yes |
| 17 | Weight (lbs.) | 0.84 | 11 |
| 18 | Dimensions (in.) H x W x L | 14.1x2.5x3.8 | 4.9x13.6x10.5 |
### Basic technological characteristics, new device vs. Predicate device
## 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The CT2011 SonicStimu Pain Relief Device did not conduct, nor rely upon, clinical tests to determine substantial equivalence. Nonclinical testing was performed in order to validate the design according with the company's specified design requirements, and to assure conformance with the following voluntary design standards:
Section 05-Page 4 of 5
{4}------------------------------------------------
File No: WMI-04-CT2011-FDA-05 Version: 1.1
- IEC 60601-1 "Medical electrical equipment Part 1: General requirements for safety". A
- A IEC 60601-1-2 "Medical electrical equipment - Part 1-2: General requirements for safety -Collateral Standard"
- ) IEC 60601-2-10 "Medical electrical equipment - Part 2: Particular requirements for the safety of nerve and muscle stimulators"
- ♪ IEC 60601-2-5 "Medical electrical equipment - Particular requirements for the safety of ultrasonic physiotherapy equipment ED "
#### Discussion of Clinical Tests Performed: 8.
Not applicable
#### 9. Conclusions:
The CT2011 SonicStimu Pain Relief Device has the same intended use and similar characteristics as the predicate device, the Sonicator Plus 940, Model ME940 device. Moreover, bench testing, safety report and Risk Analysis Report documentation supplied in this submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. In the other words, those engineering difference do not affect the intended use or alter the fundamental scientific technology of the device. Thus, The CT2011 SonicStimu Pain Relief Device is substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an abstract image of a bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
JAN 1 4 2010
Shenzhen Dongdixin Technology Co., Ltd % Mr. Zhigang Zhao No. 3 Bldg. Xiliyangguang Industrial Estate Xilixiaobaimang Nanshan District Shenzhen, CN-44, Guangdong Province People's Republic of China 518108
Re: K092627
Trade/Device Name: CT2011 SonicStimu Pain Relief Device Regulation Number: 21 CFR 890.5860 Regulation Name: Ultrasound and muscle stimulator. Regulatory Class: Class II Product Code: IMG, GZJ Dated: January 8, 2010 Received: January 8, 2010
## Dear Mr. Zhao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Mr. Zhigang Zhao
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For
Mark N. Malkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic. and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
File No: WMI-04-CT2011-FDA-04 Version: 1.1
## Indications for Use Statement
510(k) Number (
):
Device Name: CT2011 SonicStimu Pain Relief Device
## Indications for Use:
Therapeutic Ultrasound
- 】. Pain relief
- 2. Reduction of muscle spasm
- Joint contractures 3.
### Transcutaneous Electrical Nerve Stimulation
- 1. Symptomatic relief of chronic intractable pain
- 2. Post-traumatic pain
- 3. Post-surgical pain
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
WilliAM R. CARY FOR M. MELKERSON
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K092627
Section 04 -Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.