X-SMART EASY

K092614 · Dentsply Interntional · EKX · Dec 2, 2009 · Dental

Device Facts

Record IDK092614
Device NameX-SMART EASY
ApplicantDentsply Interntional
Product CodeEKX · Dental
Decision DateDec 2, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4200
Device ClassClass 1
AttributesTherapeutic

Indications for Use

X-Smart Easy is indicated for use by dentists in standard endodontic procedures using rotary endodontic files and rotary endodontic drills (Gates-Glidden).

Device Story

X-Smart Easy is an electric motor-driven handpiece for root canal preparation. Operated by dentists in clinical settings, it powers rotary endodontic files and drills to facilitate endodontic procedures. The device transforms electrical energy into mechanical rotation to assist in cleaning and shaping root canals. It is used as a tool for the clinician to perform standard endodontic tasks, potentially improving procedural efficiency and patient outcomes during root canal therapy.

Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (ISO 10993), electrical safety (UL60601-1), and electromagnetic compatibility (IEC 60601-1-2, IEC 61000-3-2).

Technological Characteristics

Electric motor-driven handpiece. Patient contact components meet biocompatibility standards. Conforms to UL60601-1, IEC 60601-1-2, and IEC 61000-3-2. Includes hardware and software components.

Indications for Use

Indicated for use by dentists performing standard endodontic procedures using rotary endodontic files and rotary endodontic drills (Gates-Glidden).

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a partial view of the Dell logo. The letters "DE" are visible in bold, black font. The logo is simple and recognizable, representing the Dell brand. # K092614 # DENTSPLY International World Headquarters Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405-0872 (717) 845-7511 {voice} (717) 849-4343 (fax) www.dentsply.com DEC - 2 2009 ## 510(k) Summary for X-Smart Easy - 1. Submitter Information: DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405 | Contact Person: | Helen Lewis | |-------------------|--------------| | Telephone Number: | 717-849-4229 | | Fax Number: | 717-849-4343 | Date Prepared: 18 August 2009 - Device Name: 2. Proprietary Name: X-Smart Easy Classification Name: Dental Handpiece and Accessories 872.4200 CFR Number: Device Class: I EKX Product Code: - 3. Predicate Device: Prophy-Mate and Endo Mate Motor Handpiece (K990682) - Description of Device: 4. The X-Smart Easy is an electric motor-driven handpiece intended for root canal preparation procedures in endodontic industry. - న్. Indications for Use: X-Smart Easy is designed for use by dentists in standard endodontic procedures using rotary endodontic files and rotary endodontic drills (Gates-Glidden). - Description of Safety and Substantial Equivalence 6. ## Technological Characteristics Patient contact elements of the components found in X-Smart Easy have been used in legally marketed devices and/or were found safe for dental use. DENTSPLY International | X-Smart Easy 18 August 2009 9 - ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ Image /page/0/Picture/21 description: The image contains the number 000015 in a bold, sans-serif font. The numbers are black against a white background. The numbers are evenly spaced and aligned horizontally. {1}------------------------------------------------ ## Non-Clinical Performance Biocompatibility Testing Cytotoxicity, sensitization and irritation testing were performed in accordance with ISO 10993. All patient contact components demonstrated biocompatibility. ### Electromagnetic Compatibility and Electrical Safetv The X-Smart East conforms to UL60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety; IEC 60601-1-2 Medical Electrical Equipment, Part 1-2: General Requirements for Safety Collateral Standard Electromagnetic Compatibility; and IEC 61000-3-2 Electromagnetic Compatibility (EMC) Part 3: Limits - Section 3: Limitations of Voltage Fluctuations. ### Conclusion as to Substantial Equivalence The X-Smart Easy is substantially equivalent to the Prophy Mate and Endo Mate Handpiece (K990682) based on equivalence of the intended use, similar features and technical characteristics. Performance testing was performed to validate the safety and effectiveness of the X-Smart Easy, which included electrical safety, electromagnetic compatibility, and validation testing of both hardware and software functions. The X-Smart Easy does not raise any new issues of safety, effectiveness, or performance of the product. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a family of three figures, with the central figure slightly taller than the others. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 Ms. Helen Lewis Director DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street, Suite 60 York, Pennsylvania 17404 DEC - 2 2009 Re: K092614 Trade/Device Name: X-Smart Easy Regulation Number: 21CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EKX Dated: November 24, 2009 Received: November 25, 2009 Dear Ms. Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2- Ms. Lewis Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, for Susan Runner, D.D.S., M.A. Acting Division Director . Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 1092614 510(k) Number (if known): Device Name: X-Smart Easy Indications for Use: X-Smart Easy is indicated for use by dentists in standard endodontic procedures using rotary endodontic files and rotary endodontic drills (Gates-Glidden). Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Kain Hurley for NSR (Division Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital Division of Andelhoental Devices 510(k) Number:
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...