K092581 · Mitsmed Medical Instrument Technology Co., Ltd. · LLZ · Oct 30, 2009 · Radiology
Device Facts
Record ID
K092581
Device Name
MITS DICOM GATEWAY AND IMAGE MANAGER
Applicant
Mitsmed Medical Instrument Technology Co., Ltd.
Product Code
LLZ · Radiology
Decision Date
Oct 30, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The MITS DICOM Gateway and Image Manager a device that captures 2 D images and data or receives images and data from various medical imaging sources ( i.e. ultrasound system, R/F units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data (2 D or 3 D) can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Typical users of this system are trained medical professionals, including physicians, nurses, technicians and computer system professionals. To support the diagnostic interpretation of mammography studies, ImageSVR PACS the full fidelity DICOM image in a non-compressed will display format. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation.
Device Story
MITS DICOM Gateway and Image Manager is a client-server software suite for medical image management; functions include archiving, retrieving, and displaying images. Inputs include 2D/3D images/data from imaging sources like ultrasound, R/F units, and radiographic devices. System components include MITS DICOM Gateway (converter), MITS Image Manager SERVER, and MITS Image Manager CLIENT. Operated by trained medical professionals in clinical environments; facilitates image communication across networks. Healthcare providers use the system to view images for diagnostic interpretation, including mammography. Device does not contact patients or control life-sustaining equipment; relies on human intervention for clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only; software validation performed to assess risks and potential hazards.
Technological Characteristics
Software server application; client-server architecture. Does not contact patients. No energy source or physical dimensions. Connectivity via computer networks. Software-based image management system.
Indications for Use
Indicated for trained medical professionals (physicians, nurses, technicians) to store, communicate, process, and display 2D/3D medical images from various sources (ultrasound, R/F, CR/DR, scanners). Supports diagnostic interpretation of mammography studies using non-compressed, full-fidelity DICOM images on FDA-approved 5Mpixel monitors. Contraindicated for primary diagnosis or interpretation using lossy compressed mammographic images or digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for MITS, which is a telecommunications company. The logo consists of the letters "MITS" in a bold, sans-serif font. The letters are arranged horizontally, with the "M" and "I" connected. A circle surrounds the letters, and there is a small line underneath the circle.
### 2. ADMINISTRATIVE INFORMATION
K092581
- 2.1 510(k) Summary of Safety and Effectiveness
(Per 21 CFR 807.92)
General Information Establishment
OCT 3 0 2000
- MITSMED Medical Instrument Technology Co., Ltd. . Manufacturer:
- . Address: ' 4F-6, No. 99, Sec 1, NanKan Rd., Luzhu shiang, Taoyua County, 33858. Taiwan, R.O.C.
Owner Number: 10026488 I Contact Person: E-mail: ceirs.jen@msa.hinet.net Dr. Jen, Ke-Min (official correspondent) 886-3-5208829 (Tel); 886-3-5209783 (Fax) Address: No.58, Fu Chiun Street, Hsin Chu City, 30067, Taiwan, ROC
. Date Prepared: August 12, 2009
Device Information
| ● | Proprietary Name: | MITS DICOM Gateway and Image Manager |
|---|----------------------|-----------------------------------------------------|
| ● | Classification Name: | SYSTEM, IMAGE PROCESSING, RADIOLOGICAL,<br>Class II |
| ● | Regulation Number: | 892.2050 |
| ● | Product Code: | LLZ |
Safety and Effectiveness Information
- . Predicate Device:
Claim of Substantial Equivalence (SE) is made to CSIST DICOM Gateway and Image Manager, K012327.
- . Device Description:
MITS DICOM Gateway and Image Manager is a suite of applications designed and produced by MITSMED Medical Instrument Technology CO., Ltd. It is a client-server based software system manages archiving, retrieving and displaying medical images. The DICOM converter application is MITS DICOM Gateway. The server side is MITS Image Manager SERVER. And the client side is MITS Image Manager CLIENT,
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters 'MIT' inside of a circle. The letters are in a bold, sans-serif font. There is a small graphic to the left of the circle, which appears to be a stylized representation of a building or structure. The logo is simple and clean, with a focus on the acronym 'MIT'.
#### 0 Intended Use:
The MITS DICOM Gateway and Image Manager a device that captures 2 D images and data or receives images and data from various medical imaging sources ( i.e. ultrasound system, R/F units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data (2 D or 3 D) can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.
Typical users of this system are trained medical professionals, including physicians, nurses, technicians and computer system professionals.
To support the diagnostic interpretation of mammography studies, ImageSVR PACS the full fidelity DICOM image in a non-compressed will display format. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation.
#### Technological Characteristics .
The MITS DICOM Gateway and Image Manager are a software server application and do not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.
#### Substantial Equivalence (SE) .
A claim of substantial equivalence is made to CSIST DICOM Gateway and Image Manager, K012327. Both of them have the same working principle and technologies. The differences are due to the feature design aspects, not relating to the safety or Besides, the submission contains the results of software effectiveness aspects. validation that the risks analysis and the potential hazards have been classified Minor. Thus they are substantially equivalent.
Keter Teon
Dr. Jen. Ke-Min official correspondent for MITSMED Medical Instrument Technology Co., Ltd.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is in black and white.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
MITSMED Medical Instrument Technology Co., Ltd % Dr. Jen, Ke-Min Official Correspondent ROC Chinese-European Industry Research Society No. 58, Fu Chiun Street, Hsin Chu City, 30067 TATWAN ROC
OCT 3 0 2009
Re: K092581
Trade/Device Name: MITS DICOM Gateway and Image Manager Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: П Product Code: LLZ Dated: October 11, 2009 Received: October 21, 2009
Dear Dr. Jen, Ke-Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 -
Enclosure
device-related adverse events) (21 CFR 803), good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Lucian M. Morris
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
K192581 510(k) Number:
Device Name: MITSMED Medical Instrument Technology Co., Ltd. MITS DICOM Gateway and Image Manager
Indications for Use :
- · The MITS DICOM Gateway and Image Manager is a device that captures 2 dimensional images and data or receives images and data from various medical imaging sources ( i.e. ultrasound system, R/F units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data (2 dimensional or 3 dimensional) can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.
Typical users of this system are trained medical professionals, including physicians, nurses, technicians and computer system professionals.
- · To support the diagnostic interpretation of mammography studies the full fidelity DICOM image in a non-compressed format. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation.
Prescription Use V
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D) .
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.