← Product Code [KNS](/productcode/KNS) · K092571

# GYRUS ACMI BIOCOAG PROBE (K092571)

_Gyrus Acmi, Inc. · KNS · Oct 20, 2010 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K092571

## Device Facts

- **Applicant:** Gyrus Acmi, Inc.
- **Product Code:** [KNS](/productcode/KNS.md)
- **Decision Date:** Oct 20, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Gyrus ACMI Inc. BiCoag Hemostasis Probe is intended for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract during endoscopic procedures. The device also can be used for irrigation.

## Device Story

Bipolar electrosurgical probe for endoscopic use; designed for coagulation of active bleeding lesions in upper or lower GI tract. Device features central irrigation port; available in 10 Fr (3.3mm) and 7 Fr (2.3mm) diameters for use through endoscope working channels (min 3.7mm and 2.8mm respectively). Operated by physician during endoscopic procedures; requires connection to external bipolar high-frequency generator. Provides localized hemostasis via bipolar electrical energy; irrigation capability assists visualization or site preparation. Benefits patient by enabling controlled coagulation of GI bleeding.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use.

## Technological Characteristics

Bipolar electrosurgical probe; 10 Fr and 7 Fr configurations; central irrigation port. Materials and principles of operation are consistent with predicate bipolar hemostasis probes. Energy source: external bipolar high-frequency generator.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Biosearch Hemostatic probe ([K912129](/device/K912129.md))
- Everest Hemostatic Bipolar Probe ([K892895](/device/K892895.md))
- Bard Biprobe ([K912601](/device/K912601.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Gyrus ACMI Inc. BiCoag Hemostasis Probe Gyrus ACMI Inc 136 Turnpike Road Southborough, MA 01772

Traditional 510(k) Notification Summary of Safety and Effectiveness

| 510(k) Summary of Safety and Effectiveness |
|--------------------------------------------|
| Gyrus ACMI Incorporated                    |
| Gyrus ACMI Inc. BiCoag Hemostasis Probe    |

#### General Information

Manufacturer:

UCT 2 0 2010

Gyrus ACMI Incorporated 136 Turnpike Rd. Southborough, MA 01772-2104

Lorraine Calzetta

August 19, 2009

Contact Person:

Date Prepared:

#### Device Description

Classification Name:

Endoscopic electrosurgical unit and accessories (21CFR 876.4300) Class II Gastroenterology/Urology

Trade Name:

Gyrus ACMI Inc. BiCoag Hemostasis Probe

Generic/Common Name:

Endoscopic electrosurgical accessories.

### Predicate Device

Biosearch Hemostatic probe cleared under K912129

Everest Hemostatic Bipolar Probe cleared under K892895

Bard Biprobe cleared under K912601

#### Intended Uses

The Gyrus ACMI Inc. BiCoag Hemostasis Probe is intended for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract during endoscopic procedures. The device also can be used for irrigation.

### Device Description

The BiCoag Hemostasis Probe is intended to be used for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract when used with a bipolar high frequency generator. It contains a central irrigation port. The probe is to be available in 10 Fr and 7 Fr (3.3mm and 2.3mm respectively) outer diameter configurations to allow its use through a minimum 3.7mm and 2.8mm endoscope working channel, respectively.

## Technological Characteristics and Substantial Equivalence

The Gyrus ACMI Inc. BiCoag Hemostasis Probe, as described in this submission, is substantially equivalent to the predicates in intended use, materials, principles of operation and fundamental scientific technology.

{1}------------------------------------------------

/

# Summary of Safety and Effectiveness

The Gyrus ACMI Inc. BiCoag Hemostasis Probe, as described in this submission, is substantially equivalent to the predicates in intended use, materials, principles of operation and fundamental scientific technology and raises no new issues of safety and effectiveness.

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest wings or feathers.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Gyrus ACMI Incorporated % Ms. Lorraine Calzetta 136 Turnpike Road Southborough, Massachusetts 01772

Re: K092571

Trade/Device Name: Gyrus ACMI Inc. BiCoag Hemostasis Probe Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: Class II Product Code: KNS Dated: October 14, 2010 Received: October 15, 2010

OCT 2 0 2010

Dear Ms. Calzetta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Lorraine Calzetta

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdagov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark A. Milliken

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K092571

Gyrus ACMI Inc. BiCoag Hemostasis probe Gyrus ACMI Inc 136 Turnpike Road Southborough, MA 01772

Traditional 510(k) Notification Statement of Intended Use

Device Name: Gyrus ACMI Inc. BiCoag Hemostasis Probe 510(k) Number: OCT 2 0 2010 Indications for use: The Gyrus ACMI Inc. BiCoag Hemostasis Probe is intended for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract during endoscopic procedures.

Prescription Use: __ X

Over-the-Counter Use:

(Per 21 CFR 801.109)

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Only, Orthopedic, Division of Suices

510(k) Number K092571

---

**Source:** [https://fda-staging.innolitics.com/device/K092571](https://fda-staging.innolitics.com/device/K092571)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
