The ALATUS Vaginal Balloon Packing System is a single use, non-sterile, disposable, flexible, inflatable, non-powered positioning device, intended to be used on a daily treatment basis for the temporary positioning of the vaginal wall and adjacent structural anatomies. The purpose of the device is to displace and stabilize the vaginal wall during computed tomography (CT) exam, x-ray, or radiation treatment (RT) therapy. The placement of the balloon requires a physician or physician directed healthcare professional, and it is performed as a separate procedure outside of the standard (CT) exam and (RT) treatment. This device is not intended to be inserted into the uterine cavity.
Device Story
ALATUS Vaginal Balloon Packing System is a non-powered, single-use, disposable positioning device. Components include a polyurethane balloon, catheter tubing, connector, flex tubing, and a 1-way stopcock with a 90-degree turn handle. Used by physicians or physician-directed healthcare professionals in clinical settings (CT, x-ray, or radiation therapy suites). The device is inserted into the vagina to displace and stabilize the vaginal wall and adjacent anatomy, ensuring predictable and reproducible positioning during imaging or radiation treatment. It is placed as a separate procedure prior to the standard exam or treatment. The device is latex-free.
Indicated for temporary positioning, displacement, and stabilization of the vaginal wall and adjacent structural anatomies during CT, x-ray, or radiation therapy. For use by physicians or physician-directed healthcare professionals. Contraindicated for insertion into the uterine cavity.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
{0}------------------------------------------------
K092534
## SECTION 5 - 510(k) SUMMARY
# Submission Correspondent
Emergo Group, Inc.
www.emergogroup.com
DEC - 7 2009
Address 1705 S. Capital of Texas Hwy Suite 500 Austin, TX 78746
## Phone
(512) 327-9997
### Fax
(512) 327-9998
## Contact
Stuart R. Goldman
## Submission Sponsor
- Radiadyne 3801 Kirby Suite 710 Houston, TX 77098 Tel: 713.807.9100 Fax: 713.807.8030
### www.Radiadyne.com/
### Date Prepared
July 29, 2009
### Trade Name
ALATUS Vaginal Balloon Packing System
## Regulation Name(s)
{1}------------------------------------------------
Medical charged-particle radiation therapy system (primary) Obstetric-gynecologic general manual instrument (secondary)
#### Requlation Number(s)
892.5050 (primary) 884.4520 (secondary)
### Classification Name(s)
System, planning, radiation therapy treatment (primary) Retractor, vaginal (secondary)
## Product Code(s)
MUJ (primary) HDL (secondary)
### Classification Panel
Radiology (primary) Obstetrics/Gynecology (secondary)
## Regulatory Class
Class II
### Device Description
The ALATUS Vaginal Balloon Packing System is designed as an immobilizer to assist in positioning and displacing the vaginal wall in a more predictable and reproducible location during computed tomography (CT) exams and radiation treatment (RT) therapy. The proposed device is a latex free balloon made from Polyurethane consisting of the following components:
- 1. Balloon
- 2. Catheter tubing
- 3. Connector
- 4. Flex tubing
- 5. 1-way stopcock, 90° turn handle
### Intended Use
{2}------------------------------------------------
The ALATUS Vaginal Balloon Packing System is a single use, non-sterile, disposable, flexible, inflatable, non-powered positioning device, intended to be used on a daily treatment basis for the temporary positioning of the vaginal wall and adjacent structural anatomies. The purpose of the device is to displace and stabilize the vaginal wall during computed tomography (CT) exam, x-ray, or radiation treatment (RT) therapy. The placement of the balloon requires a physician or physician directed healthcare professional, and it is performed as a separate procedure outside of the standard (CT) exam and (RT) treatment. This device is not intended to be inserted into the uterine cavity.
### Predicate Device(s)
1. Mick Radio-Nuclear Instruments, Inc. - Vaginal Applicator Shielded (K001544) 2. Sterilis, Inc. - Vaginal Port Inflatable Speculum (K002912)
### Safety and Effectiveness
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, and it can be demonstrated that the device is as safe and effective as the predicate device, and that the new device does not raise different questions regarding safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the differences between the ALATUS Vaginal Balloon Packing System and the predicate devices listed do not raise any questions regarding its safety and effectiveness. The subject device, as designed and manufactured, is as safe and effective as the predicate devices for its intended application; that is, as a temporary vaginal wall positioning device, and therefore is determined to be substantially equivalent to the referenced predicate devices in the context of that application.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or other bird-like figure with outstretched wings.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD . 20993-0002
Radiadyne % Mr. Stuart R. Goldman Senior Consultant Emergo Group, Inc. 1705 S. Capital of Texas Hwy, Suite 500 AUSTIN TX 78746
## DEC - 7 2009
Re: K092534
Trade/Device Name: ALATUS Vaginal Balloon Packing System Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radio-nuclide applicator system Regulatory Class: II Product Code: JAQ and IYE Dated: November 20, 2009 Received: November 23, 2009
Dear Mr. Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 -
Enclosure
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.html
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
## INDICATIONS FOR USE
092534
510(k) Number (if known):
#### Device Name
## ALATUS Vaginal Balloon Packing System
### Indications for Use
The ALATUS Vaginal Balloon Packing System is a single use, non-sterile, disposable, flexible, inflatable, non-powered positioning device, intended to be used on a daily treatment basis for the temporary positioning of the vaginal wall and adjacent structural anatomies. The purpose of the device is to displace and stabilize the vaginal wall during computed tomography (CT) exam, x-ray, or radiation treatment (RT) therapy. The placement of the balloon requires a physician or physician directed healthcare professional, and it is performed as a separate procedure outside of the standard (CT) exam and (RT) treatment. This device is not intended to be inserted into the uterine cavity.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Page of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.