K092484 · Cardinal Health 207, Inc. · BZR · Apr 2, 2010 · Anesthesiology
Device Facts
Record ID
K092484
Device Name
HELIOX LOW FLOW SENTRY BLENDER
Applicant
Cardinal Health 207, Inc.
Product Code
BZR · Anesthesiology
Decision Date
Apr 2, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5330
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Heliox Low Flow Sentry Blender is designed to provide a continuous heliox/oxygen gas mixture to infant, pediatric, and adult patients. It is intended as a device to mix an 79% Helium / 21% Oxygen gas source with a 100% oxygen gas source to provide low flow mixtures containing 20% to 100% oxygen with the balance of the gas being Helium for patients who are spontaneously breathing. It is a medical device intended for use by qualified, trained personnel, under the direction of a physician, in institutional environments where the delivery and monitoring of helium/oxygen mixture is required.
Device Story
Device mixes 79% Helium/21% Oxygen source with 100% Oxygen source to deliver heliox/oxygen mixtures; includes integral battery-operated oxygen concentration analyzer with user-adjustable alarms. Used in institutional environments by qualified, trained personnel under physician direction. Input: gas sources; Output: heliox/oxygen mixture and O2 concentration monitoring. Provider uses output to adjust gas delivery; benefits patient by providing controlled heliox therapy. Modified from Sentry air/oxygen microblender (K911962) with updated poppet orifice to account for helium density and modified dial knob labeling.
Clinical Evidence
Bench testing only. Performance verification included dimensional envelope, weight, interface, supply/operating pressure, O2 control, flow characteristics, blender safety features, and monitor/analyzer performance (display, controls, alarms, system accuracy). All tests passed.
Technological Characteristics
Gas mixing valve with integral O2 analyzer. Modified poppet orifice for helium density. Battery-operated monitor with digital display, user-adjustable high/low O2 alarms, calibration, and lock/unlock controls. Standalone device for institutional use.
Indications for Use
Indicated for infant, pediatric, and adult patients who are spontaneously breathing and require delivery and monitoring of heliox/oxygen gas mixtures (20-100% O2, balance Helium) in institutional settings.
Regulatory Classification
Identification
A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.
{0}------------------------------------------------
K092484
Image /page/0/Picture/1 description: The image contains the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The text "CardinalHealth" is written in a simple, sans-serif font.
APR - 2 2010
# 510(K) Summary
# 510 (K) Summary as required by 21 CFR 807.92
| 510 (K) Submitter: | Cardinal Health 207, Inc.<br>Yorba Linda, CA 92887<br>(714) 283 - 2228 |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Monther Abushaban<br>(714) 919 - 3660<br>Monther.Abushaban@CardinalHealth.com |
| Establishment Registration Number: | 2050001 |
| Date Prepared: | April 1, 2010 |
| Name of the Device: | Heliox Low Flow Sentry Blender |
| Common/ usual name: | Oxygen Blender for Helium/Oxygen Mixtures |
| Classification: | The Heliox Low Flow Sentry Blender is classified as<br>a class II device under the following classification<br>code |
| Product code | CFR Section | Panel |
|--------------|-----------------|----------------|
| 73 BZR | 21 CFR 868.5330 | Anesthesiology |
Predicate Device: The predicate devices are:
Sentry air/oxygen microblender (K911962) Low Flow air/oxygen blender (K883038) Helio2 Blender (K090781)
Device Description: The helium/oxygen and oxygen blender/monitor combines a precision helium/oxygen (79%/21%) and oxygen mixing valve with an integral battery operated oxygen concentration analyzer and user adjustable alarms.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The graphic is made up of three curved lines that converge at a point, resembling the wings of a bird in flight. The text "CardinalHealth" is written in a simple, sans-serif font.
### Intended Use:
The Heliox Low Flow Sentry Blender is designed to provide a continuous heliox/oxygen gas mixture to infant, pediatric, and adult patients. It is intended as a device to mix an 79% Helium / 21% Oxygen gas source with a 100% oxygen gas source to provide low flow mixtures containing 20% to 100% oxygen with the balance of the gas being Helium for patients who are spontaneously breathing. It is a medical device intended for use by qualified, trained personnel, under the direction of a physician, in institutional environments where the delivery and monitoring of helium/oxygen mixture is required.
### Modifications to the predicate device:
The Heliox Low Flow Sentry Blender is the same device as the Sentry air/oxygen microblender (K911962). Modifications to Heliox Low Flow Sentry Blender are associated with this submittal are as follows:
- Intended use: Changed the gas mixture from air/oxygen to He- 02/oxygen . mixture.
- . The orifice of the poppet has been changed to address the differences in density of the helium/oxygen mixture from the density of air.
- . The dial knob label artwork has been changed from analog numbers to a amplitude range.
- In summary, the Heliox Low Flow Sentry Blender described in this submission is, in our opinion, substantially equivalent to the predicate device(s).
#### Reason for the submission:
This is a modified device with a new intended use to be marketed by Cardinal Health 207, Inc.
#### Summary of Technological Characteristics of Device Compared to the Predicate Device:
The Heliox Low Flow Sentry Blender is substantially equivalent in safety and effectiveness to the legally marketed (predicate) currently in distribution.
## Summary of Non-clinical Testing for the Device and Conclusions:
Performance testing verified that the Heliox Low Flow Sentry Blender meets it's performance requirements that this device is substantially equivalent to medical devices currently legally marketed in the United States.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the Cardinal Health logo. The logo consists of a stylized graphic above the text "CardinalHealth". The graphic is composed of three curved lines that converge at a point, resembling a bird in flight or abstract wings.
| 91727 Rev A | | ER 3348 | | Pass or Fail |
|--------------------------|--------------------------------------|------------------------------------------------|--------------------------------|--------------|
| PRS Reference<br>Numbers | Description | Verification Test Protocol<br>Reference Number | Number of<br>Units for<br>Test | |
| 4.1.1 | Dimensional Envelope | ER3348 Section 4.1.1 | 4 | Pass |
| 4.1.2 | Weight | ER3348 Section 4.1.1 | 4 | Pass |
| 4.1.3 | Interface (gas inlets and outlets) | ER3348 Section 4.1.1 | 4 | Pass |
| 4.2.1 | Nominal Supply Pressure | ER3348 Section 4.1.1 | 4 | Pass |
| 4.2.2 | Normal Operating Pressure | ER3348 Section 4.1.1 | 4 | Pass |
| 4.3 | Environmental Withstand | Refer to Sentry Blender<br>DHF | N/A | Pass |
| 4.4.1 | % O2 Control | ER3348 Sect. 5.2 | 4 | Pass |
| 4.4.2 | Flow Characteristics | ER3348 Sect. 5.2 | 4 | Pass |
| 4.4.4 | Blender Safety Features | ER3348 Sect. 5.1, 5.3 | 4 | Pass |
| 4.5.1 | Monitor Display | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 a | Oxygen Range, Resolution, & Accuracy | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 b | Low O2 Alarm Set point | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 c | High O2 Alarm Set point | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 d | "%" Flashing during calibration | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 e | Locked Indicator | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 f | Attach Sensor | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.1 g | Low Battery Indicator | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2 | Monitor Controls | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2 a | On/Off Switch | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2 b | Lock/Unlock Switch | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2.1 | Calibrate Button | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2.2 | High Set Button | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2.3 | Low Set Button | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.2.4 | Alarm Silence Button | ER3348 Sect. 5.1 | 4 | Pass |
| 4.5.3 | Alarm/Alert Indicatons | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.3 a | Hi O2 Concentration | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.3 b | Low O2 Concentration | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.3 c | Alarm Silence | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.3 d | Alarm Loudness | ER3348 Sect. 5.3 | 4 | Pass |
| 4.5.4 | Monitor Power Source | Specified by Manufacturer<br>(not tested) | N/A | Pass |
| 4.5.5 | Oxygen Sensor | Specified by Manufacturer<br>(not tested) | N/A | Pass |
| 5.1 | Oxygen Blender Performance | ER3348 Sect. 5.2 | 4 | Pass |
| 5.2 | Monitor/Analyzer Performance | ER3348 Sect. 5.3 | 4 | Pass |
| 5.2.1 | Display, Controls, Alarms | ER3348 Sect. 5.3 | 4 | Pass |
| 5.2.2 | System Accuracy | ER3348 Sect. 5.3 | 4 | Pass |
| 6.1 | Labeling | ER3348 Section 4.1.1 | 4 | Pass |
| 6.2 | Shipping Container | ER3348 Section 4.1.1 | 4 | Pass |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Cardinal Health. The logo consists of a stylized graphic above the company name. The graphic appears to be a series of curved lines, possibly representing a bird in flight or a stylized medical symbol. The text "CardinalHealth" is written in a simple, sans-serif font.
# Conclusion:
The summary above shows that there are no new questions of safety and effectiveness of the subject device (Heliox Low Flow Sentry Blender) as compared to predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The graphic is made up of four curved lines that resemble the wings of a bird in flight. The text "CardinalHealth" is written in a simple, sans-serif font.
.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with three stripes forming its body and wing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Monther Abushaban Regulatory Affairs Manager Cardinal Health 207, Incorporated 22745 Savi Ranch Parkway Yorba Linda, California 92887
APR - 2 2010
Re: K092484
Trade/Device Name: Heliox Low Flow Sentry Blender Regulation Number: 21CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: II Product Code: BZR Dated: March 3, 2010 Received: March 31, 2010
Dear Mr. Abushaban:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Mr. Abushaban
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Th for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The text "CardinalHealth" is written in a simple, sans-serif font, with the first letter capitalized.
## Indications for Use
Device Name: Heliox Low Flow Sentry Blender
Indications for Use:
The Heliox Low Flow Sentry Blender is designed to provide a continuous heliox/oxygen qas mixture to infant, pediatric, and adult patients. It is intended as a device to mix an 79% Helium / 21% Oxygen gas source with a 100% oxygen gas source to provide low flow mixtures containing 20% to 100% oxygen with the balance of the gas being Helium for patients who are spontaneously breathing. It is a medical device intended for use by qualified, trained personnel, under the direction of a physician, in institutional environments where the delivery and monitoring of helium/oxygen mixture is required.
X AND/ OR Prescription Use (Part 21CFR 801 subpart D)
Over the Counter Use: (Part 21CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Division Sign-Off) ivision of Anesthesiology, General Hospital fection Control, Dental Devices
10(k) Number: K092484
Traditional 510(k) Submission for Heliox Low Flow Sentry Blender
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.