K092438 · Ziehm Imaging, Inc. · OWB · Dec 23, 2009 · Radiology
Device Facts
Record ID
K092438
Device Name
ZIEHM SOLO
Applicant
Ziehm Imaging, Inc.
Product Code
OWB · Radiology
Decision Date
Dec 23, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Ziehm Solo is intended to provide contactless fluoroscopy Imaging, capturing, temporarily storing, and display of non-invasive x-ray imaging of the patient using pulsed and continuous fluoroscopy, during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room fluoroscopy procedures. Radiographic film examinations can be made with an accessory cassette device when attached to the Image Intensifier. This device does not support and is not intended for use in performing mammographic imaging.
Device Story
Mobile C-arm system for intraoperative fluoroscopic imaging; utilizes 2.02 kW high-frequency generator and single-focus fixed-anode X-ray tube. Inputs: X-ray radiation through patient anatomy. Processing: Pulsed/continuous fluoroscopy and digital radiography (snapshot) modes; virtual collimation; image capture and temporary storage. Outputs: Real-time X-ray images displayed on high-resolution flat-panel LCD monitors mounted on a mechanical arm. Operated by physicians (orthopedic, vascular, neuro-vascular, radiologists) and technologists in ORs, clinics, and emergency rooms. Provides clinicians with immediate visualization of anatomical structures to guide surgical/interventional procedures; benefits include real-time guidance for minimally invasive interventions.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including adherence to FDA and IEC standards for diagnostic X-ray systems, safety, electromagnetic compatibility, and radiation protection.
Technological Characteristics
Mobile C-arm with 2.02 kW high-frequency generator; single-focus fixed-anode X-ray tube; mono-block generator housing. Features virtual collimator, touch-screen SoloCenter interface, and mechanical arm for LCD display mounting. Connectivity via external monitor ports, USB, and DVD. Standards: 21 CFR 1020.30-32, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-4, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, IEC 60825-1, ISO 14971.
Indications for Use
Indicated for diagnostic, surgical, and interventional fluoroscopic imaging in patients including pediatric, adolescent, and pregnant populations (use with caution). Clinical applications include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures. Contraindicated for mammographic imaging.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K0924358
#### Appendix 7 - 510(k) Summary Ziehm Solo
# DEC 2 3 2009
The following information is being submitted in accordance with the requirements of 21 CFR §807.92
| Preparation Date: | Dec 17, 2009 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name of Submitter: | Ziehm Imaging, Inc.<br>4181 Latham Street<br>Riverside, CA 92501<br>(951) 718-2020 |
| Contact Person: | Richard L. Westrich<br>Vice President of Regulatory Affairs and Quality Assurance<br>4181 Latham Street<br>Riverside, CA 92501<br>Office phone: +951-781-2020 ext 140<br>Fax: +951-781-6457 |
| Device Trade Name: | Ziehm Solo |
| Classification Name: | 21 CFR 892.1650 System, X-Ray, Fluoroscopic, Image Intensifier,<br>Product Code: JAA |
| Common/Usual Names: | Digital Mobile C-Arm<br>Mobile Surgical C-Arm<br>Mobile C-Arm |
| Predicate Device: | ZIEHM VISION2 K073346, 21CFR 892.1650, System, X-Ray, Fluoroscopic,<br>Image Intensifier,<br>Product Code JAA, IZL |
| Device Description: | The Ziehm Solo Mobile Stand incorporates a single arm monitor assembly<br>providing a more adaptable positioning of the display monintors to the clinician<br>and the C-arm in relation to the patient. The C-profile provides fixed distance<br>mounting of the generator and Image Intensifier and manual rotation around a<br>non iso-centric location. The mobile stand allows the manual rotational and linear<br>movements of the C-arm with a motorized vertical movement for positioning the<br>c-arm at various angles and distances for visualization of patient's anatomical<br>structures. Optional 120kV and or 20mA Fluoro operations are available. The<br>2.02 kW high frequency generator incorporates a single focus fixed anode x-ray<br>tube, and high voltage generator is a single mono-block generator tube housing<br>assembly. The virtual collimator and x-ray control mount to the generator<br>housing assemble and provides pulsed and continuous fluoroscopy operations<br>including a special digital radiography (snapshot) mode. The SoloCenter is a<br>centralized touch screen providing the user/operator with a graphical user |
Interface including the x-ray control panel. The Ziehm Solo does not have a conventional Monitor Cart, the workstation consists of a mechanical arm assembly supporting a single or dual high-resolution flat panel LCD display monitor(s). Ceiling or wall monitors can be added by means of two external monitor connectors. Other conventional external interfaces are provided on the mobile stand for preipherial devices such as video printers, and storage devices such as USB and DVD.
Section 7 - 510(k) Summary
K092438
Page 1 of 3-SUM
{1}------------------------------------------------
The Ziehm Solo is intended to provide contactless fluoroscopy Indications of Use: Imaging, capturing, temporarily storing, and display of non-invasive xray imaging of the patient using pulsed and continuous fluoroscopy, during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room fluoroscopy procedures. Radiographic film examinations can be made with an accessory cassette device when attached to the Image Intensifier. This device does not support and is not intended for use in performing mammographic imaging.
Contraindications to The exposure of humans to ionizing radiation must always be medically justified. Especially when used on pregnant women, adolescents, children, the use of X-rays: and pediatric patients, all procedures using ionizing radiation should be used with caution or be avoided altogether. However, the final decision and responsibility lies with the attending physician or attending surgeon in deciding whether to use the system with such patients.
> The system may only be used in heights up to 6561.7 ft (2000 m) above sea level and must be used within the limits defined by the technical specification. The use of the system is only allowed in rooms used for medical purposes in accordance 60601-1-2 with EMC class A as well as with protective earth conductor. The system may only be used in an environment with an oxygen saturation <25%. The system may only be used in faultless condition and in accordance with the terms set forth by the operating instructions.
The Ziehm Solo Employs the same fundamental scientific technology as Technology: the predicate device.
User Characteristics: The Ziehm Solo does not require nor is it intended to be used in contact with patients.
> In some circumstances however, as part of its use in clinical environments the patient may come in contact with the device when operator moves or positions the device. The device is intended for use by health care professionals such as but not limited to physicians, orthopedic surgeons, vascular surgeons, neuro-vascular surgeons, radiologists, or other clinical physicians and technologists in hospitals, emergency rooms, out-patient clinics, and other clinical environments. Ziehm Imaging anticipates the device will be used on a nearly daily basis. Ziehm Imaging applications specialists and/or qualified site personnel provide on site operator training in the proper use of the device.
The Ziehm Solo mobile x-ray devices shall be tested and be shown to meet the Standards: appropriate requirements of the following standards prior to being marketed.
| 21 CFR 1020.30-32 | Federal Performance Standard for Diagnostic X-ray Systems |
|-------------------|----------------------------------------------------------------------------------------------|
| IEC 60601-1, | Medical Electrical Equipment, General Requirements for Safety |
| IEC 60601-1-2 | Medical Electrical Equipment, General Requirements for Safety, Electromagnetic Compatibility |
| IEC 60601-1-3, | Medical Electrical Equipment, Radiation Protection in Diagnostic X-ray Equipment |
Section 7 - Ziehm Solo 510(k) Summary
Page 2 of 3-SUM
{2}------------------------------------------------
| IEC 60601-1-4, | General requirements for safety, Programmable electrical<br>medical systems. |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-2-7, | Medical Electrical Equipment, Safety of HV/X-ray<br>Generators (FDA/CDRH Recognition Number 12-34: IEC<br>60601-2-7 (1998), Medical electrical equipment - Part 2-<br>7: Particular requirements for the safety of high-voltage<br>generators of diagnostic X-ray generators. (Radiology)) |
| IEC 60601-2-28, | Medical Electrical Equipment Part 2: Particular<br>Requirements for the Safety of X-Ray Source Assemblies<br>and X-Ray Tube Assemblies for Medical Diagnosis |
| IEC 60601-2-32, | Medical Electrical Equipment, Safety of Associated X-ray<br>Equipment |
| IEC 60825-1, | Safety of laser products, Equipment Safety, requirements,<br>and user guide |
| IEC 14971 | Risk Management |
Determination of Substantial Equivalence:
The demonstration of substantial equivalence is based on a comparison of features to the predicate device and an assessment of non-clinical performance test data. Information is included with this submission that supports this determination.
The comparison table of the predicate device in Appendix 3 and the Performance Conclusion: testing, Clinical Evaluation Addendum, and Clinical Evaluation Ziehm Vision Family included in Appendix 5 demonstrates that the Ziehm Solo does not raise new questions of safety or effectiveness and performs in an equivalent manner to the predicate device and in accordance with its labeling.
End of 510(k) Summary
Richard L. Westrich Vice President Regulatory Affairs and Quality Assurance Ziehm Imaging, Inc. 4181 Latham St. Riverside, CA 92501
#### Section 7 - Ziehm Solo 510(k) Summary
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
· Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
### DEC 2 3 2009
Mr. Richard L. Westrich V. P. Regulatory Affairs & Quality Assurance Ziehm Imaging, Inc. Division of Ziehm LLC (based in the USA) 4181 Latham Street RIVERSIDE CA 92501
Re: K092438
Trade/Device Name: Ziehm Solo Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: October 29, 2009 Received: November 9, 2009
Dear Mr. Westrich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean . that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 -
Enclosure
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Ziehm Imaging. The logo consists of a circular graphic with a stylized human figure inside, next to the text "ziehm imaging". The text is in a simple, sans-serif font and is horizontally aligned with the graphic.
## Indications for Use Statement
Applicant: Ziehm Imaging, Inc.
K092438 510(k) Number (if known):
Device Name: Ziehm Solo
Indications for Use:
The Ziehm Solo is intended to provide contactless fluoroscopy Imaging, capturing, temporarily storing, and display of non-invasive x-ray imaging of the patient using pulsed and continuous fluoroscopy, during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room fluoroscopy procedures. Radiographic film examinations can be made with an accessory cassette device when attached to the Image Intensifier. This device does not support and is not intended for use in performing mammographic imaging.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | AND/OR | Over-The-Counter Use | |
|------------------------------------------------------------|--------|------------------------|--|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 801 Subpart C) | |
| (Division Sign-Off<br>Division of Reproductive, Abdominal, | | | |
| and Radiological Devices 7<br>510(k) Number_ | | 1097438 | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.