OSV II LUMBAR VALVE SYSTEM AND OSV II LOW FLOW LUMBAR VALVE SYSTEM
Applicant
Integra Lifesciences Corp.
Product Code
JXG · Neurology
Decision Date
Nov 19, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OSV II® Lumbar Valve System is an implantable system used in the treatment of patients with communicating hydrocephalus to shunt CSF from the lumbar subarachnoid region to the peritoneal cavity. The OSV II Low Flow Lumbar Valve System is an implantable system used in the treatment of patients with communicating hydrocephalus to shunt CSF from the lumbar subarachnoid region to the peritoneal cavity.
Device Story
Implantable variable resistance valve system for controlled cerebrospinal fluid (CSF) drainage from lumbar subarachnoid space to peritoneal cavity. Device utilizes silicone elastomer diaphragm, synthetic ruby seat, and notched pin to regulate flow based on differential pressure (DP). Operates in three stages: low DP (5-18 ml/hr), flow regulation (120-300 mm H2O DP), and pressure relief mode for rapid flow during abrupt intraventricular pressure elevation. System includes 14G tuohy needle, F5 lumbar catheter, guidewire, and connectors. Used by neurosurgeons in clinical settings to manage hydrocephalus by maintaining constant drainage rates (approx. 20ml/hr for OSV II; 10ml/hr for Low Flow). Output is physical CSF diversion; helps normalize intracranial pressure and treat hydrocephalus symptoms.
Clinical Evidence
No clinical data provided; substantial equivalence based on identical design, materials, and principle of operation to predicate devices.
Indicated for patients with communicating hydrocephalus requiring CSF shunting from the lumbar subarachnoid region to the peritoneal cavity.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
{0}------------------------------------------------
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
### Submitter's Name and Address:
NOV 1 9 2009
... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Integra LifeSciences Corporation 311 Enterprise Drive Plainsboro, NJ 08536, USA Telephone: (609) 654-2873 Facsimile: (609) 275-9445
### Contact Person and Telephone Number:
Donna Millisky Regulatory Associate II Integra LifeSciences Corporation Integra NeuroSciences Division Telephone: (609) 936-2454 Facsimile: (609) 275-9445 E-mail: donna.millisky(@Integra-LS.com
Date Summary was Prepared: July 31, 2009
Name of the Device:
Trade Names:
Common Name: Classification Name: Product Code: Classification Panel:
OSV II® Lumbar Valve System OSV II® Low Flow Lumbar Valve System Central Nervous System Fluid Shunt and Components Hydrocephalus Valve JXG Neurology
### Device Description:
The OSV II® Lumbar Valve System (OSV II Valve) and the OSV II® Low Flow Lumbar Valve System (OSV II Low Flow Valve) are implantable devices for controlled cerebrospinal fluid drainage (CSF) from the lumbar subarachnoid region to the peritoneal cavity. Unlike conventional valves, they are variable resistance valves which maintain a drainage rate constant within the physiological range (for the specified populations and disorders) of intracranial pressure. They maintain a drainage rate around 20ml/hr for OSV II Valves and around 10ml/hr for OSV II Low Flow Valves.
The OSV II Lumbar Valve System and OSV II Low Flow Lumbar Valve System are provided with accessories including a 14G tuohy needle, a F5 lumbar catheter, a guidewire, a luer lock connector and a F8-F5 stepdown connector, which are required for using the valves in a lumbar approach. The designs of the accessories are identical to the accessories currently provided with the Integra H-V Lumbar Valve System.
{1}------------------------------------------------
# Indications for Use:
Both the OSV II Lumbar Valve System and the OSV II Low Flow Lumbar Valve System are indicated as implantable systems used in the treatment of patients with communicating hydrocephalus to shunt CSF from the lumbar subarachnoid region to the peritoneal cavity.
# Principle of Operation:
The Principle of Operation of both the OSV II Lumbar Valve System and OSV II Low Flow Lumbar Valve System are as follows:
The variable flow restrictor consists of a silicone elastomer diaphragm, synthetic ruby seat, and notched pin. The diaphragm reacts to differential pressure (DP) variations. The seat is inserted into the center of the diaphragm. As differential pressure varies, the clearance between the seat and the pin increases depending on seat movement along the pin. Changes in the seat and pin clearance vary flow rates.
Note - Differential pressure/flow characteristics for each valve are verified to be within specification at time of manufacture.
The three stages of operation of the OSV II® Lumbar Valve System and the OSV II Low Flow Lumbar Valve System are defined below:
# Stage I - Low Differential Pressure
This stage begins when the flow rate through the valve reaches 5ml/hr (DP will be between 30 and 80 mm H20). The valve remains in Stage I with CSF flow rates up to 18 ml/hr (up to 8 ml/hr for the OSV II Low Flow) (DP will be between 40 and 120 mm and 120 mm H20).
#### Stage II - Flow Regulation
When the DP increases, the valve operates as a variable resistance flow regulator. At DP ranges between 120 and 300 mm H20, the valve restricts flow between 18 and 30 ml/hr (between 8 and 17 ml/hr for the OSV II Low Flow).
#### Stage III - Pressure Relief Mode
Should the intraventricular pressure (IVP) elevate abruptly, the shunt operates in a rapid flow rate mode to facilitate IVP normalization. The valve then reverts to Stage II or I, depending upon conditions.
# Materials:
The materials of composition for both the OSV II Lumbar Valve System and OSV II Low Flow Lumbar Valve System are identical: silicone elastomer with or without barium sulfate, polysulfone, synthetic ruby, silicone adhesive, tantalum, polypropylene and epoxy.
The accessories include a lumbar catheter and suture clamp made of silicone elastomer with
{2}------------------------------------------------
# Integra LifeSciences Corporation-Traditional 510(k) Bundled OSV II and OSV II Low Flow Valve Systems - Lumbo-Periotoneal
barium sulfate. The guidewire used to introduce the catheter is made of stainless steel, coated with polytetrafluoroethylene (PTFE), the stepdown connector is made of polypropylene with barium sulfate and the luer connector (to flush catheter) is made of polycarbonate. The 14G tuohy needle is made of stainless steel, with polycarbonate wings. The materials used in the accessories are identical to those materials used in the H-V Lumbar Valve accessories. except the F8-F5 stepdown connector is made of stainless steel in the H-V Lumbar Valve and is made of polypropylene with barium sulfate for the proposed OSV II and OSV II Low Flow valves. The current OSV II and Integra NPH Low Flow Valve systems already provide straight connectors made of the same polypropylene with barium sulfate and a lumbar catheter made with silicone elastomer.
# Substantial Equivalence:
The proposed OSV II and OSV II Low Flow Valves are substantially equivalent in function to the currently approved OSV II and Integra NPH Low Flow Valve Systems cleared to market by FDA under K971799, K081773, and K042192. This bundled 510(k) supports the use of both valve systems using a lumbo-peritoneal approach. The accessories provided allow for a lumbar approach and are substantially equivalent to the accessories provided with the Integra H-V Lumbar Valve System, which was cleared by FDA under K944595 for the same intended use. These accessories include a 14G tuohy needle, F5 lumbar catheter, guidewire, stepdown connector, and suture clamp.
The OSV II Lumbar Valve System and OSV II Low Flow Lumbar Valve System are designed the same as the valves currently used with the OSV II and Integra NPH Low Flow Valve Systems provided for the ventricular approach. Since the valves are the same, the product and performance specifications are the same as the predicate valves.
The Principle of Operation, intended use, design, materials of composition and manufacturing processes for both the OSV II Lumbar Valve System and OSV II Low Flow Lumbar Valve System are the same as for the current OSV II and Integra NPH Low Flow hydrocephalus valves. The labeling has been revised to include the indications for treatment of patients with communicating hydrocephalus to shunt CSF from the lumbar subarachnoid region to the peritoneal cavity. Based upon the stated information and supported by the clinical evidence included in this 510(k), Integra concludes that both systems are equivalent to the predicates defined in this submission.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Integra LifeSciences Corporation c/o Ms. Donna Millisky Regulatory Associate II 311 Enterprise Drive Plainsboro, NJ 08536
# NOV 1 9 2009
Re: K092395
Trade/Device Name: OSV II® Lumbar Valve System, OSV II® Low Flow Lumbar Valve System Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: II Product Code: JXG Dated: October 23, 2009 Received: October 26, 2009
Dear Ms. Millisky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Malvina B. Eydelman, M.D.
Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number:
K092395
Device Name: OSV II Lumbar Valve System
Indications for Use:
The OSV II® Lumbar Valve System is an implantable system used in the treatment of patients with communicating hydrocephalus to shunt CSF from the lumbar subarachnoid region to the peritoneal cavity.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (Part 21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
JOE HUTTER
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K092395
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
K092395 510(k) Number: Device Name: OSV It Low Flow Lumbar Valve System Indications for Use: The OSV II Low Flow Lumbar Valve System is an implantable system used in the treatment of patients with communicating hydrocephalus to shunt CSF from the lumbar subarachnoid region to the peritoneal cavity.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter-Use_ (Part 21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Joe HUTTER
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K092395
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.