K092366 · Thermal Therapeutic Systems, Inc. · LGZ · Mar 4, 2010 · General Hospital
Device Facts
Record ID
K092366
Device Name
TTS100 PORTABLE HYPERTHERMIC PERFUSION DEVICE
Applicant
Thermal Therapeutic Systems, Inc.
Product Code
LGZ · General Hospital
Decision Date
Mar 4, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the TTS100 Portable Hyperthermic Perfusion Device is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating, warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician.
Device Story
The TTS100 is a prescription-use portable hyperthermic perfusion device consisting of a durable console and a sterile, single-use Hyperthermic Lavage Tubing Set. The console features a heater, roller-type fluid pump, microprocessor, and touchscreen interface. The disposable set includes a fluid reservoir, inlet/outlet tubing, and auxiliary temperature sensors; portions of the accessory kit may be implanted during lavage. The device circulates sterile fluid through the cavity; fluid is heated in the reservoir and pumped to the patient, returning via gravity. The system monitors fluid temperature and pressure, providing real-time data and alarms on the touchscreen. Software automatically responds to unsafe conditions or sensor failures, while independent protection circuits ensure safety during software faults. Used in clinical settings, the device allows physicians to control cavity temperature, potentially benefiting patients by maintaining therapeutic hyperthermia. The closed-loop disposable design prevents fluid contact with console components, facilitating rapid post-procedure processing.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including software verification/validation, biocompatibility testing (ISO 10993), electrical safety/EMC testing (IEC 60601), and human factors studies to mitigate user-related risks.
Technological Characteristics
Durable console with heater, roller pump, and touchscreen; sterile single-use tubing set. Materials comply with ISO 10993. Connectivity includes sensor monitoring of fluid path. Sterilization via ethylene oxide (ISO 11135). Software-controlled monitoring of temperature and pressure with independent hardware protection circuits. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for patients requiring thoracic or peritoneal cavity lavage with warmed, physiologically compatible sterile solution to achieve a target temperature, as determined by a physician-selected protocol.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
## MAR - 4 2010
ThermalTherapeuticSystems, Inc.
POB 82670 Pittsburgh, PA 15218 Phone: 412.571.2622 www.thermaltherapeutics.com
## 510(k) Summary
## Model TTS100 Portable Hyperthermic Perfusion Device
#### 510(k) Number: K092366
Submitted in accordance with the requirements of SMDA 1990 and 21CFR807.92.
## 1.0 21CFR 807.92(A)(1)
1.1 Applicant's Information
J. Michael Fausset BS CP Chief Technical Officer ThermalTherapeutic Systems, Inc. Mobile: 765 427.3308 Fax: 412 291.1951 E: imf@thermaltheraneutics.com I: http://www.thermaltherapeutics.com Medical Establishment Registration No .: pending
1.2 Submitter's Information
James Jochen Rogers General Manager Coastal Consulting Group, Ltd. P.O. Box 961 Moon Township, PA 15108 Mobile: 724 713.2298 Fax: 412 291.1951 E: ir@coastalco.com I: http://www.coastakcg.com
- 1.3 Date
March 1, 2010
## 2.0 21CFR 807.92(A)(2)
2.1 Device Information
Trade/Proprietary Name: TTS100 Portable Hyperthermic Perfusion Device Common Name: TTS100 Portable Hyperthermic Perfusion Device Model Number: TTS100 ·Regulation Number: None Regulation Name: None Regulatory Class: Unclassified, pre-amendment device; non-exempt from 510(k) Device Classification Name: Warmer, thermal, Infusion fluid Classification Panel: General Hospital Classification Product Code(s): LGZ
2.2 Device Classification
Infusion fluid thermal warmers and their accessories are unclassified, Pre-Amendment devices. 510(k) premarket notification submissions for these devices are reviewed by the General Hospital Devices Branch, Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices,
{1}------------------------------------------------
## ThermalTherapeuticSystems, Inc.
## 3.0 21CFR 807.92(A)(3)
3.1 Predicate Device Information
| PREDICATE DEVICES | | | |
|-------------------|---------------------------|--------------------------|--------------------------|
| 510(k) # | DEVICE | 510(k) SPONSOR | 510(k)<br>CLEARANCE DATE |
| K993330 | ThermoChem-HT System | Hemocleanse, Inc. | 12/30/1999 |
| K070654 | Belmont Hyperthermia Pump | Belmont Instrument, Inc. | 6/8/2007 |
## 4.0 21CFR 807.92(A)(4)
#### 4.1 Device Description
The TTS100 Portable Hyperthermic Perfusion Device is a prescription device comprised of two components - one durable and one disposable. The TTS L00 Console is the durable component - a portable, unit containing heater, pump, user touchscreen, microprocessor, and interface electronics. The primary user interface of the TTS100 is a large, color touchscreen display that receives operational commands from the clinical user, and presents normal operating conditions, flow rate, output fluid temperature, patient temperature, target temperature, line pressure, alarm and status messages, various timers and auxiliary temperatures. The TTS100 software monitors various sensors in the fluid path to ensure safe operation, automatically reacts to unsafe conditions, and alerts the clinical user to those conditions for resolution. Independent protection circuits prevent unsafe operation in the event of system software fault.
The disposable, single-use component is the Hyperthermic Lavage Tubing Set, which has two subcomponents - a Circulation Loop and an Accessory Kit. The Circulation Loop contains the filtered soft-shell fluid reservoir, inlet ports, vent lines, integrated distributed sensors, patient inlet circulation loop tubing, patient outlet circulation loop tubing, and drain line. The Accessory Kit - an integral component of the Hyperthermic Lavage Tubing Set - contains patient inlet and patient outlet tubing, tubing connectors, and auxiliary temperature sensors. Portions of the Accessory Kit can be implanted into the patient for use during the lavage procedure. The Circulation Loop does not have direct patient contact - only the circulating fluid within the loop contacts the patient. The Hyperthermic Lavage Tubing set is designed such that fluid travels through the Disposable Set without wetting the heater, pump or other TTS system components (e.g., only the internal surfaces of the Disposable Set are wetted), enabling rapid and straightforward post-processing of the TTS100 Console between patients. The Hyperthermic Lavage Tubing Set is supplied sterile, packaged in a proprietary-design container, and is only intended to be used in conjunction with the TTS100 Portable Hyperthermic Perfusion Device.
### 5.0 21CFR 807.92(A)(5)
- 5.1 Indications for Use
The intended use of the TTS100 Portable Hyperthermic Perfusion Device is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating, warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician.
#### 6.0 21CFR 807.92(A)(6)
#### 6.1 Technological Characteristics Comparison to Predicate Device(s)
As compared to the predicate devices, the TTS100 Portable Hyperthermic Perfusion Device has the same operating principle, energy type, environmental specifications, and performance specifications. All devices use a roller-type fluid pump, touchscreen-based clinical user Interface for machine set-up and control, sensor monitoring of various fluid temperatures, and a proprietary-design disposable set including large fluid reservoir to circulate sterile fluid into and out of the body cavity. For the TTS100, fluid in the reservoir is heated via direct heat to the desired temperature, and the heater drain tubing passes through a roller pump and out to the patient inlet. Fluid returns from the patient outlet wia gravity drain back to the fluid reservoir. The TTS100 monitors circulating fluid temperature and pressure, and automatically responds to ineffective or unsafe operating conditions.
{2}------------------------------------------------
## ThermalTherapeuticSystems, Inc.
## 7.0 21CFR 807.92(B)(1)
#### 7.1 Performance Data - Discussion of Non-Clinical Tests
- 7.1.1 FDA Recognized Consensus Standards
The TTS100 is designed to, complies with, and has been tested as part of verification and validation activities to, the FDA Recognized Consensus Standards listed in the table below, as applicable to device features and components:
| REQUIREMENT | REFERENCE # | TITLE |
|---------------------------|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | AAMI/ANSI/ISO 10993-1:2003(E) | Biological evaluation of medical devices - Part<br>1: Evaluation and testing |
| | AAMI/ANSI/ISO 10993-5:1999 | Biological evaluation of medical devices - Part<br>5: Tests for in vitro cytotoxicity |
| | AAMI / ANSI / ISO 10993-<br>7:1995(R)2001 | Biological Evaluation of Medical Devices - Part<br>7: Ethylene Oxide Sterilization Residuals. |
| | AAMI/ANSI/ISO 10993-10:2002 | Biological evaluation of medical devices -<br>Part 10: tests for irritation and delayed-type<br>hypersensitization |
| | AAMI/ANSI/ISO 10993-11:2006 | Biological evaluation of medical devices - Part<br>11: Tests for systematic toxicity |
| Electrical<br>Safety, EMC | IEC<br>60601-1:1988+A1:1991+A2:1995<br>(IEC 60601-1+A1+A2) | Medical electrical equipment - Part 1: General<br>Requirements for Safety, 1988; Amendment 1,<br>1991-11, Amendment 2, 1995 |
| | IEC 60601-1-2:2001+A1:2004<br>(IEC 60601-1-2+A1) | Medical Electrical Equipment - Part 1-2: General<br>Requirements for Safety - Collateral standard:<br>Electromagnetic Compatibility - Requirements<br>and Tests (Ed. 2.1) |
| Sterility | AAMI / ANSI / ISO 11135-1:2007 | Sterilization of health care products -<br>Ethylene oxide - Part 1: Requirements for the<br>development, validation, and routine control of<br>a sterilization process for medical devices. |
| Packaging | ASTM D4169-05:2005 | Standard Practice for Performance Testing of<br>Shipping Containers and Systems |
#### 7.1.2 Software Verification/Validation Testing
As part of software verification/validation activities, and in support of substantial equivalence, the following tests were carried out the TTS100 to evaluate its ability to meet product performance specifications. Specifically, the TTS100 was tested to assess ...
- I The ability of the TTS100 to heat fluids over the full range of fluid flowrates.
- The ability of the TTS100 to maintain fluid temperature over the full range of fluid flowrates. E
- E The ability of the TTS100 to detect and automatically respond to unsafe or ineffective operating conditions, as caused by the failure of the system sensors, excessive fluid temperature, excessive fluid pressure, or system-internal component failures.
- The ability of the TTS100 to inform the clinical user of those unsafe or ineffective operating conditions via notifications, alerts, alarms, and/or system faults.
- D The ability of the TTS100 to mitigate against known or predictable operator errors.
- The ability of the TTS100 to store treatment parameters and data in non-volatile memory. D
- ם Measured temperature accuracy over the full range of fluid flowrates within the operating pressure range.
- D Measured pressure accuracy over the full range of fluid flowrates within the operating temperature range.
- Fluid flowrate accuracy over the full range of fluid flowrates within the operating temperature and pressure range.
The TTS100 performed within specification in all of the above tests.
{3}------------------------------------------------
# K092366
## ThermalTherapeuticSystems, Inc.
#### 7.1.3 Human Factors
A human factors non-cilnical study evaluated user- and patient-safety risks associated with the design of the TTS100, and demonstrated that the task-risks have been identified, addressed and mitigated.
## 8.0 21CFR 807.92(s)(2)
- 8.1 Performance Data Discussion of Clinical Tests
N/A
- 9.0 21CFR 807.92(B)(3)
#### 9.1 Conclusions
Based upon safety and performance testing and compliance with voluntary standards, ThermalTherapeutic Systems, Inc. believes that the TTS100 Portable Hyperthermic Perfusion Device is substantially equivalent to the predicate devices, and does not raise any new questions of safety or effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three curved lines.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Thermal Therapeutic Systems, Incorporated C/O Mr. James Jochen Rogers General Manager Coastal Consulting Group, Limited P.O. Box 961 Moon Township, Pennsylvania 15108
JAN 1 0 2017
Re: K092366
Trade/Device Name: TTS100 Portable Hyperthermic Perfusion Device Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LGZ Dated: March 3, 2010 Received: March 4, 2010
Dear Mr. Rogers:
This letter corrects our substantially equivalent letter of March 4, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mr. James Jochen Rogers
CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K092366
Device Name: TTS100 Portable Hyperthermic Perfusion Device
Indications For Use:
The intended use of the TTS100 Portable Hyperthermic Perfusion Device is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating, warmed, physiologically compatible sterile solution, according lo a protocol to be selected by the physician.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE-ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Charles Schwab
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K092366
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.