K092349 · Inverness Medical, Inc. · PSZ · May 10, 2010 · Microbiology
Device Facts
Record ID
K092349
Device Name
CLEARVIEW EXACT II INFLUENZA A & B TEST
Applicant
Inverness Medical, Inc.
Product Code
PSZ · Microbiology
Decision Date
May 10, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3328
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Clearview Exact II Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasal swab specimens collected from symptomatic patients. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. It is recommended that negative test results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions.
Device Story
Immunochromatographic membrane assay; detects influenza A and B nucleoprotein antigens in nasal swab specimens. Input: nasal swab sample added to extraction reagent in coated reaction tube. Operation: test strip placed in extracted sample; monoclonal antibodies immobilized on membrane capture antigens. Output: visual interpretation of pink-to-purple sample lines at 10 minutes; blue control line indicates valid test. Used in clinical settings; performed by healthcare professionals. Results aid rapid differential diagnosis of influenza A and B infections. Benefits: rapid identification of viral infection to support clinical management.
Clinical Evidence
Prospective multi-center study (7 U.S. sites, 2008-2009 season, N=486). Compared to viral culture. Influenza A: 94% sensitivity (95% CI: 83-98%), 94% specificity (95% CI: 91-96%). Influenza B: 78% sensitivity (95% CI: 68-86%), 97% specificity (95% CI: 95-98%). Reproducibility study conducted across 3 sites using blind-coded panels; high detection rates for positive samples and 100% negative agreement.
Technological Characteristics
Lateral flow immunochromatographic assay. Components: monoclonal antibodies and control protein immobilized on membrane support (test strip), extraction reagent, coated reaction tube. Visual readout (colorimetric). Standalone, non-electronic, point-of-care diagnostic. No software or energy source required.
Indications for Use
Indicated for qualitative detection of influenza A and B nucleoprotein antigens in nasal swab specimens from symptomatic patients of all ages. Not for use as sole basis for treatment or management decisions; negative results require confirmation by cell culture.
Regulatory Classification
Identification
An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
{0}------------------------------------------------
MAY 1 0 2010
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _ K092349
To establish substantial equivalence to the predicate, the Clearview® Exact II Influenza A & B Test was compared to the BinaxNOW® Influenza A & B Test (510(k) # K062109).
### SUBMITTER
Binax, Inc., d/b/a Invemess Medical 10 Southgate Road Scarborough, Maine 04074 (207) 730-5739 (Office) (207) 730-5717 (FAX) Establishment Registration Number: 1221359
CONTACT PERSON Anne Jepson anne.jepson@invmed.com (email)
ALTERNATE CONTACT PERSON Angela Drysdale angela.drysdale@invmed.com (email)
DATE PREPARED
May 7, 2010
#### TRADE NAME
Clearview® Exact II Influenza A & B Test
COMMON NAME Not applicable
CLASSIFICATION NAME Antigen, Cf (including Cf Control), Influenza Virus A, B, C (per 21 CFR 866.3330)
PREDICATE DEVICE BinaxNOW® Influenza A & B Test, K062109
#### DEVICE DESCRIPTION
The Clearview® Exact II Influenza A & B Test is an immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect influenza type A and B nucleoprotein antigens in respiratory swab specimens. These antibodies and a control protein are immobilized onto a membrane support as three distinct lines and are combined with other reagents/pads to construct a Test Strip.
{1}------------------------------------------------
Nasal swab samples are added to a Coated Reaction Tube to which an extraction reagent has been added. A Clearview Exact II Influenza A & B Test Strip is then placed in the Coated Reaction Tube holding the extracted liquid sample. Test results are interpreted at 10 minutes based on the presence of pink-to-purple colored Sample Lines. The yellow Control Line tums blue in a valid test.
# INTENDED USE
The Clearview Exact II Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasal swab specimens collected from symptomatic patients. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. It is recommended that negative test results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions.
# TECHNOLOGICAL CHARACTERISTICS
The Clearview® Exact II Influenza A & B Test and the predicate, the BinaxNOW® Influenza A & B Test, both use lateral flow immunochromatographic technology. Both tests are rapid immunoassays that employ specific antibodies immobilized onto solid phases to capture and visualize influenza nucleoprotein antigens.
### PERFORMANCE SUMMARY
# CLINICAL STUDY
# Clearview Exact II Influenza A & B Test Performance vs. Viral Culture -Prospective Study
The clinical performance of the Clearview® Exact II Influenza A & B Test was established in a multi-center, prospective, clinical study conducted at seven U.S. trial sites during the 2008-2009 respiratory season.
A total of 486 prospective specimens, collected from children (less than 18 years of age) and adults (18 years or older), were evaluated in the Clearview® Exact II Influenza A & B Test and compared to viral culture. Evaluated specimens were nasal swabs collected from patients presenting with influenza-like symptoms. Forty-four percent (44%) of the population tested was < 5 years of age, 31% was 5 - < 18 years of age, and 25% was > 18 years. A/H3 and A/H1 were the predominant influenza A subtypes observed during this time.
Clearview® Exact II Influenza A & B Test performance versus viral culture, including 95% confidence intervals, is detailed below.
{2}------------------------------------------------
# Clearview Exact II Influenza A & B Test Performance vs. Culture
| Influenza Type A | | | |
|------------------|-----------|-----------|-----|
| | Culture + | Culture - | |
| Clearview + | 45 | 26 | 71 |
| Clearview - | 3 | 412 | 415 |
| | 48 | 438 | 486 |
| Influenza Type B | | | |
|------------------|-----------|-----------|-----|
| | Culture + | Culture - | |
| Clearview + | 68 | 13 | 81 |
| Clearview - | 19* | 386 | 405 |
| | 87 | 399 | 486 |
Sensitivity: 45/48 = 94% (95% CI: 83-98%) Specificity: 412/438 = 94% (95% Cl: 91-96%) PPV: 45/71 = 63% (95% Cl: 52-74%) NPV: 412/415 = 99% (95% CI: 98-100%)
Sensitivity: 68/87 = 78% (95% CI: 68-86%) Specificity: 386/399 = 97% (95% CI: 95-98%) PPV: 68/81 = 84% (95% Cl = 74-90%) NPV: 386/405 = 95% (95% CI = 93-97%)
* The nineteen samples that tested positive on culture for influenza B, but were negative on the Clearview Exact II Test, were also tested on an investigational RT-PCR assay. Ten (10) of these samples were negative for influenza B by PCR.
ANALYTICAL STUDIES
# ANALYTICAL SENSITIVITY
The Clearview® Exact II Test limit of detection (LOD or Cos), defined as the concentration of influenza virus that produces positive Clearview® Exact II Test results approximately 95% of the time, was identified by evaluating different concentrations of 2 subtypes of live influenza A and 2 strains of live influenza B in the Clearview® Exact II Test. Multiple operators tested each concentration of the four influenza strains multiple times. The concentrations identified as the LOD (Cgs) levels for each strain tested are listed below.
| Influenza Strain | Concentration<br>(TCID50/ml) | # Detected per<br>Total Tests | % Detected |
|--------------------------------|------------------------------|-------------------------------|------------|
| Influenza A/HongKong/8/68 | $2.37 x 10^4$ | 64/66 | 97% |
| Influenza A/PuertoRico/8/34 | $3.16 x 10^5$ | 37/42 | 88% |
| Influenza B/Malaysia/2506/2004 | $3.00 x 10^6$ | 19/20 | 95% |
| Influenza B/Lee/40 | $4.20 x 10^5$ | 19/20 | 95% |
### ANALYTICAL REACTIVITY TESTING
The influenza A and B strains listed tested positive in the Clearview® Exact II Influenza A & B Test at the concentrations specified. Although the specific influenza strains causing infection in humans can vary year to year, all contain the conserved nucleoproteins targeted by the Clearview® Exact II test. 1 Performance characteristics of the Clearview® Exact II Influenza A & B Test for detecting influenza A virus from human specimens was established when H1 and H3 subtypes were prevalent. Performance characteristics of the test when other influenza A virus subtypes are emerqing as human pathogens have not been established.
### Influenza Strain
Flu A/Port Chalmers/1/73 (H3N2) Flu A/WS/33 (H1N1) Flu A/Aichi/2/68 (H3N2) Flu A/Malaya/302/54 (H1N1) Flu A/New Jersey/8/76 (H1N1) Flu A/Denver/1/57 (H1N1)
#### Concentration
- 5.6 x 105 TCID50/ml 5.0 x 104 TCID50/ml 3.0 x 104 TCID50/ml 6.0 x 103 TCID50/ml 2.8 x 105 TCID50/ml 8.9 x 103 TCID50/ml
{3}------------------------------------------------
1 8 x 104 TCID50/ml Flu A/Victoria/3/75 (H3N2) 1.5 x 105 TCID50/ml Flu A/Solomon İslands/3/2006 (H1N1) 2.5 x 106 EiU50/ml Flu A/Brisbane/10/07 (H3N2) 5.6 x 105 TCID50/ml Flu A/PuertoRico/8/34 (H1N1) 1.3 x 105 TCID50/ml Flu A/Wisconsin/67/2005 (H3N2) Flu A/Hong Kong/8/68 (H3N2) 7.9 x 103 TCID50/ml 1.4 x 105 TCID50/ml Flu A/California/04/2009 (H1N1) 1.4 x 104 TCID50/ml Flu B/Florida/02/2006 7.1 x 104 TCID50/ml Flu B/Florida/04/2006 8.5 x 104 TCID50/ml Flu B/Florida/07/04 1.5 x 106 TCID50/ml Flu B/Malavsia/2506/04 1.7 x 104 TCID50/ml Flu B/Panama/45/90 5.0 x 105 TCID50/ml Flu B/R75 2.2 x 10° TCID50/ml Flu B/Russia/69 1.0 x 105 TCIDso/ml
1.5 x 105 TCIDso/ml Flu B/Taiwan/2/62 Flu B/Mass/3/66 1.8 x 10° TCID50/ml Flu B/Lee/40
Although this test has been shown to detect the Flu A/California/04/2009 (H1N1) virus cultured from a positive human specimen, the performance characteristics of this device with human specimens infected with the 2009 H1N1 influenza virus have not been established. The Clearview Exact II Influenza A & B test can distinguish between influenza A and B viruses, but it does not differentiate seasonal influenza A virus from the novel influenza A (2009 H1N1) virus, and the test's ability to detect human infection with the novel influenza A 2009 H1N1 virus in clinical specimens is unknown.
### ANALYTICAL SPECIFICITY (CROSS-REACTIVITY)
To determine the analytical specificity of the Clearview® Exact II Influenza A & B Test, 54 commensal and pathogenic microorganisms (38 bacteria, 15 viruses and 1 yeast) that may be present in the nasal cavity or nasopharynx were tested. All of the following microorganisms were negative when tested at concentrations ranging from 10° to 10° cells/ml, CFU/ml or IFU/ml (bacteria), 105 to 10° TCIDso/ml or CEIDso/ml (viruses), and 109 cells/ml (veast).
#### Bacteria
Acinetobacter calcoaceticus Bacteroides fragilis Bordetella pertussis Chlamydia pneumoniae Corynebacterium diphtheria Enterococcus faecalis Escherichia coli Gardnerella vaginalis Haemophilus influenzae Klebsiella pneumoniae Lactobacillus casei Lactobacillus plantarum Legionella pneumophila Listeria monocytogenes Moraxella catarrhalis Mycobacterium avium
Yeast Viruses Adenovirus type 1 Adenovirus type 7 Coronavirus OC43 Coronavirus 229E Coxsackievirus B4 Cytomegalovirus (CMV) (Herpes V) Epstein Barr Virus Human metapneumovirus Measles (Edmonston) Mumps (Enders) Parainfluenza 1 Parainfluenza 2 Parainfluenza 3 Respiratory Syncytial Virus type B Rhinovirus type 1A
Candida albicans
{4}------------------------------------------------
Mycobacterium intracellulare Mycobacterium tuberculosis Mycoplasma pneumoniae Neisseria gonorrhoeae Neissena meningitidis Neisseria sicca Neisseria subflava Proteus vulgaris Pseudomonas aeruginosa Serratia marcescens Staphylococcus aureus Staphylococcus aureus (Cowan protein A producing strain) Staphylococcus epidermidis Streptococcus, Group A Streptococcus, Group B Streptococcus, Group C Streptococcus, Group F Streptococcus, Group G Streptococcus mutans Streptococcus pneumoniae Streptococcus salivaris Streptococcus sanguis
#### INTERFERING SUBSTANCES
The following substances, naturally present in respiratory specimens or that may be artificially introduced into the nasal cavity or nasopharynx, were evaluated in the Clearview Exact II Influenza A & B Test at the concentrations listed below and were found not to affect test performance. Whole blood (1%) did not interfere with the interpretation of negative Clearview Exact II Test results, but did interfere with the interpretation of influenza A LOD (Cgs) positive samples. Therefore, visibly bloody samples may not be appropriate for use in this test.
| Substance | Concentration |
|-------------------------|---------------|
| 3 OTC nasal sprays | 10% |
| 3 OTC mouthwashes | 10% |
| 3 OTC throat drops | 10% |
| 4-acetamidophenol | 10 mg/ml |
| Acetylsalicylic acid | 20 mg/ml |
| Albuterol | 20 mg/ml |
| Chlorpheniramine | 5 mg/ml |
| Dexamethasone | 5 mg/ml |
| Dextromethorphan | 10 mg/ml |
| Diphenhydramine | 5 mg/ml |
| Doxylamine succinate | 1 mg/ml |
| Flunisolide | 3 mg/ml |
| Guaiacol glycerol ether | 20 mg/ml |
| Mucin | 1% |
| Mupirocin | 250 µg/ml |
| Oxymetazoline | 10 mg/ml |
| Phenylephrine | 10 mg/ml |
| Phenylpropanolamine | 20 mg/ml |
{5}------------------------------------------------
Rebetol® (Ribavirin)
Relenza® (Zanamivir) Rimantadine Tamiflu® (Oseltamivir) Tobramvcin Triamcinolone
500 ng/ml 20 mg/ml 500 ng/ml 100 mg/ml 40 mg/ml 14 mg/ml
#### REPRODUCIBILITY
A reproducibility study of the Clearview® Exact II Influenza A & B Test was conducted by operators from 3 sites using panels of blind coded specimens containing negative, high negative (below the limit of detection), low positive (at the limit of detection), and moderate positive (above the limit of detection) influenza A and B viral samples. Participants tested each sample multiple times on 5 different days. The detection rates for the influenza A moderate positive, low positive, and high negative samples were 99.2% (119/120), 94.2 % (113/120) and 9.2% (11/120), respectfully. The detection rates for the influenza B moderate positive, low positive, and high negative samples were 99.2% (119/120); 96.7 % (116/120); and 7:5 % (9/120); respectfully. All of the negative samples (118) generated negative test results.
Signed Angela Drysdale Date 5/7/20
Director, Clinical Affairs Binax, Inc., d/b/a Inverness Medical
1 Dowdle, W.R. Kendal, A.P., and Noble, G.R. 1980. Influenza Virus, p 836-884. Ma Clinical Microbiology, 3rd edition, In Lennette, et. Al (ed.). American Society for Microbiolo Washington, D.C.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
Ms. Anne Jepson Manager, Clinical Affairs Binax, Inc. Inverness Medical Inc. 10 Southgate Road Scarborough, ME 04074
Y 1.0 .20
Re: k092349
Trade/Device Name: Binax Clearview® Exact II Influenza A &B Test. Regulation Number: 21 CFR 866.3300 Regulation Name: Influenza Virus Serological Reagents Regulatory Class: Class I Product Code: GNX Dated: April 30, 2010 Received: May 3, 2010
Dear Ms. Jepson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
### Page 2 - Anne Jepson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely vours.
Saly ator
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{8}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
K092349 510(k) Number (if known):
Clearview® Exact II Influenza A & B Test Device Name:
# Indications For Use:
The Clearview Exact II Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasal swab specimens collected from symptomatic patients. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. It is recommended that negative test results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions.
> Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - (CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
| 510(k) | k092349 |
|--------|---------|
|--------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.