The Randox Laboratories Ltd. Methamphetamine Assay is an in vitro diagnostic test for the qualitative and semi-quantitative detection of Methamphetamine in numan urine at the cut off of 1000ng/ml. The assay is calibrated against methamphetamine. Qualitative and semi-quantitative results can be utilized in the diagnosis and treatment of Methamine use or overdose. The Randox Methamphetamine Assay has been developed for use on the form analysers, which includes the And the annound " and the annound" This in vitro diagnostic device is intended for prescription use only. The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GCMS or (2) permitting laboratories to establish quality control procedures. This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatograph/Mass Spectrometry(GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
Device Story
In vitro diagnostic assay for methamphetamine detection in human urine; utilizes competitive enzyme immunoassay (EIA) principle. Reagents contain mouse monoclonal antibodies and G6PDH-labeled drug conjugate; enzyme activity inhibited by drug-antibody binding; free drug in sample displaces conjugate, increasing enzyme activity. Activity measured spectrophotometrically at 340 nm via NAD to NADH conversion. Used on RX Daytona and RX Imola analyzers in clinical laboratories. Provides preliminary qualitative/semi-quantitative results; aids in determining specimen dilution for GC/MS confirmation. Benefits include rapid screening for drug use or overdose; supports clinical decision-making regarding treatment.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
In vitro diagnostic immunoassay; liquid-based reagents, calibrators, and controls. Designed for use on automated clinical chemistry analyzers. Qualitative and semi-quantitative detection of methamphetamine in urine via immunoassay principle. Prescription use only.
Indications for Use
Indicated for the qualitative and semi-quantitative detection of methamphetamine in human urine at a 1000 ng/mL cutoff. Intended for prescription use in clinical settings to aid in the diagnosis and treatment of methamphetamine use or overdose. Provides preliminary results requiring confirmation by GC/MS.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k092266
B. Purpose for Submission:
New device
C. Measurand:
Methamphetamine
D. Type of Test:
Qualitative and semi-quantitative enzyme immunoassay
E. Applicant:
Randox Laboratories
F. Proprietary and Established Names:
Randox Methamphetamine Assay
Randox Multi Drug Calibrator Set
Randox Multi Drug Control Level I and II
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LAF | II | 862.3619- Methamphetamine test system | 91-Toxicology |
| DKB | II | 862.3200– Clinical toxicology calibrator | 91- Toxicology |
| DIF | I, reserved | 862.3280–Clinical toxicology control material | 91- Toxicology |
H. Intended Use:
1. Intended use(s):
See indications for use below
{1}
2. Indication(s) for use:
Randox Methamphetamine Assay
The Randox Laboratories Ltd. Methamphetamine Assay is an in vitro diagnostic test for the qualitative and semi-quantitative analysis of Methamphetamine in human urine at the cutoff of 1000 ng/mL. The assay is calibrated against methamphetamine. Qualitative and semi-quantitative results can be utilized in the diagnosis and treatment of Methamphetamine use or overdose. The Randox Methamphetamine Assay has been developed for use on the RXseries analyzers, which includes the Rx Daytona and the Rx Imola analyzers. This in vitro diagnostic device is intended for prescription use only.
The semi-quantitative mode is for purposes of
(1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GCMS or
(2) permitting laboratories to establish quality control procedures.
This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatograph/Mass Spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
Randox Multidrug Calibrator Set
The Randox Multidrug Calibrator Set consists of liquid calibrators containing Methamphetamine, Oxazepam and Methadone. There are 5 levels of calibrator. They have been developed for use in the calibration of Methamphetamine Benzodiazepines and Methadone assays for use on the RXseries analyzers, which includes the RX Daytona and the RX Imola analyzers. This in vitro diagnostic device is intended for prescription use only.
Randox Multidrug Controls, Level 1 & 2
The Randox Multidrug Controls level 1 and 2 are liquid controls containing Methamphetamine, Oxazepam and Methadone. There are 2 levels of controls. They have been developed for use in the quality control of Methamphetamines, Benzodiazepines and Methadone assays for use on the RXseries analyzers, which includes the RX Daytona and the RX Imola analyzers. This in vitro diagnostic device is intended for prescription use only.
3. Special conditions for use statement(s):
Prescription use only
{2}
4. Special instrument requirements:
The studies were performed on the X Daytona and the X Imola analyzers.
I. Device Description:
The assay consists of ready-to-use liquid reagents. Reagent 1 contains mouse monoclonal methamphetamine antibodies, glucose-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers and sodium azide <0.1% w/v. Reagent 2 contains amphetamine-labeled glucose-6-phosphate dehydrogenase (G6PDH) in buffer and sodium azide <0.1% w/v. The calibrators and controls are ready to use human urine-based liquid. The calibrators and controls are sold separately and previously cleared for benzodiazepine and methadone.
J. Substantial Equivalence Information:
1. Predicate device name(s):
DRI Multi-Drug Calibrators and Controls, Microgenics Corporation
2. Predicate 510(k) number(s):
k040758
3. Comparison with predicate:
| Similarities/Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Qualitative and semi-quantitative analysis of methamphetamine in human urine | Same |
| Test Principle | A competitive enzyme immunoassay based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. In the absence of drug in the sample the drug-labeled G6PDH conjugate is bound to antibody and enzyme activity is inhibited. When free drug is present in the sample, antibody binds to the free drug and the unbound drug-labeled G6PDH exhibits its maximum enzyme activity. Active enzyme converts NAD to NADH resulting in an absorbance change measured spectrophotometrically at 340nm. | Same |
| Cutoff | 1000 ng/mL | 500 ng/mL and 1000 ng/mL |
{3}
| Similarities/Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Matrix | Human Urine | Human Urine |
| Type of reagent | Liquid Ready to use, two reagent assay | Same |
| Calibrators | Liquid Ready to use (0, 300, 500, 1000, 2000 ng/mL) | Liquid ready to use (500, 1000, 1500, 2000 ng/mL) |
| Controls | Liquid ready to use (+/- 25% from cutoffs) | Same |
K. Standard/Guidance Document Referenced (if applicable):
None were referenced
L. Test Principle:
The Randox Methamphetamine assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. In the absence of drug in the sample, the drug-labeled G6PDH conjugate is bound to antibody, thus the enzyme activity is inhibited. When free drug is present in the sample, the antibody will bind to the free drug and the unbound amphetamine-labeled G6PDH exhibits its maximal enzyme activity. The G6PDH activity is measured spectrophotometrically at 340 nm because of conversion of NAD to NADH.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies were performed by spiking a d-methamphetamine into negative urine at various concentrations (-75%, -50%, -25%, +25% and 50%, 75% and 200% of the cutoff and negative). Concentrations were confirmed by GC-MS. Studies were performed on 2 RX Daytona systems and 2 RX Imola systems. Testing was performed twice a day for 20 non-consecutive days for all concentrations.
During the study, a calibration was performed twice on each of 4 analyzers and there were two operators. The results are presented in the table below:
{4}
| Qualitative Methamphetamine | | RX Daytona | RX Imola |
| --- | --- | --- | --- |
| Sample concentration (ng/mL) | No. Observations | | |
| | | # Neg/#Pos | # Neg/#Pos |
| 0 (negative) | 80 | 80/0 | 80/0 |
| 250 (-75% c/o) | 80 | 80/0 | 80/0 |
| 500 (-50% c/o) | 80 | 80/0 | 80/0 |
| 750 (-25% c/o) | 80 | 80/0 | 80/0 |
| 1250 (+25% c/o) | 80 | 0/80 | 0/80 |
| 1500 (+50% c/o) | 80 | 0/80 | 0/80 |
| 1750 (+75% c/o) | 80 | 0/80 | 0/80 |
| 2000 (+100% c/o) | 80 | 0/80 | 0/80 |
| Semi-quantitative Methamphetamine | | RX Daytona | RX Imola |
| --- | --- | --- | --- |
| Sample concentration (ng/mL) | No. Observations | | |
| | | # Neg/#Pos | # Neg/#Pos |
| 0 (negative) | 80 | 80/0 | 80/0 |
| 250 (-75% c/o) | 80 | 80/0 | 80/0 |
| 500 (-50% c/o) | 80 | 80/0 | 80/0 |
| 750 (-25% c/o) | 80 | 80/0 | 80/0 |
| 1250 (+25% c/o) | 80 | 0/80 | 0/80 |
| 1500 (+50% c/o) | 80 | 0/80 | 0/80 |
| 1750 (+75% c/o) | 80 | 0/80 | 0/80 |
| 2000 (+100% c/o) | 80 | 0/80 | 0/80 |
b. Linearity/assay reportable range:
Linearity across the range was confirmed by spiking a drug free urine pool with drug and serially diluting the sample into 20 additional concentrations. Each sample was assayed on both the Daytona and Imola analyzers in the semi-quantitative mode. The dilutions were assayed in triplicate then averaged and compared to the expected result and the percent recovery was calculated. Results are presented in the table below:
{5}
| Methamphetamine | Daytona Analyzer | | Imola Analyzer | |
| --- | --- | --- | --- | --- |
| Expected | Observed | % Recovery | Observed | % Recovery |
| 0 | 6 | n/a | 0 | n/a |
| 20 | 14.1 | 70.32 | 7.76 | 38.78 |
| 40 | 12.4 | 30.91 | 18.49 | 46.23 |
| 60 | 0 | 0.00 | 86.59 | 144.32 |
| 80 | 26.12 | 32.65 | 86.59 | 108.24 |
| 100 | 84.2 | 84.17 | 133.07 | 133.07 |
| 120 | 86.9 | 72.45 | 176.23 | 146.86 |
| 140 | 55.8 | 39.86 | 107.47 | 76.77 |
| 160 | 114.97 | 71.86 | 191.03 | 119.40 |
| 180 | 170.69 | 94.83 | 201.79 | 112.11 |
| 200 | 178.37 | 89.19 | 192.68 | 96.34 |
| 400 | 417.50 | 104.37 | 430.76 | 107.69 |
| 600 | 581.39 | 96.90 | 638.26 | 106.38 |
| 800 | 758.99 | 94.87 | 906.14 | 113.27 |
| 1000 | 936.01 | 93.60 | 1002.05 | 100.21 |
| 1200 | 1024.22 | 85.35 | 1166.54 | 97.21 |
| 1400 | 1256.26 | 89.73 | 1506.47 | 107.61 |
| 1600 | 1506.05 | 94.13 | 1499.85 | 93.74 |
| 1800 | 1834.94 | 101.94 | 1836.57 | 102.03 |
| 2000 | 2247.40 | 112.37 | 2159.67 | 107.98 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Calibrators and controls for are traceable to Cerilliant Reference Standards and confirmed by GC/MS.
An intermediate concentration of calibrators (designated as master calibrator) was made by adding Methamphetamine Standard to buffered human urine matrix. This master calibrator was value-assigned by GC/MS. The master calibrators are then gravimetrically diluted using synthetic urine to various concentration levels. The concentration values of diluted calibrators/controls are assigned using GC/MS. These calibrators/controls become internal reference standards. These internal reference standards are aliquoted and stored at below $-80^{\circ}\mathrm{C}$ environments and are used only for testing working calibrators (i.e. calibrators used to produce products). The production protocol for internal reference standards and working calibrators are identical except that the working calibrators are tested against the internal reference standards.
Stability: Shelf-life and open-vial stability studies were performed for the calibrators and controls using both accelerated and on-going real-time studies. The protocols were found to be adequate. The Randox Methamphetamine assay reagents, controls and calibrators are stable for 18 months when stored unopened at $+2$ to $8^{\circ}\mathrm{C}$ and for 28 days once opened and stored at $+2$ to $8^{\circ}\mathrm{C}$.
{6}
Calibrators and controls are sold separately. In the labeling the sponsor recommends that users follow federal, state and local guidelines for testing external quality control materials.
d. Detection limit:
Performance at low drug concentrations in the semi-quantitative assay was characterized by the determination of recovery (see section b above).
e. Analytical specificity:
Cross-reactivity was established by spiking various concentrations of structurally related into drug-free urine. Results are expressed as a minimum concentration of metabolite or compound required to produce a response approximately equivalent to the cutoff concentration of the assay. The percent cross-reactivity of those compounds are presented below:
Structurally related:
| Daytona | Qualitative | | Semi-quantitative | |
| --- | --- | --- | --- | --- |
| Compound | Equivalent to 1000 ng/mL | % Cross reactivity | Equivalent to 1000 ng/mL | % Cross reactivity |
| l-amphetamine | 58,650 | 1.17 | 53,059 | 1.88 |
| l-methamphetamine | 10,272 | 9.74 | 15,799 | 6.33 |
| d,l-methamphetamine | 2,170 | 46.08 | 1,987 | 50.34 |
| (+/-)-amphetamine | 1,512 | 66.13 | 1,550 | 64.53 |
| d-methamphetamine | 1,000 | 100 | 1,000 | 100 |
| d-amphetamine | 909 | 110.05 | 843 | 118.6 |
| MDA | 2,252 | 44.41 | 1,950 | 51.3 |
| MDMA | 4,131 | 24.21 | 2,519 | 39.7 |
| MDEA | 54,161 | 1.85 | 44,479 | 2.25 |
| MBDB | 4,356 | 22.96 | 4,056 | 24.66 |
| BDB | 7,631 | 13.1 | 7,520 | 13.3 |
| B-phenylethylamine | 56,133 | 1.78 | 54,329 | 1.84 |
| Tyramine | 100,000 | 0.27 | 100,000 | 0.27 |
| Phentermine | 36,740 | 2.72 | 31,822 | 3.14 |
| Phenylpropanolamine | 100,000 | 0.29 | 100,000 | 0.29 |
| Chlorpheniramine | 100,000 | 0 | 100,000 | 0 |
{7}
8
| Imola | Qualitative | | Semi-quantitative | |
| --- | --- | --- | --- | --- |
| Compound | Equivalent to 1000 ng/mL | % Cross reactivity | Equivalent to 1000 ng/mL | % Cross reactivity |
| l-amphetamine | 34,273 | 2.92 | 51,967 | 1.92 |
| l-methamphetamine | 13,942 | 7.17 | 14,676 | 6.81 |
| d,l-methamphetamine | 2,119 | 47.20 | 2,844 | 35.16 |
| (+/-)-amphetamine | 1,169 | 85.53 | 1,495 | 66.91 |
| d-methamphetamine | 1,000 | 100 | 1,000 | 100 |
| d-amphetamine | 698 | 143.26 | 905 | 110.47 |
| MDA | 1,412 | 70.82 | 1,944 | 51.45 |
| MDMA | 2,189 | 45.88 | 3,547 | 39.27 |
| MDEA | 28.283 | 3.54 | 53,076 | 1.88 |
| MBDB | 3,665 | 27.29 | 3,481 | 28.73 |
| BDB | 7,035 | 14.22 | 8,178 | 12.23 |
| B-phenylethylamine | 24,085 | 2.93 | 60,528 | 1.65 |
| Tyramine | 100,000 | 0.27 | 100,000 | 0.27 |
| Phentermine | 18,306 | 5.46 | 24,516 | 4.08 |
| Phenylpropanolamine | 100,000 | 0.29 | 100,000 | 0.29 |
| Chlorpheniramine | 100,000 | 0 | 100,000 | 0 |
Structurally un-related
This study was performed by spiking structurally unrelated compounds into urine samples containing d-methamphetamine. The concentration of drug in the urine sample were +25% and -25% of the cut-off concentration of 1000 ng/mL. The results are presented below:
| | | Daytona | | | |
| --- | --- | --- | --- | --- | --- |
| Compound | Conc. ng/mL | -25% of 1000 ng/mL d-methamphetamine cutoff | | +25% of 1000 ng/mL d-methamphetamine cutoff | |
| | | Qualitative | Semi-quant. | Qualitative | Semi-quant. |
| 11-nor9-carboxy-delta9-THC | 100,000 | NEG | NEG | POS | POS |
| 11-hydroxy-delta9-THC | 100,000 | NEG | NEG | POS | POS |
| 6 Acetyl morphine | 100,000 | NEG | NEG | POS | POS |
| Amitriptyline | 100,000 | NEG | NEG | POS | POS |
| Amobarbital | 100,000 | NEG | NEG | POS | POS |
| Ascorbic acid | 100,000 | NEG | NEG | POS | POS |
| Aspirin | 100,000 | NEG | NEG | POS | POS |
| Benzoylecgonine | 100,000 | NEG | NEG | POS | POS |
| Caffeine | 100,000 | NEG | NEG | POS | POS |
| Cannabidiol | 100,000 | NEG | NEG | POS | POS |
| Cocaethylene | 100,000 | NEG | NEG | POS | POS |
{8}
| | | Daytona | | | |
| --- | --- | --- | --- | --- | --- |
| Compound | Conc. ng/mL | -25% of 1000 ng/mL d-methamphetamine cutoff | | +25% of 1000 ng/mL d-methamphetamine cutoff | |
| Cocaine | 70,000 | NEG | NEG | POS | POS |
| Codeine | 70,000 | NEG | NEG | POS | POS |
| Cotinine | 100,000 | NEG | NEG | POS | POS |
| delta9-THC | 70,000 | NEG | NEG | POS | POS |
| d-Ephedrine | 100,000 | NEG | NEG | POS | POS |
| Diazepam | 100,000 | NEG | NEG | POS | POS |
| Dihydrocodeine | 100,000 | NEG | NEG | POS | POS |
| d,l-ephedrine | 100,000 | NEG | NEG | POS | POS |
| Ecgonine methyl ester | 100,000 | NEG | NEG | POS | POS |
| EDDP | 100,000 | NEG | NEG | POS | POS |
| EMDP | 100,000 | NEG | NEG | POS | POS |
| Heroin | 40,000 | NEG | NEG | POS | POS |
| Ibuprofen | 100,000 | NEG | NEG | POS | POS |
| l-ephedrine | 10,000 | NEG | NEG | POS | POS |
| LAAM | 100,000 | NEG | NEG | POS | POS |
| Methadone | 100,000 | NEG | NEG | POS | POS |
| Morphine | 100,000 | NEG | NEG | POS | POS |
| Oxycodone | 100,000 | NEG | NEG | POS | POS |
| Paracetamol | 100,000 | NEG | NEG | POS | POS |
| Phendimetrazine | 25,000 | NEG | NEG | POS | POS |
| R,R (-) Pseudoephedrine | 100,000 | NEG | NEG | POS | POS |
| S,S (+) Pseudoephedrine | 8,000 | NEG | NEG | POS | POS |
| Temazepam | 100,000 | NEG | NEG | POS | POS |
| | | Imola | | | |
| --- | --- | --- | --- | --- | --- |
| Compound | Conc. ng/mL | -25% of 1000 ng/mL d-methamphetamine cutoff | | +25% of 1000 ng/mL d-methamphetamine cutoff | |
| | | Qualitative | Semi-quant. | Qualitative | Semi-quant. |
| 11-nor9-carboxy-delta9-THC | 100,000 | NEG | NEG | POS | POS |
| 11-hydroxy-delta9-THC | 100,000 | NEG | NEG | POS | POS |
| 6 Acetyl morphine | 100,000 | NEG | NEG | POS | POS |
| Amitriptyline | 100,000 | NEG | NEG | POS | POS |
| Amobarbital | 100,000 | NEG | NEG | POS | POS |
| Ascorbic acid | 100,000 | NEG | NEG | POS | POS |
| Aspirin | 100,000 | NEG | NEG | POS | POS |
| Benzoylecgonine | 100,000 | NEG | NEG | POS | POS |
{9}
| | | Imola | | | |
| --- | --- | --- | --- | --- | --- |
| Compound | Conc. ng/mL | -25% of 1000 ng/mL d-methamphetamine cutoff | | +25% of 1000 ng/mL d-methamphetamine cutoff | |
| Caffeine | 100,000 | NEG | NEG | POS | POS |
| Cannabidiol | 100,000 | NEG | NEG | POS | POS |
| Cocaethylene | 100,000 | NEG | NEG | POS | POS |
| Cocaine | 70,000 | NEG | NEG | POS | POS |
| Codeine | 70,000 | NEG | NEG | POS | POS |
| Cotinine | 100,000 | NEG | NEG | POS | POS |
| delta9-THC | 70,000 | NEG | NEG | POS | POS |
| d-Ephedrine | 100,000 | NEG | NEG | POS | POS |
| Diazepam | 100,000 | NEG | NEG | POS | POS |
| Dihydrocodeine | 100,000 | NEG | NEG | POS | POS |
| d,l-ephedrine | 100,000 | NEG | NEG | POS | POS |
| Ecgonine methyl ester | 100,000 | NEG | NEG | POS | POS |
| EDDP | 100,000 | NEG | NEG | POS | POS |
| EMDP | 100,000 | NEG | NEG | POS | POS |
| Heroin | 40,000 | NEG | NEG | POS | POS |
| Ibuprofen | 100,000 | NEG | NEG | POS | POS |
| l-ephedrine | 5,000 | NEG | NEG | POS | POS |
| LAAM | 100,000 | NEG | NEG | POS | POS |
| Methadone | 100,000 | NEG | NEG | POS | POS |
| Morphine | 100,000 | NEG | NEG | POS | POS |
| Oxycodone | 100,000 | NEG | NEG | POS | POS |
| Paracetamol | 100,000 | NEG | NEG | POS | POS |
| Phendimetrazine | 25,000 | NEG | NEG | POS | POS |
| R,R (-) Pseudoephedrine | 100,000 | NEG | NEG | POS | POS |
| S,S (+) Pseudoephedrine | 5,000 | NEG | NEG | POS | POS |
| Temazepam | 100,000 | NEG | NEG | POS | POS |
Endogenous Compounds:
The sponsor prepared urine-free samples spiked with d-methamphetamine at concentration levels (±25% of the cutoff) and evaluated the possible interference the following endogenous compounds; Total Bilirubin, Direct Bilirubin, Creatinine, Urea, Glucose, H.S.A., Ethanol, Acetone, Gamma globulin, Oxalic acid, Riboflavin, sodium Chloride, Boric acid, Sodium azide, Sodium fluoride. No positive or negative interference was seen.
To test for possible positive and/or negative interference from specific gravity, the sponsor prepared samples containing d-methamphetamine at concentration levels (±25% of the cutoff) with specific gravities ranging from 1.000 to
{10}
1.030. No positive or negative interference due to specific gravity was observed.
To test for potential positive or negative interference from pH the sponsor prepared samples containing d-methamphetamine at concentration levels $(\pm 25\%$ of the cutoff) with pH values of 3, 4.5, 5, 6, 7, 8 and 11. No negative or positive interference due to pH was observed.
f. Assay cut-off:
Analytical performance of the device around the claimed cutoff is described in the precision section (1 a.) above.
# 2. Comparison studies:
a. Method comparison with predicate device:
The sponsor conducted a method comparison study to evaluate the performance of the device for detection of methamphetamine. In, the method comparison study, 91 unaltered clinical samples (45 negative and 46 positive) were tested with the Randox Methamphetamine assay and compared against the results obtained with GC/MS. The results of the studies are presented below:
Qualitative Methamphetamine
| | | Negative | Low Negative by GC/MS (less than -50%) | Near Cutoff Negative by GC/MS (Between -50% and cutoff) | Near Cutoff Positive by GC/MS (Between the cutoff and +50%) | High Positive by GC/MS (greater than +50%) | % Agreement |
| --- | --- | --- | --- | --- | --- | --- | --- |
| RX Daytona | Positive | 0 | 0 | 1 | 16 | 30 | 100% |
| | Negative | 25 | 9 | 10 | 0 | 0 | 97.8% |
| RX Imola | Positive | 0 | 0 | 1 | 16 | 30 | 100% |
| | Negative | 25 | 9 | 10 | 0 | 0 | 97.8% |
Semi-quantitative Methamphetamine
| | | Negative | Low Negative by GC/MS (less than -50%) | Near Cutoff Negative by GC/MS (Between -50% and cutoff) | Near Cutoff Positive by GC/MS (Between the cutoff and +50%) | High Positive by GC/MS (greater than +50%) | % Agreement |
| --- | --- | --- | --- | --- | --- | --- | --- |
| RX Daytona | Positive | 0 | 0 | 1 | 16 | 30 | 100% |
| | Negative | 25 | 9 | 10 | 0 | 0 | 97.8% |
| RX Imola | Positive | 0 | 0 | 1 | 15 | 30 | 97.8% |
| | Negative | 25 | 9 | 10 | 1 | 0 | 97.8% |
{11}
12
Discordant
| | RX Daytona | |
| --- | --- | --- |
| Cutoff Value (ng/mL) | Randox Methamphetamine Assay (POS/NEG) | Drug/Metabolite GC/MS value (ng/mL) |
| | Qualitative | |
| 1000 | POS | 904 Methamphetamine |
| | Semi-quantitative | |
| 1000 | POS | 904 Methamphetamine |
Discordant
| | RX Imola | |
| --- | --- | --- |
| Cutoff Value (ng/mL) | Randox Methamphetamine Assay (POS/NEG) | Drug/Metabolite GC/MS value (ng/mL) |
| Qualitative | | |
| 1000 | POS | 904 Methamphetamine |
| Semi-quantitative | | |
| 1000 | POS | 904 Methamphetamine |
| 1000 | NEG | 1138 Methamphetamine |
b. Matrix comparison:
Not applicable. The test is only for urine specimens.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
{12}
Not applicable
5. Expected values/Reference range:
Not applicable
**N. Proposed Labeling:**
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
**O. Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.