The Randox Laboratories Ltd. Methadone Metabolite Assay is an in vitro diagnostic test for the qualitative and semi-quantitative detection of 2-ethylidene-1,5-diphenylpyrrolidine (EDDP) in human urine. The cut off for both the qualitative and semi-quantitative modes of the assay is 300ng/ml for EDDP. Qualitative and semi-quantitative results can be utilized in the diagnosis and treatment of EDDP use or overdose. The Randox Metabolite Assay has been developed for use on the Rx series analysers, which includes the Rx Daytona and the Rx Imola. This in vitro diagnostic device is intended for prescription use only. The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GCMS or (2) permitting laboratories to establish quality control procedures. This assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatograph/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
Device Story
Randox Methadone Metabolite (EDDP) assay; competitive enzyme immunoassay; uses murine monoclonal antibody and enzyme-drug conjugate (G6PDH). Input: human urine sample. Principle: drug in sample competes with EDDP-labeled G6PDH for antibody binding sites; unbound conjugate converts NAD to NADH; absorbance measured spectrophotometrically at 340nm. Output: qualitative or semi-quantitative EDDP concentration. Used on RX Daytona and RX Imola analyzers in clinical laboratory settings. Results provide preliminary analytical data; requires GC/MS confirmation. Assists clinicians in diagnosing/treating EDDP use or overdose.
Clinical Evidence
No clinical data. Performance established via bench testing. Method comparison performed using 40 positive and 40 negative natural urine samples analyzed on RX Daytona and RX Imola versus GC/MS. Precision evaluated over 20 days (n=80) at ±25%, ±50%, ±75%, and ±100% of cutoff. Analytical specificity tested against endogenous, structurally related, and unrelated compounds; no significant interference observed.
Technological Characteristics
Two-reagent system; murine monoclonal antibody; G6PDH enzyme-drug conjugate. Spectrophotometric detection at 340nm. Liquid, ready-to-use reagents. Calibrators (0-1000 ng/mL) and controls (±25% of 300 ng/mL cutoff) derived from human urine. Analyzers: RX Daytona and RX Imola.
Indications for Use
Indicated for the qualitative and semi-quantitative detection of 2-ethylidene-1,5-diphenylpyrrolidine (EDDP) in human urine to assist in the diagnosis and treatment of EDDP use or overdose. Intended for prescription use only.
Regulatory Classification
Identification
A methadone test system is a device intended to measure methadone, an addictive narcotic pain-relieving drug, in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of methadone use or overdose and to determine compliance with regulations in methadone maintenance treatment.
Special Controls
*Classification.* Class II (special controls). A methadone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k092258
B. Purpose for Submission:
New device
C. Measurand:
Methadone metabolite 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP)
D. Type of Test:
Semi-quantitative and qualitative enzyme immunoassay
E. Applicant:
Randox Laboratories, Limited
F. Proprietary and Established Names:
Randox Methadone Metabolite (EDDP)
G. Regulatory Information:
Regulation section:
| Regulation | Regulation Name | Class | Product Code | Panel |
| --- | --- | --- | --- | --- |
| 21 CFR§862.3620 | Methadone test system | II | DJR | (91) Toxicology |
| 21 CFR§862.3200 | Clinical toxicology calibrator | II | DKB | (91) Toxicology |
| 21 CFR§862.3280 | Clinical toxicology control material | I, reserved | DIF | (91) Toxicology |
H. Intended Use:
1. Intended use(s):
{1}
See Indications for use below.
2. **Indication(s) for use:**
The Randox Laboratories Ltd. Methadone Metabolite Assay is an in vitro diagnostic test for the qualitative and semi-quantitative detection of 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine. The cut off for both the qualitative and semi-quantitative modes of the assay is 300 ng/mL for EDDP. Qualitative and semi-quantitative results can be utilized in the diagnosis and treatment of EDDP use or overdose. The Randox Methadone Metabolite Assay has been developed for use on the RX series analyzers, which includes the RX Daytona and RX Imola. This in vitro diagnostic device is intended for prescription use only. The semi-quantitative mode is for purposes of
1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GC/MS, or,
2) permitting laboratories to establish quality control procedures.
This assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatograph/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
**Randox EDDP Calibrator Set**
The Randox EDDP Calibrator Set consists of liquid calibrators containing EDDP. There are 5 levels of calibrator. They have been developed for use in the calibration of EDDP assays on the RX series analyzers, which includes the RX Daytona and the RX Imola. This *in vitro* diagnostic device is intended for prescription use only.
**Randox EDDP Controls, Level 1 and 2**
The Randox EDDP Controls, level 1 and 2 are liquid controls containing EDDP. There are 2 levels of controls. They have been developed for use in the quality control of EDDP assays on the RX series analyzers, which includes the RX Daytona and the RX Imola. This *in vitro* diagnostic device is intended for prescription use only.
3. **Special conditions for use statement(s):**
**Prescription use only**
This assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatograph/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment
{2}
should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
4. Special instrument requirements:
The Rx Daytona and Rx Imola analyzers were used to conduct performance studies below.
I. Device Description:
The Randox Methadone Metabolite (EDDP) is calibrated against EDDP. The assay is a two reagent system consisting of a murine monoclonal antibody/substrate reagent and the enzyme-drug conjugate reagent. The Randox EDDP Calibrator Set consists of 5 calibrators with EDDP concentrations from 0 to 1,000 ng/mL in human urine. They are ready to use. The Randox EDDP Controls, Level 1 and Level 2, are derived from human urine and are ready to use. The control values are set at approximately ±25% of cutoff (300 ng/mL).
J. Substantial Equivalence Information:
1. Predicate device name(s):
DRI® Methadone Metabolite Enzyme Assay
2. Predicate K number(s):
k023617
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate (k023617) |
| Intended Use/Indications for use | For the qualitative and semi-quantitative determination of Methadone Metabolite, (2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine) EDDP in human urine. The values are utilized in the diagnosis and treatment of EDDP use or overdose | Same |
| Test Principle | A competitive enzyme immunoassay based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase | A homogeneous enzyme immunoassay based on competition of an enzyme glucose-6-phosphate dehydrogenase (G6PDH) labeled drug |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate (k023617) |
| | (G6PDH) for a fixed amount of antibody in the reagent. In the absence of drug in the sample the EDDP labeled G6PDH conjugate is bound to antibody and enzyme activity is inhibited. When free drug is present in the sample, antibody binds to the free drug and the unbound EDDP labeled G6PDH exhibits its maximum enzyme activity. Active enzyme converts NAD to NADH resulting in an absorbance change measured spectrophotometrically at 340nm. | and drug from urine sample for a fixed amount of specific antibody binding sites.
Direct relationship between drug concentration in urine and the enzyme activity.
Enzyme activity is determined spectrophotometrically at 340nm by measuring its ability to convert NAD to NADH. |
| Sample type | Human urine | Same |
| Type of reagent | Liquid Ready to use
Two reagent assay | Same |
| Antibody | Monoclonal | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Cutoff | 300ng/mL | 300ng/mL and 1000ng/mL |
| Calibrator materials | Liquid ready to use (0, 200, 300, 600, 1000ng/mL) | Liquid ready to use (0, 150, 300, 1000, 2000ng/mL) |
| Control materials | 225, 375ng/ml | 75, 125ng/ml |
## K. Standard/Guidance Document Referenced (if applicable):
None were referenced.
## L. Test Principle:
The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, methadone metabolite-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. However, when free drug is present in the sample, antibody would bind to free drug; the
{4}
unbound methadone metabolite-labelled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
An EDDP negative urine pool was spiked with EDDP to concentrations of ±25%, ±50%, ±75% and ±100% of the cutoff of 300 ng/mL as confirmed by GC/MS. Two samples were tested on one RX Daytona and one RX Imola systems twice per day for 20 non-consecutive days (n=80) in the qualitative and semi-quantitative modes. The samples were tested randomly and in a masked order. Two calibrations were performed through the course the studies. A summary of the results is below:
| Sample Concentration % of cutoff | Qualitative Results (#Neg/# Pos) | Semi-Quantitative Results (#Neg/# Pos) |
| --- | --- | --- |
| -100% | 80 Neg | 80 Neg |
| -75% | 80 Neg | 80 Neg |
| -50% | 80 Neg | 80 Neg |
| -25% | 80 Neg | 80 Neg |
| +25% | 80 Pos | 80 Pos |
| +50% | 80 Pos | 80 Pos |
| +75% | 80 Pos | 80 Pos |
| +100% | 80 Pos | 80 Pos |
| Sample Concentration % of cutoff | Qualitative Results (#Neg/# Pos) | Semi-Quantitative Results (#Neg/# Pos) |
| --- | --- | --- |
| -100% | 80 Neg | 80 Neg |
| -75% | 80 Neg | 80 Neg |
| -50% | 80 Neg | 80 Neg |
| -25% | 80 Neg | 80 Neg |
| +25% | 80 Pos | 80 Pos |
| +50% | 80 Pos | 80 Pos |
| +75% | 80 Pos | 80 Pos |
| +100% | 80 Pos | 80 Pos |
#### b. Linearity/assay reportable range:
A drug free urine pool was spiked with pure EDDP to 1000 ng/mL and
{5}
serially diluted in increments of 10% with EDDP negative urine. Each level was tested in triplicate in the semi-quantitative mode on the Rx Daytona and Rx Imola. % recoveries of the target concentrations were calculated for each instrument. Linear regression was also calculated by plotting the observed result against the expected result. See the tables below:
| | Rx Daytona | | Rx Imola | |
| --- | --- | --- | --- | --- |
| Target (ng/mL) | Result (ng/mL) | % Recovery | Result (ng/mL) | % Recovery |
| 0 | 0.00 | 0.00 | 0.00 | 0.00 |
| 10 | 12.63 | 126.27 | 0.00 | 0.00 |
| 20 | 49.20 | 246.02 | 72.84 | 364.22 |
| 30 | 57.13 | 197.11 | 106.34 | 354.46 |
| 40 | 64.46 | 161.16 | 89.40 | 223.50 |
| 50 | 66.57 | 133.15 | 104.25 | 208.49 |
| 60 | 81.94 | 136.57 | 99.63 | 166.06 |
| 70 | 78.46 | 112.08 | 99.71 | 142.45 |
| 80 | 82.30 | 102.88 | 57.88 | 72.35 |
| 90 | 93.40 | 103.78 | 115.55 | 128.39 |
| 100 | 92.15 | 92.15 | 93.66 | 93.66 |
| 200 | 197.55 | 98.78 | 194.74 | 97.37 |
| 300 | 297.74 | 99.25 | 306.86 | 102.29 |
| 400 | 383.71 | 95.93 | 405.18 | 101.30 |
| 500 | 533.19 | 106.64 | 514.53 | 102.91 |
| 600 | 612.15 | 102.03 | 589.23 | 98.21 |
| 700 | 722.72 | 103.25 | 686.34 | 98.05 |
| 800 | 772.88 | 96.61 | 750.34 | 93.79 |
| 900 | 802.11 | 89.12 | 834.48 | 92.72 |
| 1000 | 983.05 | 98.31 | 928.02 | 92.80 |
| | Rx Daytona | Rx Imola |
| --- | --- | --- |
| Slope | 0.95 | 0.91 |
| Intercept | 15.91 | 32.26 |
| R | 0.994 | 0.993 |
| Sy x | 24.75 | 25.87 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability and Value Assignment: The Randox Methadone Metabolite (EDDP) assay has 4 calibrator materials and two control materials. The master calibrator and control stock solutions are gravimetrically prepared using EDDP from an outside supplier and buffered human urine. The purity of the EDDP is determined by GC/FID, HPLC and NMR. The EDDP concentration for the master lot is determined by GC/MS. Value assignment
{6}
for control materials and calibrator materials is determined against the current master lot by performing nested testing of the new control materials and the new calibrator materials. The new lots are analyzed in replicates of 10. Results need to meet the sponsor’s pre-determined acceptance criteria %CV and % recovery against the current master lot.
Calibrator materials and control materials are sold separately.
Stability: Accelerated and Real-time stability studies were performed with the control and calibrator materials at 2-8° C, 25° C and 37° C on the Rx Imola and Rx Daytona. EDDP recoveries from the stressed material were compared to fresh controls or calibrators stored at 2-8° C. Based on the real time studies, the sponsor determined that calibrators and controls are stable for 18 months at 2-8° C. Open vials of calibrators and controls were stored at 2-8° C and analyzed daily on the Rx Imola and Rx Daytona until the EDDP recovery exceeded the pre-determined acceptance criteria. Open vials of control materials and calibrator materials are stable for 28 days at 2-8° C.
d. Detection limit:
Performance at low drug concentrations in the semi-quantitative assay was characterized by the determination of recovery (see section b above).
e. Analytical specificity:
The cross-reactivity of the parent drug with endogenous compounds, structurally related, and structurally unrelated compounds was evaluated by adding known amounts of each potential interferent to drug-free urine spiked with EDDP to ± 25% of the cutoff (300 ng/mL). Samples were assayed in quintuplets for the qualitative and semi-quantitative modes on the RX Daytona and Rx Imola for endogenous and structurally unrelated compounds and in triplicate for structurally related compounds. The results were compared to the appropriate sample containing no interferent. Interference was defined as the concentration of each compound that had a response approximately equal to that of the cutoff calibrator (as positive), or the maximum concentration of the compound tested that remained negative. Any compounds demonstrating interference were retested at lower concentrations of the interferent to determine where interference began.
The interfering concentrations of endogenous, structurally unrelated, and structurally related compounds that were equivalent to the cutoff (300 ng/mL) were determined by comparing the absorbance readings of the samples containing the interferent(s) to the absorbance of the calibrator at the cutoff (300 ng/mL). Results are summarized below:
7
{7}
# Structurally Related Compounds Equivalent to the Cutoff--Rx Daytona
| | | Qualitative | | Semi- Quantitative | | |
| --- | --- | --- | --- | --- | --- | --- |
| Compound | Tested Conc. (ng/mL) | Response equivalent to 300 ng/mL cutoff | Cross-reactivity % | Tested Conc. (ng/mL) | Response equivalent to 300 ng/mL cutoff | Cross-reactivity % |
| EDDP | 300 | POS | 100.00% | 300 | POS | 100.00% |
| EMDP | 995840 | POS | 0.03% | 857402 | POS | 0.03% |
| (-) alpha Methadol | 60661 | POS | 0.49% | 40770 | POS | 0.74% |
| Methadone | 108243 | POS | 0.28% | 89714 | POS | 0.33% |
| LAAM | 571494 | POS | 0.05% | 506356 | POS | 0.06% |
| d, l-Alpha methadol | 485119 | POS | 0.06% | 349296 | POS | 0.09% |
# Structurally Related Compounds Equivalent to the Cutoff-- Rx Imola
| | | Qualitative | | Semi- Quantitative | | |
| --- | --- | --- | --- | --- | --- | --- |
| Compound | Tested Conc. (ng/mL) | Response equivalent to 300 ng/mL cutoff | Cross-reactivity % | Tested Conc. (ng/mL) | Response equivalent to 300 ng/mL cutoff | Cross-reactivity % |
| EDDP | 300 | POS | 100.00% | 300 | POS | 100.00% |
| EMDP | 1210432 | POS | 0.02% | 942693 | POS | 0.03% |
| (-) alpha Methadol | 117542 | POS | 0.26% | 81334 | POS | 0.37% |
| Methadone | 103210 | POS | 0.29% | 72649 | POS | 0.41% |
| LAAM | 929952 | POS | 0.03% | 489669 | POS | 0.06% |
| d, l-Alpha methadol | 1232051 | POS | 0.02% | 780475 | POS | 0.04% |
Structurally unrelated compounds were tested by spiking each of them into urine samples that contain EDDP at $\pm 25\%$ of cutoff for the Rx Imola and Rx Daytona. The following is the list of compounds tested; 11-hydroxy-delta9-THC, 11-nor9-carboxy-delta9-THC, 6 Acetyl-morphine, Amitriptyline Amobarbital, $(+/-)$ -Amphetamine, Aspirin, Ascorbic acid, Benzoylecgonine, $\beta$ -phenylethylamine, Buprenorphine, Caffeine, Cannabidiol, Chlorpheniramine, Cocaethylene, Cocaine, Codeine, Cotinine, delta9-THC, Diazepam, Dihydrocodeine, Doxylamine, Ecgonine methyl ester, D, l-Ephedrine, l-Ephedrine, d-Ephedrine, S,S (+) Pseudoephedrine, R,R (-) Pseudoephedrine, Heroin, MBDB, MDA, MDEA, MDMA, d-Methamphetamine, Morphine, Oxycodone, Paracetamol, Temazepam, Ibuprofen. Results are summarized in the labeling. Interference at the cutoff is summarized below:
{8}
Structurally Unrelated Compounds Equivalent to the Cutoff--Rx Daytona
| | | Response 300 ng/mL | equivalent to cutoff | Cross- reactivity % | |
| --- | --- | --- | --- | --- | --- |
| Compound | Tested Conc. (ng/mL) | Qualitative | Semi-quantitative | Qualitative | Semi-quantitative |
| 11-hydroxy-delta9-THC | 100,000 | NEG | NEG | 0% | 0% |
| 11-nor9-carboxy-delta9-THC | 100,000 | NEG | NEG | 0% | 0% |
| 6 Acetyl-morphine | 100,000 | NEG | NEG | 0% | 0% |
| Amitriptyline | 100,000 | NEG | NEG | 0.10% | 0.15% |
| Amobarbital | 100,000 | NEG | NEG | 0% | 0% |
| (+/-)-Amphetamine | 100,000 | NEG | NEG | 0% | 0% |
| Aspirin | 100,000 | NEG | NEG | 0% | 0% |
| Ascorbic acid | 100,000 | NEG | NEG | 0% | 0% |
| Benzoylecgonine | 100,000 | NEG | NEG | 0% | 0% |
| β-phenylethylamine | 100,000 | NEG | NEG | 0% | 0.04% |
| Buprenorphine | 100,000 | NEG | NEG | 0% | 0% |
| Caffeine | 100,000 | NEG | NEG | 0% | 0% |
| Cannabidiol | 100,000 | NEG | NEG | 0% | 0% |
| Chlorpheniramine | 100,000 | NEG | NEG | 0.22% | 0.25% |
| Cocaethylene | 100,000 | NEG | NEG | 0.05% | 0.12% |
| Cocaine | 100,000 | NEG | NEG | 0% | 0% |
| Codeine | 100,000 | NEG | NEG | 0% | 0% |
| Cotinine | 100,000 | NEG | NEG | 0% | 0% |
| delta9-THC | 100,000 | NEG | NEG | 0% | 0% |
| Diazepam | 100,000 | NEG | NEG | 0% | 0% |
| Dihydrocodeine | 100,000 | NEG | NEG | 0% | 0% |
| Doxylamine | 100,000 | NEG | NEG | 0.04% | 0.08% |
| Ecgonine methyl ester | 100,000 | NEG | NEG | 0% | 0% |
| D,l-Ephedrine | 100,000 | NEG | NEG | 0% | 0% |
| l-Ephedrine | 100,000 | NEG | NEG | 0% | 0% |
| d-Ephedrine | 100,000 | NEG | NEG | 0% | 0% |
| S,S (+) Pseudoephedrine | 100,000 | NEG | NEG | 0% | 0% |
| R,R (-) Pseudoephedrine | 100,000 | NEG | NEG | 0% | 0% |
| Heroin | 100,000 | NEG | NEG | 0% | 0% |
| MBDB | 100,000 | NEG | NEG | 0% | 0% |
| MDA | 100,000 | NEG | NEG | 0% | 0% |
| MDEA | 100,000 | NEG | NEG | 0% | 0% |
| MDMA | 100,000 | NEG | NEG | 0% | 0% |
| d-Methamphetamine | 100,000 | NEG | NEG | 0.05% | 0.08% |
| Morphine | 100,000 | NEG | NEG | 0% | 0% |
| Oxycodone | 100,000 | NEG | NEG | 0% | 0% |
| Paracetamol | 100,000 | NEG | NEG | 0% | 0% |
| Temazepam | 100,000 | NEG | NEG | 0% | 0% |
| Ibuprofen | 100,000 | NEG | NEG | 0% | 0% |
{9}
Structurally Unrelated Compounds Equivalent to the Cutoff--Rx Imola
| | | Response 300 ng/mL | equivalent to cutoff | Cross- | reactivity % |
| --- | --- | --- | --- | --- | --- |
| Compound | Tested Conc. (ng/mL) | Qualitative | Semi-quantitative | Qualitative | Semi-quantitative |
| 11-hydroxy-delta9-THC | 100,000 | NEG | NEG | 0% | 0% |
| 11-nor9-carboxy-delta9-THC | 100,000 | NEG | NEG | 0% | 0% |
| 6 Acetyl-morphine | 100,000 | NEG | NEG | 0% | 0% |
| Amitriptyline | 100,000 | NEG | NEG | 0.10% | 0.15% |
| Amobarbital | 100,000 | NEG | NEG | 0% | 0% |
| (+/-)-Amphetamine | 100,000 | NEG | NEG | 0% | 0% |
| Aspirin | 100,000 | NEG | NEG | 0% | 0% |
| Ascorbic acid | 100,000 | NEG | NEG | 0% | 0% |
| Benzoylecgonine | 100,000 | NEG | NEG | 0% | 0% |
| β-phenylethylamine | 100,000 | NEG | NEG | 0% | 0% |
| Buprenorphine | 100,000 | NEG | NEG | 0% | 0% |
| Caffeine | 100,000 | NEG | NEG | 0% | 0% |
| Cannabidiol | 100,000 | NEG | NEG | 0% | 0% |
| Chlorpheniramine | 100,000 | NEG | NEG | 0.27% | 0.19% |
| Cocaethylene | 100,000 | NEG | NEG | 0.04% | 0.13% |
| Cocaine | 100,000 | NEG | NEG | 0% | 0% |
| Codeine | 100,000 | NEG | NEG | 0% | 0% |
| Cotinine | 100,000 | NEG | NEG | 0% | 0% |
| delta9-THC | 100,000 | NEG | NEG | 0% | 0% |
| Diazepam | 100,000 | NEG | NEG | 0% | 0% |
| Dihydrocodeine | 100,000 | NEG | NEG | 0% | 0% |
| Doxylamine | 100,000 | NEG | NEG | 0.04% | 0.06% |
| Ecgonine methyl ester | 100,000 | NEG | NEG | 0% | 0% |
| D,l-Ephedrine | 100,000 | NEG | NEG | 0% | 0.03% |
| l-Ephedrine | 100,000 | NEG | NEG | 0% | 0% |
| d-Ephedrine | 100,000 | NEG | NEG | 0% | 0% |
| S,S (+) Pseudoephedrine | 100,000 | NEG | NEG | 0% | 0% |
| R,R (-) Pseudoephedrine | 100,000 | NEG | NEG | 0% | 0% |
| Heroin | 100,000 | NEG | NEG | 0% | 0% |
| MBDB | 100,000 | NEG | NEG | 0% | 0% |
| MDA | 100,000 | NEG | NEG | 0% | 0% |
| MDEA | 100,000 | NEG | NEG | 0% | 0% |
| MDMA | 100,000 | NEG | NEG | 0% | 0% |
| d-Methamphetamine | 100,000 | NEG | NEG | 0.13% | 0.05% |
| Morphine | 100,000 | NEG | NEG | 0% | 0% |
| Oxycodone | 100,000 | NEG | NEG | 0% | 0% |
| Paracetamol | 100,000 | NEG | NEG | 0% | 0% |
| Temazepam | 100,000 | NEG | NEG | 0% | 0% |
| Ibuprofen | 100,000 | NEG | NEG | 0% | 0% |
{10}
Endogenous interfering compounds were tested at $\pm 25\%$ of cutoff for the Rx Imola and Rx Daytona for Total Bilirubin, Direct Bilirubin, Hemoglobin, Creatinine, Urea, Glucose, H.S.A., Ethanol, Acetone, Gamma globulin, Oxalic acid, Riboflavin, Sodium chloride, Boric acid, Sodium azide, Sodium fluoride. Results are summarized in the labeling. Interference at the cutoff is in the table below. No interference was observed at the cutoff.
Endogenous Interference Equivalent to the Cutoff—Rx Daytona
| | | Response 300 ng/mL | equivalent to cutoff | Cross- reactivity % | |
| --- | --- | --- | --- | --- | --- |
| Compound | Tested Conc. (ng/mL) | Qualitative | Semi-quantitative | Qualitative | Semi-quantitative |
| Total Bilirubin | 15 | NEG | NEG | 0% | 0% |
| Direct Bilirubin | 5 | NEG | NEG | 0% | 0% |
| Haemoglobin | 115 | NEG | NEG | 0% | 0% |
| Creatinine | 30 | NEG | NEG | 0% | 0% |
| Urea | 258 | NEG | NEG | 0% | 0% |
| Glucose | 2000 | NEG | NEG | 0% | 0% |
| H.S.A. | 500 | NEG | NEG | 0% | 0% |
| Ethanol | 1000 | NEG | NEG | 0% | 0% |
| Acetone | 1000 | NEG | NEG | 0% | 0% |
| Gamma globulin | 500 | NEG | NEG | 0% | 0% |
| Oxalic acid | 100 | NEG | NEG | 0% | 0% |
| Riboflavin | 7.5 | NEG | NEG | 0% | 0% |
| Sodium chloride | 6000 | NEG | NEG | 0% | 0% |
| Boric acid | 1000 | NEG | NEG | 0% | 0% |
| Sodium azide | 1000 | NEG | NEG | 0% | 0% |
| Sodium fluoride | 1000 | NEG | NEG | 0% | 0% |
Endogenous Interference Equivalent to the Cutoff—Rx Imola
| | | Response 300 ng/mL | equivalent to cutoff | Cross- reactivity % | |
| --- | --- | --- | --- | --- | --- |
| Compound | Tested Conc. (ng/mL) | Qualitative | Semi-quantitative | Qualitative | Semi-quantitative |
| Total Bilirubin | 15 | NEG | NEG | 0% | 0% |
| Direct Bilirubin | 5 | NEG | NEG | 0% | 0% |
| Haemoglobin | 115 | NEG | NEG | 0% | 0% |
| Creatinine | 30 | NEG | NEG | 0% | 0% |
| Urea | 258 | NEG | NEG | 0% | 0% |
| Glucose | 2000 | NEG | NEG | 0% | 0% |
| H.S.A. | 500 | NEG | NEG | 0% | 0% |
| Ethanol | 1000 | NEG | NEG | 0% | 0% |
| Acetone | 1000 | NEG | NEG | 0% | 0% |
| Gamma globulin | 500 | NEG | NEG | 0% | 0% |
| Oxalic acid | 100 | NEG | NEG | 0% | 0% |
| Riboflavin | 7.5 | NEG | NEG | 0% | 0% |
| Sodium chloride | 6000 | NEG | NEG | 0% | 0% |
| Boric acid | 1000 | NEG | NEG | 0% | 0% |
| Sodium azide | 1000 | NEG | NEG | 0% | 0% |
| Sodium fluoride | 1000 | NEG | NEG | 0% | 0% |
Specific Gravity studies were performed on the Rx Daytona and Rx Imola.
{11}
Two EDDP negative urine pools were spiked with EDDP to + and – 25% cutoff (300 ng/mL). The pools were further divided and had sodium chloride added to achieve specific gravities of 1.000 (water blank), 1.010, 1.020, 1.030. Specific gravities were confirmed by a commercial method. Each sample was analyzed in replicates of 10 in the semi-quantitative and qualitative modes. Recoveries were compared to the spiked target samples and were expected to be ≤ 10%. Specific gravity from 1.000-1.030 had no significant (≤10%) effect on EDDP results for the Rx Daytona or Rx Imola.
pH studies were performed using the spiked pools from the specific gravity studies. pH was adjusted by adding either 1M HCl or 1M NaOH to obtain samples with pH levels of 3, 5, 7, 9, 11. pH was confirmed by a pH meter. Each sample was analyzed in replicates of 10 in the semi-quantitative and qualitative modes. Recoveries were compared to the pH 7 sample and were expected to be ≤ 10% of the target values. pH from 3-11 had no significant (≤10%) effect on EDDP results for the Rx Daytona or Rx Imola.
f. Assay cut-off:
Analytical performance of the device around the claimed cutoff is described in precision section (1 a.) above.
2. Comparison studies:
a. Method comparison with predicate device:
40 positive and 40 negative unaltered natural samples were analyzed on the Rx Daytona and Rx Imola and by GC/MS. The results with discordant tables are as follows:
| RX Daytona Semi-quantitative | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | | |
| Candidate Device Results | Negative | Less than half the cut-off (concentration by GC/MS analysis) | Near Cutoff Negative (Between 50% below the cut-off and the cut-off concentration) | Near Cutoff Positive (Between the cut-off and 50% above the cut-off concentration) | High Positive (greater than 50% above the cut-off concentration) | % Agreement with GC/MS |
| Positive | - | - | 2 | 8 | 32 | 100.00% |
| Negative | 31 | - | 7 | - | - | 95.00% |
| Cutoff Value | Semi-quantitative EDDP Rx Daytona | Drug/Metabolite GC/MS EDDP (ng/mL) |
| --- | --- | --- |
| 300 ng/mL | Pos | 272 |
| | Pos | 290 |
{12}
13
| RX Daytona Qualitative | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | | |
| Candidate Device Results | Negative | Less than half the cut-off (concentration by GC/MS analysis) | Near Cutoff Negative (Between 50% below the cut-off and the cut-off concentration) | Near Cutoff Positive (Between the cut-off and 50% above the cut-off concentration) | High Positive (greater than 50% above the cut-off concentration) | % Agreement with GC/MS |
| Positive | - | - | 3 | 8 | 32 | 100.00% |
| Negative | 31 | - | 6 | - | - | 92.50% |
| Cutoff Value | Qualitative EDDP Rx Daytona | Drug/Metabolite GC/MS EDDP (ng/mL) |
| --- | --- | --- |
| 300 ng/mL | Pos | 272 |
| | Pos | 290 |
| | Pos | 292 |
| RX Imola Semi-quantitative | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | | |
| Candidate Device Results | Negative | Less than half the cut-off (concentration by GC/MS analysis) | Near Cutoff Negative (Between 50% below the cut-off and the cut-off concentration) | Near Cutoff Positive (Between the cut-off and 50% above the cut-off concentration) | High Positive (greater than 50% above the cut-off concentration) | % Agreement with GC/MS |
| Positive | - | - | 4 | 8 | 32 | 100.00% |
| Negative | 31 | - | 5 | - | - | 90.00% |
| Cutoff Value | Qualitative EDDP Rx Imola | Drug/Metabolite GC/MS EDDP (ng/mL) |
| --- | --- | --- |
| 300 ng/mL | Pos | 272 |
| | Pos | 283 |
| | Pos | 290 |
| | Pos | 292 |
| RX Imola Qualitative | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | | |
| Candidate Device Results | Negative | Less than half the cut-off (concentration by GC/MS analysis) | Near Cutoff Negative (Between 50% below the cut-off and the cut-off concentration) | Near Cutoff Positive (Between the cut-off and 50% above the cut-off concentration) | High Positive (greater than 50% above the cut-off concentration) | % Agreement with GC/MS |
| Positive | - | - | 5 | 8 | 32 | 100.00% |
| Negative | 31 | - | 4 | - | - | 87.50% |
| Cutoff Value | Qualitative EDDP Rx Imola | Drug/Metabolite GC/MS EDDP (ng/mL) |
| --- | --- | --- |
| 300 ng/mL | Pos | 247 |
| | Pos | 272 |
| | Pos | 283 |
| | Pos | 290 |
| | Pos | 292 |
{13}
b. Matrix comparison:
Not applicable. This device is for urine only.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.