The VOLUX consists of an x-ray device that uses a cone beam with a rotational sequence, providing two dimensional images and three dimensional volume reconstructions of the head area, which includes dental and maxillofacial areas, for use in planning and diagnostic support. VOLUX comprises a package of PC software modules to expand TRIANA capabilities to handling 3D data. This includes 3D reconstructions, storage, retrieval, viewing and processing of 3D-image data.
Device Story
VOLUX is a digital extraoral cone beam X-ray system; generates conical X-ray beam rotating around patient head; captures 2D exposures; reconstructs 3D volume images. System includes VOLUX hardware, reconstruction server, and TRIANA 2D/3D viewing client. Used in clinical settings by dental/maxillofacial professionals for diagnostic support and treatment planning. Clinicians use TRIANA software to navigate, view, and process 3D data. Output assists in clinical decision-making by providing anatomical visualization of dental and maxillofacial structures.
Clinical Evidence
The document states that bench and clinical evaluation were performed to support the safety and effectiveness of the device. No specific quantitative metrics (e.g., sensitivity, specificity, AUC) or study sample sizes are provided in the summary.
Technological Characteristics
Cone beam X-ray system; 60-85kV, 5-7mA; pulse mode; image intensifier detector; 0.5mm focal spot; 2.8mm Al equivalent total filtration. Reconstruction via cone beam algorithm. PC-based software (TRIANA) for 3D data handling. Dimensions: 1500x1750x1950mm; weight: 220kg.
Indications for Use
Indicated for dental and maxillofacial imaging in patients requiring 2D or 3D head area diagnostic support and treatment planning.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
#### Exhibit 5 510(k) Summary
Digital Extraoral Source X-Ray System / Model: VOLUX
# APR 2 6 2010
| 1. | Submitter and US Official Correspondent |
|----------------|-----------------------------------------|
| Submitter: | GENORAY Co., Ltd. |
| Address: | #512, Byucksan Technopia |
| | 434-6, Sangdaewon 1-dong, |
| | Jungwon-gu, Seongnam-city |
| | Gyeonggi-do, 462-716, Korea |
| Telephone No.: | +82-31-737-8020 |
| Fax: | +82-31-737-8025 |
### Official Correspondent (U.S): Jae Kim - Business Manager
| Correspondent : | GENORAY America Inc. |
|-----------------|-----------------------|
| Address: | 1073 N. Batavia St. |
| | Orange, CA 92867, USA |
| Telephone No .: | 714-289-8020 |
| Fax: | 714-453-9661 |
| Email: | jac@genorav.com |
## 2. Establishment Registration Number 3005843418
#### 3. Device Information
| Proprietary/Trade Name: | Digital Extraoral Source X-Ray System / Model: VOLUX |
|-------------------------|------------------------------------------------------|
| Common/Usual Name: | Digital Extraoral Source X-Ray System |
| Classification Name: | System , X-ray, Extraoral Source, Digital |
| Product Code: | MUH |
| Device Class: | Class II per regulation 21 CFR 872.1800 |
#### 4. Equivalent Legally Marketed Device
| | < CB MercuRay > |
|-----------------|--------------------------------------------------------------------------------------------|
| Manufacturer: | Hitachi Medical Systems America, Inc. |
| Device Name: | CB MercuRay |
| 510(k) Number: | K033248 (Decision Date - October 20, 2003) |
| Classification: | System, X-Ray, Extraoral Source, Digital : MUH,<br>Class II per regulation 21 CFR 872.1800 |
.
{1}------------------------------------------------
| < Galileos > | |
|-----------------|-------------------------------------------------|
| Manufacturer: | Sirona Dental Systems GmbH, |
| Device Name: | Galileos |
| 510(k) Number: | K060892 (Decision Date - April 14, 2006) |
| Classification: | System, X-Ray, Extraoral Source, Digital : MUH, |
| | Class II per regulation 21 CFR 872.1800 |
### 5. Description of the Device
The VOLUX comprises of the VOLUX-device, the reconstruction server and the 2D and 3D viewing client TRIANA. The VOLUX device generates a conical x-ray beam that rotates round the patient's head within a ccrtain angle. From the obtained exposures a three dimensional image is reconstructed and can be viewed. The VOLUX features the navigation within this displayed volume and special views may be selected, calculated and eventually displayed.
| Product | VOLUX |
|---------------------------|------------------------------------------------------|
| Items | |
| kV range | 60~85kV |
| mA range | 5~7mA |
| mAs range | 30~42mAs |
| X-ray Source Mode | Pulse mode |
| Image Detector | Image Intensifier |
| Voxel Size | 6": 0.1667mm<br>4": 0.1111mm |
| Camera Pixel | 1004(H) x 1004(V) |
| Image Acquistion | 220° |
| Exposure time | 10sec |
| Scan time | 20sec |
| Image reconstruction time | 3 min |
| Reconstruction type | Cone beam |
| Focal spot size (small) | 0.5mm |
| Total filtration | 2.8mm Al equivalent<br>(inherent filtration 0.8mmAl) |
| Source to skin distance | 200mm |
| Main unit Dimensions (mm) | 1500 X 1750 X 1950 |
| Total weight | 220kg |
{2}------------------------------------------------
### 6. Indications for use
The VOLUX consists of an x-ray device that uses a cone beam with a rotational sequence, providing two dimensional images and three dimensional volume reconstructions of the head area, which includes dental and maxillofacial areas, for use in planning and diagnostic support. VOLUX comprises a package of PC software modules to expand TRIANA capabilities to handling 3D data. This includes 3D reconstructions, storage, retrieval, viewing and processing of 3D-image data.
### 7. Safety and Effectiveness, comparison to Predicate
The result of bench and clinical evaluation indicates that the new device is as safe and effective as the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
## APR 2 6 2010
GENORAY Co., Ltd. % Mr. Jae Kim Business Development Manager GENORAY America, Inc. · 1073 N. Batavia St. ORANGE CA 92867
Re: K092171
Trade/Device Name: Digital Extraoral Source X-Ray System (Models: VOLUX) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: March 23, 2010 Received: March 24, 2010
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Donald J. Trump
Donald J. St Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for use Exhibit 4
510(k) number (if known): K 092171
Device Name: Digital Extraoral Source X-Ray System (Models: VOLUX)
Indications for Use:
The VOLUX consists of an x-ray device that uses a cone beam with a rotational sequence, providing two dimensional images and three dimensional volume reconstructions of the head area, which includes dental and maxillofacial areas, for use in planning and diagnostic support.
VOLUX comprises a package of PC software modules to expand TRIANA capabilities to handling 3D data. This includes 3D reconstructions, storage, retrieval, viewing and processing of 3D-image data.
Prescription Use > (Part 21 CFR 801 Subpari D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Deviee Exaluation (ODE) ODE) OF 1
Robert Beckerf
(Division Sign-Off)
Division of Radiological Devices/
Office of In
510K K092 N1
GENORAY Co., Ltd.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.