The Presidio PACS software is intended for the manipulation, management, and display of medical images. It can manage and display images from different modalities and interfaces and can distribute those images to various workstation, image storage and printing devices using DICOM or similar standards. Typical users of this system are trained medical professionals, including physicians, nurses, technicians and computer system professionals. To support the diagnostic interpretation of mammography studies, Presidio PACS will display the full fidelity DICOM image in a non-compressed format. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation. It is the user's responsibility to ensure quality, ambient light conditions and image compression ratios are consistent with the clinical application.
Device Story
Presidio PACS is a software-only Picture Archiving and Communication System (PACS) comprising workstation, enterprise, and RIS modules. It accepts medical images from various modalities via DICOM or similar standards. The system manages, stores (uncompressed, lossless, or lossy), and distributes images to workstations and printing devices. It provides tools for viewing, editing, annotating, analyzing, and processing images (e.g., window/level, pan, zoom, cine display, segmentation, smoothing). Changes to image presentation are saved as display definitions, preserving original pixel data. Used in clinical environments or remote locations by trained medical professionals. The system supports primary mammography interpretation on FDA-approved 5MP monitors. It facilitates electronic scheduling and exam management via the RIS module. The software runs on off-the-shelf PC/server hardware. It does not provide diagnostic assistance; clinical decisions remain the responsibility of the physician.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by software validation testing demonstrating the device meets its intended use and specifications.
Technological Characteristics
Software-only PACS; runs on off-the-shelf PC/server hardware. Supports DICOM standard protocol for image storage, retrieval, and transmission. Features modular software architecture with plug-in support. Includes JPEG and JPEG 2000 compression (lossy/lossless). Provides image processing tools: window/level, pan, zoom, cine display, segmentation, and smoothing. No patient contact; no life-sustaining function.
Indications for Use
Indicated for trained medical professionals (physicians, nurses, technicians, computer system professionals) for the manipulation, management, and display of medical images from various modalities. Supports primary interpretation of digital mammography images when using an FDA-approved 5MP monitor. Contraindicated for primary diagnosis or interpretation of lossy compressed mammographic images and digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K092125
# 5. 510(k) Summary of Safety and Effectiveness
Reaonix Security, LLC 2434 Malden Park Dr. Buford, GA 30519
DEC 1 0 2009
Date Summary Prepared: June 5, 2009
Contact Person: Gerald Jackson Reaonix Security, LLC Telephone: 678-985-3434 Fax: 678-985-3434
Device Name: Presidio PACS Trade Name(s): Presidio PACS Classification Name: Picture Archiving and Communications System Classification Regulation: Class II - 892.2050 Panel Code Device: Radiology Product Code: LLZ
Predicate Device Information: Device Name: IMAGESVR PACS Manufacturer: JAGUAR TECHNOLOGY LIMITED Reference: K082808
Device Name: DATACOM DC PACS Manufacturer: DATACOM TECHNOLOGY CORP. Reference: K083182
Device Description:
The Presidio PACS include the following major components: Workstation, Enterprise and RIS.
The Presidio PACS workstation is used to view, edit, manipulate, annotate, analyze, store and distribute images and data that are stored and managed in the Presidio PACS Enterprise for diagnosis. This software-based product provides capabilities for the acceptance, transmission, printing, display, storage, editing and digital processing of medical images and associated data.
All acquired image data is preserved in the format in which it is received. Changes may be made to the presentation of the images. These changes are saved as display definitions only and do not alter the acquired image pixel data. Any and all display definitions applied to an image can always be reversed to the acquired state.
The Presidio PACS workstation may also be used in a remote location, away from the healthcare facility, as long as the workstation has the ability to connect, via network, to the primary healthcare facility where the Presidio PACS Enterprise is located.
The Presidio PACS Workstation extends its diagnostic and productivity capabilities into the mammography reading environment and may also be used for the primary interpretation of digital mammography images with a FDA approve 5MP monitor.
{1}------------------------------------------------
The Presidio PACS Workstation has a modular software architecture which allows adding addition feature or enhancements in the form of plug-ins without any modification to the existing software code base.
Presidio PACS Enterprise software delivers a complete, scalable storage solution for diagnostic images. Images can be stored in uncompressed, lossless or lossy. The system also has the ability to send data to DICOM ready devices via the DICOM standard protocol. It is a DICOM compliant solution for image storage, retrieval and transmission. The Enterprise Archive provides redundancy in long-term storage in several ways, including Redundant Archives, Media Copy, and Application Service Provider (ASP) Archive.
Presidio RIS provides a hospital or clinic with automated tools to electronically schedule and manages patient exam information. The Presidio RIS utilizes a Workstation client that provides referring physicians the added flexibility to instantly view and schedule their own patients' radiographic services.
The Presidio PACS software application is a software only solution and will use 'off the shelf' PC and server hardware technology that meets defined minimum specifications.
#### Intended Use:
The Presidio PACS software is intended for the manipulation, management, and display of medical images. It can manage and display images from different modalities and interfaces and can distribute those images to various workstation, image storage and printing devices using DICOM or similar standards. Typical users of this system are trained medical professionals, including physicians, nurses, technicians and computer system professionals.
To support the diagnostic interpretation of mammography studies, Presidio PACS will display the full fidelity DICOM image in a non-compressed format. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation.
It is the user's responsibility to ensure quality, ambient light conditions and image compression ratios are consistent with the clinical application.
{2}------------------------------------------------
Comparison to Predicate Device(s):
| Product Name | Presidio PACS | ImageSVR PACS | DATACOM DC-PACS |
|---------------------------------------------------------------------------------|---------------|---------------|-----------------|
| 510(k) Number | | K082808 | K083182 |
| Communications | TCP/IP | TCP/IP | TCP/IP |
| Image Archive | YES | YES | YES |
| Image Processing | Yes | YES | YES |
| Image Edit | Yes | YES | YES |
| Edit Patient Demographics | YES | YES | YES |
| Add and remove images | YES | YES | YES |
| Combine studies | YES | YES | YES |
| Edit Patient Orientation Information | YES | YES | YES |
| Set and Edit Routing Informatiion | YES | YES | YES |
| JPEG Lossy/Lossless Compression | YES | YES | YES |
| JPEG 2000 Lossy/Lossless Compression | YES | YES | NO |
| Image Processing (Image segmentation, image smoothing) | YES | YES | YES |
| Image Processing (Window Level, Pan, Zoom,Variable Smooth Filter, Cine Display) | YES | YES | YES |
| DICOM Print | YES | YES | YES |
### SUBSTANTIAL EQUIVALENCE CONCLUSIONS:
A claim of substantial equivalence is made to ImageSVR (K082808) and DATACOM DC-PACS (K083182). All of them have the same working principle and technologies. The differences are due to the feature design aspects, not relating to the safety or effectiveness aspects. Besides, the submission contains the results of software validation that the risks analysis and the potential hazards have been classified Minor.
Thus they are substantially equivalent.
#### Technological Characteristics:
The device does not contact the patient, nor does it control any life sustaining devices. The software does not provide any diagnostic assistance to the physician. Any diagnostic determination or treatment is solely determined by a physician and not the software. A physician, providing ample opportunity for competent human intervention, interprets images and information being displayed and printed.
{3}------------------------------------------------
#### SAFETY INFORMATION:
The Presidio PACS has no patient contact and is utilized only by trained professionals. Trained professionals allow sufficient review to give identification and intervention in the event of a malfunction. Patient data is limited to authorized individuals. The Presidio PACS utilize "off-the-shelf" servers and Personal computers. It complies with applicable electrical safety standards for standard hardware and peripherals.
#### TESTING AND CONCLUSIONS:
ﻨ
Software validation has established the device's ability to meet its intended use and established specifications. The information provided in this traditional premarket notification submission has shown that the Presidio PACS is substantially equivalent to the predicate device(s).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three heads facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle. The logo is black and white.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002.
# DEC 1 0 2009
Mr. Gerald Jackson CEO Reaonix Security, LLC 2434 Malden Park Dr. BUFORD GA 30519
Re: K092125
Trade/Device Name: Presidio PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 20, 2009 Received: November 23, 2009
Dear Mr. Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical
{5}------------------------------------------------
Enclosure
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
## 4. Indication for Use Statement
Indications for Use 510(k) Number (if known): _780 12125 Device Name: _____________ Presidio PACS
Indications for Use:
The Presidio PACS software is intended for the manipulation, management, and display of medical images. It can manage and display images from different modalities and interfaces and can distribute those images to various workstation, image storage and printing devices using DICOM or similar standards. Typical users of this system are trained medical professionals, including physician, nurses, technicians and computer system professionals.
To support the diagnostic interpretation of mammography studies, Presidio PACS will display the full fidelity DICOM image in a non-compressed format. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation.
It is the user's responsibility to ensure quality, ambient light conditions and image compression ratios are consistent with the clinical application.
Prescription Use _ XXX Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
H.R. Lewis
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.