K092117 · Ritm Okb Zao · GZJ · May 28, 2010 · Neurology
Device Facts
Record ID
K092117
Device Name
SCENAR, MODEL 10
Applicant
Ritm Okb Zao
Product Code
GZJ · Neurology
Decision Date
May 28, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SCENAR is indicated for: - . relief and management of chronic and acute pain - adjunctive treatment in the management of post-surgical and post-traumatic . pain.
Device Story
SCENAR devices are transcutaneous electrical nerve stimulators used for chronic/acute pain relief and post-surgical/post-traumatic pain management. Operated by medical professionals in clinical settings or by patients at home via doctor's prescription. Device generates low/medium frequency bipolar electric pulses; pulse shape varies dynamically based on skin electrophysiological characteristics. Output delivered via built-in stainless steel electrode. Healthcare providers use device to provide independent or adjunctive pain therapy. Benefits include non-invasive pain management.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through comparative bench testing against predicate device and adherence to international safety standards (IEC 60601 series).
Technological Characteristics
Transcutaneous electrical nerve stimulator. Materials: stainless steel (electrode), ABS (case). Energy source: alkaline batteries. Output: low/medium frequency bipolar pulses (14-350 Hz, 7-2000 μS). Dimensions: 140x55x35 mm to 190x70x40 mm. Safety standards: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2.
Indications for Use
Indicated for relief and management of chronic and acute pain, and as adjunctive treatment for post-surgical and post-traumatic pain in patients requiring pain management.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the text 'K092117' enclosed in a partial circle. The text is in a simple, sans-serif font and is horizontally centered. The circle begins above the 'K' and ends after the last digit, suggesting an identifier or code within a circular marking.
# Закрытое акционерное общество "ОКБ "РИТМ" RITM OKB ZAO
MAY 2 8 2010
ЗАО "ОКБ "РИТМ", ул.Петровская, 99, г.Таганрог, Россия, 347900 Тел /факс: (8634) 62-31-79 E-mail: medso@scenar.com.ru RITM OKB ZAO Petrovskava St., 99 Taganrog, Russia, 347900 Tel. ffax: {8634} 62-31-79 E-mail: medsc@scenar.com.ru
# 510(k) Summary
#### Submitter Information:
RITM OKB ZAO Address: 99, Petrovskaya str., Taganrog, Russia, 347900 Phone/fax: +7 (8634) 623-179 Contact person: Larisa Shpungina Date of Summary Preparation: November 11, 2009
Trade Name: SCENAR
Common Name: TENS device
#### Classification Name:
TRANSCUTANEOUS ELECTRICAL NERVE STIMULATOR FOR PAIN RELIEF (21 CFR 882.5890, Product Code GZJ)
#### Predicate Device:
InterX5000 (K042912)
#### Device Description:
### Purpose
SCENAR transcutaneous electrostimulators are designed for delivering general noninvasive treatment via human skin areas for relief and management of chronic and acute pain, as adjunctive treatment in the management of post-surgical and post-traumatic pain.
SCENAR devices can be used:
- at home on a doctor's prescription to relief pain after various injuries and traumas;
- by medical professionals in medical and prophylactic institutions - as an independent remedy as well as combined with other treatments.
#### Action: How it works
SCENAR devices generate low and medium frequency bipolar electric pulses. The shape of the pulses dynamically varies with changing electrophysiological characteristics of the skin.
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the text "K092117" inside of an oval shape. The text is in a simple, sans-serif font and is centered within the oval. The oval is outlined in black and appears to be hand-drawn.
# List of produced devices
SCENAR devices produced by RITM OKB ZAO are listed in the Table 1. The devices are united into series since their purpose, application, mechanism of action, utilized materials, basic technical characteristics, functionality and manufacturing processes are similar.
Table 1
| Device Model and Version |
|--------------------------------------------------------------|
| SCENAR-1-NT (version 02) – Transcutaneous electrostimulator. |
| CHANS-01-SCENAR - Transcutaneous electrostimulator. |
| CHANS-01-SCENAR-M - Transcutaneous electrostimulator. |
# Specifications
Specifications of SCENAR devices are given in Table 2.
Table 2
| Parameter | SCENAR-1-NT (version 02) | CHANS-01-SCENAR<br>CHANS-01-SCENAR-M |
|-------------------------------------------------------------|----------------------------|--------------------------------------|
| Supply voltage | 4.0 to 6.4 V | 9 V |
| Maximum supply<br>current | not more than 650 mA | not more than 85 mA |
| Amplitude of the<br>stimulating pulse at<br>a standard load | not more than 150 V | not more than 150 V |
| Pulse frequency | 15 to 350 Hz | 14, 60, 90, 340 Hz |
| Pulse Width | 7 - 2000 μS | 7 - 1000 μS |
| Maximum Output<br>RMS Voltage | 12 V @ 500 Ω | 15 V @ 500 Ω |
| | 15 V @ 2 kΩ | 20 V @ 2 kΩ |
| | 15 V @ 10 kΩ | 20 V @ 10 kΩ |
| Maximum Output<br>RMS Current | 25 mA @ 500 Ω | 30 mA @ 500 Ω |
| | 8 mA @ 2 kΩ | 10 mA @ 2 kΩ |
| | 1.5 mA @ 10 kΩ | 2 mA @ 10 kΩ |
| Weight | not more than 0.4 kg | not more than 0.2 kg |
| Dimensions | not more than 190x70x40 mm | not more than 140x55x35 mm |
# Design Description
The device has an upper cover, case with a built-in electrode and a battery cover. All components except for the batteries are located on the printed circuit board inside the device's case. The device's controls and visual indicators are located on the upper side of its case.
# Materials
Stainless steel - electrode, ABS - case.
Power supply
Alkaline batteries.
# Intended Use
SCENAR is indicated for:
- . relief and management of chronic and acute pain
- adjunctive treatment in the management of post-surgical and post-traumatic . pain.
{2}------------------------------------------------
The comparison of SCENAR and INTERX5000 Specifications is given in Table 3.
| Table 3 | | InterX5000 |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Parameter | SCENAR | |
| Supply voltage | CHANS-01-SCENAR, CHANS-01-SCENAR-M - 9 V<br>SCENAR-1-NT (versions 02) - 4.0 to 6.4 V | 9 V |
| Maximum supply current | CHANS-01-SCENAR, CHANS-01-SCENAR-M - not more than 85 mA<br>SCENAR-1-NT (versions 02) – not more than 650 mA | data not available |
| Amplitude of the stimulating pulse at a standard load | not more than 150 V | not more than 450 V<br>(measuring conditions not available) |
| Pulse frequency | CHANS-01-SCENAR, CHANS-01-SCENAR-M – 14, 60, 90, 340 Hz<br>SCENAR-1-NT (version 02) - 15 to 350 Hz | 15 to 350 Hz |
| Weight | CHANS-01-SCENAR, CHANS-01-SCENAR-M - not more than 0.2 kg<br>SCENAR-1-NT (versions 02) – not more than 0.4 kg | not more than 0.3 kg |
| Dimensions | CHANS-01-SCENAR, CHANS-01-SCENAR-M - not more than 140x55x35 mm<br>SCENAR-1-NT (versions 02) - not more than 190x70x40 mm | 220x60x45 mm |
Moreover, to support our determination of substantial equivalence, evaluation tests have been performed. The results of these tests demonstrate that technological characteristics and performance of SCENAR devices are substantially equivalent to those of InterX5000. The test reports have been provided as part of this premarket notification.
The software verification has been conducted according to the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff- May 11, 2005.
RITM OKB ZAO has received test reports certifying that SCENAR was tested and found to be in conformity with IEC 60601-2-10:2001 + A1:2001 and IEC 60601-1:1988 + Al:1991 + A2:1995 and EMC: IEC 60601-1-2:2001 (ed. 2) international safety standards.
The test reports have been provided as part of this premarket notification.
SCENAR has received the CE Mark in the category Impulse Therapy devices type SCENAR for the devices in this submission certifying that the devices meet the EC-Directive 93/42/EEC.
The Certificate has been provided as part of this premarket notification.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the text 'K092117' inside of an oval shape. The text is centered within the oval. The oval is outlined in black.
#### Conclusions
The devices have intended uses and technological characteristics that are substantially equivalent to those of the predicate device InterX5000 (K042912).
Verification and validation tests as well as certificates and test reports contained in this submission demonstrate that the submitted models are equivalent to the safety and effectiveness as that of the cleared devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the address for the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Mail Center - WO66-G609, Silver Spring, MD 20993-0002. The text is left-aligned and uses a serif font.
RITM OKB ZAO c/o Ms. Larisa Shpungina Certification and Licensing Engineer 99 Petrovskaya str., Taganrog, 347900 Russia
Re: K092117
MAY 2 8 2010
Trade Name(s): CHANS-01-SCENAR,
> CHANS-01-SCENAR - M. SCENAR-1-NT (version 02).
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ Dated: April 08, 2010 Received: April 23, 2010
Dear Ms. Shpungina:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
الات Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K092117
Device Name:
CHANS-01-SCENAR, CHANS-01-SCENAR - M, SCENAR-1-NT (version 02).
Indications for Use:
For relief and management of chronic and acute pain, as adjunctive treatment in the management of post-surgical and post-traumatic pain.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (21 CFR Part 801 Subpart D)
i.
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Aesm
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K092117
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.